Published in: BMC Sports Science, Medicine and Rehabilitation 1/2015

Open Access 01-12-2015 | Research article

Reliability and validity of ten consumer activity trackers

Authors: Thea J. M. Kooiman, Manon L. Dontje, Siska R. Sprenger, Wim P. Krijnen, Cees P. van der Schans, Martijn de Groot

Published in: BMC Sports Science, Medicine and Rehabilitation | Issue 1/2015

Login to get access

Abstract

Background

Activity trackers can potentially stimulate users to increase their physical activity behavior. The aim of this study was to examine the reliability and validity of ten consumer activity trackers for measuring step count in both laboratory and free-living conditions.

Method

Healthy adult volunteers (n = 33) walked twice on a treadmill (4.8 km/h) for 30 min while wearing ten different activity trackers (i.e. Lumoback, Fitbit Flex, Jawbone Up, Nike+ Fuelband SE, Misfit Shine, Withings Pulse, Fitbit Zip, Omron HJ-203, Yamax Digiwalker SW-200 and Moves mobile application). In free-living conditions, 56 volunteers wore the same activity trackers for one working day. Test-retest reliability was analyzed with the Intraclass Correlation Coefficient (ICC). Validity was evaluated by comparing each tracker with the gold standard (Optogait system for laboratory and ActivPAL for free-living conditions), using paired samples t-tests, mean absolute percentage errors, correlations and Bland-Altman plots.

Results

Test-retest analysis revealed high reliability for most trackers except for the Omron (ICC .14), Moves app (ICC .37) and Nike+ Fuelband (ICC .53). The mean absolute percentage errors of the trackers in laboratory and free-living conditions respectively, were: Lumoback (−0.2, −0.4), Fibit Flex (−5.7, 3.7), Jawbone Up (−1.0, 1.4), Nike+ Fuelband (−18, −24), Misfit Shine (0.2, 1.1), Withings Pulse (−0.5, −7.9), Fitbit Zip (−0.3, 1.2), Omron (2.5, −0.4), Digiwalker (−1.2, −5.9), and Moves app (9.6, −37.6). Bland-Altman plots demonstrated that the limits of agreement varied from 46 steps (Fitbit Zip) to 2422 steps (Nike+ Fuelband) in the laboratory condition, and 866 steps (Fitbit Zip) to 5150 steps (Moves app) in the free-living condition.

