Comparison of the biometric values obtained by two different A-mode ultrasound devices (Eye Cubed vs. PalmScan): A Transversal, descriptive, and comparative study

Authors: Raul Velez-Montoya, Eugene Mark Shusterman, Miriam Jessica López-Miranda, Mariana Mayorquin-Ruiz, Guillermo Salcedo-Villanueva, Hugo Quiroz-Mercado, Virgilio Morales-Cantón

Published in: BMC Ophthalmology | Issue 1/2010

Login to get access

Abstract

Background

To assess the reliability of the measurements obtained with the PalmScan™, when compared with another standardized A-mode ultrasound device, and assess the consistency and correlation between the two methods.

Methods

Transversal, descriptive, and comparative study. We recorded the axial length (AL), anterior chamber depth (ACD) and lens thickness (LT) obtained with two A-mode ultrasounds (PalmScan™ A2000 and Eye Cubed™) using an immersion technique. We compared the measurements with a two-sample t-test. Agreement between the two devices was assessed with Bland-Altman plots and 95% limits of agreement.

Results

70 eyes of 70 patients were enrolled in this study. The measurements with the Eye Cubed™ of AL and ACD were shorter than the measurements taken by the PalmScan™. The differences were not statistically significant regarding AL (p < 0.4) but significant regarding ACD (p < 0.001). The highest agreement between the two devices was obtained during LT measurement. The PalmScan™ measurements were shorter, but not statistically significantly (p < 0.2).

