Published in: BMC Anesthesiology 1/2014

Open Access 01-12-2014 | Research article

Extubation force depends upon angle of force application and fixation technique: a study of 7 methods

Authors: Jennifer L Wagner, Robin Shandas, Craig J Lanning

Published in: BMC Anesthesiology | Issue 1/2014

Login to get access

Abstract

Background

Endotracheal tubes are frequently used to establish alternate airways. Precise placement of the tubes must be maintained to prevent serious complications. Several methods for fixation of endotracheal tubes are available. Available methods vary widely in form and functionality. Due to the unpredictable and dynamic nature of circumstances surrounding intubation, thorough evaluation of tube restraints may help reduce airway accidents such as tube dislodgement and unplanned extubation.

Methods

Seven different tube-restraint combinations were compared against themselves and one another at a series of discrete angles (test points) covering a hemisphere on the plane of the face. Force values for tube motion of 2 cm and 5 cm (or failure) were recorded for 3 pull tests, at each angle, for each method of tube fixation.

Results

All methods showed variation in the force required for tube motion with angle of force application. When forces were averaged over all test points, for each fixation technique, differences as large as 132 N (30 lbf) were observed (95% CI 113 N to 152 N). Compared to traditional methods of fixation, only 1 of the 3 commercially available devices consistently required a higher average force to displace the tube 2 cm and 5 cm. When ranges of force values for 5 cm displacement were compared, devices span from 80–290 N (18–65 lbf) while traditional methods span from 62–178 N (14–40 lbf), highlighting the value of examining forces at the different angles of application. Significant differences in standard deviations were also observed between the 7 techniques indicating that some methods may be more reproducible than others.

