Assessment of a novel biomechanical fracture model for distal radius fractures

Authors: Sebastian Felix Baumbach, Enrico Dall’Ara, Patrick Weninger, Anna Antoni, Hannes Traxler, Martin Dörr, Philippe K Zysset

Published in: BMC Musculoskeletal Disorders | Issue 1/2012

Login to get access

Abstract

Background

Distal radius fractures (DRF) are one of the most common fractures and often need surgical treatment, which has been validated through biomechanical tests. Currently a number of different fracture models are used, none of which resemble the in vivo fracture location. The aim of the study was to develop a new standardized fracture model for DRF (AO-23.A3) and compare its biomechanical behavior to the current gold standard.

Methods

Variable angle locking volar plates (ADAPTIVE, Medartis) were mounted on 10 pairs of fresh-frozen radii. The osteotomy location was alternated within each pair (New: 10 mm wedge 8 mm / 12 mm proximal to the dorsal / volar apex of the articular surface; Gold standard: 10 mm wedge 20 mm proximal to the articular surface). Each specimen was tested in cyclic axial compression (increasing load by 100 N per cycle) until failure or −3 mm displacement. Parameters assessed were stiffness, displacement and dissipated work calculated for each cycle and ultimate load. Significance was tested using a linear mixed model and Wald test as well as t-tests.

Results

7 female and 3 male pairs of radii aged 74 ± 9 years were tested. In most cases (7/10), the two groups showed similar mechanical behavior at low loads with increasing differences at increasing loads. Overall the novel fracture model showed a significant different biomechanical behavior than the gold standard model (p < 0,001). The average final loads resisted were significantly lower in the novel model (860 N ± 232 N vs. 1250 N ± 341 N; p = 0.001).

