Published in: BMC Anesthesiology 1/2012

Open Access 01-12-2012 | Research article

Use of a blood gas analyzer and a laboratory autoanalyzer in routine practice to measure electrolytes in intensive care unit patients

Authors: Yasemin U Budak, Kagan Huysal, Murat Polat

Published in: BMC Anesthesiology | Issue 1/2012

Login to get access

Abstract

Background

Electrolyte values are measured in most critically ill intensive care unit (ICU) patients using both an arterial blood gas analyzer (ABG) and a central laboratory auto-analyzer (AA). The aim of the present study was to investigate whether electrolyte levels assessed using an ABG and an AA were equivalent; data on sodium and potassium ion concentrations were examined.

Methods

We retrospectively studied patients hospitalized in the ICU between July and August 2011. Of 1,105 test samples, we identified 84 instances of simultaneous sampling of arterial and venous blood, where both Na+ and K+ levels were measured using a pHOx Stat Profile Plus L blood gas analyzer (Nova Biomedical, Waltham MA, USA) and a Roche Modular P autoanalyzer (Roche Diagnostics, Mannheim, Germany). Statistical measures employed to compare the data included Spearman's correlation coefficients, paired Student’s t-tests, Deming regression analysis, and Bland-Altman plots.

Results

The mean sodium concentration was 138.1 mmol/L (SD 10.2 mmol/L) using the ABG and 143.0 mmol/L (SD 10.5) using the AA (p < 0.001). The mean potassium level was 3.5 mmol/L (SD 0.9 mmol/L) using the ABG and 3.7 mmol/L (SD 1.0 mmol/L) using the AA (p < 0.001). The extent of inter-analyzer agreement was unacceptable for both K+ and Na+, with biases of 0.150-0.352 and −0.97-10.05 respectively; the associated correlation coefficients were 0.88 and 0.90.

