Abstract

Background

Intramedullary nailing is a standard surgical procedure for fixation of proximal femoral fractures, but is associated with considerable radiation exposure for controlling the implant placement, due to the percutaneous insertion technique.
The aim of this study was the evaluation of potential benefits of 2D-fluoroscopic based navigation focused on the reduction of radiation exposure, a decrease of procedure time, as well as an increase of accuracy for Gamma3 nail insertions.

Methods

Twenty randomized Gamma3 nail insertions were performed in non-fractured synthetic femora according to the manufactures operation guidelines (group I) or with use of a 2D-fluoroscopic based navigation system (group II). Time of different steps of the procedure and the radiation exposure were measured, as well as the accuracy evaluated in postoperative CT scans.

Results and discussion

All Gamma3 nails were placed without any technical problems. Independent of the used procedure, the overall operating time (group I: 584 ± 99.2 sec; group II: 662 ± 64.9 sec; p=0.06) and accuracy of the final nail-positions were equivalent, but the radiation exposure was significantly reduced (92% reduction in fluoroscopic images and 91% reduction in fluoroscopic time, p< 0.01), using the 2D fluoroscopic based navigation procedure.

Conclusions

2D-fluoroscopic based navigation for Gamma3 nail insertion facilitates a relevant reduction of radiation exposure with equivalent accuracy of the final implant position and no prolonged operating time. This promising procedure modification is independent of different cephalomedullary implant manufacturers and specific implant designs, but needs to be evaluated in further clinical settings.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bundesauswertung zum Verfahrensjahr: Hüftgelenksnahe Frakturen. 2009, Gottingen: AQUA – Institut für angewandte Qualitätsförderung und Forschung im Gesundheitswesen GmbH 2010 Bundesauswertung zum Verfahrensjahr: Hüftgelenksnahe Frakturen. 2009, Gottingen: AQUA – Institut für angewandte Qualitätsförderung und Forschung im Gesundheitswesen GmbH 2010
2.
go back to reference Berrington de Gonzalez A, Darby S: Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet. 2004, 363 (9406): 345-351. 10.1016/S0140-6736(04)15433-0.CrossRefPubMed Berrington de Gonzalez A, Darby S: Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet. 2004, 363 (9406): 345-351. 10.1016/S0140-6736(04)15433-0.CrossRefPubMed
3.
go back to reference Chong KW, Wong MK, Rikhraj IS, Howe TS: The use of computer navigation in performing minimally invasive surgery for intertrochanteric hip fractures–The experience in Singapore. Injury. 2006, 37 (8): 755-762. 10.1016/j.injury.2006.03.002.CrossRefPubMed Chong KW, Wong MK, Rikhraj IS, Howe TS: The use of computer navigation in performing minimally invasive surgery for intertrochanteric hip fractures–The experience in Singapore. Injury. 2006, 37 (8): 755-762. 10.1016/j.injury.2006.03.002.CrossRefPubMed
4.
go back to reference Hafez MA, Smith RM, Matthews SJ, Kalap G, Sherman KP: Radiation exposure to the hands of orthopaedic surgeons: are we underestimating the risk?. Arch Orthop Trauma Surg. 2005, 125 (5): 330-335. 10.1007/s00402-005-0807-5.CrossRefPubMed Hafez MA, Smith RM, Matthews SJ, Kalap G, Sherman KP: Radiation exposure to the hands of orthopaedic surgeons: are we underestimating the risk?. Arch Orthop Trauma Surg. 2005, 125 (5): 330-335. 10.1007/s00402-005-0807-5.CrossRefPubMed
5.
go back to reference Mehlman CT, DiPasquale TG: Radiation exposure to the orthopaedic surgical team during fluoroscopy: "how far away is far enough?". J Orthop Trauma. 1997, 11 (6): 392-398. 10.1097/00005131-199708000-00002.CrossRefPubMed Mehlman CT, DiPasquale TG: Radiation exposure to the orthopaedic surgical team during fluoroscopy: "how far away is far enough?". J Orthop Trauma. 1997, 11 (6): 392-398. 10.1097/00005131-199708000-00002.CrossRefPubMed
6.
go back to reference Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77 (7): 1058-1064.PubMed Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77 (7): 1058-1064.PubMed
7.
go back to reference Gras F, Marintschev I, Klos K, Muckley T, Hofmann GO, Kahler DM: Screw placement for acetabular fractures: which navigation modality (2-dimensional vs. 3-dimensional) should be used? An experimental study. J Orthop Trauma. 2012, 26 (8): 466-473. 10.1097/BOT.0b013e318234d443.CrossRefPubMed Gras F, Marintschev I, Klos K, Muckley T, Hofmann GO, Kahler DM: Screw placement for acetabular fractures: which navigation modality (2-dimensional vs. 3-dimensional) should be used? An experimental study. J Orthop Trauma. 2012, 26 (8): 466-473. 10.1097/BOT.0b013e318234d443.CrossRefPubMed
8.
go back to reference Gras F, Marintschev I, Wilharm A, Lindner R, Klos K, Muckley T, Hofmann GO: Sustentaculum tali screw placement for calcaneus fractures–different navigation procedures compared to the conventional technique. Z Orthop Unfall. 2010, 148 (3): 309-318. 10.1055/s-0029-1240973.CrossRefPubMed Gras F, Marintschev I, Wilharm A, Lindner R, Klos K, Muckley T, Hofmann GO: Sustentaculum tali screw placement for calcaneus fractures–different navigation procedures compared to the conventional technique. Z Orthop Unfall. 2010, 148 (3): 309-318. 10.1055/s-0029-1240973.CrossRefPubMed