Conclusions

Clinically, forces can be applied to endotracheal tubes from various directions. Efficacies of different fixation techniques are sensitive to the angle of force application. Standard deviations, which could be used as a measure of fixator reliability, also vary with angle of force application and method of tube restraint. Findings presented in this study may be used to advance clinical implementation of current methods as well as fixator device design in an effort to reduce the incidence of unplanned extubation.
Appendix
Available only for authorised users
Literature
1.
go back to reference Neumar RW, Otto CW, Link MS, Kronick SL, Shuster M, Callaway CW, Kudenchuck PJ, Ornato JP, McNally B, Silvers SM, Passman RS, White RD, Hess EP, Tang W, Davis D, Sinz E, Morrison LJ: Part 8: adult advanced cardiovascular life support: 2010 American heart association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2010, 122 (Suppl 3): S729-S767.CrossRefPubMed Neumar RW, Otto CW, Link MS, Kronick SL, Shuster M, Callaway CW, Kudenchuck PJ, Ornato JP, McNally B, Silvers SM, Passman RS, White RD, Hess EP, Tang W, Davis D, Sinz E, Morrison LJ: Part 8: adult advanced cardiovascular life support: 2010 American heart association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2010, 122 (Suppl 3): S729-S767.CrossRefPubMed
2.
go back to reference Roberts JR, Hedges JR, Chanmugam AS: Clinical Procedures in Emergency Medicine. 2004, Philadelphia, PA: W.B. Saunders Roberts JR, Hedges JR, Chanmugam AS: Clinical Procedures in Emergency Medicine. 2004, Philadelphia, PA: W.B. Saunders
3.
go back to reference Kaplow R, Bookbinder M: A comparison of four endotracheal tube holders. Heart Lung. 1994, 23: 59-66.PubMed Kaplow R, Bookbinder M: A comparison of four endotracheal tube holders. Heart Lung. 1994, 23: 59-66.PubMed
4.
go back to reference Luria S, Vidan A, Nahtomi O, Khanin A, Alcalay M: Proposed technique for evaluation of endotracheal tube fixation and comparison of four fixation methods. Mil Med. 2001, 166: 82-PubMed Luria S, Vidan A, Nahtomi O, Khanin A, Alcalay M: Proposed technique for evaluation of endotracheal tube fixation and comparison of four fixation methods. Mil Med. 2001, 166: 82-PubMed
5.
go back to reference Kapadia FN, Bajan KB, Raje KV: Airway accidents in intubated intensive care unit patients: an epidemiological study. Crit Care Med. 2000, 28: 659-664. 10.1097/00003246-200003000-00010.CrossRefPubMed Kapadia FN, Bajan KB, Raje KV: Airway accidents in intubated intensive care unit patients: an epidemiological study. Crit Care Med. 2000, 28: 659-664. 10.1097/00003246-200003000-00010.CrossRefPubMed
6.
go back to reference Birkett KM, Southerland KA, Leslie GD: Reporting unplanned extubation. Intensive Crit Care Nurs. 2005, 21: 65-75. 10.1016/j.iccn.2004.07.012.CrossRefPubMed Birkett KM, Southerland KA, Leslie GD: Reporting unplanned extubation. Intensive Crit Care Nurs. 2005, 21: 65-75. 10.1016/j.iccn.2004.07.012.CrossRefPubMed
7.
go back to reference Chang L, Wang K, Chao Y: Influence of physical restraint on unplanned extubation of adult intensive care patients: a case–control study. Am J Crit Care. 2008, 17: 408-415.PubMed Chang L, Wang K, Chao Y: Influence of physical restraint on unplanned extubation of adult intensive care patients: a case–control study. Am J Crit Care. 2008, 17: 408-415.PubMed
8.
go back to reference Boulain T: Unplanned extubations in the adult intensive care unit: a prospective multicenter study. Am J Respir Crit Care Med. 1998, 157: 1131-1137. 10.1164/ajrccm.157.4.9702083.CrossRefPubMed Boulain T: Unplanned extubations in the adult intensive care unit: a prospective multicenter study. Am J Respir Crit Care Med. 1998, 157: 1131-1137. 10.1164/ajrccm.157.4.9702083.CrossRefPubMed
9.
go back to reference Tindol GA, DiBenedetto RJ, Kosciuk L: Unplanned extubations. CHEST J. 1994, 105: 1804-1807. 10.1378/chest.105.6.1804.CrossRef Tindol GA, DiBenedetto RJ, Kosciuk L: Unplanned extubations. CHEST J. 1994, 105: 1804-1807. 10.1378/chest.105.6.1804.CrossRef
10.
go back to reference Betbese AJ, Perez M, Bak E, Rialp G, Mancebo J: A prospective study of unplanned endotracheal extubation in intensive care unit patients. Crit Care Med. 1998, 26: 1180-1186. 10.1097/00003246-199807000-00016.CrossRefPubMed Betbese AJ, Perez M, Bak E, Rialp G, Mancebo J: A prospective study of unplanned endotracheal extubation in intensive care unit patients. Crit Care Med. 1998, 26: 1180-1186. 10.1097/00003246-199807000-00016.CrossRefPubMed
11.
go back to reference Stauffer JL, Olson DE, Petty TL: Complications and consequences of endotracheal intubation and tracheotomy: a prospective study of 150 critically ill adult patients. Am J Med. 1981, 70: 65-76. 10.1016/0002-9343(81)90413-7.CrossRefPubMed Stauffer JL, Olson DE, Petty TL: Complications and consequences of endotracheal intubation and tracheotomy: a prospective study of 150 critically ill adult patients. Am J Med. 1981, 70: 65-76. 10.1016/0002-9343(81)90413-7.CrossRefPubMed