Conclusion

The reliability and validity of most trackers for measuring step count is good. The Fitbit Zip is the most valid whereas the reliability and validity of the Nike+ Fuelband is low.
Literature
1.
go back to reference Lee IM, Shiroma EJ, Lobelo F, Puska P, Blair SN, Katzmarzyk PT, et al. Effect of physical inactivity on major non-communicable diseases worldwide: an analysis of burden of disease and life expectancy. Lancet. 2012;380(9838):219–29.CrossRefPubMedPubMedCentral Lee IM, Shiroma EJ, Lobelo F, Puska P, Blair SN, Katzmarzyk PT, et al. Effect of physical inactivity on major non-communicable diseases worldwide: an analysis of burden of disease and life expectancy. Lancet. 2012;380(9838):219–29.CrossRefPubMedPubMedCentral
3.
go back to reference Haskell WL, Lee IM, Pate RR, Blair SN, Franklin BA, Macera CA, et al. Physical activity and public health: updated recommendation for adults from the American College of Sports Medicine and the American Heart Association. Med Sci Sports Exerc. 2007;39(8):1423–34.CrossRefPubMed Haskell WL, Lee IM, Pate RR, Blair SN, Franklin BA, Macera CA, et al. Physical activity and public health: updated recommendation for adults from the American College of Sports Medicine and the American Heart Association. Med Sci Sports Exerc. 2007;39(8):1423–34.CrossRefPubMed
4.
go back to reference Tudor-Locke C, Craig CL, Brown WJ, Clemes SA, De Cocker K, Giles-Corti B, et al. How many steps/day are enough? For adults. Int J Behav Nutr Phys Act. 2011;8:79-5868-8-79. Tudor-Locke C, Craig CL, Brown WJ, Clemes SA, De Cocker K, Giles-Corti B, et al. How many steps/day are enough? For adults. Int J Behav Nutr Phys Act. 2011;8:79-5868-8-79.
5.
go back to reference Godino JG, Watkinson C, Corder K, Sutton S, Griffin SJ, Van Sluijs EM. Awareness of physical activity in healthy middle-aged adults: a cross-sectional study of associations with sociodemographic, biological, behavioural, and psychological factors. BMC Public Health. 2014;14(1):421.CrossRefPubMedPubMedCentral Godino JG, Watkinson C, Corder K, Sutton S, Griffin SJ, Van Sluijs EM. Awareness of physical activity in healthy middle-aged adults: a cross-sectional study of associations with sociodemographic, biological, behavioural, and psychological factors. BMC Public Health. 2014;14(1):421.CrossRefPubMedPubMedCentral
6.
go back to reference Vooijs M, Alpay LL, Snoeck-Stroband JB, Beerthuizen T, Siemonsma PC, Abbink JJ, et al. Validity and usability of low-cost accelerometers for internet-based self-monitoring of physical activity in patients with chronic obstructive pulmonary disease. Interact J Med Res. 2014;3(4):e14.CrossRefPubMedPubMedCentral Vooijs M, Alpay LL, Snoeck-Stroband JB, Beerthuizen T, Siemonsma PC, Abbink JJ, et al. Validity and usability of low-cost accelerometers for internet-based self-monitoring of physical activity in patients with chronic obstructive pulmonary disease. Interact J Med Res. 2014;3(4):e14.CrossRefPubMedPubMedCentral
7.
go back to reference Bravata DM, Smith-Spangler C, Sundaram V, Gienger AL, Lin N, Lewis R, et al. Using pedometers to increase physical activity and improve health: a systematic review. JAMA. 2007;298(19):2296–304.CrossRefPubMed Bravata DM, Smith-Spangler C, Sundaram V, Gienger AL, Lin N, Lewis R, et al. Using pedometers to increase physical activity and improve health: a systematic review. JAMA. 2007;298(19):2296–304.CrossRefPubMed
8.
go back to reference El-Gayar O, Timsina P, Nawar N, Eid W. A systematic review of IT for diabetes self-management: are we there yet? Int J Med Inform. 2013;82(8):637–52.CrossRefPubMed El-Gayar O, Timsina P, Nawar N, Eid W. A systematic review of IT for diabetes self-management: are we there yet? Int J Med Inform. 2013;82(8):637–52.CrossRefPubMed
9.
go back to reference Adam Noah J, Spierer DK, Gu J, Bronner S. Comparison of steps and energy expenditure assessment in adults of Fitbit Tracker and Ultra to the Actical and indirect calorimetry. J Med Eng Technol. 2013;37(7):456–62.CrossRefPubMed Adam Noah J, Spierer DK, Gu J, Bronner S. Comparison of steps and energy expenditure assessment in adults of Fitbit Tracker and Ultra to the Actical and indirect calorimetry. J Med Eng Technol. 2013;37(7):456–62.CrossRefPubMed
10.