Conclusion

The novel biomechanical fracture model for DRF more closely mimics the in vivo fracture site and shows a significantly different biomechanical behavior with increasing loads when compared to the current gold standard.
Appendix
Available only for authorised users
Literature
1.
go back to reference Henry MH: Distal radius fractures: current concepts. J Hand Surg Am. 2008, 33: 1215-1227. 10.1016/j.jhsa.2008.07.013.CrossRefPubMed Henry MH: Distal radius fractures: current concepts. J Hand Surg Am. 2008, 33: 1215-1227. 10.1016/j.jhsa.2008.07.013.CrossRefPubMed
2.
go back to reference Orbay JL, Touhami A: Current concepts in volar fixed-angle fixation of unstable distal radius fractures. Clin Orthop Relat Res. 2006, 445: 58-67.PubMed Orbay JL, Touhami A: Current concepts in volar fixed-angle fixation of unstable distal radius fractures. Clin Orthop Relat Res. 2006, 445: 58-67.PubMed
3.
go back to reference Lozano-Calderón SA, Souer S, Mudgal C, Jupiter JB, Ring D: Wrist mobilization following volar plate fixation of fractures of the distal part of the radius. J Bone Joint Surg Am. 2008, 90: 1297-1304. 10.2106/JBJS.G.01368.CrossRefPubMed Lozano-Calderón SA, Souer S, Mudgal C, Jupiter JB, Ring D: Wrist mobilization following volar plate fixation of fractures of the distal part of the radius. J Bone Joint Surg Am. 2008, 90: 1297-1304. 10.2106/JBJS.G.01368.CrossRefPubMed
4.
go back to reference Liporace FA, Gupta S, Jeong GK, Stracher M, Kummer F, Egol KA, Koval KJ: A biomechanical comparison of a dorsal 3.5-mm T-plate and a volar fixed-angle plate in a model of dorsally unstable distal radius fractures. J Orthop Trauma. 2005, 19: 187-191. 10.1097/00005131-200503000-00006.CrossRefPubMed Liporace FA, Gupta S, Jeong GK, Stracher M, Kummer F, Egol KA, Koval KJ: A biomechanical comparison of a dorsal 3.5-mm T-plate and a volar fixed-angle plate in a model of dorsally unstable distal radius fractures. J Orthop Trauma. 2005, 19: 187-191. 10.1097/00005131-200503000-00006.CrossRefPubMed
5.
go back to reference Liporace FA, Kubiak EN, Jeong GK, Iesaka K, Egol KA, Koval KJ: A biomechanical comparison of two volar locked plates in a dorsally unstable distal radius fracture model. J Trauma. 2006, 61: 668-672. 10.1097/01.ta.0000234727.51894.7d.CrossRefPubMed Liporace FA, Kubiak EN, Jeong GK, Iesaka K, Egol KA, Koval KJ: A biomechanical comparison of two volar locked plates in a dorsally unstable distal radius fracture model. J Trauma. 2006, 61: 668-672. 10.1097/01.ta.0000234727.51894.7d.CrossRefPubMed
6.
go back to reference Blythe M, Stoffel K, Jarrett P, Kuster M: Volar versus dorsal locking plates with and without radial styloid locking plates for the fixation of dorsally comminuted distal radius fractures: A biomechanical study in cadavers. J Hand Surg Am. 2006, 31: 1587-1593. 10.1016/j.jhsa.2006.09.011.CrossRefPubMed Blythe M, Stoffel K, Jarrett P, Kuster M: Volar versus dorsal locking plates with and without radial styloid locking plates for the fixation of dorsally comminuted distal radius fractures: A biomechanical study in cadavers. J Hand Surg Am. 2006, 31: 1587-1593. 10.1016/j.jhsa.2006.09.011.CrossRefPubMed
7.
go back to reference Kandemir U, Matityahu A, Desai R, Puttlitz C: Does a volar locking plate provide equivalent stability as a dorsal nonlocking plate in a dorsally comminuted distal radius fracture?: a biomechanical study. J Orthop Trauma. 2008, 22: 605-610. 10.1097/BOT.0b013e318186006f.CrossRefPubMed Kandemir U, Matityahu A, Desai R, Puttlitz C: Does a volar locking plate provide equivalent stability as a dorsal nonlocking plate in a dorsally comminuted distal radius fracture?: a biomechanical study. J Orthop Trauma. 2008, 22: 605-610. 10.1097/BOT.0b013e318186006f.CrossRefPubMed
8.
