Published in: Journal of Cardiovascular Magnetic Resonance 1/2008

Open Access 01-12-2008 | Technical notes

Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla

Authors: Frank G Shellock, Samuel Valencerina

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2008

Login to get access

Abstract

Purpose

To evaluate MRI issues at 3-Tesla for a ventricular assist device (VAD).

Methods

The AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached (Abiomed, Inc., Danvers, MA) was evaluated for magnetic field interactions, heating, and artifacts at 3-Tesla. MRI-related heating was assessed with the device in a gelled-saline-filled, head/torso phantom using a transmit/received RF body coil while performing MRI at a whole body averaged SAR of 3-W/kg for 15-min. Artifacts were assessed for the main metallic component of this VAD (atrial cannula) using T1-weighted, spin echo and gradient echo pulse sequences.

Results

The AB5000 Ventricle with the prototype In-Flow Cannula and Out-Flow Cannula attached showed relatively minor magnetic field interactions that will not cause movement in situ. Heating was not excessive (highest temperature change, +0.8°C). Artifacts may create issues for diagnostic imaging if the area of interest is in the same area or close to the implanted metallic component of this VAD (i.e., the venous cannula).

Conclusion

The results of this investigation demonstrated that it would be acceptable for a patient with this VAD (AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached) to undergo MRI at 3-Tesla or less. Notably, it is likely that the operation console for this device requires positioning a suitable distance (beyond the 100 Gauss line or in the MR control room) from the 3-Tesla MR system to ensure proper function of the VAD.
Appendix
Available only for authorised users
Literature
1.
go back to reference Moazami N, Smedira NG: Temporary mechanical support. J Card Surg. 2001, 16: 193-202. 10.1111/j.1540-8191.2001.tb00508.x.CrossRefPubMed Moazami N, Smedira NG: Temporary mechanical support. J Card Surg. 2001, 16: 193-202. 10.1111/j.1540-8191.2001.tb00508.x.CrossRefPubMed
2.
go back to reference Leshnower BG, Gleason TG, O'Hara ML, Pochettino A, Woo YJ, Morris RJ, Gardner TJ, Acker MA: Safety and efficacy of left ventricular assist device support in postmyocardial infarction cardiogenic shock. Ann Thorac Surg. 2006, 81: 1365-70. 10.1016/j.athoracsur.2005.11.040.CrossRefPubMed Leshnower BG, Gleason TG, O'Hara ML, Pochettino A, Woo YJ, Morris RJ, Gardner TJ, Acker MA: Safety and efficacy of left ventricular assist device support in postmyocardial infarction cardiogenic shock. Ann Thorac Surg. 2006, 81: 1365-70. 10.1016/j.athoracsur.2005.11.040.CrossRefPubMed
3.
go back to reference Samuels LE, Holmes EC, Garwood P, Ferdinand F: Initial experience with the Abiomed AB5000 ventricular assist device system. Ann Thorac Surg. 2005, 80: 309-12. 10.1016/j.athoracsur.2004.07.086.CrossRefPubMed Samuels LE, Holmes EC, Garwood P, Ferdinand F: Initial experience with the Abiomed AB5000 ventricular assist device system. Ann Thorac Surg. 2005, 80: 309-12. 10.1016/j.athoracsur.2004.07.086.CrossRefPubMed
4.
go back to reference Zhang L, Kapetanakis EI, Cooke RH, Sweet LC, Boyce SW: Bi-ventricular circulatory support with the Abiomed AB5000 system in a patient with idiopathic refractory ventricular fibrillation. Ann Thorac Surg. 2007, 83 (1): 298-300. 10.1016/j.athoracsur.2006.05.034.CrossRefPubMed Zhang L, Kapetanakis EI, Cooke RH, Sweet LC, Boyce SW: Bi-ventricular circulatory support with the Abiomed AB5000 system in a patient with idiopathic refractory ventricular fibrillation. Ann Thorac Surg. 2007, 83 (1): 298-300. 10.1016/j.athoracsur.2006.05.034.CrossRefPubMed
5.
go back to reference Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD: Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization. Circulation. 2007, 116: 2878-91. 10.1161/CIRCULATIONAHA.107.187256.CrossRefPubMed Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD: Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization. Circulation. 2007, 116: 2878-91. 10.1161/CIRCULATIONAHA.107.187256.CrossRefPubMed
6.
go back to reference Shellock FG: Reference Manual for Magnetic Resonance Safety, Implants, and Devices: 2008 Edition. 2008, Biomedical Research Publishing Group, Los Angeles, CA Shellock FG: Reference Manual for Magnetic Resonance Safety, Implants, and Devices: 2008 Edition. 2008, Biomedical Research Publishing Group, Los Angeles, CA
7.
go back to reference American Society for Testing and Materials (ASTM) International, Designation: F2503-05. Standard practice for marking medical devices and other items for safety in the magnetic resonance environment. 2005, ASTM International, West Conshohocken, Pennsylvania American Society for Testing and Materials (ASTM) International, Designation: F2503-05. Standard practice for marking medical devices and other items for safety in the magnetic resonance environment. 2005, ASTM International, West Conshohocken, Pennsylvania