go back to reference Dannecker KL, Sazonova NA, Melanson EL, Sazonov ES, Browning RC. A comparison of energy expenditure estimation of several physical activity monitors. Med Sci Sports Exerc. 2013;45(11):2105–12.CrossRefPubMedPubMedCentral Dannecker KL, Sazonova NA, Melanson EL, Sazonov ES, Browning RC. A comparison of energy expenditure estimation of several physical activity monitors. Med Sci Sports Exerc. 2013;45(11):2105–12.CrossRefPubMedPubMedCentral
11.
go back to reference Fortune E, Lugade V, Morrow M, Kaufman K. Validity of using tri-axial accelerometers to measure human movement - Part II: Step counts at a wide range of gait velocities. Med Eng Phys. 2014;36(6):659–69.CrossRefPubMedPubMedCentral Fortune E, Lugade V, Morrow M, Kaufman K. Validity of using tri-axial accelerometers to measure human movement - Part II: Step counts at a wide range of gait velocities. Med Eng Phys. 2014;36(6):659–69.CrossRefPubMedPubMedCentral
12.
go back to reference Lee JM, Kim Y, Welk GJ. Validity of consumer-based physical activity monitors. Med Sci Sports Exerc. 2014;46(9):1840–8.CrossRefPubMed Lee JM, Kim Y, Welk GJ. Validity of consumer-based physical activity monitors. Med Sci Sports Exerc. 2014;46(9):1840–8.CrossRefPubMed
13.
go back to reference Stahl ST, Insana SP. Caloric expenditure assessment among older adults: criterion validity of a novel accelerometry device. J Health Psychol. 2014;19(11):1382–7.CrossRefPubMed Stahl ST, Insana SP. Caloric expenditure assessment among older adults: criterion validity of a novel accelerometry device. J Health Psychol. 2014;19(11):1382–7.CrossRefPubMed
14.
go back to reference Takacs J, Pollock CL, Guenther JR, Bahar M, Napier C, Hunt MA. Validation of the Fitbit One activity monitor device during treadmill walking. Interact J Med Res. 2014;17(5):496–500. Takacs J, Pollock CL, Guenther JR, Bahar M, Napier C, Hunt MA. Validation of the Fitbit One activity monitor device during treadmill walking. Interact J Med Res. 2014;17(5):496–500.
15.
go back to reference Tully MA, McBride C, Heron L, Allen W, Hunter RF. The validation of Fibit ZipTM physical activity monitor as a measure of free-living physical activity. BMC Res Notes. 2014;7(1):952.CrossRefPubMedPubMedCentral Tully MA, McBride C, Heron L, Allen W, Hunter RF. The validation of Fibit ZipTM physical activity monitor as a measure of free-living physical activity. BMC Res Notes. 2014;7(1):952.CrossRefPubMedPubMedCentral
16.
go back to reference Case MA, Burwick HA, Volpp KG, Patel MS. Accuracy of smartphone applications and wearable devices for tracking physical activity data. JAMA. 2015;313(6):625–6.CrossRefPubMed Case MA, Burwick HA, Volpp KG, Patel MS. Accuracy of smartphone applications and wearable devices for tracking physical activity data. JAMA. 2015;313(6):625–6.CrossRefPubMed
17.
go back to reference Ferguson T, Rowlands AV, Olds T, Maher C. The validity of consumer-level, activity monitors in healthy adults worn in free-living conditions: a cross-sectional study. Int J Behav Nutr Phys Act. 2015;12:42-015-0201-9.CrossRef Ferguson T, Rowlands AV, Olds T, Maher C. The validity of consumer-level, activity monitors in healthy adults worn in free-living conditions: a cross-sectional study. Int J Behav Nutr Phys Act. 2015;12:42-015-0201-9.CrossRef
18.
go back to reference Dontje ML, de Groot M, Lengton RR, van der Schans CP, Krijnen WP. Measuring steps with the Fitbit activity tracker: an inter-device reliability study. J Med Eng Technol. 2015;39(5):286–90.CrossRefPubMed Dontje ML, de Groot M, Lengton RR, van der Schans CP, Krijnen WP. Measuring steps with the Fitbit activity tracker: an inter-device reliability study. J Med Eng Technol. 2015;39(5):286–90.CrossRefPubMed
19.
go back to reference Lee M, Song C, Lee K, Shin D, Shin S. Agreement between the spatio-temporal gait parameters from treadmill-based photoelectric cell and the instrumented treadmill system in healthy young adults and stroke patients. Med Sci Monit. 2014;20:1210–9.CrossRefPubMedPubMedCentral Lee M, Song C, Lee K, Shin D, Shin S. Agreement between the spatio-temporal gait parameters from treadmill-based photoelectric cell and the instrumented treadmill system in healthy young adults and stroke patients. Med Sci Monit. 2014;20:1210–9.CrossRefPubMedPubMedCentral
20.
go back to reference Dahlgren G, Carlsson D, Moorhead A, Hager-Ross C, McDonough SM. Test-retest reliability of step counts with the ActivPAL device in common daily activities. Gait Posture. 2010;32(3):386–90.CrossRefPubMed Dahlgren G, Carlsson D, Moorhead A, Hager-Ross C, McDonough SM. Test-retest reliability of step counts with the ActivPAL device in common daily activities. Gait Posture. 2010;32(3):386–90.CrossRefPubMed