go back to reference Strauss EJ, Banerjee D, Kummer FJ, Tejwani NC: Evaluation of a novel, nonspanning external fixator for treatment of unstable extra-articular fractures of the distal radius: biomechanical comparison with a volar locking plate. J Trauma. 2008, 64: 975-981. 10.1097/TA.0b013e3180eea9f0.CrossRefPubMed Strauss EJ, Banerjee D, Kummer FJ, Tejwani NC: Evaluation of a novel, nonspanning external fixator for treatment of unstable extra-articular fractures of the distal radius: biomechanical comparison with a volar locking plate. J Trauma. 2008, 64: 975-981. 10.1097/TA.0b013e3180eea9f0.CrossRefPubMed
9.
go back to reference Rausch S, Klos K, Stephan H, Hoffmeier K, Gras F, Windolf M, Gueorguiev B, Hofmann GO, Mückley T: Evaluation of a polyaxial angle-stable volar plate in a distal radius C-fracture model - A biomechanical study. Injury. 2011, 42 (11): 1248-1252. 10.1016/j.injury.2010.12.005.CrossRefPubMed Rausch S, Klos K, Stephan H, Hoffmeier K, Gras F, Windolf M, Gueorguiev B, Hofmann GO, Mückley T: Evaluation of a polyaxial angle-stable volar plate in a distal radius C-fracture model - A biomechanical study. Injury. 2011, 42 (11): 1248-1252. 10.1016/j.injury.2010.12.005.CrossRefPubMed
10.
go back to reference Ekenstam F, Hagert CG: The distal radio ulnar joint. The influence of geometry and ligament on simulated Colles' fracture. An experimental study. Scand J Plast Reconstr Surg. 1985, 19: 27-31. 10.3109/02844318509052862.CrossRef Ekenstam F, Hagert CG: The distal radio ulnar joint. The influence of geometry and ligament on simulated Colles' fracture. An experimental study. Scand J Plast Reconstr Surg. 1985, 19: 27-31. 10.3109/02844318509052862.CrossRef
11.
go back to reference Dunning CE, Lindsay CS, Bicknell RT, Patterson SD, Johnson JA, King GJ: Supplemental pinning improves the stability of external fixation in distal radius fractures during simulated finger and forearm motion. J Hand Surg Am. 1999, 24: 992-1000. 10.1053/jhsu.1999.0992.CrossRefPubMed Dunning CE, Lindsay CS, Bicknell RT, Patterson SD, Johnson JA, King GJ: Supplemental pinning improves the stability of external fixation in distal radius fractures during simulated finger and forearm motion. J Hand Surg Am. 1999, 24: 992-1000. 10.1053/jhsu.1999.0992.CrossRefPubMed
12.
go back to reference Koh S, Morris RP, Patterson RM, Kearney JP, Buford WL, Viegas SF: Volar fixation for dorsally angulated extra-articular fractures of the distal radius: a biomechanical study. J Hand Surg Am. 2006, 31: 771-779. 10.1016/j.jhsa.2006.02.015.CrossRefPubMed Koh S, Morris RP, Patterson RM, Kearney JP, Buford WL, Viegas SF: Volar fixation for dorsally angulated extra-articular fractures of the distal radius: a biomechanical study. J Hand Surg Am. 2006, 31: 771-779. 10.1016/j.jhsa.2006.02.015.CrossRefPubMed
13.
go back to reference Gondusky JS, Carney J, Erpenbach J, Robertson C, Mahar A, Oka R, Thompson M, Mazurek M: Biomechanical comparison of locking versus nonlocking volar and dorsal T-plates for fixation of dorsally comminuted distal radius fractures. J Orthop Trauma. 2011, 25: 44-50. 10.1097/BOT.0b013e3181d7a3a6.CrossRefPubMed Gondusky JS, Carney J, Erpenbach J, Robertson C, Mahar A, Oka R, Thompson M, Mazurek M: Biomechanical comparison of locking versus nonlocking volar and dorsal T-plates for fixation of dorsally comminuted distal radius fractures. J Orthop Trauma. 2011, 25: 44-50. 10.1097/BOT.0b013e3181d7a3a6.CrossRefPubMed
14.
go back to reference Peine R, Rikli DA, Hoffmann R, Duda G, Regazzoni P: Comparison of three different plating techniques for the dorsum of the distal radius: a biomechanical study. J Hand Surg Am. 2000, 25: 29-33. 10.1053/jhsu.2000.jhsu025a0029.CrossRefPubMed Peine R, Rikli DA, Hoffmann R, Duda G, Regazzoni P: Comparison of three different plating techniques for the dorsum of the distal radius: a biomechanical study. J Hand Surg Am. 2000, 25: 29-33. 10.1053/jhsu.2000.jhsu025a0029.CrossRefPubMed
