Is the rotation of the femural head a potential initiation for cutting out? A theoretical and experimental approach

Authors: Andreas Lenich, Samuel Bachmeier, Lukas Prantl, Michael Nerlich, Jochen Hammer, Edgar Mayr, Amir Andreas Al-Munajjed, Bernd Füchtmeier

Published in: BMC Musculoskeletal Disorders | Issue 1/2011

Login to get access

Abstract

Background

Since cut-out still remains one of the major clinical challenges in the field of osteoporotic proximal femur fractures, remarkable developments have been made in improving treatment concepts. However, the mechanics of these complications have not been fully understood.
We hypothesize using the experimental data and a theoretical model that a previous rotation of the femoral head due to de-central implant positioning can initiate a cut-out.

Methods

In this investigation we analysed our experimental data using two common screws (DHS/Gamma 3) and helical blades (PFN A/TFN) for the fixation of femur fractures in a simple theoretical model applying typical gait pattern on de-central positioned implants. In previous tests during a forced implant rotation by a biomechanical testing machine in a human femoral head the two screws showed failure symptoms (2-6Nm) at the same magnitude as torques acting in the hip during daily activities with de-central implant positioning, while the helical blades showed a better stability (10-20Nm).
To calculate the torque of the head around the implant only the force and the leverarm is needed (N [Nm] = F [N] * × [m]). The force F is a product of the mass M [kg] multiplied by the acceleration g [m/s2]. The leverarm is the distance between the center of the head of femur and the implant center on a horizontal line.

Results

Using 50% of 75 kg body weight a torque of 0.37Nm for the 1 mm decentralized position and 1.1Nm for the 3 mm decentralized position of the implant was calculated. At 250% BW, appropriate to a normal step, torques of 1.8Nm (1 mm) and 5.5Nm (3 mm) have been calculated.
Comparing of the experimental and theoretical results shows that both screws fail in the same magnitude as torques occur in a more than 3 mm de-central positioned implant.

