Morphometric evaluation of subaxial cervical spine using multi-detector computerized tomography (MD-CT) scan: the consideration for cervical pedicle screws fixation

Authors: Pongsthorn Chanplakorn, Chaiwat Kraiwattanapong, Kitti Aroonjarattham, Pittavat Leelapattana, Gun Keorochana, Suphaneewan Jaovisidha, Wiwat Wajanavisit

Published in: BMC Musculoskeletal Disorders | Issue 1/2014

Login to get access

Abstract

Background

Cervical pedicle screw (CPS) insertion is a technically demanding procedure. The quantitative understanding of cervical pedicle morphology, especially the narrowest part of cervical pedicle or isthmus, would minimize the risk of catastrophic damage to surrounding neurovascular structures and improve surgical outcome. The aim of this study was to investigate morphology and quantify cortical thickness of the cervical isthmus by using Multi-detector Computerized Tomography (MD-CT) scan.

Methods

The cervical CT scans were performed in 74 patients (37 males and 37 females) with 1-mm slice thickness and then retro-reconstructed into sagittal and coronal planes to measure various cervical parameters as follows: outer pedicle width (OPW), inner pedicle width (IPW), outer pedicle height (OPH), inner pedicle height (IPH), pedicle cortical thickness, pedicle sagittal angle (PSA), and pedicle transverse angle (PTA).

Results

Total numbers of 740 pedicles were measured in this present study. The mean OPW and IPW significantly increased from C3 to C7 while the mean OPH and IPH of those showed non-significant difference between any measured levels. The medial-lateral cortical thickness was significantly smaller than the superior-inferior one. PTA in the upper cervical spine was significantly wider than the lower ones. The PSA changed from upward inclination at upper cervical spine to the downward inclination at lower cervical spine.