8.
go back to reference Shellock FG, Fischer L, Fieno DS: Cardiac pacemakers and implantable cardioverter defibrillators: in vitro evaluation of MRI safety at 1.5-Tesla. Journal of Cardiovascular Magnetic Resonance. 2007, 9: 21-31. 10.1080/10976640600897237.CrossRefPubMed Shellock FG, Fischer L, Fieno DS: Cardiac pacemakers and implantable cardioverter defibrillators: in vitro evaluation of MRI safety at 1.5-Tesla. Journal of Cardiovascular Magnetic Resonance. 2007, 9: 21-31. 10.1080/10976640600897237.CrossRefPubMed
9.
go back to reference Shellock FG, Forder J: Drug eluting coronary stent: In vitro evaluation of magnet resonance safety at 3-Tesla. Journal of Cardiovascular Magnetic Resonance. 2005, 7: 415-419. 10.1081/JCMR-200053588.CrossRefPubMed Shellock FG, Forder J: Drug eluting coronary stent: In vitro evaluation of magnet resonance safety at 3-Tesla. Journal of Cardiovascular Magnetic Resonance. 2005, 7: 415-419. 10.1081/JCMR-200053588.CrossRefPubMed
10.
go back to reference Shellock FG, Gounis M, Wakhloo A: Detachable coil for cerebral aneurysms: In vitro evaluation of magnet field interactions, heating, and artifacts at 3-Tesla. American Journal of Neuroradiology. 2005, 26: 363-366.PubMed Shellock FG, Gounis M, Wakhloo A: Detachable coil for cerebral aneurysms: In vitro evaluation of magnet field interactions, heating, and artifacts at 3-Tesla. American Journal of Neuroradiology. 2005, 26: 363-366.PubMed
11.
go back to reference Shellock FG, Habibi R, Knebel J: Programmable CSF shunt valve: In vitro assessment of MRI safety at 3-Tesla. American Journal of Neuroradiology. 2006, 27: 661-665.PubMed Shellock FG, Habibi R, Knebel J: Programmable CSF shunt valve: In vitro assessment of MRI safety at 3-Tesla. American Journal of Neuroradiology. 2006, 27: 661-665.PubMed
12.
go back to reference Shellock FG, Begnaud J, Inman DM: VNS Therapy System: In vitro evaluation of MRI-related heating and function at 1.5- and 3-Tesla. Neuromodulation. 2006, 9: 204-213. 10.1111/j.1525-1403.2006.00061.x.CrossRefPubMed Shellock FG, Begnaud J, Inman DM: VNS Therapy System: In vitro evaluation of MRI-related heating and function at 1.5- and 3-Tesla. Neuromodulation. 2006, 9: 204-213. 10.1111/j.1525-1403.2006.00061.x.CrossRefPubMed
13.
go back to reference Shellock FG, Fieno DS, Thomson TJ, Talavage TM, Berman DS: Cardiac pacemaker: in vitro assessment of MR safety at 1.5-Tesla. American Heart Journal. 2006, 151: 436-443. 10.1016/j.ahj.2005.04.015.CrossRefPubMed Shellock FG, Fieno DS, Thomson TJ, Talavage TM, Berman DS: Cardiac pacemaker: in vitro assessment of MR safety at 1.5-Tesla. American Heart Journal. 2006, 151: 436-443. 10.1016/j.ahj.2005.04.015.CrossRefPubMed
14.
go back to reference American Society for Testing and Materials (ASTM) Designation: F 2052-02 Standard test method for measurement of magnetically induced displacement force on passive implants in the magnetic resonance environment. Annual Book of ASTM Standards, Section 13, Medical Devices and Services, Volume 13.01 Medical Devices; Emergency Medical Services. West Conshohocken, PA. 1576-1580. American Society for Testing and Materials (ASTM) Designation: F 2052-02 Standard test method for measurement of magnetically induced displacement force on passive implants in the magnetic resonance environment. Annual Book of ASTM Standards, Section 13, Medical Devices and Services, Volume 13.01 Medical Devices; Emergency Medical Services. West Conshohocken, PA. 1576-1580.
15.
go back to reference Shellock FG, Crivelli R, Venugopalan R: Programmable infusion pump and catheter: evaluation using 3-Tesla MRI. Neuromodulation. Shellock FG, Crivelli R, Venugopalan R: Programmable infusion pump and catheter: evaluation using 3-Tesla MRI. Neuromodulation.
16.
go back to reference Baker KB, Tkach JA, Phillips MD, Rezai AR: Variability in RF-induced heating of a deep brain stimulation implant across MR systems. J Magn Reson Imaging. 2006, 24: 1236-42. 10.1002/jmri.20769.CrossRefPubMed Baker KB, Tkach JA, Phillips MD, Rezai AR: Variability in RF-induced heating of a deep brain stimulation implant across MR systems. J Magn Reson Imaging. 2006, 24: 1236-42. 10.1002/jmri.20769.CrossRefPubMed
17.
go back to reference American Society for Testing and Materials International, Designation F 2182-02a Standard test method for measurement of radio frequency induced heating near passive implants during magnetic resonance imaging . ASTM International. American Society for Testing and Materials International, Designation F 2182-02a Standard test method for measurement of radio frequency induced heating near passive implants during magnetic resonance imaging . ASTM International.
Metadata
Title
Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla
Authors
Frank G Shellock
Samuel Valencerina
Publication date
01-12-2008
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2008
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-10-23

Other articles of this Issue 1/2008

Journal of Cardiovascular Magnetic Resonance 1/2008 Go to the issue