Published in: BMC Ophthalmology 1/2012

Open Access 01-12-2012 | Research article

Prospective study of toric IOL outcomes based on the Lenstar LS 900® dual zone automated keratometer

Authors: Kjell Gunnar Gundersen, Richard Potvin

Published in: BMC Ophthalmology | Issue 1/2012

Login to get access

Abstract

Background

To establish clinical expectations when using the Lenstar LS 900® dual-zone automated keratometer for surgery planning of toric intraocular lenses.

Methods

Fifty eyes were measured with the Lenstar LS 900® dual-zone automated keratometer . Surgical planning was performed with the data from this device and the known surgically induced astigmatism of the surgeon. Post-operative refractions and visual acuity were measured at 1 month and 3 months.

Results

Clinical outcomes from 43 uncomplicated surgeries showed an average post-operative refractive astigmatism of 0.44D ±0.25D. Over 70% of eyes had 0.50D or less of refractive astigmatism and no eye had more than 1.0D of refractive astigmatism. Uncorrected visual acuity was 20/32 or better in all eyes at 3 months, with 70% of eyes 20/20 or better. A significantly higher number of eyes had 0.75D or more of post-operative refractive astigmatism when the standard deviation of the pre-operative calculated corneal astigmatism angle, reported by the keratometer, was > 5 degrees.