Conclusions

This study has demonstrated that cervical vertebra has relatively small and narrow inner pedicle canal with thick outer pedicle cortex and also shows a variable in pedicle width and inconsistent transverse angle. To enhance the safety of CPS insertion, the entry point and trajectories should be determined individually by using preoperative MD-CT scan and the inner pedicle width should be a key parameter to determine the screw dimensions.
Appendix
Available only for authorised users
Literature
1.
go back to reference An HS, Gordin R, Renner K: Anatomic considerations for plate-screw fixation of the cervical spine. Spine. 1991, 16 (Suppl 10): 548-551.CrossRef An HS, Gordin R, Renner K: Anatomic considerations for plate-screw fixation of the cervical spine. Spine. 1991, 16 (Suppl 10): 548-551.CrossRef
2.
go back to reference Kotani Y, Cunningham BW, Albumi K, McAfee PC: Biomechanic analysis of cervical stabilization system: an assessment of tranpedicular screw fixation in the cervical spine. Spine. 1994, 19: 2529-2539. 10.1097/00007632-199411001-00007.CrossRefPubMed Kotani Y, Cunningham BW, Albumi K, McAfee PC: Biomechanic analysis of cervical stabilization system: an assessment of tranpedicular screw fixation in the cervical spine. Spine. 1994, 19: 2529-2539. 10.1097/00007632-199411001-00007.CrossRefPubMed
3.
go back to reference Jones EL, Heller JG, Silicox DH, Hutton WC: Cervical pedicle screws versus lateral mass screws. Anatomic fesibility and biomechanical comparison. Spine. 1997, 22: 977-982. 10.1097/00007632-199705010-00009.CrossRefPubMed Jones EL, Heller JG, Silicox DH, Hutton WC: Cervical pedicle screws versus lateral mass screws. Anatomic fesibility and biomechanical comparison. Spine. 1997, 22: 977-982. 10.1097/00007632-199705010-00009.CrossRefPubMed
4.
go back to reference Johnston TL, Karaikovic EE, Lautenschlager EP, Marcu D: Cervical pedicle screws vs. lateral mass screws: uniplanar fatique analysis and residual pullout strengths. Spine J. 2006, 6: 667-672. 10.1016/j.spinee.2006.03.019.CrossRefPubMed Johnston TL, Karaikovic EE, Lautenschlager EP, Marcu D: Cervical pedicle screws vs. lateral mass screws: uniplanar fatique analysis and residual pullout strengths. Spine J. 2006, 6: 667-672. 10.1016/j.spinee.2006.03.019.CrossRefPubMed
5.
go back to reference Schmidt R, Koller H, Wilke HJ, Brade J, Zenner J, Meiero O, Ferraris L, Mayer M: The impact of cervical screw for primary stability in multilevel posterior cervical stabilizations. Spine. 2010, 35: E1167-E1171. 10.1097/BRS.0b013e3181e6bc59.CrossRefPubMed Schmidt R, Koller H, Wilke HJ, Brade J, Zenner J, Meiero O, Ferraris L, Mayer M: The impact of cervical screw for primary stability in multilevel posterior cervical stabilizations. Spine. 2010, 35: E1167-E1171. 10.1097/BRS.0b013e3181e6bc59.CrossRefPubMed
6.
go back to reference Albumi K, Itoh H, Taneichi H, Kaneda K: Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. J Spinal Disord. 1994, 7: 19-28. 10.1097/00002517-199407010-00003.CrossRef Albumi K, Itoh H, Taneichi H, Kaneda K: Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. J Spinal Disord. 1994, 7: 19-28. 10.1097/00002517-199407010-00003.CrossRef
7.
go back to reference Albumi K, Takada T, Shono Y, Keneda K, Fujiya M: Posterior ocipitalcervical reconstruction using cervical pedicle screws and plate-rod system. Spine. 1999, 24: 1425-1434. 10.1097/00007632-199907150-00007.CrossRef Albumi K, Takada T, Shono Y, Keneda K, Fujiya M: Posterior ocipitalcervical reconstruction using cervical pedicle screws and plate-rod system. Spine. 1999, 24: 1425-1434. 10.1097/00007632-199907150-00007.CrossRef
8.
go back to reference Albumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K: Complication of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine. 2000, 25: 962-969. 10.1097/00007632-200004150-00011.CrossRef Albumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K: Complication of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine. 2000, 25: 962-969. 10.1097/00007632-200004150-00011.CrossRef
9.
