Published in: BMC Medical Imaging 1/2005

Open Access 01-12-2005 | Research article

Noise correlation in PET, CT, SPECT and PET/CT data evaluated using autocorrelation function: a phantom study on data, reconstructed using FBP and OSEM

Authors: Pasha Razifar, Mattias Sandström, Harald Schnieder, Bengt Långström, Enn Maripuu, Ewert Bengtsson, Mats Bergström

Published in: BMC Medical Imaging | Issue 1/2005

Login to get access

Abstract

Background

Positron Emission Tomography (PET), Computed Tomography (CT), PET/CT and Single Photon Emission Tomography (SPECT) are non-invasive imaging tools used for creating two dimensional (2D) cross section images of three dimensional (3D) objects. PET and SPECT have the potential of providing functional or biochemical information by measuring distribution and kinetics of radiolabelled molecules, whereas CT visualizes X-ray density in tissues in the body. PET/CT provides fused images representing both functional and anatomical information with better precision in localization than PET alone.
Images generated by these types of techniques are generally noisy, thereby impairing the imaging potential and affecting the precision in quantitative values derived from the images. It is crucial to explore and understand the properties of noise in these imaging techniques. Here we used autocorrelation function (ACF) specifically to describe noise correlation and its non-isotropic behaviour in experimentally generated images of PET, CT, PET/CT and SPECT.

Methods

Experiments were performed using phantoms with different shapes. In PET and PET/CT studies, data were acquired in 2D acquisition mode and reconstructed by both analytical filter back projection (FBP) and iterative, ordered subsets expectation maximisation (OSEM) methods. In the PET/CT studies, different magnitudes of X-ray dose in the transmission were employed by using different mA settings for the X-ray tube. In the CT studies, data were acquired using different slice thickness with and without applied dose reduction function and the images were reconstructed by FBP. SPECT studies were performed in 2D, reconstructed using FBP and OSEM, using post 3D filtering. ACF images were generated from the primary images, and profiles across the ACF images were used to describe the noise correlation in different directions. The variance of noise across the images was visualised as images and with profiles across these images.

Results

The most important finding was that the pattern of noise correlation is rotation symmetric or isotropic, independent of object shape in PET and PET/CT images reconstructed using the iterative method. This is, however, not the case in FBP images when the shape of phantom is not circular. Also CT images reconstructed using FBP show the same non-isotropic pattern independent of slice thickness and utilization of care dose function. SPECT images show an isotropic correlation of the noise independent of object shape or applied reconstruction algorithm. Noise in PET/CT images was identical independent of the applied X-ray dose in the transmission part (CT), indicating that the noise from transmission with the applied doses does not propagate into the PET images showing that the noise from the emission part is dominant. The results indicate that in human studies it is possible to utilize a low dose in transmission part while maintaining the noise behaviour and the quality of the images.

