Effects of disinfection and sterilization on the dimensional changes and mechanical properties of 3D printed surgical guides for implant therapy – pilot study

Authors: Gréta Török, Péter Gombocz, Eszter Bognár, Péter Nagy, Elek Dinya, Barbara Kispélyi, Péter Hermann

Published in: BMC Oral Health | Issue 1/2020

Login to get access

Abstract

Background

The purpose of this research was to investigate the effects of disinfection and three different sterilization methods on the dimensional changes and mechanical properties of three-dimensional (3D) printed surgical guide for implant therapy. The objective was to assess the effects of sterilization procedures in 3D printed drill guide templates with destructive and non-destructive material testing.

Methods

Fifteen identical drill guide templates were produced using a 3D printer. The surgical guides were classified into five groups: three controls, three disinfected (4% Gigasept®, 60 min), three plasma sterilized, three autoclave sterilized (+ 1 bar, 121 °C, 20 min), and three autoclave sterilized (+ 2 bar, 134 °C, 10 min).
The templates were digitalized with a Steinbichler SCAN ST 3D scanner. Length was measured under an SZX16 stereomicroscope. A scanning electron microscope was used to study the surface morphology of the drill templates. The hardness, and flexural and compressive strength were measured to assess any changes in the physical characteristics of the material caused by sterilization. The drill guide templates were also examined with a Dage XiDAT 6600 X-ray. During the X-ray examinations, the following parameters were used: 100 kV voltage, 128 AVG averaging, 0.8 W power. One-way analysis of variance (ANOVA) was used to detect the difference between groups.

Results

Evaluation of the hardness measurements of the various specimens shows that the hardness of the material was not changed by the plasma sterilization (p = 0.0680), steam sterilization on 121 °C (p = 0.6033) or disinfection process (p = 0.1399). The statistical analysis revealed significant difference in hardness strength of the autoclave sterilized (134 °C) specimens (p = 0.0002). There was no significant difference between the goups regarding the scanning electron microscopic and stereomicroscopic examinations. There was no significant difference regarding the X-ray visibility of the templates to the effect of the disinfection (p = 0.7844), plasma sterilization (p = 0.4091) and steam sterilization on 121 °C (p = 0.9277) and steam sterilization on 131 °C (p = 0.093). The effect of the sterilization was the same in case of both flexural and compressive strength of the material.

