Published in: Journal of NeuroEngineering and Rehabilitation 1/2022

Open Access 01-12-2022 | Research

Human-prosthesis cooperation: combining adaptive prosthesis control with visual feedback guided gait

Authors: Bretta L. Fylstra, I-Chieh Lee, Minhan Li, Michael D. Lewek, He Huang

Published in: Journal of NeuroEngineering and Rehabilitation | Issue 1/2022

Login to get access

Abstract

Background

Personalizing prosthesis control is often structured as human-in-the-loop optimization. However, gait performance is influenced by both human control and intelligent prosthesis control. Hence, we need to consider both human and prosthesis control, and their cooperation, to achieve desired gait patterns. In this study, we developed a novel paradigm that engages human gait control via user-fed visual feedback (FB) of stance time to cooperate with automatic prosthesis control tuning. Three initial questions were studied: (1) does user control of gait timing (via visual FB) help the prosthesis tuning algorithm to converge faster? (2) in turn, does the prosthesis control influence the user’s ability to reach and maintain the target stance time defined by the feedback? and (3) does the prosthesis control parameters tuned with extended stance time on prosthesis side allow the user to maintain this potentially beneficial behavior even after feedback is removed (short- and long-term retention)?

Methods

A reinforcement learning algorithm was used to achieve prosthesis control to meet normative knee kinematics in walking. A visual FB system cued the user to control prosthesis-side stance time to facilitate the prosthesis tuning goal. Seven individuals without amputation (AB) and four individuals with transfemoral amputation (TFA) walked with a powered knee prosthesis on a treadmill. Participants completed prosthesis auto-tuning with three visual feedback conditions: no FB, self-selected stance time FB (SS FB), and increased stance time FB (Inc FB). The retention of FB effects was studied by comparing the gait performance across three different prosthesis controls, tuned with different visual FB.

Results

(1) Human control of gait timing reduced the tuning duration in individuals without amputation, but not for individuals with TFA. (2) The change of prosthesis control did not influence users’ ability to reach and maintain the visual FB goal. (3) All participants increased their prosthesis-side stance time with the feedback and maintain it right after feedback was removed. However, in the post-test, the prosthesis control parameters tuned with visual FB only supported a few participants with longer stance time and better stance time symmetry.

