Intraoperative detection of 18F-FDG-avid tissue sites using the increased probe counting efficiency of the K-alpha probe design and variance-based statistical analysis with the three-sigma criteria

Authors: Stephen P Povoski, Gregg J Chapman, Douglas A Murrey, Robert Lee, Edward W Martin, Nathan C Hall

Published in: BMC Cancer | Issue 1/2013

Login to get access

Abstract

Background

Intraoperative detection of 18F-FDG-avid tissue sites during 18F-FDG-directed surgery can be very challenging when utilizing gamma detection probes that rely on a fixed target-to-background (T/B) ratio (ratiometric threshold) for determination of probe positivity. The purpose of our study was to evaluate the counting efficiency and the success rate of in situ intraoperative detection of 18F-FDG-avid tissue sites (using the three-sigma statistical threshold criteria method and the ratiometric threshold criteria method) for three different gamma detection probe systems.

Methods

Of 58 patients undergoing 18F-FDG-directed surgery for known or suspected malignancy using gamma detection probes, we identified nine 18F-FDG-avid tissue sites (from amongst seven patients) that were seen on same-day preoperative diagnostic PET/CT imaging, and for which each 18F-FDG-avid tissue site underwent attempted in situ intraoperative detection concurrently using three gamma detection probe systems (K-alpha probe, and two commercially-available PET-probe systems), and then were subsequently surgical excised.

Results

The mean relative probe counting efficiency ratio was 6.9 (± 4.4, range 2.2–15.4) for the K-alpha probe, as compared to 1.5 (± 0.3, range 1.0–2.1) and 1.0 (± 0, range 1.0–1.0), respectively, for two commercially-available PET-probe systems (P < 0.001). Successful in situ intraoperative detection of 18F-FDG-avid tissue sites was more frequently accomplished with each of the three gamma detection probes tested by using the three-sigma statistical threshold criteria method than by using the ratiometric threshold criteria method, specifically with the three-sigma statistical threshold criteria method being significantly better than the ratiometric threshold criteria method for determining probe positivity for the K-alpha probe (P = 0.05).