Conclusion

The combined effect of noise correlation for asymmetric objects and a varying noise variance across the image field significantly complicates the interpretation of the images when statistical methods are used, such as with statistical estimates of precision in average values, use of statistical parametric mapping methods and principal component analysis. Hence it is recommended that iterative reconstruction methods are used for such applications. However, it is possible to calculate the noise analytically in images reconstructed by FBP, while it is not possible to do the same calculation in images reconstructed by iterative methods. Therefore for performing statistical methods of analysis which depend on knowing the noise, FBP would be preferred.
Appendix
Available only for authorised users
Literature
1.
go back to reference Swindell W, Barrett HH: Computerized tomography: taking sectional x-rays. Phys Today. 1977, 30 (12): 32-41. Reprinted in Optics today, J. Howard, ed. (American Institute of Physics, New York, 1986)CrossRef Swindell W, Barrett HH: Computerized tomography: taking sectional x-rays. Phys Today. 1977, 30 (12): 32-41. Reprinted in Optics today, J. Howard, ed. (American Institute of Physics, New York, 1986)CrossRef
2.
go back to reference Townsend DW, Beyer T, Blodgett TM: PET/CT scanners: a hardware approach to image fusion. Semin Nucl Med. 2003, 33: 193-204. 10.1053/snuc.2003.127314.CrossRefPubMed Townsend DW, Beyer T, Blodgett TM: PET/CT scanners: a hardware approach to image fusion. Semin Nucl Med. 2003, 33: 193-204. 10.1053/snuc.2003.127314.CrossRefPubMed
3.
go back to reference Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, Jerin J, Young J, Byars L, Nutt R: A combined PET/CT scanner for clinical oncology. J Nucl Med. 2000, 41: 1369-1379.PubMed Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, Jerin J, Young J, Byars L, Nutt R: A combined PET/CT scanner for clinical oncology. J Nucl Med. 2000, 41: 1369-1379.PubMed
4.
go back to reference Beyer T, Townsend DW, Blodgett TM: Dual-modality PET/CT tomography for clinical oncology. Q J Nucl Med. 2002, 46: 24-34.PubMed Beyer T, Townsend DW, Blodgett TM: Dual-modality PET/CT tomography for clinical oncology. Q J Nucl Med. 2002, 46: 24-34.PubMed
5.
go back to reference Messa C, Bettinardi V, Picchio M, Pelosi E, Landoni C, Gianolli L, Gilardi MC, Fazio F: PET/CT in diagnostic oncology. Q J Nucl Med Mol Imaging. 2004, 48 (2): 66-75.PubMed Messa C, Bettinardi V, Picchio M, Pelosi E, Landoni C, Gianolli L, Gilardi MC, Fazio F: PET/CT in diagnostic oncology. Q J Nucl Med Mol Imaging. 2004, 48 (2): 66-75.PubMed
6.
go back to reference Hutton BF, Braun M, Thurfjell L, Lau DY: Image registration: an essential tool for nuclear medicine. Eur J Nucl Med Mol Imaging. 2002, 29: 559-577. 10.1007/s00259-001-0700-6.CrossRefPubMed Hutton BF, Braun M, Thurfjell L, Lau DY: Image registration: an essential tool for nuclear medicine. Eur J Nucl Med Mol Imaging. 2002, 29: 559-577. 10.1007/s00259-001-0700-6.CrossRefPubMed
7.
go back to reference Hudson HM, Larkin RS: Accelerated image reconstruction using Ordered Subsets of Projection data. IEEE Transactions on medical imaging. 1994, 13 (4): 601-609. 10.1109/42.363108.CrossRefPubMed Hudson HM, Larkin RS: Accelerated image reconstruction using Ordered Subsets of Projection data. IEEE Transactions on medical imaging. 1994, 13 (4): 601-609. 10.1109/42.363108.CrossRefPubMed
8.
go back to reference Brooks RA, Chiro GD: Principles of computer assisted tomography (CAT) in radiographic and radioisotopic imaging. Phys Med Biol. 1976, 21 (5): 689-732. 10.1088/0031-9155/21/5/001.CrossRefPubMed Brooks RA, Chiro GD: Principles of computer assisted tomography (CAT) in radiographic and radioisotopic imaging. Phys Med Biol. 1976, 21 (5): 689-732. 10.1088/0031-9155/21/5/001.CrossRefPubMed
9.
go back to reference Faulkner K, Moores BM: Analysis of x-ray computed tomography images using the noise power spectrum and autocorrelation function. Phys Med Biol. 1984, 29 (11): 1343-1352. 10.1088/0031-9155/29/11/003.CrossRefPubMed Faulkner K, Moores BM: Analysis of x-ray computed tomography images using the noise power spectrum and autocorrelation function. Phys Med Biol. 1984, 29 (11): 1343-1352. 10.1088/0031-9155/29/11/003.CrossRefPubMed
10.
go back to reference Hatton RL, Hutton BF, Angelides S, Choong KK, Larcos G: Improved tolerance to missing data in myocardial perfusion SPET using OSEM reconstruction. Eur J Nucl Med Mol Imaging. 2004, 31: 857-861. 10.1007/s00259-003-1450-4.CrossRefPubMed Hatton RL, Hutton BF, Angelides S, Choong KK, Larcos G: Improved tolerance to missing data in myocardial perfusion SPET using OSEM reconstruction. Eur J Nucl Med Mol Imaging. 2004, 31: 857-861. 10.1007/s00259-003-1450-4.CrossRefPubMed