Conclusions

The study provides novel insights on human-prosthesis interaction when cooperating in walking, which may guide the future successful adoption of this paradigm in prosthesis control personalization or human-in-the-loop optimization to improve the prosthesis user’s gait performance.
Literature
1.
go back to reference Au SK, Herr H, Weber J, Martinez-Villalpando EC. Powered ankle-foot prosthesis for the improvement of amputee ambulation. In: 2007 29th annual international conference of the IEEE engineering in medicine and biology society 2007. 2007. p. 3020–6. Au SK, Herr H, Weber J, Martinez-Villalpando EC. Powered ankle-foot prosthesis for the improvement of amputee ambulation. In: 2007 29th annual international conference of the IEEE engineering in medicine and biology society 2007. 2007. p. 3020–6.
2.
go back to reference Sup F, Bohara A, Goldfarb M. Design and control of a powered transfemoral prosthesis. Int J Robot Res. 2008;27(2):263–73.CrossRef Sup F, Bohara A, Goldfarb M. Design and control of a powered transfemoral prosthesis. Int J Robot Res. 2008;27(2):263–73.CrossRef
3.
go back to reference Liu M, Zhang F, Datseris P, Huang HH. Improving finite state impedance control of active-transfemoral prosthesis using dempster-shafer based state transition rules. J Intell Robot Syst. 2014;76(3):461–74.CrossRef Liu M, Zhang F, Datseris P, Huang HH. Improving finite state impedance control of active-transfemoral prosthesis using dempster-shafer based state transition rules. J Intell Robot Syst. 2014;76(3):461–74.CrossRef
4.
go back to reference Martinez-Villalpando EC, Herr H. Agonist-antagonist active knee prosthesis: a preliminary study in level-ground walking. J Rehabil Res Dev. 2009;46(3):361–73.CrossRef Martinez-Villalpando EC, Herr H. Agonist-antagonist active knee prosthesis: a preliminary study in level-ground walking. J Rehabil Res Dev. 2009;46(3):361–73.CrossRef
5.
go back to reference Rouse EJ, Mooney LM, Herr HM. Clutchable series-elastic actuator: implications for prosthetic knee design. Int J Robot Res. 2014;33(13):1611–25.CrossRef Rouse EJ, Mooney LM, Herr HM. Clutchable series-elastic actuator: implications for prosthetic knee design. Int J Robot Res. 2014;33(13):1611–25.CrossRef
6.
go back to reference Mendez J, Hood S, Gunnel A, Lenzi T. Powered knee and ankle prosthesis with indirect volitional swing control enables level-ground walking and crossing over obstacles. Sci Robot. 2020;5(44): eaba6635.CrossRef Mendez J, Hood S, Gunnel A, Lenzi T. Powered knee and ankle prosthesis with indirect volitional swing control enables level-ground walking and crossing over obstacles. Sci Robot. 2020;5(44): eaba6635.CrossRef
7.
go back to reference Elery T, Rezazadeh S, Nesler C, Gregg RD. Design and validation of a powered knee-ankle prosthesis with high-torque, low-impedance actuators. IEEE Trans Robot. 2020;36(6):1649–68.CrossRef Elery T, Rezazadeh S, Nesler C, Gregg RD. Design and validation of a powered knee-ankle prosthesis with high-torque, low-impedance actuators. IEEE Trans Robot. 2020;36(6):1649–68.CrossRef
8.
go back to reference Quesada RE, Caputo JM, Collins SH. Increasing ankle push-off work with a powered prosthesis does not necessarily reduce metabolic rate for transtibial amputees. J Biomech. 2016;49(14):3452–9.CrossRef Quesada RE, Caputo JM, Collins SH. Increasing ankle push-off work with a powered prosthesis does not necessarily reduce metabolic rate for transtibial amputees. J Biomech. 2016;49(14):3452–9.CrossRef
9.
go back to reference Varol HA, Sup F, Goldfarb M. Multiclass real-time intent recognition of a powered lower limb prosthesis. IEEE Trans Biomed Eng. 2010;57(3):542–51.CrossRef Varol HA, Sup F, Goldfarb M. Multiclass real-time intent recognition of a powered lower limb prosthesis. IEEE Trans Biomed Eng. 2010;57(3):542–51.CrossRef
10.
go back to reference Huang H, Crouch DL, Liu M, Sawicki GS, Wang D. A cyber expert system for auto-tuning powered prosthesis impedance control parameters. Ann Biomed Eng. 2016;44(5):1613–24.CrossRef Huang H, Crouch DL, Liu M, Sawicki GS, Wang D. A cyber expert system for auto-tuning powered prosthesis impedance control parameters. Ann Biomed Eng. 2016;44(5):1613–24.CrossRef
11.
go back to reference Welker CG, Voloshina AS, Chiu VL, Collins SH. Shortcomings of human-in-the-loop optimization of an ankle-foot prosthesis emulator: a case series. R Soc Open Sci. 2021;8(5): 202020.CrossRef Welker CG, Voloshina AS, Chiu VL, Collins SH. Shortcomings of human-in-the-loop optimization of an ankle-foot prosthesis emulator: a case series. R Soc Open Sci. 2021;8(5): 202020.CrossRef