go back to reference Ishikawa Y, Kanemura T, Yoshida G, Matsumoto A, Ito Z, Tauchi R, Ohno S, Nishimura Y: Intraoperative, full-rotation, three-dimensional image (O-arm)- based navigation system for cervical pedicle screw insertion. J Neurosurg Spine. 2011, 15: 472-478. 10.3171/2011.6.SPINE10809.CrossRefPubMed Ishikawa Y, Kanemura T, Yoshida G, Matsumoto A, Ito Z, Tauchi R, Ohno S, Nishimura Y: Intraoperative, full-rotation, three-dimensional image (O-arm)- based navigation system for cervical pedicle screw insertion. J Neurosurg Spine. 2011, 15: 472-478. 10.3171/2011.6.SPINE10809.CrossRefPubMed
10.
go back to reference Jarvers JS, Katscher S, Franck A, Glasmacher S, Schmidt C, Blattert T, Josten C: 3D- based navigation in posterior stabilizations of the cervical and thoracic spine: problems and benefits. Result of 451 screws. Eur J Trauma Emerg Surg. 2011, 37: 109-119. 10.1007/s00068-011-0098-1.CrossRefPubMed Jarvers JS, Katscher S, Franck A, Glasmacher S, Schmidt C, Blattert T, Josten C: 3D- based navigation in posterior stabilizations of the cervical and thoracic spine: problems and benefits. Result of 451 screws. Eur J Trauma Emerg Surg. 2011, 37: 109-119. 10.1007/s00068-011-0098-1.CrossRefPubMed
11.
go back to reference Rajasekaran S, Vidyadhara S, Shetty AP: Iso-C3D fluoroscopy-based navigation in direct pedicle screw fixation of hangman fracture: a case report. J Spinal Disord Tech. 2007, 20: 616-619. 10.1097/BSD.0b013e318074f978.CrossRefPubMed Rajasekaran S, Vidyadhara S, Shetty AP: Iso-C3D fluoroscopy-based navigation in direct pedicle screw fixation of hangman fracture: a case report. J Spinal Disord Tech. 2007, 20: 616-619. 10.1097/BSD.0b013e318074f978.CrossRefPubMed
12.
go back to reference Ito Y, Sugimoto Y, Tomioka M, Hasegawa Y, Nakago K, Yagata Y: Clinical accuracy of 3D fluoroscopy-assisted cervical pedicle screw insertion. J Neurosurg Spine. 2008, 9: 450-453. 10.3171/SPI.2008.9.11.450.CrossRefPubMed Ito Y, Sugimoto Y, Tomioka M, Hasegawa Y, Nakago K, Yagata Y: Clinical accuracy of 3D fluoroscopy-assisted cervical pedicle screw insertion. J Neurosurg Spine. 2008, 9: 450-453. 10.3171/SPI.2008.9.11.450.CrossRefPubMed
13.
go back to reference Ugur HC, Attar A, Uz A, Tekdemir I, Egemen N, Caglar S, Genc Y: Surgical anatomic evaluation of the cervical pedicle and adjacent neural structures. Neurosurg. 2000, 47: 1162-1168. 10.1097/00006123-200011000-00029.CrossRef Ugur HC, Attar A, Uz A, Tekdemir I, Egemen N, Caglar S, Genc Y: Surgical anatomic evaluation of the cervical pedicle and adjacent neural structures. Neurosurg. 2000, 47: 1162-1168. 10.1097/00006123-200011000-00029.CrossRef
14.
go back to reference Neo M, Sakamoto T, Fujibayashi S, Nakamura T: The clinical risk of vertebral artery injury from cervical pedicle screws inserted in degenerative vertebrae. Spine. 2005, 30: 2800-2805. 10.1097/01.brs.0000192297.07709.5d.CrossRefPubMed Neo M, Sakamoto T, Fujibayashi S, Nakamura T: The clinical risk of vertebral artery injury from cervical pedicle screws inserted in degenerative vertebrae. Spine. 2005, 30: 2800-2805. 10.1097/01.brs.0000192297.07709.5d.CrossRefPubMed
15.
go back to reference Reinhold M, Bach C, Audiģe L, Bale R, Attal R, Blauth M, Magerl F: Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literature. Eur Spine J. 2008, 17: 564-575. 10.1007/s00586-008-0584-2.CrossRefPubMedPubMedCentral Reinhold M, Bach C, Audiģe L, Bale R, Attal R, Blauth M, Magerl F: Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literature. Eur Spine J. 2008, 17: 564-575. 10.1007/s00586-008-0584-2.CrossRefPubMedPubMedCentral
16.
go back to reference Miller RM, Ebraheim NA, Xu R, Yeasting RA: Anatomic consideration of transpedicular screw placement in the cervical spine. An analysis of two approaches. Spine. 1996, 21: 2317-2322. 10.1097/00007632-199610150-00003.CrossRefPubMed Miller RM, Ebraheim NA, Xu R, Yeasting RA: Anatomic consideration of transpedicular screw placement in the cervical spine. An analysis of two approaches. Spine. 1996, 21: 2317-2322. 10.1097/00007632-199610150-00003.CrossRefPubMed
17.
