Published in: BMC Musculoskeletal Disorders 1/2007

Open Access 01-12-2007 | Research article

Custom-designed orthopedic implants evaluated using finite element analysis of patient-specific computed tomography data: femoral-component case study

Authors: Ola LA Harrysson, Yasser A Hosni, Jamal F Nayfeh

Published in: BMC Musculoskeletal Disorders | Issue 1/2007

Login to get access

Abstract

Background

Conventional knee and hip implant systems have been in use for many years with good success. However, the custom design of implant components based on patient-specific anatomy has been attempted to overcome existing shortcomings of current designs. The longevity of cementless implant components is highly dependent on the initial fit between the bone surface and the implant. The bone-implant interface design has historically been limited by the surgical tools and cutting guides available; and the cost of fabricating custom-designed implant components has been prohibitive.

Methods

This paper describes an approach where the custom design is based on a Computed Tomography scan of the patient's joint. The proposed design will customize both the articulating surface and the bone-implant interface to address the most common problems found with conventional knee-implant components. Finite Element Analysis is used to evaluate and compare the proposed design of a custom femoral component with a conventional design.

Results

The proposed design shows a more even stress distribution on the bone-implant interface surface, which will reduce the uneven bone remodeling that can lead to premature loosening.

Conclusion

The proposed custom femoral component design has the following advantages compared with a conventional femoral component. (i) Since the articulating surface closely mimics the shape of the distal femur, there is no need for resurfacing of the patella or gait change. (ii) Owing to the resulting stress distribution, bone remodeling is even and the risk of premature loosening might be reduced. (iii) Because the bone-implant interface can accommodate anatomical abnormalities at the distal femur, the need for surgical interventions and fitting of filler components is reduced. (iv) Given that the bone-implant interface is customized, about 40% less bone must be removed. The primary disadvantages are the time and cost required for the design and the possible need for a surgical robot to perform the bone resection. Some of these disadvantages may be eliminated by the use of rapid prototyping technologies, especially the use of Electron Beam Melting technology for quick and economical fabrication of custom implant components.
Appendix
Available only for authorised users
Literature
1.
go back to reference Keenan J, Chakrabarty G, Newman JH: Treatment of supracondylar femoral fracture above total knee replacement by custom made hinged prosthesis. Knee. 2000, 7: 165-170. 10.1016/S0968-0160(00)00041-7.CrossRefPubMed Keenan J, Chakrabarty G, Newman JH: Treatment of supracondylar femoral fracture above total knee replacement by custom made hinged prosthesis. Knee. 2000, 7: 165-170. 10.1016/S0968-0160(00)00041-7.CrossRefPubMed
2.
go back to reference Sathasivam S, Walker PS, Pinder IM, Cannon SR, Briggs TWR: Custom constrained condylar total knees using CAD-CAM. Knee. 1999, 6: 49-53. 10.1016/S0968-0160(98)00022-2.CrossRef Sathasivam S, Walker PS, Pinder IM, Cannon SR, Briggs TWR: Custom constrained condylar total knees using CAD-CAM. Knee. 1999, 6: 49-53. 10.1016/S0968-0160(98)00022-2.CrossRef
3.
go back to reference Harrysson OLA, Robertsson O, Nayfeh JF: Higher cumulative revision rate of knee arthroplasties in younger patients with osteoarthritis. Clin Orthop Relat Res. 2004, 421: 162-168. 10.1097/01.blo.0000127115.05754.ce.CrossRefPubMed Harrysson OLA, Robertsson O, Nayfeh JF: Higher cumulative revision rate of knee arthroplasties in younger patients with osteoarthritis. Clin Orthop Relat Res. 2004, 421: 162-168. 10.1097/01.blo.0000127115.05754.ce.CrossRefPubMed
4.
go back to reference Taylor J, Rorabeck CH, Bourne RB, Inman K: Total knee arthroplasty in patients under the age of 50: long-term follow-up. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000, Paper No. 185: Taylor J, Rorabeck CH, Bourne RB, Inman K: Total knee arthroplasty in patients under the age of 50: long-term follow-up. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000, Paper No. 185:
5.
go back to reference Krackow KA: The Technique of Total Knee Arthroplasty. 1990, St. Louis: The C. V. Mosby Company Krackow KA: The Technique of Total Knee Arthroplasty. 1990, St. Louis: The C. V. Mosby Company
6.
go back to reference Chaffin DB, Andersson GBJ, Martin BJ: Occupational Biomechanics. 1999, New York: John Wiley & Sons, Inc Chaffin DB, Andersson GBJ, Martin BJ: Occupational Biomechanics. 1999, New York: John Wiley & Sons, Inc
7.
go back to reference Taylor M, Tanner KE, Freeman M: Finite element analysis of the implanted proximal tibia: a relationship between the initial cancellous bone stresses and implant migration. J Biomech. 1998, 31 (4): 303-310. 10.1016/S0021-9290(98)00022-0.CrossRefPubMed Taylor M, Tanner KE, Freeman M: Finite element analysis of the implanted proximal tibia: a relationship between the initial cancellous bone stresses and implant migration. J Biomech. 1998, 31 (4): 303-310. 10.1016/S0021-9290(98)00022-0.CrossRefPubMed
8.