12.
go back to reference Li M, Wen Y, Gao X, Si J, Huang H. Towards expedited impedance tuning of a robotic prosthesis for personalized gait assistance by reinforcement learning control. IEEE Trans Robot. 2021;38:1–10. Li M, Wen Y, Gao X, Si J, Huang H. Towards expedited impedance tuning of a robotic prosthesis for personalized gait assistance by reinforcement learning control. IEEE Trans Robot. 2021;38:1–10.
13.
go back to reference Wen Y, Si J, Brandt A, Gao X, Huang HH. Online reinforcement learning control for the personalization of a robotic knee prosthesis. IEEE Trans Cybern. 2020;50(6):2346–56.CrossRef Wen Y, Si J, Brandt A, Gao X, Huang HH. Online reinforcement learning control for the personalization of a robotic knee prosthesis. IEEE Trans Cybern. 2020;50(6):2346–56.CrossRef
14.
go back to reference Fylstra BL, Lee IC, Huang S, Brandt A, Lewek MD, Huang HH. Human-prosthesis coordination: a preliminary study exploring coordination with a powered ankle-foot prosthesis. Clin Biomech. 2020;80: 105171. CrossRef Fylstra BL, Lee IC, Huang S, Brandt A, Lewek MD, Huang HH. Human-prosthesis coordination: a preliminary study exploring coordination with a powered ankle-foot prosthesis. Clin Biomech. 2020;80: 105171. CrossRef
15.
go back to reference Malcolm P, Quesada RE, Caputo JM, Collins SH. The influence of push-off timing in a robotic ankle-foot prosthesis on the energetics and mechanics of walking. J Neuroeng Rehabil. 2015;12:21.CrossRef Malcolm P, Quesada RE, Caputo JM, Collins SH. The influence of push-off timing in a robotic ankle-foot prosthesis on the energetics and mechanics of walking. J Neuroeng Rehabil. 2015;12:21.CrossRef
16.
go back to reference Huang HH, Si J, Brandt A, Li M. Taking both sides: seeking symbiosis between intelligent prostheses and human motor control during locomotion. Curr Opin Biomed Eng. 2021;20: 100314.CrossRef Huang HH, Si J, Brandt A, Li M. Taking both sides: seeking symbiosis between intelligent prostheses and human motor control during locomotion. Curr Opin Biomed Eng. 2021;20: 100314.CrossRef
17.
go back to reference Highsmith MJ, Andrews CR, Millman C, Fuller A, Kahle JT, Klenow TD, et al. Gait training interventions for lower extremity amputees: a systematic literature review. Technol Innov. 2016;18(2–3):99–113.CrossRef Highsmith MJ, Andrews CR, Millman C, Fuller A, Kahle JT, Klenow TD, et al. Gait training interventions for lower extremity amputees: a systematic literature review. Technol Innov. 2016;18(2–3):99–113.CrossRef
18.
go back to reference Escamilla-Nunez R, Michelini A, Andrysek J. Biofeedback systems for gait rehabilitation of individuals with lower-limb amputation: a systematic review. Sensors. 2020;20(6):1628.CrossRef Escamilla-Nunez R, Michelini A, Andrysek J. Biofeedback systems for gait rehabilitation of individuals with lower-limb amputation: a systematic review. Sensors. 2020;20(6):1628.CrossRef
19.
go back to reference Brandt A, Riddick W, Stallrich J, Lewek M, Huang HH. Effects of extended powered knee prosthesis stance time via visual feedback on gait symmetry of individuals with unilateral amputation: a preliminary study. J Neuroeng Rehabil. 2019;16(1):112.CrossRef Brandt A, Riddick W, Stallrich J, Lewek M, Huang HH. Effects of extended powered knee prosthesis stance time via visual feedback on gait symmetry of individuals with unilateral amputation: a preliminary study. J Neuroeng Rehabil. 2019;16(1):112.CrossRef
20.
go back to reference Brandt A, Huang HH. Effects of extended stance time on a powered knee prosthesis and gait symmetry on the lateral control of balance during walking in individuals with unilateral amputation. J Neuroeng Rehabil. 2019;16(1):151.CrossRef Brandt A, Huang HH. Effects of extended stance time on a powered knee prosthesis and gait symmetry on the lateral control of balance during walking in individuals with unilateral amputation. J Neuroeng Rehabil. 2019;16(1):151.CrossRef
21.
go back to reference Wen Y, Li M, Si J, Huang H. Wearer-prosthesis interaction for symmetrical gait: a study enabled by reinforcement learning prosthesis control. IEEE Trans Neural Syst Rehabil Eng. 2020;28(4):904–13.CrossRef Wen Y, Li M, Si J, Huang H. Wearer-prosthesis interaction for symmetrical gait: a study enabled by reinforcement learning prosthesis control. IEEE Trans Neural Syst Rehabil Eng. 2020;28(4):904–13.CrossRef
22.