15.
go back to reference Cooper EO, Segalman KA, Parks BG, Sharma KM, Nguyen A: Biomechanical stability of a volar locking-screw plate versus fragment-specific fixation in a distal radius fracture model. Am J Orthop. 2007, 36: E46-E49.PubMed Cooper EO, Segalman KA, Parks BG, Sharma KM, Nguyen A: Biomechanical stability of a volar locking-screw plate versus fragment-specific fixation in a distal radius fracture model. Am J Orthop. 2007, 36: E46-E49.PubMed
16.
go back to reference Rogge RD, Adams BD, Goel VK: An analysis of bone stresses and fixation stability using a finite element model of simulated distal radius fractures. J Hand Surg Am. 2002, 27: 86-92. 10.1053/jhsu.2002.29485.CrossRefPubMed Rogge RD, Adams BD, Goel VK: An analysis of bone stresses and fixation stability using a finite element model of simulated distal radius fractures. J Hand Surg Am. 2002, 27: 86-92. 10.1053/jhsu.2002.29485.CrossRefPubMed
17.
go back to reference Augat P, Reeb H, Claes LE: Prediction of fracture load at different skeletal sites by geometric properties of the cortical shell. J Bone Miner Res. 1996, 11: 1356-1363.CrossRefPubMed Augat P, Reeb H, Claes LE: Prediction of fracture load at different skeletal sites by geometric properties of the cortical shell. J Bone Miner Res. 1996, 11: 1356-1363.CrossRefPubMed
18.
go back to reference Augat P, Iida H, Jiang Y, Diao E, Genant HK: Distal radius fractures: mechanisms of injury and strength prediction by bone mineral assessment. J Orthop Res. 1998, 16: 629-635. 10.1002/jor.1100160517.CrossRefPubMed Augat P, Iida H, Jiang Y, Diao E, Genant HK: Distal radius fractures: mechanisms of injury and strength prediction by bone mineral assessment. J Orthop Res. 1998, 16: 629-635. 10.1002/jor.1100160517.CrossRefPubMed
19.
go back to reference Baumbach SF, Schmidt R, Varga P, Heinz T, Vécsei V, Zysset PK: Where is the distal fracture line location of dorsally displaced distal radius fractures?. J Orthop Res. 2011, 29: 489-494. 10.1002/jor.21268.CrossRefPubMed Baumbach SF, Schmidt R, Varga P, Heinz T, Vécsei V, Zysset PK: Where is the distal fracture line location of dorsally displaced distal radius fractures?. J Orthop Res. 2011, 29: 489-494. 10.1002/jor.21268.CrossRefPubMed
20.
go back to reference Varga P, Baumbach S, Pahr D, Zysset PK: Validation of an anatomy specific finite element model of Colles' fracture. J Biomech. 2009, 42: 1726-1731. 10.1016/j.jbiomech.2009.04.017.CrossRefPubMed Varga P, Baumbach S, Pahr D, Zysset PK: Validation of an anatomy specific finite element model of Colles' fracture. J Biomech. 2009, 42: 1726-1731. 10.1016/j.jbiomech.2009.04.017.CrossRefPubMed
21.
go back to reference Wolfram U, Wilke H-J, Zysset PK: Damage accumulation in vertebral trabecular bone depends on loading mode and direction. J Biomech. 2011, 44: 1164-1169. 10.1016/j.jbiomech.2011.01.018.CrossRefPubMed Wolfram U, Wilke H-J, Zysset PK: Damage accumulation in vertebral trabecular bone depends on loading mode and direction. J Biomech. 2011, 44: 1164-1169. 10.1016/j.jbiomech.2011.01.018.CrossRefPubMed
22.
go back to reference Cartner JL, Hartsell ZM, Ricci WM, Tornetta P: Can We Trust Ex Vivo Mechanical Testing of Fresh-Frozen Cadaveric Specimens? The Effect of Postfreezing Delays. J Orthop Trauma. 2011, 25 (8): 459-461. 10.1097/BOT.0b013e318225b875.CrossRefPubMed Cartner JL, Hartsell ZM, Ricci WM, Tornetta P: Can We Trust Ex Vivo Mechanical Testing of Fresh-Frozen Cadaveric Specimens? The Effect of Postfreezing Delays. J Orthop Trauma. 2011, 25 (8): 459-461. 10.1097/BOT.0b013e318225b875.CrossRefPubMed
23.