21.
go back to reference Dowd KP, Harrington DM, Donnelly AE. Criterion and concurrent validity of the activPAL professional physical activity monitor in adolescent females. PLoS One. 2012;7(10):e47633.CrossRefPubMedPubMedCentral Dowd KP, Harrington DM, Donnelly AE. Criterion and concurrent validity of the activPAL professional physical activity monitor in adolescent females. PLoS One. 2012;7(10):e47633.CrossRefPubMedPubMedCentral
22.
go back to reference Ryan CG, Grant PM, Tigbe WW, Granat MH. The validity and reliability of a novel activity monitor as a measure of walking. Br J Sports Med. 2006;40(9):779–84.CrossRefPubMedPubMedCentral Ryan CG, Grant PM, Tigbe WW, Granat MH. The validity and reliability of a novel activity monitor as a measure of walking. Br J Sports Med. 2006;40(9):779–84.CrossRefPubMedPubMedCentral
23.
go back to reference Tudor-Locke C, Sisson SB, Lee SM, Craig CL, Plotnikoff RC, Bauman A. Evaluation of quality of commercial pedometers. Can J Public Health 2006;97:S10-S15.PubMed Tudor-Locke C, Sisson SB, Lee SM, Craig CL, Plotnikoff RC, Bauman A. Evaluation of quality of commercial pedometers. Can J Public Health 2006;97:S10-S15.PubMed
24.
go back to reference Portney L, Watkins M. Foundations of clinical research: applications to practice. Upper Saddle River, NJ: Pearson/Prentice Hall; 2009. Portney L, Watkins M. Foundations of clinical research: applications to practice. Upper Saddle River, NJ: Pearson/Prentice Hall; 2009.
26.
go back to reference Melanson EL, Knoll JR, Bell ML, Hill JO, Nysse LJ, Lanningham-Foster L, et al. Commercially available pedometers: considerations for accurate step counting. Prev Med. 2004;39(2):361–8.CrossRefPubMed Melanson EL, Knoll JR, Bell ML, Hill JO, Nysse LJ, Lanningham-Foster L, et al. Commercially available pedometers: considerations for accurate step counting. Prev Med. 2004;39(2):361–8.CrossRefPubMed
27.
go back to reference De Cocker KA, De Meyer J, De Bourdeaudhuij IM, Cardon GM. Non-traditional wearing positions of pedometers: validity and reliability of the Omron HJ-203-ED pedometer under controlled and free-living conditions. J Sci Med Sport. 2012;15(5):418–24.CrossRefPubMed De Cocker KA, De Meyer J, De Bourdeaudhuij IM, Cardon GM. Non-traditional wearing positions of pedometers: validity and reliability of the Omron HJ-203-ED pedometer under controlled and free-living conditions. J Sci Med Sport. 2012;15(5):418–24.CrossRefPubMed
28.
go back to reference Beets MW, Patton MM, Edwards S. The accuracy of pedometer steps and time during walking in children. Med Sci Sports Exerc. 2005;37(3):513–20.CrossRefPubMed Beets MW, Patton MM, Edwards S. The accuracy of pedometer steps and time during walking in children. Med Sci Sports Exerc. 2005;37(3):513–20.CrossRefPubMed
29.
go back to reference Martin Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.CrossRef Martin Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.CrossRef
30.
go back to reference Atallah L, Lo B, King R, Yang G. Sensor positioning for activity recognition using wearable accelerometers. IEEE Trans Biomed Circuits Syst. 2011;5(4):320–9.CrossRefPubMed Atallah L, Lo B, King R, Yang G. Sensor positioning for activity recognition using wearable accelerometers. IEEE Trans Biomed Circuits Syst. 2011;5(4):320–9.CrossRefPubMed
31.
go back to reference Mannini A, Intille SS, Rosenberger M, Sabatini AM, Haskell W. Activity recognition using a single accelerometer placed at the wrist or ankle. Med Sci Sports Exerc. 2013;45(11):2193–203.CrossRefPubMedPubMedCentral Mannini A, Intille SS, Rosenberger M, Sabatini AM, Haskell W. Activity recognition using a single accelerometer placed at the wrist or ankle. Med Sci Sports Exerc. 2013;45(11):2193–203.CrossRefPubMedPubMedCentral
Metadata
Title
Reliability and validity of ten consumer activity trackers
Authors
Thea J. M. Kooiman
Manon L. Dontje
Siska R. Sprenger
Wim P. Krijnen
Cees P. van der Schans
Martijn de Groot
Publication date
01-12-2015
Publisher
BioMed Central
Published in
BMC Sports Science, Medicine and Rehabilitation / Issue 1/2015
Electronic ISSN: 2052-1847
DOI
https://doi.org/10.1186/s13102-015-0018-5

Other articles of this Issue 1/2015

BMC Sports Science, Medicine and Rehabilitation 1/2015 Go to the issue