Conclusions

Our results suggest that the improved probe counting efficiency of the K-alpha probe design used in conjunction with the three-sigma statistical threshold criteria method can allow for improved detection of 18F-FDG-avid tissue sites when a low in situ T/B ratio is encountered.
Literature
1.
go back to reference Povoski SP, Neff RL, Mojzisik CM, O'Malley DM, Hinkle GH, Hall NC, Murrey DA, Knopp MV, Martin EW: A comprehensive overview of radioguided surgery using gamma detection probe technology. World J Surg Oncol. 2009, 7: 11-10.1186/1477-7819-7-11.CrossRefPubMedPubMedCentral Povoski SP, Neff RL, Mojzisik CM, O'Malley DM, Hinkle GH, Hall NC, Murrey DA, Knopp MV, Martin EW: A comprehensive overview of radioguided surgery using gamma detection probe technology. World J Surg Oncol. 2009, 7: 11-10.1186/1477-7819-7-11.CrossRefPubMedPubMedCentral
2.
go back to reference Povoski SP, Hall NC, Murrey DA, Chow AZ, Gaglani JR, Bahnson EE, Mojzisik CM, Kuhrt MP, Hitchcock CL, Knopp MV, Martin EW: Multimodal imaging and detection approach to 18F-FDG-directed surgery for patients with known or suspected malignancies: a comprehensive description of the specific methodology utilized in a single-institution cumulative retrospective experience. World J Surg Oncol. 2011, 9: 152-10.1186/1477-7819-9-152.CrossRefPubMedPubMedCentral Povoski SP, Hall NC, Murrey DA, Chow AZ, Gaglani JR, Bahnson EE, Mojzisik CM, Kuhrt MP, Hitchcock CL, Knopp MV, Martin EW: Multimodal imaging and detection approach to 18F-FDG-directed surgery for patients with known or suspected malignancies: a comprehensive description of the specific methodology utilized in a single-institution cumulative retrospective experience. World J Surg Oncol. 2011, 9: 152-10.1186/1477-7819-9-152.CrossRefPubMedPubMedCentral
3.
go back to reference Desai DC, Arnold M, Saha S, Hinkle G, Soble D, Fry J, DePalatis LR, Mantil J, Satter M, Martin EW: Correlative whole-body FDG-PET and intraoperative gamma detection of FDG distribution in colorectal cancer. Clin Positron Imaging. 2000, 3: 189-196. 10.1016/S1095-0397(00)00052-2.CrossRefPubMed Desai DC, Arnold M, Saha S, Hinkle G, Soble D, Fry J, DePalatis LR, Mantil J, Satter M, Martin EW: Correlative whole-body FDG-PET and intraoperative gamma detection of FDG distribution in colorectal cancer. Clin Positron Imaging. 2000, 3: 189-196. 10.1016/S1095-0397(00)00052-2.CrossRefPubMed
4.
go back to reference Zervos EE, Desai DC, DePalatis LR, Soble D, Martin EW: 18F-labeled fluorodeoxyglucose positron emission tomography-guided surgery for recurrent colorectal cancer: a feasibility study. J Surg Res. 2001, 97: 9-13. 10.1006/jsre.2001.6092.CrossRefPubMed Zervos EE, Desai DC, DePalatis LR, Soble D, Martin EW: 18F-labeled fluorodeoxyglucose positron emission tomography-guided surgery for recurrent colorectal cancer: a feasibility study. J Surg Res. 2001, 97: 9-13. 10.1006/jsre.2001.6092.CrossRefPubMed
5.
go back to reference Sarikaya I, Povoski SP, Al-Saif OH, Kocak E, Bloomston M, Marsh S, Cao Z, Murrey DA, Zhang J, Hall NC, Knopp MV, Martin EW: Combined use of preoperative 18F FDG-PET imaging and intraoperative gamma probe detection for accurate assessment of tumor recurrence in patients with colorectal cancer. World J Surg Oncol. 2007, 5: 80-10.1186/1477-7819-5-80.CrossRefPubMedPubMedCentral Sarikaya I, Povoski SP, Al-Saif OH, Kocak E, Bloomston M, Marsh S, Cao Z, Murrey DA, Zhang J, Hall NC, Knopp MV, Martin EW: Combined use of preoperative 18F FDG-PET imaging and intraoperative gamma probe detection for accurate assessment of tumor recurrence in patients with colorectal cancer. World J Surg Oncol. 2007, 5: 80-10.1186/1477-7819-5-80.CrossRefPubMedPubMedCentral
6.
go back to reference Hall NC, Povoski SP, Murrey DA, Knopp MV, Martin EW: Combined approach of perioperative 18F-FDG PET/CT imaging and intraoperative 18F-FDG handheld gamma probe detection for tumor localization and verification of complete tumor resection in breast cancer. World J Surg Oncol. 2007, 5: 143-10.1186/1477-7819-5-143.CrossRefPubMedPubMedCentral Hall NC, Povoski SP, Murrey DA, Knopp MV, Martin EW: Combined approach of perioperative 18F-FDG PET/CT imaging and intraoperative 18F-FDG handheld gamma probe detection for tumor localization and verification of complete tumor resection in breast cancer. World J Surg Oncol. 2007, 5: 143-10.1186/1477-7819-5-143.CrossRefPubMedPubMedCentral
