Published in: BMC Medical Imaging 1/2012

Open Access 01-12-2012 | Research article

Is single reading with computer-aided detection (CAD) as good as double reading in mammography screening? A systematic review

Authors: Edward Azavedo, Sophia Zackrisson, Ingegerd Mejàre, Marianne Heibert Arnlind

Published in: BMC Medical Imaging | Issue 1/2012

Login to get access

Abstract

Background

In accordance with European guidelines, mammography screening comprises independent readings by two breast radiologists (double reading). CAD (computer-aided detection) has been suggested to complement or replace one of the two readers (single reading + CAD).
The aim of this systematic review is to address the following question: Is the reading of mammographic x-ray images by a single breast radiologist together with CAD at least as accurate as double reading?

Methods

The electronic literature search included the databases Pub Med, EMBASE and The Cochrane Library. Two independent reviewers assessed abstracts and full-text articles.

Results

1049 abstracts were identified, of which 996 were excluded with reference to inclusion and exclusion criteria; 53 full-text articles were assessed for eligibility. Finally, four articles were included in the qualitative analysis, and one in a GRADE synthesis.

Conclusions

The scientific evidence is insufficient to determine whether the accuracy of single reading + CAD is at least equivalent to that obtained in standard practice, i.e. double reading where two breast radiologists independently read the mammographic images.
Appendix
Available only for authorised users
Literature
1.
go back to reference Tabar L, Fagerberg CJ, Gad A, Baldetorp L, Holmberg LH, Grontoft O, Ljungquist U, Lundstrom B, Manson JC, Eklund G, et al: Reduction in mortality from breast cancer after mass screening with mammography. Randomised trial from the Breast Cancer Screening Working Group of the Swedish National Board of Health and Welfare. Lancet. 1985, 1 (8433): 829-832.CrossRefPubMed Tabar L, Fagerberg CJ, Gad A, Baldetorp L, Holmberg LH, Grontoft O, Ljungquist U, Lundstrom B, Manson JC, Eklund G, et al: Reduction in mortality from breast cancer after mass screening with mammography. Randomised trial from the Breast Cancer Screening Working Group of the Swedish National Board of Health and Welfare. Lancet. 1985, 1 (8433): 829-832.CrossRefPubMed
2.
go back to reference Andersson I, Aspegren K, Janzon L, Landberg T, Lindholm K, Linell F, Ljungberg O, Ranstam J, Sigfusson B: Mammographic screening and mortality from breast cancer: the Malmo mammographic screening trial. BMJ. 1988, 297 (6654): 943-948. 10.1136/bmj.297.6654.943.CrossRefPubMedPubMedCentral Andersson I, Aspegren K, Janzon L, Landberg T, Lindholm K, Linell F, Ljungberg O, Ranstam J, Sigfusson B: Mammographic screening and mortality from breast cancer: the Malmo mammographic screening trial. BMJ. 1988, 297 (6654): 943-948. 10.1136/bmj.297.6654.943.CrossRefPubMedPubMedCentral
3.
go back to reference Bjurstam N, Bjorneld L, Duffy SW, Smith TC, Cahlin E, Eriksson O, Hafstrom LO, Lingaas H, Mattsson J, Persson S, et al: The Gothenburg breast screening trial: first results on mortality, incidence, and mode of detection for women ages 39–49 years at randomization. Cancer. 1997, 80 (11): 2091-2099. 10.1002/(SICI)1097-0142(19971201)80:11<2091::AID-CNCR8>3.0.CO;2-#.CrossRefPubMed Bjurstam N, Bjorneld L, Duffy SW, Smith TC, Cahlin E, Eriksson O, Hafstrom LO, Lingaas H, Mattsson J, Persson S, et al: The Gothenburg breast screening trial: first results on mortality, incidence, and mode of detection for women ages 39–49 years at randomization. Cancer. 1997, 80 (11): 2091-2099. 10.1002/(SICI)1097-0142(19971201)80:11<2091::AID-CNCR8>3.0.CO;2-#.CrossRefPubMed
4.
go back to reference Frisell J, Glas U, Hellstrom L, Somell A: Randomized mammographic screening for breast cancer in Stockholm. Design, first round results and comparisons. Breast Cancer Res Treat. 1986, 8 (1): 45-54. 10.1007/BF01805924.CrossRefPubMed Frisell J, Glas U, Hellstrom L, Somell A: Randomized mammographic screening for breast cancer in Stockholm. Design, first round results and comparisons. Breast Cancer Res Treat. 1986, 8 (1): 45-54. 10.1007/BF01805924.CrossRefPubMed
5.
go back to reference Vainio H, Bianchini F: IARC Handbooks of Cancer Prevention- Breast Cancer Screening. 2002, IARCPress, Lyon, France, 2002- Vainio H, Bianchini F: IARC Handbooks of Cancer Prevention- Breast Cancer Screening. 2002, IARCPress, Lyon, France, 2002-
6.
go back to reference Nystrom L, Andersson I, Bjurstam N, Frisell J, Nordenskjold B, Rutqvist LE: Long-term effects of mammography screening: updated overview of the Swedish randomised trials. Lancet. 2002, 359 (9310): 909-919. 10.1016/S0140-6736(02)08020-0.CrossRefPubMed Nystrom L, Andersson I, Bjurstam N, Frisell J, Nordenskjold B, Rutqvist LE: Long-term effects of mammography screening: updated overview of the Swedish randomised trials. Lancet. 2002, 359 (9310): 909-919. 10.1016/S0140-6736(02)08020-0.CrossRefPubMed