Conclusions

We conclude that the ABG and AA do not yield equivalent Na+ and K+ data. Concordance between ABG and AA should be established prior to introduction of new ABG systems.
Appendix
Available only for authorised users
Literature
1.
go back to reference Murray MJ: 168. Michael James, American Society of Critical Care Anesthesiologists. 2002, Lippincott Williams & Wilkins, Philadelphia, PA, 168-169. Murray MJ: 168. Michael James, American Society of Critical Care Anesthesiologists. 2002, Lippincott Williams & Wilkins, Philadelphia, PA, 168-169.
2.
go back to reference Eachempati SR: Approach to the Critically Ill Patient. Critical Care Medicine. Edited by: Porter RS. , The Merck Manuals. Merck Sharp & Dohme Corp., a subsidiary of Merck & Co., Inc, Whitehouse Station, N.J., U.S.A, 2010-2011. Eachempati SR: Approach to the Critically Ill Patient. Critical Care Medicine. Edited by: Porter RS. , The Merck Manuals. Merck Sharp & Dohme Corp., a subsidiary of Merck & Co., Inc, Whitehouse Station, N.J., U.S.A, 2010-2011.
3.
go back to reference Scott MG, LeGrys VA, Klutts JS: Electrolytes and blood gases. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. Edited by: Burtis DE, Ashwood ER, Bruns DE. 2006, Elsevier, St Louis, MO, 983-1018. 4 Scott MG, LeGrys VA, Klutts JS: Electrolytes and blood gases. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. Edited by: Burtis DE, Ashwood ER, Bruns DE. 2006, Elsevier, St Louis, MO, 983-1018. 4
4.
go back to reference D’Orazio P, Miller WG, Myers GL, Doumas BT, Eckfeldt JH, Evans SA, Graham GA, Parsons PJ, Stanton NV: Standardization of Sodium and Potassium Ion-Selective Electrode Systems to the Flame Photometric Reference Method; Approved Standard—Second Edition, C29-A2 [electronic document]. CLSI. 2000, 20: 1-22. D’Orazio P, Miller WG, Myers GL, Doumas BT, Eckfeldt JH, Evans SA, Graham GA, Parsons PJ, Stanton NV: Standardization of Sodium and Potassium Ion-Selective Electrode Systems to the Flame Photometric Reference Method; Approved Standard—Second Edition, C29-A2 [electronic document]. CLSI. 2000, 20: 1-22.
5.
go back to reference Ladenson JH: Direct potentiometric measurement of sodium and potassium in whole blood. Clin Chem. 1977, 23 (9): 10- Ladenson JH: Direct potentiometric measurement of sodium and potassium in whole blood. Clin Chem. 1977, 23 (9): 10-
6.
go back to reference Kost GJ, Ihrmeyer SS, Chernow B, Winkelman JW, Zaloga GP, Dellinger PR, Shirey T: The laboratory- clinical interface: Point-of-care testing. Chest. 1999, 115: 1140-1154. 10.1378/chest.115.4.1140.CrossRefPubMed Kost GJ, Ihrmeyer SS, Chernow B, Winkelman JW, Zaloga GP, Dellinger PR, Shirey T: The laboratory- clinical interface: Point-of-care testing. Chest. 1999, 115: 1140-1154. 10.1378/chest.115.4.1140.CrossRefPubMed
7.
go back to reference St-Louis P: Status of point-of-care testing: promise, realities, and possibilities. Clin Biochem. 2000, 33 (6): 427-440. 10.1016/S0009-9120(00)00138-7.CrossRefPubMed St-Louis P: Status of point-of-care testing: promise, realities, and possibilities. Clin Biochem. 2000, 33 (6): 427-440. 10.1016/S0009-9120(00)00138-7.CrossRefPubMed
8.
go back to reference Bland JM, Altmann DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, i: 307-310.CrossRef Bland JM, Altmann DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, i: 307-310.CrossRef
9.
go back to reference Fraser CG, Hyltoft Petersen P, Ricos C, Haeckel R: Proposed quality specifications for the imprecision and inaccuracy of analytical systems for clinical chemistry. Eur J Clin Chem Clin Biochem. 1992, 30: 311-317.PubMed Fraser CG, Hyltoft Petersen P, Ricos C, Haeckel R: Proposed quality specifications for the imprecision and inaccuracy of analytical systems for clinical chemistry. Eur J Clin Chem Clin Biochem. 1992, 30: 311-317.PubMed
10.
go back to reference Ehrmeyer SS, Laessig RH, Leinweber JE, Oryall JJ: 1990 Medicare/CLIA final rules for proficiency testing: minimum intralaboratory performance characteristics (CV and bias) needed to pass. Clin Chem. 1990, 36: 1736-1740.PubMed Ehrmeyer SS, Laessig RH, Leinweber JE, Oryall JJ: 1990 Medicare/CLIA final rules for proficiency testing: minimum intralaboratory performance characteristics (CV and bias) needed to pass. Clin Chem. 1990, 36: 1736-1740.PubMed
11.
go back to reference Jain A, Subhan I, Joshi M: Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes. Int J Emerg Med. 2009, 2 (2): 117-120. 10.1007/s12245-009-0091-1.CrossRefPubMedPubMedCentral Jain A, Subhan I, Joshi M: Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes. Int J Emerg Med. 2009, 2 (2): 117-120. 10.1007/s12245-009-0091-1.CrossRefPubMedPubMedCentral