7.
go back to reference Sun D, Bloomston M, Hinkle G, Al-Saif OH, Hall NC, Povoski SP, Arnold MW, Martin EW: Radioimmunoguided surgery (RIGS), PET/CT image-guided surgery, and fluorescence image-guided surgery: past, present, and future. J Surg Oncol. 2007, 96: 297-308. 10.1002/jso.20869.CrossRefPubMed Sun D, Bloomston M, Hinkle G, Al-Saif OH, Hall NC, Povoski SP, Arnold MW, Martin EW: Radioimmunoguided surgery (RIGS), PET/CT image-guided surgery, and fluorescence image-guided surgery: past, present, and future. J Surg Oncol. 2007, 96: 297-308. 10.1002/jso.20869.CrossRefPubMed
8.
go back to reference Povoski SP, Hall NC, Martin EW, Walker MJ: Multimodality approach of perioperative 18F-FDG PET/CT imaging, intraoperative 18F-FDG handheld gamma probe detection, and intraoperative ultrasound for tumor localization and verification of resection of all sites of hypermetabolic activity in a case of occult recurrent metastatic melanoma. World J Surg Oncol. 2008, 6: 1-10.1186/1477-7819-6-1.CrossRefPubMedPubMedCentral Povoski SP, Hall NC, Martin EW, Walker MJ: Multimodality approach of perioperative 18F-FDG PET/CT imaging, intraoperative 18F-FDG handheld gamma probe detection, and intraoperative ultrasound for tumor localization and verification of resection of all sites of hypermetabolic activity in a case of occult recurrent metastatic melanoma. World J Surg Oncol. 2008, 6: 1-10.1186/1477-7819-6-1.CrossRefPubMedPubMedCentral
9.
go back to reference Cohn DE, Hall NC, Povoski SP, Seamon LG, Farrar WB, Martin EW: Novel perioperative imaging with 18F-FDG PET/CT and intraoperative 18F-FDG detection using a handheld gamma probe in recurrent ovarian cancer. Gynecol Oncol. 2008, 110: 152-157. 10.1016/j.ygyno.2008.04.026.CrossRefPubMed Cohn DE, Hall NC, Povoski SP, Seamon LG, Farrar WB, Martin EW: Novel perioperative imaging with 18F-FDG PET/CT and intraoperative 18F-FDG detection using a handheld gamma probe in recurrent ovarian cancer. Gynecol Oncol. 2008, 110: 152-157. 10.1016/j.ygyno.2008.04.026.CrossRefPubMed
10.
go back to reference Moffatt-Bruce SD, Povoski SP, Sharif S, Hall NC, Ross P, Johnson MA, Martin EW: A novel approach to positron emission tomography in lung cancer. Ann Thorac Surg. 2008, 86: 1355-1357. 10.1016/j.athoracsur.2008.03.072.CrossRefPubMed Moffatt-Bruce SD, Povoski SP, Sharif S, Hall NC, Ross P, Johnson MA, Martin EW: A novel approach to positron emission tomography in lung cancer. Ann Thorac Surg. 2008, 86: 1355-1357. 10.1016/j.athoracsur.2008.03.072.CrossRefPubMed
11.
go back to reference Agrawal A, Hall NC, Ringel MD, Povoski SP, Martin EW: Combined use of perioperative TSH-stimulated 18F-FDG PET/CT imaging and gamma probe radioguided surgery to localize and verify resection of iodine scan-negative recurrent thyroid carcinoma. Laryngoscope. 2008, 118: 2190-2194. 10.1097/MLG.0b013e3181845738.CrossRefPubMed Agrawal A, Hall NC, Ringel MD, Povoski SP, Martin EW: Combined use of perioperative TSH-stimulated 18F-FDG PET/CT imaging and gamma probe radioguided surgery to localize and verify resection of iodine scan-negative recurrent thyroid carcinoma. Laryngoscope. 2008, 118: 2190-2194. 10.1097/MLG.0b013e3181845738.CrossRefPubMed
12.
go back to reference Hall NC, Povoski SP, Murrey DA, Knopp MV, Martin EW: Bringing advanced medical imaging into the operative arena could revolutionize the surgical care of cancer patients. Expert Rev Med Devices. 2008, 5: 663-667. 10.1586/17434440.5.6.663.CrossRefPubMed Hall NC, Povoski SP, Murrey DA, Knopp MV, Martin EW: Bringing advanced medical imaging into the operative arena could revolutionize the surgical care of cancer patients. Expert Rev Med Devices. 2008, 5: 663-667. 10.1586/17434440.5.6.663.CrossRefPubMed
13.
go back to reference Frangioni JV: The problem is background, not signal. Mol Imaging. 2009, 8: 303-304.PubMed Frangioni JV: The problem is background, not signal. Mol Imaging. 2009, 8: 303-304.PubMed
14.
go back to reference Yasuda S, Makuuchi H, Fujii H, Nakasaki H, Mukai M, Sadahiro S, Tajima T, Ide M, Shohtsu A, Suzuki Y: Evaluation of a surgical gamma probe for detection of 18F-FDG. Tokai J Exp Clin Med. 2000, 25: 93-99.PubMed Yasuda S, Makuuchi H, Fujii H, Nakasaki H, Mukai M, Sadahiro S, Tajima T, Ide M, Shohtsu A, Suzuki Y: Evaluation of a surgical gamma probe for detection of 18F-FDG. Tokai J Exp Clin Med. 2000, 25: 93-99.PubMed
