Open Access 01-12-2016 | Research

The accuracy of blood pressure measured by arterial line and non-invasive cuff in critically ill children

Authors: Rachel Joffe, Jonathan Duff, Gonzalo Garcia Guerra, Jodie Pugh, Ari R. Joffe

Published in: Critical Care | Issue 1/2016

Login to get access

Abstract

Background

The accuracy of arterial lines (AL) using the flush test or stopcock test has not been described in children, nor has the difference between invasive arterial blood pressure (IABP) versus non-invasive cuff (NIBP) blood pressure.

Methods

After ethics approval and consent, we performed the flush test and stopcock test on AL (to determine over damping, under damping, and optimal damping), and determined the difference (NIBP–IABP) in systolic, diastolic, and mean blood pressure (ΔSBP, ΔDBP, and ΔMAP). The primary outcome was incidence (95 % CI) of optimally damped AL. Predictors of ΔBP (effect size (95 % CI)) were determined using multiple linear regression.

Results

There were 147 AL tests in 100 enrolled patients with mean age 44.7 (SD 56) months, weight 16.8 (SD 18.3) kg, male 59 %, postoperative-cardiovascular 52 %, peripheral-AL 78 %, inotropes 29 %, vasodilators 15 %, and ventilated 73 %. The flush test performed in 66 patients (45 %) showed optimal damping in 30 (46 %; 95 % CI 34, 57 %), over damping in 25 (38 %) and under damping in 11 patients (17 %). The stopcock test was over-damped in 128/146 patients (88 %), with the same damping as the flush test in 24/64 (38 %). In optimally damped (flush test) AL, ΔSBP, ΔDBP, and ΔMAP were 0.8 (SD 12.2), −5.2 (SD 8.7), and −4.9 (7.6) respectively. A second set of AL tests was done 2 h later on the same day in 62 patients; AL damping often changed (10/28 flush tests) and ΔBPs correlated poorly (r = 0.31–0.55). Predictors (effect size) of ΔDBP were vasodilator infusion (15.6 (2.9 to 28.3); p = 0.016) and optimal damping (−7.2 (−12.2 to 2.2); p = 0.005); and of ΔMAP were vasodilator infusion (10.0 (−0.3 to 20.4); p = 0.057) and optimal damping (−4.0 (−8 to 0.1); p = 0.058). There were no independent predictors of damping category (n = 66 flush tests).