12.
go back to reference Morimatsu H, Rocktäschel J, Bellomo R, Uchino S, Goldsmith D, Gutteridge G: Comparison of point-of-care versus central laboratory measurement of electrolyte concentrations on calculations of the anion gap and the strong ion difference. Anesthesiology. 2003, 98: 1077-1084. 10.1097/00000542-200305000-00009.CrossRefPubMed Morimatsu H, Rocktäschel J, Bellomo R, Uchino S, Goldsmith D, Gutteridge G: Comparison of point-of-care versus central laboratory measurement of electrolyte concentrations on calculations of the anion gap and the strong ion difference. Anesthesiology. 2003, 98: 1077-1084. 10.1097/00000542-200305000-00009.CrossRefPubMed
13.
go back to reference Vlah SH, Dvornik S, Grdović D: Analytical performance of the Gem® PremierTM 4000 – a comparison study. Biochemia Medica. 2009, 19 (2): 192-198.CrossRef Vlah SH, Dvornik S, Grdović D: Analytical performance of the Gem® PremierTM 4000 – a comparison study. Biochemia Medica. 2009, 19 (2): 192-198.CrossRef
14.
go back to reference Chow E, Fox N, Gama R: Effect of low serum total protein on sodium and potassium measurement by ion-selective electrodes in critically ill patients. Br J Biomed Sci. 2008, 65 (3): 128-131.CrossRefPubMed Chow E, Fox N, Gama R: Effect of low serum total protein on sodium and potassium measurement by ion-selective electrodes in critically ill patients. Br J Biomed Sci. 2008, 65 (3): 128-131.CrossRefPubMed
15.
go back to reference Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, Minchinela J, Perich C, Simon M: Current databases on biologic variation: pros, cons and progress. Scand J Clin Lab Invest. 1999, 59: 491-500. 10.1080/00365519950185229.CrossRefPubMed Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, Minchinela J, Perich C, Simon M: Current databases on biologic variation: pros, cons and progress. Scand J Clin Lab Invest. 1999, 59: 491-500. 10.1080/00365519950185229.CrossRefPubMed
16.
go back to reference Scott MG, LeGrys VA, Klufts JC: Electrolytes and Blood Gases. Ch. 27. Tietz Textbook of Clinical Chemistry. Edited by: Burtis CA, Ashwood E, Bruns DE. 2006, Elsevier, Missouri USA, 985-4 Scott MG, LeGrys VA, Klufts JC: Electrolytes and Blood Gases. Ch. 27. Tietz Textbook of Clinical Chemistry. Edited by: Burtis CA, Ashwood E, Bruns DE. 2006, Elsevier, Missouri USA, 985-4
17.
go back to reference Weaver DK, Miller D, Leventhal EA, Tropeano V: Evaluation of a computer-directed pneumatic-tube system for pneumatic transport of blood specimens. Am J Clin Pathol. 1978, 70: 400-405.CrossRefPubMed Weaver DK, Miller D, Leventhal EA, Tropeano V: Evaluation of a computer-directed pneumatic-tube system for pneumatic transport of blood specimens. Am J Clin Pathol. 1978, 70: 400-405.CrossRefPubMed
18.
go back to reference Burritt MF: Current analytical approaches to measuring blood analytes. Clin Chem. 1990, 36 (8 Pt 2): 1562-6.PubMed Burritt MF: Current analytical approaches to measuring blood analytes. Clin Chem. 1990, 36 (8 Pt 2): 1562-6.PubMed
19.
go back to reference Lima-Oliveira G, Lippi G, Salvagno GL, Montagnana M, Picheth G, Guidi GC: Different manufacturers of syringes: A new source of variability in blood gas, acid–base balance and related laboratory test?. Clin Biochem. 2012, 45 (9): 683-7. 10.1016/j.clinbiochem.2012.03.007.CrossRefPubMed Lima-Oliveira G, Lippi G, Salvagno GL, Montagnana M, Picheth G, Guidi GC: Different manufacturers of syringes: A new source of variability in blood gas, acid–base balance and related laboratory test?. Clin Biochem. 2012, 45 (9): 683-7. 10.1016/j.clinbiochem.2012.03.007.CrossRefPubMed
20.
go back to reference van Berkel M, Scharnhorst V: Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes. Clin Chem Lab Med. 2011, 49 (2): 249-52.CrossRefPubMed van Berkel M, Scharnhorst V: Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes. Clin Chem Lab Med. 2011, 49 (2): 249-52.CrossRefPubMed
21.
go back to reference Rouse C: Theoretical reasons for sodium measurement error - POCT specialist -C29-A2 ISBN 1–56238–410–4 ISSN 0273–3099. Rouse C: Theoretical reasons for sodium measurement error - POCT specialist -C29-A2 ISBN 1–56238–410–4 ISSN 0273–3099.
Metadata
Title
Use of a blood gas analyzer and a laboratory autoanalyzer in routine practice to measure electrolytes in intensive care unit patients
Authors
Yasemin U Budak
Kagan Huysal
Murat Polat
Publication date
01-12-2012
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2012
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/1471-2253-12-17

Other articles of this Issue 1/2012

BMC Anesthesiology 1/2012 Go to the issue