15.
go back to reference Yamamoto S, Higashi T, Matsumoto K, Senda M: Development of a positron-imaging detector with background rejection capability. Ann Nucl Med. 2006, 20: 655-662. 10.1007/BF02984676.CrossRefPubMed Yamamoto S, Higashi T, Matsumoto K, Senda M: Development of a positron-imaging detector with background rejection capability. Ann Nucl Med. 2006, 20: 655-662. 10.1007/BF02984676.CrossRefPubMed
16.
go back to reference Gulec SA, Daghighian F, Essner R: PET-Probe: Evaluation of Technical Performance and Clinical Utility of a Handheld High-Energy Gamma Probe in Oncologic Surgery. Ann Surg Oncol. 2006, Jul 24; [Epub ahead of print] Gulec SA, Daghighian F, Essner R: PET-Probe: Evaluation of Technical Performance and Clinical Utility of a Handheld High-Energy Gamma Probe in Oncologic Surgery. Ann Surg Oncol. 2006, Jul 24; [Epub ahead of print]
17.
go back to reference Gulec SA, Hoenie E, Hostetter R, Schwartzentruber D: PET probe-guided surgery: applications and clinical protocol. World J Surg Oncol. 2007, 5: 65-10.1186/1477-7819-5-65.CrossRefPubMedPubMedCentral Gulec SA, Hoenie E, Hostetter R, Schwartzentruber D: PET probe-guided surgery: applications and clinical protocol. World J Surg Oncol. 2007, 5: 65-10.1186/1477-7819-5-65.CrossRefPubMedPubMedCentral
18.
19.
go back to reference Piert M, Burian M, Meisetschläger G, Stein HJ, Ziegler S, Nährig J, Picchio M, Buck A, Siewert JR, Schwaiger M: Positron detection for the intraoperative localisation of cancer deposits. Eur J Nucl Med Mol Imaging. 2007, 34: 1534-1544. 10.1007/s00259-007-0430-5.CrossRefPubMedPubMedCentral Piert M, Burian M, Meisetschläger G, Stein HJ, Ziegler S, Nährig J, Picchio M, Buck A, Siewert JR, Schwaiger M: Positron detection for the intraoperative localisation of cancer deposits. Eur J Nucl Med Mol Imaging. 2007, 34: 1534-1544. 10.1007/s00259-007-0430-5.CrossRefPubMedPubMedCentral
20.
go back to reference Piert M, Carey J, Clinthorne N: Probe-guided localization of cancer deposits using [18F]fluorodeoxyglucose. Q J Nucl Med Mol Imaging. 2008, 52: 37-49.PubMed Piert M, Carey J, Clinthorne N: Probe-guided localization of cancer deposits using [18F]fluorodeoxyglucose. Q J Nucl Med Mol Imaging. 2008, 52: 37-49.PubMed
21.
go back to reference Martin EW, Chapman GJ, Subramaniam VV, Povoski SP: Intraoperative detection of gamma emissions using K-alpha X-ray fluorescence. Expert Rev Med Devices. 2010, 7: 431-434. 10.1586/erd.10.33.CrossRefPubMed Martin EW, Chapman GJ, Subramaniam VV, Povoski SP: Intraoperative detection of gamma emissions using K-alpha X-ray fluorescence. Expert Rev Med Devices. 2010, 7: 431-434. 10.1586/erd.10.33.CrossRefPubMed
22.
go back to reference Currie LA: Limits for qualitative detection and quantitative determination. Application to radiochemistry. Anal Chem. 1968, 40: 586-593. 10.1021/ac60259a007.CrossRef Currie LA: Limits for qualitative detection and quantitative determination. Application to radiochemistry. Anal Chem. 1968, 40: 586-593. 10.1021/ac60259a007.CrossRef
23.
go back to reference Thurston MO: Development of the gamma-detecting probe for radioimmunoguided surgery. Radioimmunoguided Surgery (RIGS) in the Detection and Treatment of Colorectal Cancer. Edited by: Martin EW. 1994, Austin: R.G. Landes Company, 41-65. 1 Thurston MO: Development of the gamma-detecting probe for radioimmunoguided surgery. Radioimmunoguided Surgery (RIGS) in the Detection and Treatment of Colorectal Cancer. Edited by: Martin EW. 1994, Austin: R.G. Landes Company, 41-65. 1
24.
go back to reference Martin EW, Thurston MO: The use of monoclonal antibodies (MAbs) and the development of an intraoperative hand-held probe for cancer detection. Cancer Invest. 1996, 14: 560-571. 10.3109/07357909609076901.CrossRefPubMed Martin EW, Thurston MO: The use of monoclonal antibodies (MAbs) and the development of an intraoperative hand-held probe for cancer detection. Cancer Invest. 1996, 14: 560-571. 10.3109/07357909609076901.CrossRefPubMed
25.