11.
go back to reference Leong LK, Kruger RL, O'Connor MK: A comparison of the uniformity requirements for SPECT image reconstruction using FBP and OSEM techniques. J Nucl Med Technol. 2001, 29 (2): 79-83.PubMed Leong LK, Kruger RL, O'Connor MK: A comparison of the uniformity requirements for SPECT image reconstruction using FBP and OSEM techniques. J Nucl Med Technol. 2001, 29 (2): 79-83.PubMed
12.
go back to reference Pan TS, Lou DS, Kohli V, King MA: Influence of OSEM, elliptical orbits and background activity on SPECT 3D resolution recovery. Phys Med Biol. 1997, 42 (12): 2517-29. 10.1088/0031-9155/42/12/015.CrossRefPubMed Pan TS, Lou DS, Kohli V, King MA: Influence of OSEM, elliptical orbits and background activity on SPECT 3D resolution recovery. Phys Med Biol. 1997, 42 (12): 2517-29. 10.1088/0031-9155/42/12/015.CrossRefPubMed
13.
go back to reference Holm S, Toft P, Jensen M: Estimation of the noise contributions from Blank, Transmission and Emission scans in PET. IEEE trans Nucl Sci. 1996, 43 (4): 2285-2291. 10.1109/23.531893. Part 1CrossRef Holm S, Toft P, Jensen M: Estimation of the noise contributions from Blank, Transmission and Emission scans in PET. IEEE trans Nucl Sci. 1996, 43 (4): 2285-2291. 10.1109/23.531893. Part 1CrossRef
14.
go back to reference Metz CE, Beck RN: Quantitative effects of stationary linear image processing on noise and resolution of structure in radio-nuclide images. J Nucl Med. 1974, 15: 164-PubMed Metz CE, Beck RN: Quantitative effects of stationary linear image processing on noise and resolution of structure in radio-nuclide images. J Nucl Med. 1974, 15: 164-PubMed
15.
go back to reference Tsui BMW: Effects of the Recorder System on Spatial Resolution and Noise in the Nuclear Medicine. Dissertation. 1977, The University of Chicago, USA Tsui BMW: Effects of the Recorder System on Spatial Resolution and Noise in the Nuclear Medicine. Dissertation. 1977, The University of Chicago, USA
16.
go back to reference Alpert NM, Chesler DA, Correia JA, Ackerman RH, Chang JY, Finklestein S, Davis SM, Brownell GL, Taveras JM: Estimation of the Local Statistical Noise in Emission Computed Tomography. IEEE Trans Med Img, MI-1. 1982, 2: 142-146.CrossRef Alpert NM, Chesler DA, Correia JA, Ackerman RH, Chang JY, Finklestein S, Davis SM, Brownell GL, Taveras JM: Estimation of the Local Statistical Noise in Emission Computed Tomography. IEEE Trans Med Img, MI-1. 1982, 2: 142-146.CrossRef
17.
go back to reference Wilson DW, Tsui BMW: Noise properties of Filtered_Backprojection and ML-EM reconstructed Emission Tomographic Images. IEEE trans Nucl Sci. 1993, 40 (4): Wilson DW, Tsui BMW: Noise properties of Filtered_Backprojection and ML-EM reconstructed Emission Tomographic Images. IEEE trans Nucl Sci. 1993, 40 (4):
18.
go back to reference Riddell C, Carson RE, Carrasquillo JA, Libutti SK, Danforth DN, Whatley M, Bacharach SL: Noise Reduction in Oncology FDG PET Images by Iterative Reconstruction: A Quantitative Assessment. J Nucl Med. 2001, 42 (9): 1316-1323.PubMed Riddell C, Carson RE, Carrasquillo JA, Libutti SK, Danforth DN, Whatley M, Bacharach SL: Noise Reduction in Oncology FDG PET Images by Iterative Reconstruction: A Quantitative Assessment. J Nucl Med. 2001, 42 (9): 1316-1323.PubMed
19.
go back to reference Bergström M: Performance Evaluation and Improvements of Quantitation Accuracy in Transmission and Positron Emission Computer Assisted Tomography. PhD thesis. 1982, Department of Radiation Physics, University of Stockholm, Sweden Bergström M: Performance Evaluation and Improvements of Quantitation Accuracy in Transmission and Positron Emission Computer Assisted Tomography. PhD thesis. 1982, Department of Radiation Physics, University of Stockholm, Sweden
20.
go back to reference Greess H, Wolf H, Baum U, Lell M, Pirkl M, Kalender WA, Bautz WA: Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol. 2000, 10: 391-394. 10.1007/s003300050062.CrossRefPubMed Greess H, Wolf H, Baum U, Lell M, Pirkl M, Kalender WA, Bautz WA: Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol. 2000, 10: 391-394. 10.1007/s003300050062.CrossRefPubMed
21.
go back to reference Gies M, Kalender WA, Wolf H, Suess C: Dose reduction in CT by anatomically adapted tube current modulation. I Simulation studies. Med Phys. 1999, 26: 2235-2247. 10.1118/1.598779.CrossRefPubMed Gies M, Kalender WA, Wolf H, Suess C: Dose reduction in CT by anatomically adapted tube current modulation. I Simulation studies. Med Phys. 1999, 26: 2235-2247. 10.1118/1.598779.CrossRefPubMed