12.
go back to reference Tominaga GT, Rudzwick H, Scannel G, Waxman K: Decreasing unplanned extubations in the surgical intensive care unit. Am J Surg. 1995, 170: 586-590. 10.1016/S0002-9610(99)80021-X.CrossRefPubMed Tominaga GT, Rudzwick H, Scannel G, Waxman K: Decreasing unplanned extubations in the surgical intensive care unit. Am J Surg. 1995, 170: 586-590. 10.1016/S0002-9610(99)80021-X.CrossRefPubMed
13.
go back to reference Yeh SH, Lee LN, Ho TH, Chiang MC, Lin LW: Implications of nursing care in the occurrence and consequences of unplanned extubation in adult intensive care units. Int J Nurs Stud. 2004, 41: 255-262. 10.1016/S0020-7489(03)00136-6.CrossRefPubMed Yeh SH, Lee LN, Ho TH, Chiang MC, Lin LW: Implications of nursing care in the occurrence and consequences of unplanned extubation in adult intensive care units. Int J Nurs Stud. 2004, 41: 255-262. 10.1016/S0020-7489(03)00136-6.CrossRefPubMed
14.
go back to reference Christie JM, Dethlefsen M, Cane RD: Unplanned endotracheal extubation in the intensive care unit. J Clin Anesth. 1996, 8: 289-293. 10.1016/0952-8180(96)00037-2.CrossRefPubMed Christie JM, Dethlefsen M, Cane RD: Unplanned endotracheal extubation in the intensive care unit. J Clin Anesth. 1996, 8: 289-293. 10.1016/0952-8180(96)00037-2.CrossRefPubMed
15.
go back to reference Atkins PM, Mian LC, Mendelson W, Palmer RM, Franko T: Characteristics and outcomes of patients Who self-extubate from ventilatory support. Chest. 1997, 112: 1317-1323. 10.1378/chest.112.5.1317.CrossRefPubMed Atkins PM, Mian LC, Mendelson W, Palmer RM, Franko T: Characteristics and outcomes of patients Who self-extubate from ventilatory support. Chest. 1997, 112: 1317-1323. 10.1378/chest.112.5.1317.CrossRefPubMed
16.
go back to reference Coppolo P, May J: Self-extubations. A 12-month experience. CHEST J. 1990, 98: 165-169. 10.1378/chest.98.1.165.CrossRef Coppolo P, May J: Self-extubations. A 12-month experience. CHEST J. 1990, 98: 165-169. 10.1378/chest.98.1.165.CrossRef
17.
go back to reference Shimizu T, Mizutani T, Yamashita S, Hagiya K, Tanaka M: Endotracheal tube extubation force: adhesive tape versus endotracheal tube holder. Respir Care. 2011, 56: 1825-1829. 10.4187/respcare.00954.CrossRefPubMed Shimizu T, Mizutani T, Yamashita S, Hagiya K, Tanaka M: Endotracheal tube extubation force: adhesive tape versus endotracheal tube holder. Respir Care. 2011, 56: 1825-1829. 10.4187/respcare.00954.CrossRefPubMed
18.
go back to reference Carlson JN, Mayrose J, Wang HE: How much force is required to dislodge an alternate airway?. Prehosp Emerg Care. 2009, 14: 31-35.CrossRef Carlson JN, Mayrose J, Wang HE: How much force is required to dislodge an alternate airway?. Prehosp Emerg Care. 2009, 14: 31-35.CrossRef
19.
go back to reference Owen R, Castle N, Hann H, Reeves D, Naidoo R, Naidoo S: Extubation force: a comparison of adhesive tape, non-adhesive tape and a commercial endotracheal tube holder. Resuscitation. 2009, 80: 1296-1300. 10.1016/j.resuscitation.2009.08.007.CrossRefPubMed Owen R, Castle N, Hann H, Reeves D, Naidoo R, Naidoo S: Extubation force: a comparison of adhesive tape, non-adhesive tape and a commercial endotracheal tube holder. Resuscitation. 2009, 80: 1296-1300. 10.1016/j.resuscitation.2009.08.007.CrossRefPubMed
20.
go back to reference Carlson J, Mayrose J, Krause R, Jehle D: Extubation force: tape versus endotracheal tube holders. Ann Emerg Med. 2007, 50: 686-691. 10.1016/j.annemergmed.2007.05.013.CrossRefPubMed Carlson J, Mayrose J, Krause R, Jehle D: Extubation force: tape versus endotracheal tube holders. Ann Emerg Med. 2007, 50: 686-691. 10.1016/j.annemergmed.2007.05.013.CrossRefPubMed
21.
go back to reference Murdoch E, Holdgate A: A comparison of tape-tying versus a tube-holding device for securing endotracheal tubes in adults. Anaesth Intensive Care. 2007, 35: 730-735.PubMed Murdoch E, Holdgate A: A comparison of tape-tying versus a tube-holding device for securing endotracheal tubes in adults. Anaesth Intensive Care. 2007, 35: 730-735.PubMed
22.
go back to reference Tasota FJ, Hoffman LA, Zullo TG, Jamison G: Evaluation of two methods used to stabilize oral endotracheal tubes. Heart Lung. 1987, 16: 140-PubMed Tasota FJ, Hoffman LA, Zullo TG, Jamison G: Evaluation of two methods used to stabilize oral endotracheal tubes. Heart Lung. 1987, 16: 140-PubMed
Metadata
Title
Extubation force depends upon angle of force application and fixation technique: a study of 7 methods
Authors
Jennifer L Wagner
Robin Shandas
Craig J Lanning
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2014
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/1471-2253-14-74

Other articles of this Issue 1/2014

BMC Anesthesiology 1/2014 Go to the issue