Published in: BMC Ophthalmology 1/2008

Open Access 01-12-2008 | Research article

Rotational stability of the AcrySof SA60TT toric intraocular lenses: A cohort study

Authors: Bruno Zuberbuhler, Theo Signer, Richard Gale, Eduard Haefliger

Published in: BMC Ophthalmology | Issue 1/2008

Login to get access

Abstract

Background

To evaluate the rotational stability of the three types of AcrySof SA60TT toric intraocular lenses (Alcon, Switzerland) in cataract surgery after the first postoperative week.

Methods

A retrospective study of 44 eyes in 33 patients. All patients underwent similar uncomplicated phacoemulsification cataract surgery. Seven eyes with corneal astigmatism of less than 1.5 D were implanted with the AcrySof SA60T3 intraocular lens. Seventeen eyes with astigmatism between 1.5 D and 2.25 D received the SA60T4 intraocular lens, and 20 eyes with more than 2.25 D of corneal astigmatism received the SA60T5 intraocular lens. Intraoperatively, the axis of the toric lens was aligned to the steepest axis of the corneal astigmatism. Main outcome measure was the postoperative position of the lens, assessed at 1 week and 3 months, using a specially designed angle measuring eyepiece for the slit lamp.

Results

There was no significant difference in the rotational stability of the three types of toric intraocular lenses. Overall, the postoperative rotation was within 5 degrees in 95% and within 2 degrees in 68% of eyes. The mean absolute rotation was 2.2 ± 2.2 degrees. No lens showed more than 9 degrees of rotation, and no lens required secondary repositioning. There was no trend for either clockwise or anti-clockwise rotation. The surgical procedure did not change the corneal astigmatism.