Conclusions

The values of AL and LT, obtained with both devices are not identical, but within the limits of agreement. The agreement is not affected by the magnitude of the ocular dimensions (but only between range of 20 mm to 27 mm of AL and 3.5 mm to 5.7 mm of LT). A correction of about 0.5 D could be considered if an intraocular lens is being calculated. However due to the large variability of the results, the authors recommend discretion in using this conversion factor, and to adjust the power of the intraocular lenses based upon the personal experience of the surgeon.
Appendix
Available only for authorised users
Literature
1.
go back to reference Attas-Fox L, Zadok D, Gerber Y, Morad Y, Eting E, Benamou N, Pras E, Segal O, Avni I, Barkana Y: Axial length measurement in eyes with diabetic macular edema. Ophthalmology. 2007, 114: 1499-1504. 10.1016/j.ophtha.2006.10.053.CrossRefPubMed Attas-Fox L, Zadok D, Gerber Y, Morad Y, Eting E, Benamou N, Pras E, Segal O, Avni I, Barkana Y: Axial length measurement in eyes with diabetic macular edema. Ophthalmology. 2007, 114: 1499-1504. 10.1016/j.ophtha.2006.10.053.CrossRefPubMed
2.
go back to reference Sohajda Z, Papp J, Berta A, Módis L: The comparative Study of two recently developed A-scan devices: determination of central corneal thickness, anterior chamber depth and axial length. Acta Ophthalmol Scand. 2008, 86: 45-48.CrossRef Sohajda Z, Papp J, Berta A, Módis L: The comparative Study of two recently developed A-scan devices: determination of central corneal thickness, anterior chamber depth and axial length. Acta Ophthalmol Scand. 2008, 86: 45-48.CrossRef
3.
go back to reference Oliveira C, Harizman N, Girkin CA, Xie A, Tello C, Liebmann JM, Ritch R: Axial length and optic disc size in the normal eyes. Br J Ophthalmol. 2007, 91: 37-39. 10.1136/bjo.2006.102061.CrossRefPubMed Oliveira C, Harizman N, Girkin CA, Xie A, Tello C, Liebmann JM, Ritch R: Axial length and optic disc size in the normal eyes. Br J Ophthalmol. 2007, 91: 37-39. 10.1136/bjo.2006.102061.CrossRefPubMed
4.
go back to reference Jivrajka R, Shammas MC, Boenzi T, Swearingen M, Shammas HJ: Variability of axial length, anterior chamber depth, and lens thickness in the cataractous eye. J Cataract Refract Surg. 2008, 34: 289-294. 10.1016/j.jcrs.2007.10.015.CrossRefPubMed Jivrajka R, Shammas MC, Boenzi T, Swearingen M, Shammas HJ: Variability of axial length, anterior chamber depth, and lens thickness in the cataractous eye. J Cataract Refract Surg. 2008, 34: 289-294. 10.1016/j.jcrs.2007.10.015.CrossRefPubMed
5.
go back to reference Chan B, Cho P, Cheung SW: Repeatability and agreement of two A-scan ultrasonic biometers and IOLMaster in non-orthokeratology subjects and post-orthokeratology children. Clin Exp Optom. 2006, 89: 160-168. 10.1111/j.1444-0938.2006.00029.x.CrossRefPubMed Chan B, Cho P, Cheung SW: Repeatability and agreement of two A-scan ultrasonic biometers and IOLMaster in non-orthokeratology subjects and post-orthokeratology children. Clin Exp Optom. 2006, 89: 160-168. 10.1111/j.1444-0938.2006.00029.x.CrossRefPubMed
6.
go back to reference Trivedi RH, Wilson ME: Biometry data from Caucasian and African-American cataractous pediatric eyes. Invest Ophthalmol Vis Sci. 2007, 48: 4671-4678. 10.1167/iovs.07-0267.CrossRefPubMed Trivedi RH, Wilson ME: Biometry data from Caucasian and African-American cataractous pediatric eyes. Invest Ophthalmol Vis Sci. 2007, 48: 4671-4678. 10.1167/iovs.07-0267.CrossRefPubMed
7.
go back to reference Moore DB, Ben Zion I, Neely DE, Plager DA, Ofner S, Sprunger DT, Roberts GJ: Accuracy of biometry in pediatric cataract extraction with primary intraocular lens implantation. J Cataract Refract Surg. 2008, 34: 1940-1947. 10.1016/j.jcrs.2008.07.019.CrossRefPubMed Moore DB, Ben Zion I, Neely DE, Plager DA, Ofner S, Sprunger DT, Roberts GJ: Accuracy of biometry in pediatric cataract extraction with primary intraocular lens implantation. J Cataract Refract Surg. 2008, 34: 1940-1947. 10.1016/j.jcrs.2008.07.019.CrossRefPubMed
8.
go back to reference Wong TY, Foster PJ, Ng TP, Tielsch JM, Johnson GJ, Seah SK: Variations in ocular biometry in an adult Chinese population in Singapore: the Tanjong Pagar Survey. Invest Ophthalmol Vis Sci. 2001, 42: 73-80.PubMed Wong TY, Foster PJ, Ng TP, Tielsch JM, Johnson GJ, Seah SK: Variations in ocular biometry in an adult Chinese population in Singapore: the Tanjong Pagar Survey. Invest Ophthalmol Vis Sci. 2001, 42: 73-80.PubMed
9.
go back to reference Shufelt C, Fraser-Bell S, Ying-Lai M, Torres M, Varma R, Los Angeles Latino Eye Study Group: Refractive error, ocular biometry, and lens opalescence in adult population: the Los Angeles Latino Eye Study. Invest Ophthalmol Vis Sci. 2005, 46: 4450-4460. 10.1167/iovs.05-0435.CrossRefPubMed Shufelt C, Fraser-Bell S, Ying-Lai M, Torres M, Varma R, Los Angeles Latino Eye Study Group: Refractive error, ocular biometry, and lens opalescence in adult population: the Los Angeles Latino Eye Study. Invest Ophthalmol Vis Sci. 2005, 46: 4450-4460. 10.1167/iovs.05-0435.CrossRefPubMed
10.
go back to reference Kriechbaum K, Findl O, Kiss B, Sacu S, Petternel V, Drexler W: Comparison of anterior chamber depth measurement methods in phakic and pseudophakic eyes. J Cataract Refract Surg. 2003, 29: 89-94. 10.1016/S0886-3350(02)01822-9.CrossRefPubMed Kriechbaum K, Findl O, Kiss B, Sacu S, Petternel V, Drexler W: Comparison of anterior chamber depth measurement methods in phakic and pseudophakic eyes. J Cataract Refract Surg. 2003, 29: 89-94. 10.1016/S0886-3350(02)01822-9.CrossRefPubMed
11.
go back to reference Koranyi G, Lydahl E, Norrby S, Taube M: Anterior Chamber depth measurement: A-scan versus optical methods. J Cataract Refract Surg. 2002, 28: 243-247. 10.1016/S0886-3350(01)01039-2.CrossRefPubMed Koranyi G, Lydahl E, Norrby S, Taube M: Anterior Chamber depth measurement: A-scan versus optical methods. J Cataract Refract Surg. 2002, 28: 243-247. 10.1016/S0886-3350(01)01039-2.CrossRefPubMed
Metadata
Title
Comparison of the biometric values obtained by two different A-mode ultrasound devices (Eye Cubed vs. PalmScan): A Transversal, descriptive, and comparative study
Authors
Raul Velez-Montoya
Eugene Mark Shusterman
Miriam Jessica López-Miranda
Mariana Mayorquin-Ruiz
Guillermo Salcedo-Villanueva
Hugo Quiroz-Mercado
Virgilio Morales-Cantón
Publication date
01-12-2010
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2010
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/1471-2415-10-8

Other articles of this Issue 1/2010

BMC Ophthalmology 1/2010 Go to the issue