7.
go back to reference Tabar L, Vitak B, Chen TH, Yen AM, Cohen A, Tot T, Chiu SY, Chen SL, Fann JC, Rosell J, et al: Swedish two-county trial: impact of mammographic screening on breast cancer mortality during 3 decades. Radiology. 2011, 260 (3): 658-663. 10.1148/radiol.11110469.CrossRefPubMed Tabar L, Vitak B, Chen TH, Yen AM, Cohen A, Tot T, Chiu SY, Chen SL, Fann JC, Rosell J, et al: Swedish two-county trial: impact of mammographic screening on breast cancer mortality during 3 decades. Radiology. 2011, 260 (3): 658-663. 10.1148/radiol.11110469.CrossRefPubMed
8.
go back to reference Hellquist BN, Duffy SW, Abdsaleh S, Bjorneld L, Bordas P, Tabar L, Vitak B, Zackrisson S, Nystrom L, Jonsson H: Effectiveness of population-based service screening with mammography for women ages 40 to 49 years: evaluation of the Swedish Mammography Screening in Young Women (SCRY) cohort. Cancer. 2011, 117 (4): 714-722. 10.1002/cncr.25650.CrossRefPubMed Hellquist BN, Duffy SW, Abdsaleh S, Bjorneld L, Bordas P, Tabar L, Vitak B, Zackrisson S, Nystrom L, Jonsson H: Effectiveness of population-based service screening with mammography for women ages 40 to 49 years: evaluation of the Swedish Mammography Screening in Young Women (SCRY) cohort. Cancer. 2011, 117 (4): 714-722. 10.1002/cncr.25650.CrossRefPubMed
9.
go back to reference SOSSEG SOSSEG: Reduction in breast cancer mortality from organized service screening with mammography: 1. Further confirmation with extended data. Cancer Epidemiol Biomarkers Prev. 2006, 15 (1): 45-51.CrossRef SOSSEG SOSSEG: Reduction in breast cancer mortality from organized service screening with mammography: 1. Further confirmation with extended data. Cancer Epidemiol Biomarkers Prev. 2006, 15 (1): 45-51.CrossRef
10.
go back to reference Gotzsche PC, Nielsen M: Screening for breast cancer with mammography. Cochrane Database Syst Rev. 2011, CD001877-1 Gotzsche PC, Nielsen M: Screening for breast cancer with mammography. Cochrane Database Syst Rev. 2011, CD001877-1
11.
go back to reference Perry N, Broeders M, de Wolf C, Törnberg S, Holland R, von Karsa L: European guidelines for quality assurance in breast cancer screening and diagnosis. 2006, Office for Official Publications of the European Communities, Luxembourg, ISBN 2006 ISBN 92-79-01258-4: EU, 4 Perry N, Broeders M, de Wolf C, Törnberg S, Holland R, von Karsa L: European guidelines for quality assurance in breast cancer screening and diagnosis. 2006, Office for Official Publications of the European Communities, Luxembourg, ISBN 2006 ISBN 92-79-01258-4: EU, 4
12.
go back to reference Helvie M: Improving mammographic interpretation: double reading and computer-aided diagnosis. Radiol Clin North Am. 2007, 45 (5): 801-811. 10.1016/j.rcl.2007.06.004. viCrossRefPubMed Helvie M: Improving mammographic interpretation: double reading and computer-aided diagnosis. Radiol Clin North Am. 2007, 45 (5): 801-811. 10.1016/j.rcl.2007.06.004. viCrossRefPubMed
13.
go back to reference Guerriero C, Gillan MG, Cairns J, Wallis MG, Gilbert FJ: Is computer aided detection (CAD) cost effective in screening mammography? A model based on the CADET II study. BMC Health Serv Res. 2011, 11 (1): 11-10.1186/1472-6963-11-11.CrossRefPubMedPubMedCentral Guerriero C, Gillan MG, Cairns J, Wallis MG, Gilbert FJ: Is computer aided detection (CAD) cost effective in screening mammography? A model based on the CADET II study. BMC Health Serv Res. 2011, 11 (1): 11-10.1186/1472-6963-11-11.CrossRefPubMedPubMedCentral
14.
go back to reference Taylor P, Champness J, Given-Wilson R, Johnston K, Potts H: Impact of computer-aided detection prompts on the sensitivity and specificity of screening mammography. Health Technol Assess. 2005, 9 (6): 1-58. iiiCrossRefPubMed Taylor P, Champness J, Given-Wilson R, Johnston K, Potts H: Impact of computer-aided detection prompts on the sensitivity and specificity of screening mammography. Health Technol Assess. 2005, 9 (6): 1-58. iiiCrossRefPubMed
15.
go back to reference Houssami N, Given-Wilson R: Incorporating new technologies into clinical practice without evidence of effectiveness in prospective studies: computer-aided detection (CAD) in breast screening reinforces the need for better initial evaluation. Breast. 2007, 16 (3): 219-221. 10.1016/j.breast.2007.04.006.CrossRefPubMed Houssami N, Given-Wilson R: Incorporating new technologies into clinical practice without evidence of effectiveness in prospective studies: computer-aided detection (CAD) in breast screening reinforces the need for better initial evaluation. Breast. 2007, 16 (3): 219-221. 10.1016/j.breast.2007.04.006.CrossRefPubMed
16.
go back to reference Houssami N, Given-Wilson R, Ciatto S: Early detection of breast cancer: overview of the evidence on computer-aided detection in mammography screening. J Med Imaging Radiat Oncol. 2009, 53 (2): 171-176. 10.1111/j.1754-9485.2009.02062.x.CrossRefPubMed Houssami N, Given-Wilson R, Ciatto S: Early detection of breast cancer: overview of the evidence on computer-aided detection in mammography screening. J Med Imaging Radiat Oncol. 2009, 53 (2): 171-176. 10.1111/j.1754-9485.2009.02062.x.CrossRefPubMed