Conclusion

Once placed to it's position, each of the three types of the AcrySof SA60TT toric intraocular lenses demonstrate rotational stability in the capsular bag.
Appendix
Available only for authorised users
Literature
1.
2.
go back to reference Horn JD: Status of toric intraocular lenses. Curr Opin Ophthalmol. 2007, 18 (1): 58-61. 10.1097/ICU.0b013e328011f9bf.CrossRefPubMed Horn JD: Status of toric intraocular lenses. Curr Opin Ophthalmol. 2007, 18 (1): 58-61. 10.1097/ICU.0b013e328011f9bf.CrossRefPubMed
3.
go back to reference Ohmi S: Decentration associated with asymmetric capsular shrinkage and intraocular lens size. J Cataract Refract Surg. 1993, 19: 640-643.CrossRefPubMed Ohmi S: Decentration associated with asymmetric capsular shrinkage and intraocular lens size. J Cataract Refract Surg. 1993, 19: 640-643.CrossRefPubMed
4.
go back to reference Strenn K, Menapace R, Vass C: Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring. J Cataract Refract Surg. 1997, 23: 1543-1547.CrossRefPubMed Strenn K, Menapace R, Vass C: Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring. J Cataract Refract Surg. 1997, 23: 1543-1547.CrossRefPubMed
5.
go back to reference Sanders DR, Grabow HB, Shepherd J: The toric IOL. Sutureless Cataract Surgery; An Evolution Toward Minimally Invasive Technique. Edited by: Gills JP, Martin RG, Sanders DR. 1992, Thorofare, NJ, Slack, 183-197. Sanders DR, Grabow HB, Shepherd J: The toric IOL. Sutureless Cataract Surgery; An Evolution Toward Minimally Invasive Technique. Edited by: Gills JP, Martin RG, Sanders DR. 1992, Thorofare, NJ, Slack, 183-197.
6.
go back to reference De Silva DJ, Ramkissoon YD, Bloom PA: Evaluation of a toric intraocular lens with Z-haptic. J Cataract Refract Surg. 2006, 32: 1492-1498. 10.1016/j.jcrs.2006.04.022.CrossRefPubMed De Silva DJ, Ramkissoon YD, Bloom PA: Evaluation of a toric intraocular lens with Z-haptic. J Cataract Refract Surg. 2006, 32: 1492-1498. 10.1016/j.jcrs.2006.04.022.CrossRefPubMed
7.
go back to reference Warlo I, Krummenauer F, Dick HB: Rotational stability in intraocular lenses with C-haptics versus Z-haptics in cataract surgery. A prospective randomised comparison. Ophthalmologe. 2005, 102: 987-992. 10.1007/s00347-005-1213-7.CrossRefPubMed Warlo I, Krummenauer F, Dick HB: Rotational stability in intraocular lenses with C-haptics versus Z-haptics in cataract surgery. A prospective randomised comparison. Ophthalmologe. 2005, 102: 987-992. 10.1007/s00347-005-1213-7.CrossRefPubMed
8.
go back to reference Chang DF: Early rotational stability of the longer Staar toric intraocular lens: fifty consecutive cases. J Cataract Refract Surg. 2003, 29: 935-940. 10.1016/S0886-3350(02)01843-6.CrossRefPubMed Chang DF: Early rotational stability of the longer Staar toric intraocular lens: fifty consecutive cases. J Cataract Refract Surg. 2003, 29: 935-940. 10.1016/S0886-3350(02)01843-6.CrossRefPubMed
9.
go back to reference Weinand F, Jung A, Stein A, Pfotzner A, Becker R, Pavlovic S: Rotational stability of a single-piece hydrophobic acrylic intraocular lens: new method for high-precision rotation control. J Cataract Refract Surg. 2007, 33: 800-803. 10.1016/j.jcrs.2007.01.030.CrossRefPubMed Weinand F, Jung A, Stein A, Pfotzner A, Becker R, Pavlovic S: Rotational stability of a single-piece hydrophobic acrylic intraocular lens: new method for high-precision rotation control. J Cataract Refract Surg. 2007, 33: 800-803. 10.1016/j.jcrs.2007.01.030.CrossRefPubMed
10.
go back to reference Becker KA, Auffarth GU, Völcker HE: Measurement method for the determination of rotation and decentration of intraocular lenses. Ophthalmologe. 2004, 101: 600-603. 10.1007/s00347-003-0951-7.CrossRefPubMed Becker KA, Auffarth GU, Völcker HE: Measurement method for the determination of rotation and decentration of intraocular lenses. Ophthalmologe. 2004, 101: 600-603. 10.1007/s00347-003-0951-7.CrossRefPubMed
11.
go back to reference Viestenz A, Langenbucher A, Seitz B: Impact of the eye's cyclorotation on axial orientation analysis of toric intraocular lenses: recommendations for an optimized evaluation of rotational stability of toric IOLs. Klin Monatsbl Augenheilkd. 2006, 223: 134-140. 10.1055/s-2005-858715.CrossRefPubMed Viestenz A, Langenbucher A, Seitz B: Impact of the eye's cyclorotation on axial orientation analysis of toric intraocular lenses: recommendations for an optimized evaluation of rotational stability of toric IOLs. Klin Monatsbl Augenheilkd. 2006, 223: 134-140. 10.1055/s-2005-858715.CrossRefPubMed
12.
go back to reference Viestenz A, Seitz B, Langenbucher A: Evaluating the eye's rotational stability during standard photography. Effect on determining the axial orientation of toric intraocular lenses. J Cataract Refract Surg. 2005, 31: 557-561. 10.1016/j.jcrs.2004.07.019.CrossRefPubMed Viestenz A, Seitz B, Langenbucher A: Evaluating the eye's rotational stability during standard photography. Effect on determining the axial orientation of toric intraocular lenses. J Cataract Refract Surg. 2005, 31: 557-561. 10.1016/j.jcrs.2004.07.019.CrossRefPubMed
13.
go back to reference Quentin CD, Genée D: Rotationsstabilität der Silikon-HKL mit C-Haptik versus Z-Haptik im Kapselsack und mit einer C-Haptik im Sulcus ciliaris. 17. Kongress der Deutschsprachigen Gesellschaft für Intraokularlinsen-Implantation und refraktiven Chirurgie. Edited by: Auffarth GU, Welt R, Demeler U, Köln, Biermann. 175-180. Quentin CD, Genée D: Rotationsstabilität der Silikon-HKL mit C-Haptik versus Z-Haptik im Kapselsack und mit einer C-Haptik im Sulcus ciliaris. 17. Kongress der Deutschsprachigen Gesellschaft für Intraokularlinsen-Implantation und refraktiven Chirurgie. Edited by: Auffarth GU, Welt R, Demeler U, Köln, Biermann. 175-180.
14.
go back to reference Viestenz A, Walter S, Viestenz A, Behrens-Baumann W, Langenbucher A: Torische Intraokularlinse und Astigmatismuskorrektur. Ophthalmologe. 2007, 104: 620-627. 10.1007/s00347-007-1576-z.CrossRefPubMed Viestenz A, Walter S, Viestenz A, Behrens-Baumann W, Langenbucher A: Torische Intraokularlinse und Astigmatismuskorrektur. Ophthalmologe. 2007, 104: 620-627. 10.1007/s00347-007-1576-z.CrossRefPubMed
Metadata
Title
Rotational stability of the AcrySof SA60TT toric intraocular lenses: A cohort study
Authors
Bruno Zuberbuhler
Theo Signer
Richard Gale
Eduard Haefliger
Publication date
01-12-2008
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2008
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/1471-2415-8-8

Other articles of this Issue 1/2008

BMC Ophthalmology 1/2008 Go to the issue