go back to reference Greenberg JA, Warden S, Izadi KD: The effect of screw length on fracture stability in volar locked plating of distal radius fractures. Read at the AAOS Annual Meeting; 2010 March 10–14; New Orleans, LA Greenberg JA, Warden S, Izadi KD: The effect of screw length on fracture stability in volar locked plating of distal radius fractures. Read at the AAOS Annual Meeting; 2010 March 10–14; New Orleans, LA
24.
go back to reference Weninger P, Dall'ara E, Drobetz H, Nemec W, Figl M, Redl H, Hertz H, Zysset P: Multidirectional volar fixed-angle plating using cancellous locking screws for distal radius fractures - Evaluation of three screw configurations in an extra-articular fracture model. Wien Klin Wochenschr. 2010, 123 (1–2): 4-10. Weninger P, Dall'ara E, Drobetz H, Nemec W, Figl M, Redl H, Hertz H, Zysset P: Multidirectional volar fixed-angle plating using cancellous locking screws for distal radius fractures - Evaluation of three screw configurations in an extra-articular fracture model. Wien Klin Wochenschr. 2010, 123 (1–2): 4-10.
25.
go back to reference Mehling I, Müller LP, Delinsky K, Mehler D, Burkhart KJ, Rommens PM: Number and locations of screw fixation for volar fixed-angle plating of distal radius fractures: biomechanical study. J Hand Surg Am. 2010, 35: 885-891. 10.1016/j.jhsa.2010.03.027.CrossRefPubMed Mehling I, Müller LP, Delinsky K, Mehler D, Burkhart KJ, Rommens PM: Number and locations of screw fixation for volar fixed-angle plating of distal radius fractures: biomechanical study. J Hand Surg Am. 2010, 35: 885-891. 10.1016/j.jhsa.2010.03.027.CrossRefPubMed
26.
go back to reference Krukhaug Y, Gjerdet NR, Lundberg OJ, Lilleng PK, Hove LM: Different osteosyntheses for Colles' fracture: a mechanical study in 42 cadaver bones. Acta Orthop. 2009, 80: 239-244. 10.3109/17453670902947440.CrossRefPubMedPubMedCentral Krukhaug Y, Gjerdet NR, Lundberg OJ, Lilleng PK, Hove LM: Different osteosyntheses for Colles' fracture: a mechanical study in 42 cadaver bones. Acta Orthop. 2009, 80: 239-244. 10.3109/17453670902947440.CrossRefPubMedPubMedCentral
27.
go back to reference Engelke K, Libanati C, Liu Y, Wang H, Austin M, Fuerst T, Stampa B, Timm W, Genant HK: Quantitative computed tomography (QCT) of the forearm using general purpose spiral whole-body CT scanners: accuracy, precision and comparison with dual-energy X-ray absorptiometry (DXA). Bone. 2009, 45: 110-118. 10.1016/j.bone.2009.03.669.CrossRefPubMed Engelke K, Libanati C, Liu Y, Wang H, Austin M, Fuerst T, Stampa B, Timm W, Genant HK: Quantitative computed tomography (QCT) of the forearm using general purpose spiral whole-body CT scanners: accuracy, precision and comparison with dual-energy X-ray absorptiometry (DXA). Bone. 2009, 45: 110-118. 10.1016/j.bone.2009.03.669.CrossRefPubMed
28.
go back to reference Varga P, Pahr DH, Baumbach S, Zysset PK: HR-pQCT based FE analysis of the most distal radius section provides an improved prediction of Colles' fracture load in vitro. Bone. 2010, 47 (5): 982-988. 10.1016/j.bone.2010.08.002.CrossRefPubMed Varga P, Pahr DH, Baumbach S, Zysset PK: HR-pQCT based FE analysis of the most distal radius section provides an improved prediction of Colles' fracture load in vitro. Bone. 2010, 47 (5): 982-988. 10.1016/j.bone.2010.08.002.CrossRefPubMed
Metadata
Title
Assessment of a novel biomechanical fracture model for distal radius fractures
Authors
Sebastian Felix Baumbach
Enrico Dall’Ara
Patrick Weninger
Anna Antoni
Hannes Traxler
Martin Dörr
Philippe K Zysset
Publication date
01-12-2012
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2012
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-13-252

Other articles of this Issue 1/2012

BMC Musculoskeletal Disorders 1/2012 Go to the issue