17.
go back to reference Taylor P, Potts HW: Computer aids and human second reading as interventions in screening mammography: two systematic reviews to compare effects on cancer detection and recall rate. Eur J Cancer. 2008, 44 (6): 798-807. 10.1016/j.ejca.2008.02.016.CrossRefPubMed Taylor P, Potts HW: Computer aids and human second reading as interventions in screening mammography: two systematic reviews to compare effects on cancer detection and recall rate. Eur J Cancer. 2008, 44 (6): 798-807. 10.1016/j.ejca.2008.02.016.CrossRefPubMed
18.
go back to reference Bennett RL, Blanks RG, Moss SM: Does the accuracy of single reading with CAD (computer-aided detection) compare with that of double reading?: A review of the literature. Clin Radiol. 2006, 61 (12): 1023-1028. 10.1016/j.crad.2006.09.006.CrossRefPubMed Bennett RL, Blanks RG, Moss SM: Does the accuracy of single reading with CAD (computer-aided detection) compare with that of double reading?: A review of the literature. Clin Radiol. 2006, 61 (12): 1023-1028. 10.1016/j.crad.2006.09.006.CrossRefPubMed
19.
go back to reference SBU: Computer-Aided Detection (CAD) in Mammography Screening. 2011, Statens beredning för medicinsk utvärdering (SBU): SBU, Stockholm SBU: Computer-Aided Detection (CAD) in Mammography Screening. 2011, Statens beredning för medicinsk utvärdering (SBU): SBU, Stockholm
20.
go back to reference Azavedo E, Svane G: Radiologic aspects of breast cancers detected through a breast cancer screening program. Eur J Radiol. 1991, 13 (2): 88-90. 10.1016/0720-048X(91)90086-B.CrossRefPubMed Azavedo E, Svane G: Radiologic aspects of breast cancers detected through a breast cancer screening program. Eur J Radiol. 1991, 13 (2): 88-90. 10.1016/0720-048X(91)90086-B.CrossRefPubMed
21.
go back to reference Whiting P, Rutjes AW, Reitsma JB, Bossuyt PM, Kleijnen J: The development of QUADAS: a tool for the quality assessment of studies of diagnostic accuracy included in systematic reviews. BMC Med Res Methodol. 2003, 3: 25-10.1186/1471-2288-3-25.CrossRefPubMedPubMedCentral Whiting P, Rutjes AW, Reitsma JB, Bossuyt PM, Kleijnen J: The development of QUADAS: a tool for the quality assessment of studies of diagnostic accuracy included in systematic reviews. BMC Med Res Methodol. 2003, 3: 25-10.1186/1471-2288-3-25.CrossRefPubMedPubMedCentral
22.
go back to reference Guyatt GH, Oxman AD, Vist GE, Kunz R, Falck-Ytter Y, Alonso-Coello P, Schunemann HJ: GRADE: an emerging consensus on rating quality of evidence and strength of recommendations. BMJ. 2008, 336 (7650): 924-926. 10.1136/bmj.39489.470347.AD.CrossRefPubMedPubMedCentral Guyatt GH, Oxman AD, Vist GE, Kunz R, Falck-Ytter Y, Alonso-Coello P, Schunemann HJ: GRADE: an emerging consensus on rating quality of evidence and strength of recommendations. BMJ. 2008, 336 (7650): 924-926. 10.1136/bmj.39489.470347.AD.CrossRefPubMedPubMedCentral
23.
go back to reference Schunemann HJ, Oxman AD, Brozek J, Glasziou P, Bossuyt P, Chang S, Muti P, Jaeschke R, Guyatt GH: GRADE: assessing the quality of evidence for diagnostic recommendations. Evid Based Med. 2008, 13 (6): 162-163.CrossRefPubMed Schunemann HJ, Oxman AD, Brozek J, Glasziou P, Bossuyt P, Chang S, Muti P, Jaeschke R, Guyatt GH: GRADE: assessing the quality of evidence for diagnostic recommendations. Evid Based Med. 2008, 13 (6): 162-163.CrossRefPubMed
24.
go back to reference Boyer B, Balleyguier C, Granat O, Pharaboz C: CAD in questions/answers Review of the literature. Eur J Radiol. 2009, 69 (1): 24-33. 10.1016/j.ejrad.2008.07.042.CrossRefPubMed Boyer B, Balleyguier C, Granat O, Pharaboz C: CAD in questions/answers Review of the literature. Eur J Radiol. 2009, 69 (1): 24-33. 10.1016/j.ejrad.2008.07.042.CrossRefPubMed
25.
go back to reference G-S G, Chersevani R, Ciatto S, Del Favero C, Frigerio A, Giordano L, Giuseppetti G, Naldoni C, Panizza P, Petrella M, Chersevani R, Ciatto S, Del Favero C, Frigerio A, Giordano L, Giuseppetti G, Naldoni C, Panizza P, Petrella M, Gruppo di studio G-S, et al: "CADEAT": considerations on the use of CAD (computer-aided diagnosis) in mammography. La Radiologia medica. 2010, 115 (4): 563-570. 10.1007/s11547-010-0505-4.CrossRef G-S G, Chersevani R, Ciatto S, Del Favero C, Frigerio A, Giordano L, Giuseppetti G, Naldoni C, Panizza P, Petrella M, Chersevani R, Ciatto S, Del Favero C, Frigerio A, Giordano L, Giuseppetti G, Naldoni C, Panizza P, Petrella M, Gruppo di studio G-S, et al: "CADEAT": considerations on the use of CAD (computer-aided diagnosis) in mammography. La Radiologia medica. 2010, 115 (4): 563-570. 10.1007/s11547-010-0505-4.CrossRef
26.
go back to reference Kolb GR: The financial impact of computer-aided detection on the mammography practice. Applied Radiology. 2001, 30 (11 SUPPL.): 21-24. Kolb GR: The financial impact of computer-aided detection on the mammography practice. Applied Radiology. 2001, 30 (11 SUPPL.): 21-24.