go back to reference Engh GA, Rao A, Louinici S, Ammeen D, Sychterz CJ: Tibial baseplate wear: a major source of debris with contemporary modular knee implants. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000 Engh GA, Rao A, Louinici S, Ammeen D, Sychterz CJ: Tibial baseplate wear: a major source of debris with contemporary modular knee implants. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000
9.
go back to reference Walker PS, Sathasivam S: The design of guide surfaces for fixed-bearing and mobile-bearing knee replacements. J Biomech. 1999, 32 (1): 27-34. 10.1016/S0021-9290(98)00140-7.CrossRefPubMed Walker PS, Sathasivam S: The design of guide surfaces for fixed-bearing and mobile-bearing knee replacements. J Biomech. 1999, 32 (1): 27-34. 10.1016/S0021-9290(98)00140-7.CrossRefPubMed
10.
go back to reference Banks SA, Otis JC, Backus SI, Furman GL, Laskin RS, Haas SB: Function of total knee replacement during activity of daily living. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000 Banks SA, Otis JC, Backus SI, Furman GL, Laskin RS, Haas SB: Function of total knee replacement during activity of daily living. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000
11.
go back to reference Morra EA, Postak PD, Greenwald AS: The influence of mobile bearing knee geometry on the wear of UHMWPE tibial inserts III: a finite element study. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000 Morra EA, Postak PD, Greenwald AS: The influence of mobile bearing knee geometry on the wear of UHMWPE tibial inserts III: a finite element study. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000
12.
go back to reference Fraser J: Knee and hip joint replacements. Longer lasting prostheses. Aust Fam Physician. 1999, 28 (11): 1109-1111. 1114–1115PubMed Fraser J: Knee and hip joint replacements. Longer lasting prostheses. Aust Fam Physician. 1999, 28 (11): 1109-1111. 1114–1115PubMed
13.
go back to reference Emoto G, Takeshi A, DeWeese FT, Whiteside LA: Effect of femoral component conformity on clinical results with unresurfaced patellae. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000 Emoto G, Takeshi A, DeWeese FT, Whiteside LA: Effect of femoral component conformity on clinical results with unresurfaced patellae. Program and Abstracts of the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons: 15–19 March 2000; Orlando, FL. 2000
14.
go back to reference Fadda M, Marcacci M, Toksvig-Larsen S, Wang T, Meneghello R: Improving accuracy of bone resections using robotic tool holder and a high speed milling cutting tool. J Med Eng Technol. 1998, 22 (6): 280-284.CrossRefPubMed Fadda M, Marcacci M, Toksvig-Larsen S, Wang T, Meneghello R: Improving accuracy of bone resections using robotic tool holder and a high speed milling cutting tool. J Med Eng Technol. 1998, 22 (6): 280-284.CrossRefPubMed
15.
go back to reference Pokrandt P, Both A, Hekmat A: Computer assisted surgery planning and robotics by orto MAQUET. Proceedings of the 4th International Workshop on Rapid Prototyping in Medicine & Computer-Assisted Surgery: 16–18 October 1997; Erlangen, Germany. 1997 Pokrandt P, Both A, Hekmat A: Computer assisted surgery planning and robotics by orto MAQUET. Proceedings of the 4th International Workshop on Rapid Prototyping in Medicine & Computer-Assisted Surgery: 16–18 October 1997; Erlangen, Germany. 1997
16.
go back to reference Toksvig-Larsen S, Ryd L: Surface characteristics following tibial preparation during total knee arthroplasty. J Arthroplasty. 1994, 9: 63-66. 10.1016/0883-5403(94)90138-4.CrossRefPubMed Toksvig-Larsen S, Ryd L: Surface characteristics following tibial preparation during total knee arthroplasty. J Arthroplasty. 1994, 9: 63-66. 10.1016/0883-5403(94)90138-4.CrossRefPubMed
17.
go back to reference Villa T, Migliavacca F, Gastaldi D, Colombo M, Pietrabissa R: Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulations. J Biomech. 2004, 37: 45-53. 10.1016/S0021-9290(03)00255-0.CrossRefPubMed Villa T, Migliavacca F, Gastaldi D, Colombo M, Pietrabissa R: Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulations. J Biomech. 2004, 37: 45-53. 10.1016/S0021-9290(03)00255-0.CrossRefPubMed
18.
go back to reference Haut Donahue T, Hull M, Rashid M, Jacobs C: A finite element model of the human knee joint for the study of tibio-femoral contact. J Biomech Eng. 2002, 124: 273-280. 10.1115/1.1470171.CrossRef Haut Donahue T, Hull M, Rashid M, Jacobs C: A finite element model of the human knee joint for the study of tibio-femoral contact. J Biomech Eng. 2002, 124: 273-280. 10.1115/1.1470171.CrossRef
19.
go back to reference Harrysson O, Cormier D: Direct fabrication of custom orthopedic implants using electron beam melting technology. Advanced Manufacturing Technology for Medical Applications. Edited by: Gibson I. 2006, Chichester, UK: John Wiley & Sons, Ltd, 193-208. Harrysson O, Cormier D: Direct fabrication of custom orthopedic implants using electron beam melting technology. Advanced Manufacturing Technology for Medical Applications. Edited by: Gibson I. 2006, Chichester, UK: John Wiley & Sons, Ltd, 193-208.
Metadata
Title
Custom-designed orthopedic implants evaluated using finite element analysis of patient-specific computed tomography data: femoral-component case study
Authors
Ola LA Harrysson
Yasser A Hosni
Jamal F Nayfeh
Publication date
01-12-2007
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2007
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-8-91

Other articles of this Issue 1/2007

BMC Musculoskeletal Disorders 1/2007 Go to the issue