go back to reference Lee DH, Lee SW, Kang SJ, Hwang CJ, Kim NH, Bae JY, Kim YT, Lee CS, Daniel RK: Optimal entry points and trajectories for cervical pedicle screw placement in to subaxial cervical vertebrae. Eur Spine J. 2011, 20: 905-911. 10.1007/s00586-010-1655-8.CrossRefPubMedPubMedCentral Lee DH, Lee SW, Kang SJ, Hwang CJ, Kim NH, Bae JY, Kim YT, Lee CS, Daniel RK: Optimal entry points and trajectories for cervical pedicle screw placement in to subaxial cervical vertebrae. Eur Spine J. 2011, 20: 905-911. 10.1007/s00586-010-1655-8.CrossRefPubMedPubMedCentral
18.
go back to reference Panjabi MM, Duranceau J, Goel VK, Oxland T, Tanaka K: Cervical human vertebrae: quantitative three-dimensional anatomy of the middle and lower regions. Spine. 1991, 16: 861-869. 10.1097/00007632-199108000-00001.CrossRefPubMed Panjabi MM, Duranceau J, Goel VK, Oxland T, Tanaka K: Cervical human vertebrae: quantitative three-dimensional anatomy of the middle and lower regions. Spine. 1991, 16: 861-869. 10.1097/00007632-199108000-00001.CrossRefPubMed
19.
go back to reference Ebraheim NA, Xu R, Knight T, Yeasting RA: Morphometric evaluation of lower cervical pedicle and its projection. Spine. 1997, 22: 1-6. 10.1097/00007632-199701010-00001.CrossRefPubMed Ebraheim NA, Xu R, Knight T, Yeasting RA: Morphometric evaluation of lower cervical pedicle and its projection. Spine. 1997, 22: 1-6. 10.1097/00007632-199701010-00001.CrossRefPubMed
20.
go back to reference Karaikovic EE, Daubs MD, Madsen RW, Gaines RW: Morphologic characteristics of human cervical pedicles. Spine. 1997, 22: 493-500. 10.1097/00007632-199703010-00005.CrossRefPubMed Karaikovic EE, Daubs MD, Madsen RW, Gaines RW: Morphologic characteristics of human cervical pedicles. Spine. 1997, 22: 493-500. 10.1097/00007632-199703010-00005.CrossRefPubMed
21.
go back to reference Bozbuga M, Ozturk A, Ari Z, Sahinoglu K, Bayraktar B, Cecen A: A morphometric evaluation of subaxial cervical vertebrae for surgical application of transpedicular screw fixation. Spine. 2004, 29: 1876-1880. 10.1097/01.brs.0000137065.62516.01.CrossRefPubMed Bozbuga M, Ozturk A, Ari Z, Sahinoglu K, Bayraktar B, Cecen A: A morphometric evaluation of subaxial cervical vertebrae for surgical application of transpedicular screw fixation. Spine. 2004, 29: 1876-1880. 10.1097/01.brs.0000137065.62516.01.CrossRefPubMed
22.
go back to reference Yusof MI, Ming LK, Abdullah MS, Yusof MI: Computerized tomographic measurement of the cervical pedicles diameter in a Malaysian population and the feasibility for transpedicular fixation. Spine. 2006, 31: E221-E224. 10.1097/01.brs.0000210263.87578.65.CrossRefPubMed Yusof MI, Ming LK, Abdullah MS, Yusof MI: Computerized tomographic measurement of the cervical pedicles diameter in a Malaysian population and the feasibility for transpedicular fixation. Spine. 2006, 31: E221-E224. 10.1097/01.brs.0000210263.87578.65.CrossRefPubMed
23.
go back to reference Onibokun A, Khoo LLT, Bistazzoni S, Chen NF, Sassi M: Anatomical considerations for cervical pedicle screw insertion: the use of multiplanar computerized tomography measurements in 122 consecutive clinical cases. Spine J. 2009, 9: 729-734. 10.1016/j.spinee.2009.04.021.CrossRefPubMed Onibokun A, Khoo LLT, Bistazzoni S, Chen NF, Sassi M: Anatomical considerations for cervical pedicle screw insertion: the use of multiplanar computerized tomography measurements in 122 consecutive clinical cases. Spine J. 2009, 9: 729-734. 10.1016/j.spinee.2009.04.021.CrossRefPubMed
24.
go back to reference Panjabi MM, Shin EK, Chen NC, Wang JL: Internal morphology of human cervical pedicles. Spine. 2000, 25: 1197-1205. 10.1097/00007632-200005150-00002.CrossRefPubMed Panjabi MM, Shin EK, Chen NC, Wang JL: Internal morphology of human cervical pedicles. Spine. 2000, 25: 1197-1205. 10.1097/00007632-200005150-00002.CrossRefPubMed
25.
go back to reference Reinhold M, Magerl F, Rieger M, Blauth M: Cervical pedicle screw placement: feasibility and accuracy of two new insertion techniques based on morphometric data. Eur Spine J. 2007, 16: 47-56. 10.1007/s00586-006-0104-1.CrossRefPubMed Reinhold M, Magerl F, Rieger M, Blauth M: Cervical pedicle screw placement: feasibility and accuracy of two new insertion techniques based on morphometric data. Eur Spine J. 2007, 16: 47-56. 10.1007/s00586-006-0104-1.CrossRefPubMed