27.
go back to reference Nishikawa RM: Evaluation of computer-aided detection and computer detection systems. Applied Radiology. 2001, 30 (11 SUPPL.): 14-16. Nishikawa RM: Evaluation of computer-aided detection and computer detection systems. Applied Radiology. 2001, 30 (11 SUPPL.): 14-16.
28.
go back to reference Noble M, Bruening W, Uhl S, Schoelles K: Computer-aided detection mammography for breast cancer screening: systematic review and meta-analysis. Arch Gynecol Obstet. 2009, 279 (6): 881-890. 10.1007/s00404-008-0841-y.CrossRefPubMed Noble M, Bruening W, Uhl S, Schoelles K: Computer-aided detection mammography for breast cancer screening: systematic review and meta-analysis. Arch Gynecol Obstet. 2009, 279 (6): 881-890. 10.1007/s00404-008-0841-y.CrossRefPubMed
29.
go back to reference Cawson JN, Nickson C, Amos A, Hill G, Whan AB, Kavanagh AM: Invasive breast cancers detected by screening mammography: a detailed comparison of computer-aided detection-assisted single reading and double reading. J Med Imaging Radiat Oncol. 2009, 53 (5): 442-449. 10.1111/j.1754-9485.2009.02100.x.CrossRefPubMed Cawson JN, Nickson C, Amos A, Hill G, Whan AB, Kavanagh AM: Invasive breast cancers detected by screening mammography: a detailed comparison of computer-aided detection-assisted single reading and double reading. J Med Imaging Radiat Oncol. 2009, 53 (5): 442-449. 10.1111/j.1754-9485.2009.02100.x.CrossRefPubMed
30.
go back to reference van den Biggelaar FJ, Kessels AG, van Engelshoven JM, Flobbe K: Strategies for digital mammography interpretation in a clinical patient population. Int J Cancer. 2009, 125 (12): 2923-2929. 10.1002/ijc.24632.CrossRefPubMed van den Biggelaar FJ, Kessels AG, van Engelshoven JM, Flobbe K: Strategies for digital mammography interpretation in a clinical patient population. Int J Cancer. 2009, 125 (12): 2923-2929. 10.1002/ijc.24632.CrossRefPubMed
31.
go back to reference Paquerault S, Samuelson FW, Petrick N, Myers KJ, Smith RC: Investigation of reading mode and relative sensitivity as factors that influence reader performance when using computer-aided detection software. Acad Radiol. 2009, 16 (9): 1095-1107. 10.1016/j.acra.2009.03.024.CrossRefPubMed Paquerault S, Samuelson FW, Petrick N, Myers KJ, Smith RC: Investigation of reading mode and relative sensitivity as factors that influence reader performance when using computer-aided detection software. Acad Radiol. 2009, 16 (9): 1095-1107. 10.1016/j.acra.2009.03.024.CrossRefPubMed
32.
go back to reference van den Biggelaar FJ, Kessels AG, van Engelshoven JM, Boetes C, Flobbe K: Computer-aided detection in full-field digital mammography in a clinical population: performance of radiologist and technologists. Breast Cancer Res Treat. 2010, 120 (2): 499-506. 10.1007/s10549-009-0409-y.CrossRefPubMed van den Biggelaar FJ, Kessels AG, van Engelshoven JM, Boetes C, Flobbe K: Computer-aided detection in full-field digital mammography in a clinical population: performance of radiologist and technologists. Breast Cancer Res Treat. 2010, 120 (2): 499-506. 10.1007/s10549-009-0409-y.CrossRefPubMed
33.
go back to reference James JJ, Cornford EJ: Does computer-aided detection have a role in the arbitration of discordant double-reading opinions in a breast-screening programme?. Clin Radiol. 2009, 64 (1): 46-51. 10.1016/j.crad.2008.07.008.CrossRefPubMed James JJ, Cornford EJ: Does computer-aided detection have a role in the arbitration of discordant double-reading opinions in a breast-screening programme?. Clin Radiol. 2009, 64 (1): 46-51. 10.1016/j.crad.2008.07.008.CrossRefPubMed
34.
go back to reference Kim SJ, Moon WK, Cho N, Cha JH, Kim SM, Im JG: Computer-aided detection in full-field digital mammography: sensitivity and reproducibility in serial examinations. Radiology. 2008, 246 (1): 71-80.CrossRefPubMed Kim SJ, Moon WK, Cho N, Cha JH, Kim SM, Im JG: Computer-aided detection in full-field digital mammography: sensitivity and reproducibility in serial examinations. Radiology. 2008, 246 (1): 71-80.CrossRefPubMed
35.
go back to reference Brancato B, Houssami N, Francesca D, Bianchi S, Risso G, Catarzi S, Taschini R, Rosselli Del Turco M, Ciatto S: Does computer-aided detection (CAD) contribute to the performance of digital mammography in a self-referred population?. Breast Cancer Res Treat. 2008, 111 (2): 373-376. 10.1007/s10549-007-9786-2.CrossRefPubMed Brancato B, Houssami N, Francesca D, Bianchi S, Risso G, Catarzi S, Taschini R, Rosselli Del Turco M, Ciatto S: Does computer-aided detection (CAD) contribute to the performance of digital mammography in a self-referred population?. Breast Cancer Res Treat. 2008, 111 (2): 373-376. 10.1007/s10549-007-9786-2.CrossRefPubMed
36.
go back to reference Yang SK, Moon WK, Cho N, Park JS, Cha JH, Kim SM, Kim SJ, Im JG: Screening mammography-detected cancers: sensitivity of a computer-aided detection system applied to full-field digital mammograms. Radiology. 2007, 244 (1): 104-111. 10.1148/radiol.2441060756.CrossRefPubMed Yang SK, Moon WK, Cho N, Park JS, Cha JH, Kim SM, Kim SJ, Im JG: Screening mammography-detected cancers: sensitivity of a computer-aided detection system applied to full-field digital mammograms. Radiology. 2007, 244 (1): 104-111. 10.1148/radiol.2441060756.CrossRefPubMed