Conclusion

We conclude the center-center position in the head of femur of any kind of lag screw or blade is to be achieved to minimize rotation of the femoral head and to prevent further mechanical complications.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bonnaire F, Weber A, Bosl O, Eckhardt C, Schwieger K, Linke B: "Cutting out" in pertrochanteric fractures--problem of osteoporosis?. Unfallchirurg. 2007, 110: 425-32. 10.1007/s00113-007-1248-0.CrossRefPubMed Bonnaire F, Weber A, Bosl O, Eckhardt C, Schwieger K, Linke B: "Cutting out" in pertrochanteric fractures--problem of osteoporosis?. Unfallchirurg. 2007, 110: 425-32. 10.1007/s00113-007-1248-0.CrossRefPubMed
2.
go back to reference Parker M, Handoll H: Extramedullary fixation implants and external fixators for extracapsular hip fractures in adults. Cochrane Database Syst Rev. 2006, CD000339- Parker M, Handoll H: Extramedullary fixation implants and external fixators for extracapsular hip fractures in adults. Cochrane Database Syst Rev. 2006, CD000339-
3.
go back to reference Baumgaertner MR, Solberg BD: Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg Br. 1997, 79: 969-71. 10.1302/0301-620X.79B6.7949.CrossRefPubMed Baumgaertner MR, Solberg BD: Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg Br. 1997, 79: 969-71. 10.1302/0301-620X.79B6.7949.CrossRefPubMed
4.
go back to reference Lustenberger A, Ganz R: Epidemiology of trochanteric femoral fractures over 2 decades (1972-1988). Unfallchirurg. 1995, 98: 278-82.PubMed Lustenberger A, Ganz R: Epidemiology of trochanteric femoral fractures over 2 decades (1972-1988). Unfallchirurg. 1995, 98: 278-82.PubMed
5.
go back to reference Al-Munajjed AA, Hammer J, Mayr E, Nerlich M, Lenich A: Biomechanical characterisation of osteosyntheses for proximal femur fractures: helical blade versus screw. Stud Health Technol Inform. 2008, 133: 1-10.PubMed Al-Munajjed AA, Hammer J, Mayr E, Nerlich M, Lenich A: Biomechanical characterisation of osteosyntheses for proximal femur fractures: helical blade versus screw. Stud Health Technol Inform. 2008, 133: 1-10.PubMed
6.
go back to reference Lenich A, Fierlbeck J, Al-Munajjed AA, Dendorfer S, Mai R, Fuchtmeier B, Mayr E, Hammer J: First clinical and biomechanical results of the Trochanteric Fixation Nail (TFN). Technol Health Care. 2006, 14: 403-9.PubMed Lenich A, Fierlbeck J, Al-Munajjed AA, Dendorfer S, Mai R, Fuchtmeier B, Mayr E, Hammer J: First clinical and biomechanical results of the Trochanteric Fixation Nail (TFN). Technol Health Care. 2006, 14: 403-9.PubMed
7.
go back to reference Lenich A, Mayr E, Rüter A, Möckl C, Füchtmeier B: First results with the Trochanter Fixation Nail (TFN): a report on 120 cases. Arch Orthop Trauma Surg. 2006 Lenich A, Mayr E, Rüter A, Möckl C, Füchtmeier B: First results with the Trochanter Fixation Nail (TFN): a report on 120 cases. Arch Orthop Trauma Surg. 2006
8.
go back to reference Windolf M, Braunstein V, Dutoit C, Schwieger K: Is a helical shaped implant a superior alternative to the Dynamic Hip Screw for unstable femoral neck fractures? A biomechanical investigation. Clin Biomech (Bristol, Avon). 2009, 24: 59-64. 10.1016/j.clinbiomech.2008.07.004.CrossRef Windolf M, Braunstein V, Dutoit C, Schwieger K: Is a helical shaped implant a superior alternative to the Dynamic Hip Screw for unstable femoral neck fractures? A biomechanical investigation. Clin Biomech (Bristol, Avon). 2009, 24: 59-64. 10.1016/j.clinbiomech.2008.07.004.CrossRef
9.
go back to reference Al-yassari G, Langstaff RJ, Jones JW, Al-Lami M: The AO/ASIF proximal femoral nail (PFN) for the treatment of unstable trochanteric femoral fracture. Injury. 2002, 33: 395-9. 10.1016/S0020-1383(02)00008-6.CrossRefPubMed Al-yassari G, Langstaff RJ, Jones JW, Al-Lami M: The AO/ASIF proximal femoral nail (PFN) for the treatment of unstable trochanteric femoral fracture. Injury. 2002, 33: 395-9. 10.1016/S0020-1383(02)00008-6.CrossRefPubMed
10.
go back to reference Bergmann G, Graichen F, Rohlmann A: Hip joint loading during walking and running, measured in two patients. J Biomech. 1993, 26: 969-90. 10.1016/0021-9290(93)90058-M.CrossRefPubMed Bergmann G, Graichen F, Rohlmann A: Hip joint loading during walking and running, measured in two patients. J Biomech. 1993, 26: 969-90. 10.1016/0021-9290(93)90058-M.CrossRefPubMed
11.
go back to reference Dendorfer S, Meier HJ, Taylor D, Hammer J: Anisotropy of the fatigue behaviour of cancellous bone. J Biomech. 2008, 41 (3): 636-4. 10.1016/j.jbiomech.2007.09.037.CrossRefPubMed Dendorfer S, Meier HJ, Taylor D, Hammer J: Anisotropy of the fatigue behaviour of cancellous bone. J Biomech. 2008, 41 (3): 636-4. 10.1016/j.jbiomech.2007.09.037.CrossRefPubMed
12.
go back to reference Brunner A, Jöckle JA, Babst R: The PFNA proximal femur nail in treatment of unstable proximal femur fractures--3 cases of postoperative perforation of the helical blade into the hip joint. Journal of Orthopaedic Trauma. 2008, 22 (10): 731-6. 10.1097/BOT.0b013e3181893b1b.CrossRefPubMed Brunner A, Jöckle JA, Babst R: The PFNA proximal femur nail in treatment of unstable proximal femur fractures--3 cases of postoperative perforation of the helical blade into the hip joint. Journal of Orthopaedic Trauma. 2008, 22 (10): 731-6. 10.1097/BOT.0b013e3181893b1b.CrossRefPubMed
13.
go back to reference Cleveland M, Bosworth DM, Thompson FR, Wilson HJ, Ishizuka T: A ten-year analysis of intertrochanteric fractures of the femur. Am J Orthop. 1959, 41-A: 1399-408. Cleveland M, Bosworth DM, Thompson FR, Wilson HJ, Ishizuka T: A ten-year analysis of intertrochanteric fractures of the femur. Am J Orthop. 1959, 41-A: 1399-408.
14.
go back to reference Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77: 1058-64.PubMed Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77: 1058-64.PubMed
Metadata
Title
Is the rotation of the femural head a potential initiation for cutting out? A theoretical and experimental approach
Authors
Andreas Lenich
Samuel Bachmeier
Lukas Prantl
Michael Nerlich
Jochen Hammer
Edgar Mayr
Amir Andreas Al-Munajjed
Bernd Füchtmeier
Publication date
01-12-2011
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2011
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-12-79

Other articles of this Issue 1/2011

BMC Musculoskeletal Disorders 1/2011 Go to the issue