Conclusions

In this single-site study investigating the use of the keratometry from the Lenstar LS 900® for toric IOL surgical planning, clinical outcomes appear equivalent to those reported in the literature for manual keratometry and somewhat better than has been reported for some previous automated instruments. A high standard deviation in the pre-operative calculated astigmatism angle, as reported by the keratometer, appears to increase the likelihood of higher post-operative refractive astigmatism.
Appendix
Available only for authorised users
Literature
1.
go back to reference Narvaez J, Zimmerman G, Stulting RD, Chang DH: Accuracy of intraocular lens power prediction using the Hoffer Q, Holladay 1, Holladay 2, and SRK/T formulas. J Cataract Refract Surg. 2006, 32 (12): 2050-2053. 10.1016/j.jcrs.2006.09.009.CrossRefPubMed Narvaez J, Zimmerman G, Stulting RD, Chang DH: Accuracy of intraocular lens power prediction using the Hoffer Q, Holladay 1, Holladay 2, and SRK/T formulas. J Cataract Refract Surg. 2006, 32 (12): 2050-2053. 10.1016/j.jcrs.2006.09.009.CrossRefPubMed
2.
go back to reference Sarikkola AU, Uusitalo RJ, Hellstedt T, Ess SL, Leivo T, Kivelä T: Simultaneous bilateral versus sequential bilateral cataract surgery: Helsinki Simultaneous Bilateral Cataract Surgery Study Report 1. J Cataract Refract Surg. 2011, 37 (6): 992-1002. 10.1016/j.jcrs.2011.01.019.CrossRefPubMed Sarikkola AU, Uusitalo RJ, Hellstedt T, Ess SL, Leivo T, Kivelä T: Simultaneous bilateral versus sequential bilateral cataract surgery: Helsinki Simultaneous Bilateral Cataract Surgery Study Report 1. J Cataract Refract Surg. 2011, 37 (6): 992-1002. 10.1016/j.jcrs.2011.01.019.CrossRefPubMed
3.
go back to reference Holland E, Lane S, Horn JD, Ernest P, Arleo R, Miller KM: The AcrySof Toric intraocular lens in subjects with cataracts and corneal astigmatism: a randomized, subject-masked, parallel-group, 1-year study. Ophthalmology. 2010, 117 (11): 2104-2111. 10.1016/j.ophtha.2010.07.033.CrossRefPubMed Holland E, Lane S, Horn JD, Ernest P, Arleo R, Miller KM: The AcrySof Toric intraocular lens in subjects with cataracts and corneal astigmatism: a randomized, subject-masked, parallel-group, 1-year study. Ophthalmology. 2010, 117 (11): 2104-2111. 10.1016/j.ophtha.2010.07.033.CrossRefPubMed
4.
go back to reference Ahmed II, Rocha G, Slomovic AR, Climenhaga H, Gohill J, Grégoire A, Ma J: Canadian Toric Study Group. Visual function and patient experience after bilateral implantation of toric intraocular lenses. J Cataract Refract Surg. 2010, 36 (4): 609-616. 10.1016/j.jcrs.2009.10.044.CrossRefPubMed Ahmed II, Rocha G, Slomovic AR, Climenhaga H, Gohill J, Grégoire A, Ma J: Canadian Toric Study Group. Visual function and patient experience after bilateral implantation of toric intraocular lenses. J Cataract Refract Surg. 2010, 36 (4): 609-616. 10.1016/j.jcrs.2009.10.044.CrossRefPubMed
5.
go back to reference Chang M, Kang SY, Kim HM: Which Keratometer is Most Reliable for Correcting Astigmatism with Toric Intraocular Lenses?. Korean J Ophthalmol. 2012, 26 (1): 10-14. 10.3341/kjo.2012.26.1.10.CrossRefPubMedPubMedCentral Chang M, Kang SY, Kim HM: Which Keratometer is Most Reliable for Correcting Astigmatism with Toric Intraocular Lenses?. Korean J Ophthalmol. 2012, 26 (1): 10-14. 10.3341/kjo.2012.26.1.10.CrossRefPubMedPubMedCentral
6.
go back to reference Hill WE, Potvin R: Monte Carlo simulation of expected outcomes with the AcrySof toric intraocular lens. BMC Ophthalmol. 2008, 8: 22-10.1186/1471-2415-8-22.CrossRefPubMedPubMedCentral Hill WE, Potvin R: Monte Carlo simulation of expected outcomes with the AcrySof toric intraocular lens. BMC Ophthalmol. 2008, 8: 22-10.1186/1471-2415-8-22.CrossRefPubMedPubMedCentral
7.
go back to reference Jin H, Limberger IJ, Ehmer A, Guo H, Auffarth GU: Impact of axis misalignment of toric intraocular lenses on refractive outcomes after cataract surgery. J Cataract Refract Surg. 2010, 36 (12): 2061-2072. 10.1016/j.jcrs.2010.06.066.CrossRefPubMed Jin H, Limberger IJ, Ehmer A, Guo H, Auffarth GU: Impact of axis misalignment of toric intraocular lenses on refractive outcomes after cataract surgery. J Cataract Refract Surg. 2010, 36 (12): 2061-2072. 10.1016/j.jcrs.2010.06.066.CrossRefPubMed
8.
go back to reference Chang DF: Comparative rotational stability of single-piece open-loop acrylic and plate-haptic silicone toric intraocular lenses. J Cataract Refract Surg. 2008, 34 (11): 1842-1847. 10.1016/j.jcrs.2008.07.012.CrossRefPubMed Chang DF: Comparative rotational stability of single-piece open-loop acrylic and plate-haptic silicone toric intraocular lenses. J Cataract Refract Surg. 2008, 34 (11): 1842-1847. 10.1016/j.jcrs.2008.07.012.CrossRefPubMed
9.
go back to reference Hill WE, Osher R, Cooke D, Solomon K, Sandoval H, Salas-Cervantes R, Potvin R: Simulation of Toric IOL Results Comparing Manual Keratometry to Dual-zone Automated Keratometry from an Integrated Biometer. J Cataract Refract Surg. 2011, 37 (12): 2181-2187. 10.1016/j.jcrs.2011.06.028.CrossRefPubMed Hill WE, Osher R, Cooke D, Solomon K, Sandoval H, Salas-Cervantes R, Potvin R: Simulation of Toric IOL Results Comparing Manual Keratometry to Dual-zone Automated Keratometry from an Integrated Biometer. J Cataract Refract Surg. 2011, 37 (12): 2181-2187. 10.1016/j.jcrs.2011.06.028.CrossRefPubMed
10.
go back to reference Medicute J, Irgoyen C, Aramberri J, Ondarra A, Montes-Mico R: Foldable toric intraocular lens for astigmatism correction in cataract patients. J Cataract Refract Surg. 2008, 34: 601-607. 10.1016/j.jcrs.2007.11.033.CrossRef Medicute J, Irgoyen C, Aramberri J, Ondarra A, Montes-Mico R: Foldable toric intraocular lens for astigmatism correction in cataract patients. J Cataract Refract Surg. 2008, 34: 601-607. 10.1016/j.jcrs.2007.11.033.CrossRef
Metadata
Title
Prospective study of toric IOL outcomes based on the Lenstar LS 900® dual zone automated keratometer
Authors
Kjell Gunnar Gundersen
Richard Potvin
Publication date
01-12-2012
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2012
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/1471-2415-12-21