37.
go back to reference Skaane P, Kshirsagar A, Stapleton S, Young K, Castellino RA: Effect of computer-aided detection on independent double reading of paired screen-film and full-field digital screening mammograms. AJR Am J Roentgenol. 2007, 188 (2): 377-384. 10.2214/AJR.05.2207.CrossRefPubMed Skaane P, Kshirsagar A, Stapleton S, Young K, Castellino RA: Effect of computer-aided detection on independent double reading of paired screen-film and full-field digital screening mammograms. AJR Am J Roentgenol. 2007, 188 (2): 377-384. 10.2214/AJR.05.2207.CrossRefPubMed
38.
go back to reference Gilbert FJ, Astley SM, McGee MA, Gillan MG, Boggis CR, Griffiths PM, Duffy SW: Single reading with computer-aided detection and double reading of screening mammograms in the United Kingdom National Breast Screening Program. Radiology. 2006, 241 (1): 47-53. 10.1148/radiol.2411051092.CrossRefPubMed Gilbert FJ, Astley SM, McGee MA, Gillan MG, Boggis CR, Griffiths PM, Duffy SW: Single reading with computer-aided detection and double reading of screening mammograms in the United Kingdom National Breast Screening Program. Radiology. 2006, 241 (1): 47-53. 10.1148/radiol.2411051092.CrossRefPubMed
39.
go back to reference Dean JC, Ilvento CC: Improved cancer detection using computer-aided detection with diagnostic and screening mammography: prospective study of 104 cancers. AJR Am J Roentgenol. 2006, 187 (1): 20-28. 10.2214/AJR.05.0111.CrossRefPubMed Dean JC, Ilvento CC: Improved cancer detection using computer-aided detection with diagnostic and screening mammography: prospective study of 104 cancers. AJR Am J Roentgenol. 2006, 187 (1): 20-28. 10.2214/AJR.05.0111.CrossRefPubMed
40.
go back to reference Hukkinen K, Vehmas T, Pamilo M, Kivisaari L: Effect of computer-aided detection on mammographic performance: experimental study on readers with different levels of experience. Acta Radiol. 2006, 47 (3): 257-263. 10.1080/02841850500539025.CrossRefPubMed Hukkinen K, Vehmas T, Pamilo M, Kivisaari L: Effect of computer-aided detection on mammographic performance: experimental study on readers with different levels of experience. Acta Radiol. 2006, 47 (3): 257-263. 10.1080/02841850500539025.CrossRefPubMed
41.
go back to reference Ciatto S, Ambrogetti D, Collini G, Cruciani A, Ercolini E, Risso G, Rosselli Del Turco M: Computer-aided detection (CAD) of cancers detected on double reading by one reader only. Breast. 2006, 15 (4): 528-532. 10.1016/j.breast.2005.08.035.CrossRefPubMed Ciatto S, Ambrogetti D, Collini G, Cruciani A, Ercolini E, Risso G, Rosselli Del Turco M: Computer-aided detection (CAD) of cancers detected on double reading by one reader only. Breast. 2006, 15 (4): 528-532. 10.1016/j.breast.2005.08.035.CrossRefPubMed
42.
go back to reference Hukkinen K, Pamilo M: Does computer-aided detection assist in the early detection of breast cancer?. Acta Radiol. 2005, 46 (2): 135-139. 10.1080/02841850510021300.CrossRefPubMed Hukkinen K, Pamilo M: Does computer-aided detection assist in the early detection of breast cancer?. Acta Radiol. 2005, 46 (2): 135-139. 10.1080/02841850510021300.CrossRefPubMed
43.
go back to reference Destounis SV, DiNitto P, Logan-Young W, Bonaccio E, Zuley ML, Willison KM: Can computer-aided detection with double reading of screening mammograms help decrease the false-negative rate? Initial experience. Radiology. 2004, 232 (2): 578-584. 10.1148/radiol.2322030034.CrossRefPubMed Destounis SV, DiNitto P, Logan-Young W, Bonaccio E, Zuley ML, Willison KM: Can computer-aided detection with double reading of screening mammograms help decrease the false-negative rate? Initial experience. Radiology. 2004, 232 (2): 578-584. 10.1148/radiol.2322030034.CrossRefPubMed
44.
go back to reference Ciatto S, Rosselli Del Turco M, Burke P, Visioli C, Paci E, Zappa M: Comparison of standard and double reading and computer-aided detection (CAD) of interval cancers at prior negative screening mammograms: blind review. Br J Cancer. 2003, 89 (9): 1645-1649. 10.1038/sj.bjc.6601356.CrossRefPubMedPubMedCentral Ciatto S, Rosselli Del Turco M, Burke P, Visioli C, Paci E, Zappa M: Comparison of standard and double reading and computer-aided detection (CAD) of interval cancers at prior negative screening mammograms: blind review. Br J Cancer. 2003, 89 (9): 1645-1649. 10.1038/sj.bjc.6601356.CrossRefPubMedPubMedCentral
45.
go back to reference Ciatto S, Brancato B, Rosselli Del Turco M, Risso G, Catarzi S, Morrone D, Bricolo D, Zappa M: Comparison of standard reading and computer aided diagnosis (CAD) on a proficiency test of screening mammography. Radiol Med. 2003, 106 (1–2): 59-65.PubMed Ciatto S, Brancato B, Rosselli Del Turco M, Risso G, Catarzi S, Morrone D, Bricolo D, Zappa M: Comparison of standard reading and computer aided diagnosis (CAD) on a proficiency test of screening mammography. Radiol Med. 2003, 106 (1–2): 59-65.PubMed