9.
go back to reference Muller MC, Belei P, de la Fuente M, Strake M, Kabir K, Weber O, Burger C, Radermacher K, Wirtz DC: Evaluation of a 2D fluoroscopy-based navigation system for insertion of femoral neck screws: An experimental study. Unfallchirurg. 2012, 115 (9): 802-810. 10.1007/s00113-010-1915-4. SepCrossRefPubMed Muller MC, Belei P, de la Fuente M, Strake M, Kabir K, Weber O, Burger C, Radermacher K, Wirtz DC: Evaluation of a 2D fluoroscopy-based navigation system for insertion of femoral neck screws: An experimental study. Unfallchirurg. 2012, 115 (9): 802-810. 10.1007/s00113-010-1915-4. SepCrossRefPubMed
10.
go back to reference Muller MC, Belei P, de la Fuente M, Strake M, Weber O, Burger C, Radermacher K, Wirtz DC: Evaluation of a 2D Fluoroscopy-Based Navigation System for Insertion of the Dynamic Hip Screw (DHS): an Experimental Study. Rofo. 2011, 183 (6): 536-542. 10.1055/s-0031-1273329.CrossRefPubMed Muller MC, Belei P, de la Fuente M, Strake M, Weber O, Burger C, Radermacher K, Wirtz DC: Evaluation of a 2D Fluoroscopy-Based Navigation System for Insertion of the Dynamic Hip Screw (DHS): an Experimental Study. Rofo. 2011, 183 (6): 536-542. 10.1055/s-0031-1273329.CrossRefPubMed
11.
go back to reference Miller ME, Davis ML, MacClean CR, Davis JG, Smith BL, Humphries JR: Radiation exposure and associated risks to operating-room personnel during use of fluoroscopic guidance for selected orthopaedic surgical procedures. J Bone Joint Surg Am. 1983, 65 (1): 1-4.PubMed Miller ME, Davis ML, MacClean CR, Davis JG, Smith BL, Humphries JR: Radiation exposure and associated risks to operating-room personnel during use of fluoroscopic guidance for selected orthopaedic surgical procedures. J Bone Joint Surg Am. 1983, 65 (1): 1-4.PubMed
12.
go back to reference Kwong LM, Johanson PH, Zinar DM, Lenihan MR, Herman MW: Shielding of the patient's gonads during intramedullary interlocking femoral nailing. J Bone Joint Surg Am. 1990, 72 (10): 1523-1526.PubMed Kwong LM, Johanson PH, Zinar DM, Lenihan MR, Herman MW: Shielding of the patient's gonads during intramedullary interlocking femoral nailing. J Bone Joint Surg Am. 1990, 72 (10): 1523-1526.PubMed
13.
go back to reference Beldame J, Boisrenoult P, Beaufils P: Pin track induced fractures around computer-assisted TKA. Orthop Traumatol Surg Res. 2010, 96 (3): 249-255. 10.1016/j.otsr.2009.12.005. MayCrossRefPubMed Beldame J, Boisrenoult P, Beaufils P: Pin track induced fractures around computer-assisted TKA. Orthop Traumatol Surg Res. 2010, 96 (3): 249-255. 10.1016/j.otsr.2009.12.005. MayCrossRefPubMed
14.
go back to reference Kendoff D, Hufner T, Citak M, Maier C, Wesemeier F, Pearle A, Krettek C: A new parallel drill guide for navigating femoral neck screw placement. Development and evaluation. Unfallchirurg. 2006, 109 (10): 875-880. 10.1007/s00113-006-1142-1.CrossRefPubMed Kendoff D, Hufner T, Citak M, Maier C, Wesemeier F, Pearle A, Krettek C: A new parallel drill guide for navigating femoral neck screw placement. Development and evaluation. Unfallchirurg. 2006, 109 (10): 875-880. 10.1007/s00113-006-1142-1.CrossRefPubMed
15.
go back to reference Hamelinck HK, Haagmans M, Snoeren MM, Biert J, van Vugt AB, Frolke JP: Safety of computer-assisted surgery for cannulated hip screws. Clin Orthop Relat Res. 2007, 455: 241-245.CrossRefPubMed Hamelinck HK, Haagmans M, Snoeren MM, Biert J, van Vugt AB, Frolke JP: Safety of computer-assisted surgery for cannulated hip screws. Clin Orthop Relat Res. 2007, 455: 241-245.CrossRefPubMed
16.
go back to reference Liebergall M, Ben-David D, Weil Y, Peyser A, Mosheiff R: Computerized navigation for the internal fixation of femoral neck fractures. J Bone Joint Surg Am. 2006, 88 (8): 1748-1754. 10.2106/JBJS.E.00137.CrossRefPubMed Liebergall M, Ben-David D, Weil Y, Peyser A, Mosheiff R: Computerized navigation for the internal fixation of femoral neck fractures. J Bone Joint Surg Am. 2006, 88 (8): 1748-1754. 10.2106/JBJS.E.00137.CrossRefPubMed
17.
go back to reference Beckmann J, Tingart M, Perlick L, Luring C, Grifka J, Anders S: Navigated drilling for femoral head necrosis. Experimental and clinical results. Orthopade. 2007, 36 (5): 458-465. 10.1007/s00132-007-1086-3.CrossRefPubMed Beckmann J, Tingart M, Perlick L, Luring C, Grifka J, Anders S: Navigated drilling for femoral head necrosis. Experimental and clinical results. Orthopade. 2007, 36 (5): 458-465. 10.1007/s00132-007-1086-3.CrossRefPubMed
Metadata
Title
2D-fluoroscopic based navigation for Gamma 3 nail insertion versus conventional procedure- a feasibility study
Authors
A Wilharm
I Marintschev
G O Hofmann
F Gras
Publication date
01-12-2013
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2013
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-14-74

Other articles of this Issue 1/2013

BMC Musculoskeletal Disorders 1/2013 Go to the issue