Open Access 01-12-2010 | Research article
Protective effect of rifampicin and clindamycin impregnated devices against Staphylococcusspp. infection after cerebrospinal fluid diversion procedures
Published in: BMC Neurology | Issue 1/2010
Login to get accessAbstract
Background
Infection is a major complication of cerebrospinal fluid shunting procedures. The present report assesses the efficacy of such catheters in both shunts and external ventricular drains (EVDs) against infection and particularly against Staphylococcus spp. infection.
Methods
All shunt and EVD procedures performed by means of antibiotic-impregnated catheters (AICs) and non-AICs during the period of study were registered. In cases of shunt procedures, a minimal follow-up of 90 days was considered, as well as de novo insertion and catheter revisions. Single valve revisions were not included. In cases of EVD procedures, those catheters removed before the fifth post-insertion day were not included. A total of 119 cerebrospinal fluid shunting procedures performed with AICs were studied in comparison with 112 procedures performed by means of non-AICs.
Results
Antibiotic-impregnated catheters were associated with a significant decrease in both overall and staphylococcal infection (p = 0.030 and p = 0.045, respectively). The number needed to treat for AICs was 8 to prevent one infection and 14 to prevent one staphylococcal infection. When comparing with shunts, the use of EVDs was associated with a 37-fold increased likelihood of infection.
Conclusions
Antibiotic-impregnated catheters are a safe and helpful tool to reduce CSF shunting device-related infections.