46.
go back to reference Brem RF, Baum J, Lechner M, Kaplan S, Souders S, Naul LG, Hoffmeister J: Improvement in sensitivity of screening mammography with computer-aided detection: a multiinstitutional trial. AJR Am J Roentgenol. 2003, 181 (3): 687-693.CrossRefPubMed Brem RF, Baum J, Lechner M, Kaplan S, Souders S, Naul LG, Hoffmeister J: Improvement in sensitivity of screening mammography with computer-aided detection: a multiinstitutional trial. AJR Am J Roentgenol. 2003, 181 (3): 687-693.CrossRefPubMed
47.
go back to reference Quek ST, Thng CH, Khoo JB, Koh WL: Radiologists' detection of mammographic abnormalities with and without a computer-aided detection system. Australas Radiol. 2003, 47 (3): 257-260. 10.1046/j.1440-1673.2003.01173.x.CrossRefPubMed Quek ST, Thng CH, Khoo JB, Koh WL: Radiologists' detection of mammographic abnormalities with and without a computer-aided detection system. Australas Radiol. 2003, 47 (3): 257-260. 10.1046/j.1440-1673.2003.01173.x.CrossRefPubMed
48.
go back to reference Zheng B, Hardesty LA, Poller WR, Sumkin JH, Golla S: Mammography with computer-aided detection: reproducibility assessment initial experience. Radiology. 2003, 228 (1): 58-62. 10.1148/radiol.2281020489.CrossRefPubMed Zheng B, Hardesty LA, Poller WR, Sumkin JH, Golla S: Mammography with computer-aided detection: reproducibility assessment initial experience. Radiology. 2003, 228 (1): 58-62. 10.1148/radiol.2281020489.CrossRefPubMed
49.
go back to reference Karssemeijer N, Otten JD, Verbeek AL, Groenewoud JH, de Koning HJ, Hendriks JH, Holland R: Computer-aided detection versus independent double reading of masses on mammograms. Radiology. 2003, 227 (1): 192-200. 10.1148/radiol.2271011962.CrossRefPubMed Karssemeijer N, Otten JD, Verbeek AL, Groenewoud JH, de Koning HJ, Hendriks JH, Holland R: Computer-aided detection versus independent double reading of masses on mammograms. Radiology. 2003, 227 (1): 192-200. 10.1148/radiol.2271011962.CrossRefPubMed
50.
go back to reference Ciatto S, Del Turco MR, Risso G, Catarzi S, Bonardi R, Viterbo V, Gnutti P, Guglielmoni B, Pinelli L, Pandiscia A, et al: Comparison of standard reading and computer aided detection (CAD) on a national proficiency test of screening mammography. Eur J Radiol. 2003, 45 (2): 135-138. 10.1016/S0720-048X(02)00011-6.CrossRefPubMed Ciatto S, Del Turco MR, Risso G, Catarzi S, Bonardi R, Viterbo V, Gnutti P, Guglielmoni B, Pinelli L, Pandiscia A, et al: Comparison of standard reading and computer aided detection (CAD) on a national proficiency test of screening mammography. Eur J Radiol. 2003, 45 (2): 135-138. 10.1016/S0720-048X(02)00011-6.CrossRefPubMed
51.
go back to reference Malich A, Marx C, Facius M, Boehm T, Fleck M, Kaiser WA: Tumour detection rate of a new commercially available computer-aided detection system. Eur Radiol. 2001, 11 (12): 2454-2459. 10.1007/s003300101079.CrossRefPubMed Malich A, Marx C, Facius M, Boehm T, Fleck M, Kaiser WA: Tumour detection rate of a new commercially available computer-aided detection system. Eur Radiol. 2001, 11 (12): 2454-2459. 10.1007/s003300101079.CrossRefPubMed
52.
go back to reference Moberg K, Bjurstam N, Wilczek B, Rostgard L, Egge E, Muren C: Computed assisted detection of interval breast cancers. Eur J Radiol. 2001, 39 (2): 104-110. 10.1016/S0720-048X(01)00291-1.CrossRefPubMed Moberg K, Bjurstam N, Wilczek B, Rostgard L, Egge E, Muren C: Computed assisted detection of interval breast cancers. Eur J Radiol. 2001, 39 (2): 104-110. 10.1016/S0720-048X(01)00291-1.CrossRefPubMed
53.
go back to reference Warren Burhenne LJ, Wood SA, D'Orsi CJ, Feig SA, Kopans DB, O'Shaughnessy KF, Sickles EA, Tabar L, Vyborny CJ, Castellino RA: Potential contribution of computer-aided detection to the sensitivity of screening mammography. Radiology. 2000, 215 (2): 554-562.CrossRefPubMed Warren Burhenne LJ, Wood SA, D'Orsi CJ, Feig SA, Kopans DB, O'Shaughnessy KF, Sickles EA, Tabar L, Vyborny CJ, Castellino RA: Potential contribution of computer-aided detection to the sensitivity of screening mammography. Radiology. 2000, 215 (2): 554-562.CrossRefPubMed
54.
go back to reference Jiang Y, Nishikawa RM, Schmidt RA, Metz CE, Giger ML, Doi K: Improving breast cancer diagnosis with computer-aided diagnosis. Acad Radiol. 1999, 6 (1): 22-33. 10.1016/S1076-6332(99)80058-0.CrossRefPubMed Jiang Y, Nishikawa RM, Schmidt RA, Metz CE, Giger ML, Doi K: Improving breast cancer diagnosis with computer-aided diagnosis. Acad Radiol. 1999, 6 (1): 22-33. 10.1016/S1076-6332(99)80058-0.CrossRefPubMed
55.
go back to reference Thurfjell E, Thurfjell MG, Egge E, Bjurstam N: Sensitivity and specificity of computer-assisted breast cancer detection in mammography screening. Acta Radiol. 1998, 39 (4): 384-388.CrossRefPubMed Thurfjell E, Thurfjell MG, Egge E, Bjurstam N: Sensitivity and specificity of computer-assisted breast cancer detection in mammography screening. Acta Radiol. 1998, 39 (4): 384-388.CrossRefPubMed