26.
go back to reference Chazono M, Tanaka T, Kumagae Y, Sai T, Marumo K: Ethnic differences in pedicle and bony spinal canal dimensions calculated from computed tomography of the cervical spine: a review of the English-language literature. Eur Spine J. 2012, 21: 1451-1458. 10.1007/s00586-012-2295-y.CrossRefPubMedPubMedCentral Chazono M, Tanaka T, Kumagae Y, Sai T, Marumo K: Ethnic differences in pedicle and bony spinal canal dimensions calculated from computed tomography of the cervical spine: a review of the English-language literature. Eur Spine J. 2012, 21: 1451-1458. 10.1007/s00586-012-2295-y.CrossRefPubMedPubMedCentral
27.
go back to reference Ruofu Z, Huilin Y, Xiaoyun H, Xishun H, Tiansi T, Liang C, Xigong L: CT evaluation of cervical pedicle in a Chinese population for surgical application of transpedicular screw placement. Surg Radiol Anat. 2008, 30: 389-396. 10.1007/s00276-008-0339-0.CrossRefPubMed Ruofu Z, Huilin Y, Xiaoyun H, Xishun H, Tiansi T, Liang C, Xigong L: CT evaluation of cervical pedicle in a Chinese population for surgical application of transpedicular screw placement. Surg Radiol Anat. 2008, 30: 389-396. 10.1007/s00276-008-0339-0.CrossRefPubMed
28.
go back to reference Gupta R, Kapoor K, Sharma A, Kochhar S, Garg R: Morphometry of typical cervical vertebra on dry bones and CT scan and its implications in the transpedicular screw placement surgery. Surg Radio Anat. 2013, 35: 181-189. 10.1007/s00276-012-1013-0.CrossRef Gupta R, Kapoor K, Sharma A, Kochhar S, Garg R: Morphometry of typical cervical vertebra on dry bones and CT scan and its implications in the transpedicular screw placement surgery. Surg Radio Anat. 2013, 35: 181-189. 10.1007/s00276-012-1013-0.CrossRef
29.
go back to reference Hirano T, Hasegawa K, Takahashi HE, Uchiyama S, Hara T, Washio T, Sugiura T, Yokaichiya M, Ikeda M: Structural characteristics of the pedicle and its role in screw stability. Spine. 1997, 22: 2504-2510. 10.1097/00007632-199711010-00007.CrossRefPubMed Hirano T, Hasegawa K, Takahashi HE, Uchiyama S, Hara T, Washio T, Sugiura T, Yokaichiya M, Ikeda M: Structural characteristics of the pedicle and its role in screw stability. Spine. 1997, 22: 2504-2510. 10.1097/00007632-199711010-00007.CrossRefPubMed
30.
go back to reference Rao RD, Marawar SV, Stemper BD, Yoganandan N, Shender BS: Computerized tomographic morphometric analysis of subaxial cervical spine pedicles in young asymptomatic volunteers. J Bone Joint Surg Am. 2008, 90: 1914-1921. 10.2106/JBJS.G.01166.CrossRefPubMed Rao RD, Marawar SV, Stemper BD, Yoganandan N, Shender BS: Computerized tomographic morphometric analysis of subaxial cervical spine pedicles in young asymptomatic volunteers. J Bone Joint Surg Am. 2008, 90: 1914-1921. 10.2106/JBJS.G.01166.CrossRefPubMed
31.
go back to reference Karaikovic EE, Kunakorsawat S, Dauds MD, Madsen TW, Gaines RW: Surgical anatomy of the cervical pedicles: landmarks for posterior cervical pedicle entrance location. J Spinal Disord. 2000, 13: 63-72. 10.1097/00002517-200002000-00013.CrossRefPubMed Karaikovic EE, Kunakorsawat S, Dauds MD, Madsen TW, Gaines RW: Surgical anatomy of the cervical pedicles: landmarks for posterior cervical pedicle entrance location. J Spinal Disord. 2000, 13: 63-72. 10.1097/00002517-200002000-00013.CrossRefPubMed
Metadata
Title
Morphometric evaluation of subaxial cervical spine using multi-detector computerized tomography (MD-CT) scan: the consideration for cervical pedicle screws fixation
Authors
Pongsthorn Chanplakorn
Chaiwat Kraiwattanapong
Kitti Aroonjarattham
Pittavat Leelapattana
Gun Keorochana
Suphaneewan Jaovisidha
Wiwat Wajanavisit
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2014
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-15-125

Other articles of this Issue 1/2014

BMC Musculoskeletal Disorders 1/2014 Go to the issue