go back to reference Vardi Y, Shepp LA, Kaufman L: A statistical model for positron emission tomography. J Am Statist Assoc. 1985, 80: 8-20. 10.1080/01621459.1985.10477119.CrossRef Vardi Y, Shepp LA, Kaufman L: A statistical model for positron emission tomography. J Am Statist Assoc. 1985, 80: 8-20. 10.1080/01621459.1985.10477119.CrossRef
26.
go back to reference Turkheimer FE, Aston JA, Cunningham VJ: On the logic of hypothesis testing in functional imaging. Eur J Nucl Med Mol Imaging. 2004, 31: 725-732. 10.1007/s00259-003-1387-7.CrossRefPubMed Turkheimer FE, Aston JA, Cunningham VJ: On the logic of hypothesis testing in functional imaging. Eur J Nucl Med Mol Imaging. 2004, 31: 725-732. 10.1007/s00259-003-1387-7.CrossRefPubMed
27.
go back to reference Arnold MW, Young DC, Hitchcock CL, Schneebaum S, Martin EW: Radioimmunoguided surgery in primary colorectal carcinoma: an intraoperative prognostic tool and adjuvant to traditional staging. Am J Surg. 1995, 170: 315-318. 10.1016/S0002-9610(99)80295-5.CrossRefPubMed Arnold MW, Young DC, Hitchcock CL, Schneebaum S, Martin EW: Radioimmunoguided surgery in primary colorectal carcinoma: an intraoperative prognostic tool and adjuvant to traditional staging. Am J Surg. 1995, 170: 315-318. 10.1016/S0002-9610(99)80295-5.CrossRefPubMed
28.
go back to reference Bertsch DJ, Burak WE, Young DC, Arnold MW, Martin EW: Radioimmunoguided Surgery system improves survival for patients with recurrent colorectal cancer. Surgery. 1995, 118: 634-638. 10.1016/S0039-6060(05)80029-7. discussion 638–639CrossRefPubMed Bertsch DJ, Burak WE, Young DC, Arnold MW, Martin EW: Radioimmunoguided Surgery system improves survival for patients with recurrent colorectal cancer. Surgery. 1995, 118: 634-638. 10.1016/S0039-6060(05)80029-7. discussion 638–639CrossRefPubMed
29.
go back to reference Bertsch DJ, Burak WE, Young DC, Arnold MW, Martin EW: Radioimmunoguided surgery for colorectal cancer. Ann Surg Oncol. 1996, 3: 310-316. 10.1007/BF02306288.CrossRefPubMed Bertsch DJ, Burak WE, Young DC, Arnold MW, Martin EW: Radioimmunoguided surgery for colorectal cancer. Ann Surg Oncol. 1996, 3: 310-316. 10.1007/BF02306288.CrossRefPubMed
30.
go back to reference Arnold MW, Hitchcock CL, Young DC, Burak WE, Bertsch DJ, Martin EW: Intra-abdominal patterns of disease dissemination in colorectal cancer identified using radioimmunoguided surgery. Dis Colon Rectum. 1996, 39: 509-513. 10.1007/BF02058702.CrossRefPubMed Arnold MW, Hitchcock CL, Young DC, Burak WE, Bertsch DJ, Martin EW: Intra-abdominal patterns of disease dissemination in colorectal cancer identified using radioimmunoguided surgery. Dis Colon Rectum. 1996, 39: 509-513. 10.1007/BF02058702.CrossRefPubMed
31.
go back to reference Barbera-Guillem E, Arnold MW, Nelson MB, Martin EW: First results for resetting the antitumor immune response by immune corrective surgery in colon cancer. Am J Surg. 1998, 176: 339-343. 10.1016/S0002-9610(98)00192-5.CrossRefPubMed Barbera-Guillem E, Arnold MW, Nelson MB, Martin EW: First results for resetting the antitumor immune response by immune corrective surgery in colon cancer. Am J Surg. 1998, 176: 339-343. 10.1016/S0002-9610(98)00192-5.CrossRefPubMed
32.
go back to reference Knoll GF: Radiation Detection and Measurement. 2010, New York: John Wiley & Sons, 4 Knoll GF: Radiation Detection and Measurement. 2010, New York: John Wiley & Sons, 4
Metadata
Title
Intraoperative detection of 18F-FDG-avid tissue sites using the increased probe counting efficiency of the K-alpha probe design and variance-based statistical analysis with the three-sigma criteria
Authors
Stephen P Povoski
Gregg J Chapman
Douglas A Murrey
Robert Lee
Edward W Martin
Nathan C Hall
Publication date
01-12-2013
Publisher
BioMed Central
Published in
BMC Cancer / Issue 1/2013
Electronic ISSN: 1471-2407
DOI
https://doi.org/10.1186/1471-2407-13-98

Other articles of this Issue 1/2013

BMC Cancer 1/2013 Go to the issue
Webinar | 19-02-2024 | 17:30 (CET)

Keynote webinar | Spotlight on antibody–drug conjugates in cancer

Antibody–drug conjugates (ADCs) are novel agents that have shown promise across multiple tumor types. Explore the current landscape of ADCs in breast and lung cancer with our experts, and gain insights into the mechanism of action, key clinical trials data, existing challenges, and future directions.

Dr. Véronique Diéras
Prof. Fabrice Barlesi
Developed by: Springer Medicine