56.
go back to reference Sang KY, Woo KM, Cho N, Jeong SP, Joo HC, Sun MK, Seung JK, Im JG: Screening mammography-detected cancers: Sensitivity of a computer-aided detection system applied to full-field digital mammograms. Radiology. 2007, 244 (1): 104-111. 10.1148/radiol.2441060756.CrossRef Sang KY, Woo KM, Cho N, Jeong SP, Joo HC, Sun MK, Seung JK, Im JG: Screening mammography-detected cancers: Sensitivity of a computer-aided detection system applied to full-field digital mammograms. Radiology. 2007, 244 (1): 104-111. 10.1148/radiol.2441060756.CrossRef
57.
go back to reference Li JG, Li S, Xu HM, Xu K: Evaluation on the use and results of computer-aided detection for full-field digital mammorgraphy. Chinese Journal of Radiology. 2006, 40 (7): 729-732. Li JG, Li S, Xu HM, Xu K: Evaluation on the use and results of computer-aided detection for full-field digital mammorgraphy. Chinese Journal of Radiology. 2006, 40 (7): 729-732.
58.
go back to reference Birdwell RL, Bandodkar P, Ikeda DM: Computer-aided detection with screening mammography in a university hospital setting. Radiology. 2005, 236 (2): 451-457. 10.1148/radiol.2362040864.CrossRefPubMed Birdwell RL, Bandodkar P, Ikeda DM: Computer-aided detection with screening mammography in a university hospital setting. Radiology. 2005, 236 (2): 451-457. 10.1148/radiol.2362040864.CrossRefPubMed
59.
go back to reference Cupples TE, Cunningham JE, Reynolds JC: Impact of computer-aided detection in a regional screening mammography program. AJR Am J Roentgenol. 2005, 185 (4): 944-950. 10.2214/AJR.04.1300.CrossRefPubMed Cupples TE, Cunningham JE, Reynolds JC: Impact of computer-aided detection in a regional screening mammography program. AJR Am J Roentgenol. 2005, 185 (4): 944-950. 10.2214/AJR.04.1300.CrossRefPubMed
60.
go back to reference Fenton JJ, Taplin SH, Carney PA, Abraham L, Sickles EA, D'Orsi C, Berns EA, Cutter G, Hendrick RE, Barlow WE, et al: Influence of computer-aided detection on performance of screening mammography. N Engl J Med. 2007, 356 (14): 1399-1409. 10.1056/NEJMoa066099.CrossRefPubMedPubMedCentral Fenton JJ, Taplin SH, Carney PA, Abraham L, Sickles EA, D'Orsi C, Berns EA, Cutter G, Hendrick RE, Barlow WE, et al: Influence of computer-aided detection on performance of screening mammography. N Engl J Med. 2007, 356 (14): 1399-1409. 10.1056/NEJMoa066099.CrossRefPubMedPubMedCentral
61.
go back to reference Freer TW, Ulissey MJ: Screening mammography with computer-aided detection: prospective study of 12,860 patients in a community breast center. Radiology. 2001, 220 (3): 781-786. 10.1148/radiol.2203001282.CrossRefPubMed Freer TW, Ulissey MJ: Screening mammography with computer-aided detection: prospective study of 12,860 patients in a community breast center. Radiology. 2001, 220 (3): 781-786. 10.1148/radiol.2203001282.CrossRefPubMed
62.
go back to reference Gur D, Sumkin JH, Rockette HE, Ganott M, Hakim C, Hardesty L, Poller WR, Shah R, Wallace L: Changes in breast cancer detection and mammography recall rates after the introduction of a computer-aided detection system. J Natl Cancer Inst. 2004, 96 (3): 185-190. 10.1093/jnci/djh067.CrossRefPubMed Gur D, Sumkin JH, Rockette HE, Ganott M, Hakim C, Hardesty L, Poller WR, Shah R, Wallace L: Changes in breast cancer detection and mammography recall rates after the introduction of a computer-aided detection system. J Natl Cancer Inst. 2004, 96 (3): 185-190. 10.1093/jnci/djh067.CrossRefPubMed
63.
go back to reference Karssemeijer N, Bluekens AM, Beijerinck D, Deurenberg JJ, Beekman M, Visser R, van Engen R, Bartels-Kortland A, Broeders MJ: Breast cancer screening results 5 years after introduction of digital mammography in a population-based screening program. Radiology. 2009, 253 (2): 353-358. 10.1148/radiol.2532090225.CrossRefPubMed Karssemeijer N, Bluekens AM, Beijerinck D, Deurenberg JJ, Beekman M, Visser R, van Engen R, Bartels-Kortland A, Broeders MJ: Breast cancer screening results 5 years after introduction of digital mammography in a population-based screening program. Radiology. 2009, 253 (2): 353-358. 10.1148/radiol.2532090225.CrossRefPubMed
64.
go back to reference Ko JM, Nicholas MJ, Mendel JB, Slanetz PJ: Prospective assessment of computer-aided detection in interpretation of screening mammography. AJR Am J Roentgenol. 2006, 187 (6): 1483-1491. 10.2214/AJR.05.1582.CrossRefPubMed Ko JM, Nicholas MJ, Mendel JB, Slanetz PJ: Prospective assessment of computer-aided detection in interpretation of screening mammography. AJR Am J Roentgenol. 2006, 187 (6): 1483-1491. 10.2214/AJR.05.1582.CrossRefPubMed
65.
go back to reference Morton MJ, Whaley DH, Brandt KR, Amrami KK: Screening mammograms: interpretation with computer-aided detection–prospective evaluation. Radiology. 2006, 239 (2): 375-383. 10.1148/radiol.2392042121.CrossRefPubMed Morton MJ, Whaley DH, Brandt KR, Amrami KK: Screening mammograms: interpretation with computer-aided detection–prospective evaluation. Radiology. 2006, 239 (2): 375-383. 10.1148/radiol.2392042121.CrossRefPubMed