Conclusions

Optimally damped AL occur in half of critically ill children, and this is not predictable. There is much variability in ∆BP between NIBP and the gold standard IABP, and this varies even in the same patient on the same day, and is not easily predictable. In critically ill children, NIBP may not be accurate enough to guide management, and more attention to ensuring the AL is optimally damped is needed.
Appendix
Available only for authorised users
Literature
1.
go back to reference Gardner RM. Direct blood pressure measurement-dynamic response requirements. Anesthesiology. 1981;54:227–36.CrossRefPubMed Gardner RM. Direct blood pressure measurement-dynamic response requirements. Anesthesiology. 1981;54:227–36.CrossRefPubMed
2.
go back to reference Schroeder B, Barbeito A, Bar-Yosef S, Mark JB. Chapter 45: Cardiovascular monitoring. In: Miller RD, editor. Miller’s Anesthesia. 8th ed. USA: Saunders; 2014. p. 1345–95. Schroeder B, Barbeito A, Bar-Yosef S, Mark JB. Chapter 45: Cardiovascular monitoring. In: Miller RD, editor. Miller’s Anesthesia. 8th ed. USA: Saunders; 2014. p. 1345–95.
3.
go back to reference Romagnoli S, Ricci Z, Quattrone D, Tofani L, Tujjar O, Villa G, et al. Accuracy of invasive arterial pressure monitoring in cardiovascular patients: an observational study. Critical Care. 2014;18:644.CrossRefPubMedPubMedCentral Romagnoli S, Ricci Z, Quattrone D, Tofani L, Tujjar O, Villa G, et al. Accuracy of invasive arterial pressure monitoring in cardiovascular patients: an observational study. Critical Care. 2014;18:644.CrossRefPubMedPubMedCentral
4.
go back to reference Ribezzo S, Spina E, Bartolomeo SD, Sanson G. Noninvasive techniques for blood pressure measurement are not a reliable alternative to direct measurement: a randomized crossover trial in ICU. Sci World J. 2014;2014:Article ID 353628. Ribezzo S, Spina E, Bartolomeo SD, Sanson G. Noninvasive techniques for blood pressure measurement are not a reliable alternative to direct measurement: a randomized crossover trial in ICU. Sci World J. 2014;2014:Article ID 353628.
5.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.CrossRef
6.
go back to reference Holt TR, Withington DE, Mitchell E. Which pressure to believe? A comparison of direct arterial with indirect blood pressure measurement techniques in the pediatric intensive care unit. Pediatr Crit Care Med. 2011;12:e391–4.CrossRefPubMed Holt TR, Withington DE, Mitchell E. Which pressure to believe? A comparison of direct arterial with indirect blood pressure measurement techniques in the pediatric intensive care unit. Pediatr Crit Care Med. 2011;12:e391–4.CrossRefPubMed
7.
go back to reference Romagnoli S, Ricci Z, De Gaudio AR. Invasive arterial pressure: test it before believing in it. Pediatr Crit Care Med. 2012;13(2):248.CrossRefPubMed Romagnoli S, Ricci Z, De Gaudio AR. Invasive arterial pressure: test it before believing in it. Pediatr Crit Care Med. 2012;13(2):248.CrossRefPubMed
8.
go back to reference Manios E, Vemmos K, Tsivgoulis G, Barlas G, Koroboki E, Spengos K, et al. Comparison of noninvasive oscillometric and intra-arterial blood pressure measurements in hyperacute stroke. Blood Press Monit. 2007;12(3):149–56.CrossRefPubMed Manios E, Vemmos K, Tsivgoulis G, Barlas G, Koroboki E, Spengos K, et al. Comparison of noninvasive oscillometric and intra-arterial blood pressure measurements in hyperacute stroke. Blood Press Monit. 2007;12(3):149–56.CrossRefPubMed
9.
go back to reference Rutten AJ, Ilsley AH, Skowronski GA, Runciman WB. A comparative study of the measurement of mean arterial blood pressure using automatic oscillometers, arterial cannulation and auscultation. Anaesth Intensive Care. 1986;14(1):58–65.PubMed Rutten AJ, Ilsley AH, Skowronski GA, Runciman WB. A comparative study of the measurement of mean arterial blood pressure using automatic oscillometers, arterial cannulation and auscultation. Anaesth Intensive Care. 1986;14(1):58–65.PubMed
10.
go back to reference Lakhal K, Ehrmann S, Runge I, Legras A, Dequin P-F, Mercier E, et al. Tracking hypotension and dynamic changes in arterial blood pressure with brachial cuff measurements. Anesth Analg. 2009;109:494–501.CrossRefPubMed Lakhal K, Ehrmann S, Runge I, Legras A, Dequin P-F, Mercier E, et al. Tracking hypotension and dynamic changes in arterial blood pressure with brachial cuff measurements. Anesth Analg. 2009;109:494–501.CrossRefPubMed
11.
go back to reference Kimble KJ, Darnall Jr RA, Yelderman M, Ariagno RL, Ream AK. An automated oscillometric technique for estimating mean arterial pressure in critically ill newborns. Anesthesiology. 1981;54(5):423–5.CrossRefPubMed Kimble KJ, Darnall Jr RA, Yelderman M, Ariagno RL, Ream AK. An automated oscillometric technique for estimating mean arterial pressure in critically ill newborns. Anesthesiology. 1981;54(5):423–5.CrossRefPubMed
12.
go back to reference Chia F, Ang AT, Wong T-W, Tan KW, Fung K, Lee J, et al. Reliability of the Dinamap non-invasive monitor in the measurement of blood pressure of ill Asian newborns. Clin Pediatrics. 1990;29(5):262–7.CrossRef Chia F, Ang AT, Wong T-W, Tan KW, Fung K, Lee J, et al. Reliability of the Dinamap non-invasive monitor in the measurement of blood pressure of ill Asian newborns. Clin Pediatrics. 1990;29(5):262–7.CrossRef
13.
go back to reference Lang SM, Giuliano Jr JS, Carroll CL, Rosenkrantz TS, Eisenfeld L. Neonatal/infant validation study of the CAS model 740 noninvasive blood pressure monitor with the Orion/MaxIQ NIBP module. Blood Press Monit. 2014;19:180–2.CrossRefPubMed Lang SM, Giuliano Jr JS, Carroll CL, Rosenkrantz TS, Eisenfeld L. Neonatal/infant validation study of the CAS model 740 noninvasive blood pressure monitor with the Orion/MaxIQ NIBP module. Blood Press Monit. 2014;19:180–2.CrossRefPubMed
14.
go back to reference Clark JA, Lieh-Lai MW, Sarnaik A, Mattoo TK. Discrepancies between direct and indirect blood pressure measurments using various recommendations for arm cuff selection. Pediatrics. 2002;110(5):920–3.CrossRefPubMed Clark JA, Lieh-Lai MW, Sarnaik A, Mattoo TK. Discrepancies between direct and indirect blood pressure measurments using various recommendations for arm cuff selection. Pediatrics. 2002;110(5):920–3.CrossRefPubMed
Metadata
Title
The accuracy of blood pressure measured by arterial line and non-invasive cuff in critically ill children
Authors
Rachel Joffe
Jonathan Duff
Gonzalo Garcia Guerra
Jodie Pugh
Ari R. Joffe
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Critical Care / Issue 1/2016
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/s13054-016-1354-x

Other articles of this Issue 1/2016

Critical Care 1/2016 Go to the issue