22.
go back to reference Greess H, Nomayr A, Wolf H, Baum U, Lell M, Bowing B, Kalender WA, Bautz WA: Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose). Eur Radiol. 2002, 12: 1571-1576. 10.1007/s00330-001-1255-4.CrossRefPubMed Greess H, Nomayr A, Wolf H, Baum U, Lell M, Bowing B, Kalender WA, Bautz WA: Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose). Eur Radiol. 2002, 12: 1571-1576. 10.1007/s00330-001-1255-4.CrossRefPubMed
23.
go back to reference Kalender WA, Wolf H, Suess C, Gies M, Greess H, Bautz WA: Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol. 1999, 9: 323-328. 10.1007/s003300050674.CrossRefPubMed Kalender WA, Wolf H, Suess C, Gies M, Greess H, Bautz WA: Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol. 1999, 9: 323-328. 10.1007/s003300050674.CrossRefPubMed
24.
go back to reference Razifar P, Lubberink M, Schneider H, Langstrom B, Bengtsson E, Bergstrom M: Non-isotropic noise correlation in PET data reconstructed by FBP but not by OSEM demonstrated using auto-correlation function. BMC Medical Imaging. 2005, 5: 3-10.1186/1471-2342-5-3.CrossRefPubMedPubMedCentral Razifar P, Lubberink M, Schneider H, Langstrom B, Bengtsson E, Bergstrom M: Non-isotropic noise correlation in PET data reconstructed by FBP but not by OSEM demonstrated using auto-correlation function. BMC Medical Imaging. 2005, 5: 3-10.1186/1471-2342-5-3.CrossRefPubMedPubMedCentral
25.
go back to reference Brix G, Zaers J, Adam LE, Bellemann ME, Ostertag H, Trojan H, Haberkorn U, Doll J, Oberorfer F, Lorenz WJ: Performance evaluation of a whole-body PET scanner using the NEMA protocol. J Nucl Med. 1997, 38: 1614-1623.PubMed Brix G, Zaers J, Adam LE, Bellemann ME, Ostertag H, Trojan H, Haberkorn U, Doll J, Oberorfer F, Lorenz WJ: Performance evaluation of a whole-body PET scanner using the NEMA protocol. J Nucl Med. 1997, 38: 1614-1623.PubMed
26.
go back to reference Bettinardi V, Danna M, Savi A, Lecchi M, Castiglioni I, Gilardi MC, Bammer H, Lucignani G, Fazio F: Performance evaluation of the new whole-body PET/CT scanner: Discovery ST. Eur J Nucl Med Mol Imaging. 2004, 31 (6): 867-81. 10.1007/s00259-003-1444-2.CrossRefPubMed Bettinardi V, Danna M, Savi A, Lecchi M, Castiglioni I, Gilardi MC, Bammer H, Lucignani G, Fazio F: Performance evaluation of the new whole-body PET/CT scanner: Discovery ST. Eur J Nucl Med Mol Imaging. 2004, 31 (6): 867-81. 10.1007/s00259-003-1444-2.CrossRefPubMed
27.
go back to reference Knapp FFJ, Brihaye C, Callahan AP: Generators, Principles of Nuclear Medicine. Edited by: Wagner HN, Szabo Z, Buchanan JW. 1995, WB Saunders: Philadelphia, 150-165. Knapp FFJ, Brihaye C, Callahan AP: Generators, Principles of Nuclear Medicine. Edited by: Wagner HN, Szabo Z, Buchanan JW. 1995, WB Saunders: Philadelphia, 150-165.
28.
go back to reference NEMA NU-2001: Performance standards of positron emission tomographs. National Electronics Manufacturers Association NEMA NU-2001: Performance standards of positron emission tomographs. National Electronics Manufacturers Association
29.
go back to reference XU M, Cutler PD, Luk WK: Adaptive segmented attenuation correction for whole-body PET imaging. IEEE Trans Nucl Sci. 1996, 43: 331-337. 10.1109/23.485974.CrossRef XU M, Cutler PD, Luk WK: Adaptive segmented attenuation correction for whole-body PET imaging. IEEE Trans Nucl Sci. 1996, 43: 331-337. 10.1109/23.485974.CrossRef
30.
go back to reference Kamel E, Hany TF, Burger C, Treyer V, Lonn AHR, Schulthess GKV, Buck A: CT vs 68Ge attenuation correction in a combined PET/CT system- evaluation of the effect of lowering the CT tube current. Eur J Nucl Med Mol Imaging. 2002, 29: 346-350. 10.1007/s00259-001-0698-9.CrossRefPubMed Kamel E, Hany TF, Burger C, Treyer V, Lonn AHR, Schulthess GKV, Buck A: CT vs 68Ge attenuation correction in a combined PET/CT system- evaluation of the effect of lowering the CT tube current. Eur J Nucl Med Mol Imaging. 2002, 29: 346-350. 10.1007/s00259-001-0698-9.CrossRefPubMed
Metadata
Title
Noise correlation in PET, CT, SPECT and PET/CT data evaluated using autocorrelation function: a phantom study on data, reconstructed using FBP and OSEM
Authors
Pasha Razifar
Mattias Sandström
Harald Schnieder
Bengt Långström
Enn Maripuu
Ewert Bengtsson
Mats Bergström
Publication date
01-12-2005
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2005
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/1471-2342-5-5

Other articles of this Issue 1/2005

BMC Medical Imaging 1/2005 Go to the issue