go back to reference Malcolm P, Fiers P, Segers V, Van Caekenberghe I, Lenoir M, De Clercq D. Experimental study on the role of the ankle push off in the walk-to-run transition by means of a powered ankle-foot-exoskeleton. Gait Posture. 2009;30(3):322–7.CrossRef Malcolm P, Fiers P, Segers V, Van Caekenberghe I, Lenoir M, De Clercq D. Experimental study on the role of the ankle push off in the walk-to-run transition by means of a powered ankle-foot-exoskeleton. Gait Posture. 2009;30(3):322–7.CrossRef
23.
go back to reference Bellmann M, Blumentritt S, Pusch M, Schmalz T, Schönemeier M. The 3D LASAR—a new generation of static analysis for optimising prosthetic and orthotic alignment. Orthopädie Technik. 2017;10:17. Bellmann M, Blumentritt S, Pusch M, Schmalz T, Schönemeier M. The 3D LASAR—a new generation of static analysis for optimising prosthetic and orthotic alignment. Orthopädie Technik. 2017;10:17.
24.
go back to reference Kadaba MP, Ramakrishnan HK, Wootten ME. Measurement of lower extremity kinematics during level walking. J Orthop Res. 1990;8(3):383–92.CrossRef Kadaba MP, Ramakrishnan HK, Wootten ME. Measurement of lower extremity kinematics during level walking. J Orthop Res. 1990;8(3):383–92.CrossRef
25.
go back to reference Hof AL, van Bockel RM, Schoppen T, Postema K. Control of lateral balance in walking. Experimental findings in normal subjects and above-knee amputees. Gait Posture. 2007;25(2):250–8.CrossRef Hof AL, van Bockel RM, Schoppen T, Postema K. Control of lateral balance in walking. Experimental findings in normal subjects and above-knee amputees. Gait Posture. 2007;25(2):250–8.CrossRef
26.
go back to reference Li M, Zhong B, Liu Z, Lee IC, Fylstra BL, Lobaton E, et al., editors. Gaze fixation comparisons between amputees and able-bodied individuals in approaching stairs and level-ground transitions: a pilot study. In: 2019 41st annual international conference of the IEEE engineering in medicine and biology society (EMBC), July 2019; 2019. p. 23–7. Li M, Zhong B, Liu Z, Lee IC, Fylstra BL, Lobaton E, et al., editors. Gaze fixation comparisons between amputees and able-bodied individuals in approaching stairs and level-ground transitions: a pilot study. In: 2019 41st annual international conference of the IEEE engineering in medicine and biology society (EMBC), July 2019; 2019. p. 23–7.
27.
go back to reference Knaepen K, Marusic U, Crea S, Guerrero CDR, Vitiello N, Pattyn N, et al. Psychophysiological response to cognitive workload during symmetrical, asymmetrical and dual-task walking. Hum Mov Sci. 2015;40:248–63.CrossRef Knaepen K, Marusic U, Crea S, Guerrero CDR, Vitiello N, Pattyn N, et al. Psychophysiological response to cognitive workload during symmetrical, asymmetrical and dual-task walking. Hum Mov Sci. 2015;40:248–63.CrossRef
28.
go back to reference Lee IC, Pacheco MM, Newell KM. The precision demands of viewing distance modulate postural coordination and control. Hum Mov Sci. 2019;66:425–39.CrossRef Lee IC, Pacheco MM, Newell KM. The precision demands of viewing distance modulate postural coordination and control. Hum Mov Sci. 2019;66:425–39.CrossRef
29.
go back to reference Liu W, Zhong J, Wu R, Fylstra BL, Si J, Huang HH. Inferring human-robot performance objectives during locomotion using inverse reinforcement learning and inverse optimal control. IEEE Robot Autom Lett. 2022;7(2):2549–56.CrossRef Liu W, Zhong J, Wu R, Fylstra BL, Si J, Huang HH. Inferring human-robot performance objectives during locomotion using inverse reinforcement learning and inverse optimal control. IEEE Robot Autom Lett. 2022;7(2):2549–56.CrossRef
30.
go back to reference Ingraham KA, Remy CD, Rouse EJ. The role of user preference in the customized control of robotic exoskeletons. Sci Robot. 2022;7(64): eabj3487.CrossRef Ingraham KA, Remy CD, Rouse EJ. The role of user preference in the customized control of robotic exoskeletons. Sci Robot. 2022;7(64): eabj3487.CrossRef
31.
go back to reference Zhang J, Fiers P, Witte KA, Jackson RW, Poggensee KL, Atkeson CG, et al. Human-in-the-loop optimization of exoskeleton assistance during walking. Science. 2017;356(6344):1280–4.CrossRef Zhang J, Fiers P, Witte KA, Jackson RW, Poggensee KL, Atkeson CG, et al. Human-in-the-loop optimization of exoskeleton assistance during walking. Science. 2017;356(6344):1280–4.CrossRef
Metadata
Title
Human-prosthesis cooperation: combining adaptive prosthesis control with visual feedback guided gait
Authors
Bretta L. Fylstra
I-Chieh Lee
Minhan Li
Michael D. Lewek
He Huang
Publication date
01-12-2022
Publisher
BioMed Central
Published in
Journal of NeuroEngineering and Rehabilitation / Issue 1/2022
Electronic ISSN: 1743-0003
DOI
https://doi.org/10.1186/s12984-022-01118-z

Other articles of this Issue 1/2022

Journal of NeuroEngineering and Rehabilitation 1/2022 Go to the issue