66.
go back to reference Sanchez Gomez S, Torres Tabanera M, Vega Bolivar A, Sainz Miranda M, Baroja Mazo A, Ruiz Diaz M, Martinez Miravete P, Lag Asturiano E, Munoz Cacho P, Delgado Macias T: Impact of a CAD system in a screen-film mammography screening program: A prospective study. Eur J Radiol. 2010 Sanchez Gomez S, Torres Tabanera M, Vega Bolivar A, Sainz Miranda M, Baroja Mazo A, Ruiz Diaz M, Martinez Miravete P, Lag Asturiano E, Munoz Cacho P, Delgado Macias T: Impact of a CAD system in a screen-film mammography screening program: A prospective study. Eur J Radiol. 2010
67.
go back to reference James JJ, Gilbert FJ, Wallis MG, Gillan MG, Astley SM, Boggis CR, Agbaje OF, Brentnall AR, Duffy SW: Mammographic features of breast cancers at single reading with computer-aided detection and at double reading in a large multicenter prospective trial of computer-aided detection: CADET II. Radiology. 2010, 256 (2): 379-386. 10.1148/radiol.10091899.CrossRefPubMed James JJ, Gilbert FJ, Wallis MG, Gillan MG, Astley SM, Boggis CR, Agbaje OF, Brentnall AR, Duffy SW: Mammographic features of breast cancers at single reading with computer-aided detection and at double reading in a large multicenter prospective trial of computer-aided detection: CADET II. Radiology. 2010, 256 (2): 379-386. 10.1148/radiol.10091899.CrossRefPubMed
68.
go back to reference Georgian-Smith D, Moore RH, Halpern E, Yeh ED, Rafferty EA, D'Alessandro HA, Staffa M, Hall DA, McCarthy KA, Kopans DB: Blinded comparison of computer-aided detection with human second reading in screening mammography. AJR Am J Roentgenol. 2007, 189 (5): 1135-1141. 10.2214/AJR.07.2393.CrossRefPubMed Georgian-Smith D, Moore RH, Halpern E, Yeh ED, Rafferty EA, D'Alessandro HA, Staffa M, Hall DA, McCarthy KA, Kopans DB: Blinded comparison of computer-aided detection with human second reading in screening mammography. AJR Am J Roentgenol. 2007, 189 (5): 1135-1141. 10.2214/AJR.07.2393.CrossRefPubMed
69.
go back to reference Gromet M: Comparison of computer-aided detection to double reading of screening mammograms: review of 231,221 mammograms. AJR Am J Roentgenol. 2008, 190 (4): 854-859. 10.2214/AJR.07.2812.CrossRefPubMed Gromet M: Comparison of computer-aided detection to double reading of screening mammograms: review of 231,221 mammograms. AJR Am J Roentgenol. 2008, 190 (4): 854-859. 10.2214/AJR.07.2812.CrossRefPubMed
70.
go back to reference Khoo LA, Taylor P, Given-Wilson RM: Computer-aided detection in the United Kingdom National Breast Screening Programme: prospective study. Radiology. 2005, 237 (2): 444-449. 10.1148/radiol.2372041362.CrossRefPubMed Khoo LA, Taylor P, Given-Wilson RM: Computer-aided detection in the United Kingdom National Breast Screening Programme: prospective study. Radiology. 2005, 237 (2): 444-449. 10.1148/radiol.2372041362.CrossRefPubMed
71.
go back to reference Gilbert FJ, Astley SM, Gillan MG, Agbaje OF, Wallis MG, James J, Boggis CR, Duffy SW: Single reading with computer-aided detection for screening mammography. N Engl J Med. 2008, 359 (16): 1675-1684. 10.1056/NEJMoa0803545.CrossRefPubMed Gilbert FJ, Astley SM, Gillan MG, Agbaje OF, Wallis MG, James J, Boggis CR, Duffy SW: Single reading with computer-aided detection for screening mammography. N Engl J Med. 2008, 359 (16): 1675-1684. 10.1056/NEJMoa0803545.CrossRefPubMed
72.
go back to reference Azavedo E, Svane G: Radial scars detected mammographically in a breast cancer screening programme. Eur J Radiol. 1992, 15 (1): 18-21. 10.1016/0720-048X(92)90196-G.CrossRefPubMed Azavedo E, Svane G: Radial scars detected mammographically in a breast cancer screening programme. Eur J Radiol. 1992, 15 (1): 18-21. 10.1016/0720-048X(92)90196-G.CrossRefPubMed
73.
go back to reference Malich A, Azhari T, Bohm T, Fleck M, Kaiser WA: Reproducibility–an important factor determining the quality of computer-aided detection (CAD) systems. Eur J Radiol. 2000, 36 (3): 170-174. 10.1016/S0720-048X(00)00189-3.CrossRefPubMed Malich A, Azhari T, Bohm T, Fleck M, Kaiser WA: Reproducibility–an important factor determining the quality of computer-aided detection (CAD) systems. Eur J Radiol. 2000, 36 (3): 170-174. 10.1016/S0720-048X(00)00189-3.CrossRefPubMed
74.
go back to reference Luo P, Qian W, Romilly P: CAD-aided mammogram training. Acad Radiol. 2005, 12 (8): 1039-1048. 10.1016/j.acra.2005.04.011.CrossRefPubMed Luo P, Qian W, Romilly P: CAD-aided mammogram training. Acad Radiol. 2005, 12 (8): 1039-1048. 10.1016/j.acra.2005.04.011.CrossRefPubMed
Metadata
Title
Is single reading with computer-aided detection (CAD) as good as double reading in mammography screening? A systematic review
Authors
Edward Azavedo
Sophia Zackrisson
Ingegerd Mejàre
Marianne Heibert Arnlind
Publication date
01-12-2012
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2012
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/1471-2342-12-22

Other articles of this Issue 1/2012

BMC Medical Imaging 1/2012 Go to the issue