Conclusions

Our findings indicate that plasma sterilization and steam sterilization at 121 °C were both suitable for sterilizing the tested 3D printed surgical guides.
Literature
1.
go back to reference Moslehifard E, Nokar S. Designing a custom made gauge device for application in the access hole correction in the dental implant surgical guide. Indian Prosthodont Soc. 2012;12:123–9.CrossRef Moslehifard E, Nokar S. Designing a custom made gauge device for application in the access hole correction in the dental implant surgical guide. Indian Prosthodont Soc. 2012;12:123–9.CrossRef
2.
go back to reference De Vico G, Spinelli D, Bonino M, Schiavetti R, Pozzi A, Ottria L. Computer-assisted virtual treatment planning combined with flapless surgery and immediate loading in the Rehabilitaion of partial Edentulie. Oral Implantol. 2012;5:3–10. De Vico G, Spinelli D, Bonino M, Schiavetti R, Pozzi A, Ottria L. Computer-assisted virtual treatment planning combined with flapless surgery and immediate loading in the Rehabilitaion of partial Edentulie. Oral Implantol. 2012;5:3–10.
3.
go back to reference Fortin T, Isidori M, Blanchet E, Perriat M, Bouchet H, Coudert JL. An image-guided system–drilled surgical template and trephine guide pin to make treatment of completely edentulous patients easier: a clinical report on immediate loading. Clin Implant Dent Relat Res. 2004;6:111–9.CrossRef Fortin T, Isidori M, Blanchet E, Perriat M, Bouchet H, Coudert JL. An image-guided system–drilled surgical template and trephine guide pin to make treatment of completely edentulous patients easier: a clinical report on immediate loading. Clin Implant Dent Relat Res. 2004;6:111–9.CrossRef
4.
go back to reference D'haese J, Van De Velde T, Komiyama A, Hultin M, De Bruyn H. Accuracy and complications using computer-designed Stereolithographic surgical guides for Oral rehabilitation by means of dental implants: a review of the literature. Clin Implant Dent Relat Res. 2012;14:321–35.CrossRef D'haese J, Van De Velde T, Komiyama A, Hultin M, De Bruyn H. Accuracy and complications using computer-designed Stereolithographic surgical guides for Oral rehabilitation by means of dental implants: a review of the literature. Clin Implant Dent Relat Res. 2012;14:321–35.CrossRef
5.
go back to reference Sennhenn-Kirchner S, Weustermann S, Mergeryan H, Jacobs HG, von Zepelin M B, Kirchner B. Preoperative sterilization and disinfection of drillguide templates. Clin Oral Investig. 2008;12:179–87.CrossRef Sennhenn-Kirchner S, Weustermann S, Mergeryan H, Jacobs HG, von Zepelin M B, Kirchner B. Preoperative sterilization and disinfection of drillguide templates. Clin Oral Investig. 2008;12:179–87.CrossRef
6.
go back to reference Rutala WA, Weber DJ and the Healthcare Infection Control Practices Advisory Committee (HICPAC). Guideline for Disinfection and Sterilization in Healthcare Facilities. CDC. 2008. p10–12;58–67. Rutala WA, Weber DJ and the Healthcare Infection Control Practices Advisory Committee (HICPAC). Guideline for Disinfection and Sterilization in Healthcare Facilities. CDC. 2008. p10–12;58–67.
10.
14.
go back to reference Dobránszky J, Gy R, Bognár E, Kovács R, Bitay E. New method for evaluating the visibility of coronary stents. Acta Polytechnica Hungarica. 2014;11:81–94. Dobránszky J, Gy R, Bognár E, Kovács R, Bitay E. New method for evaluating the visibility of coronary stents. Acta Polytechnica Hungarica. 2014;11:81–94.
15.
go back to reference Neppelenbroek KH, Pavarina AC, Vergani CE, Giampaolo ET. Hardness of heat-polymerized acrylic resins after disinfection and long-term water immersion. J Prosthet Dent. 2005;93:171–6.CrossRef Neppelenbroek KH, Pavarina AC, Vergani CE, Giampaolo ET. Hardness of heat-polymerized acrylic resins after disinfection and long-term water immersion. J Prosthet Dent. 2005;93:171–6.CrossRef
16.
go back to reference Campanha NH, Pavarina AC, Vergani CE, Machado A. Effect of microwave sterilization and water storage on the Vickers hardness ofacrylic resin denture teeth. J Prosthet Dent. 2005;93:483–7.CrossRef Campanha NH, Pavarina AC, Vergani CE, Machado A. Effect of microwave sterilization and water storage on the Vickers hardness ofacrylic resin denture teeth. J Prosthet Dent. 2005;93:483–7.CrossRef
17.
go back to reference Phull S, Arora A, Yashendra. Sterilization & Disinfection in Prosthodontics. Indian J Dent Sci. 2014;6:112–6. Phull S, Arora A, Yashendra. Sterilization & Disinfection in Prosthodontics. Indian J Dent Sci. 2014;6:112–6.
18.
go back to reference Kohn WG, Collins AS, Cleveland JL, Harte JA, Eklund KJ, Malvitz DM. Guidelines for infection control in dental health-care settings—2003. MMWR Recomm Rep. 2003;52(RR 17):1–76.PubMed Kohn WG, Collins AS, Cleveland JL, Harte JA, Eklund KJ, Malvitz DM. Guidelines for infection control in dental health-care settings—2003. MMWR Recomm Rep. 2003;52(RR 17):1–76.PubMed
19.
go back to reference Rutala WA, Weber D. Disinfection and sterilization: an overview. Am J Infect Control. 2013;41:S2–5.CrossRef Rutala WA, Weber D. Disinfection and sterilization: an overview. Am J Infect Control. 2013;41:S2–5.CrossRef
20.
go back to reference Arısan V, Bölükbaşı N, Öksüz L. Computer-assisted flapless implant placement reduces the incidence of surgery-related bacteremia. Clin Oral Investig. 2013;17:1985–93.CrossRef Arısan V, Bölükbaşı N, Öksüz L. Computer-assisted flapless implant placement reduces the incidence of surgery-related bacteremia. Clin Oral Investig. 2013;17:1985–93.CrossRef
21.
go back to reference Rutala WA, Weber DJ. Reprocessing semicritical items: current issues and new technologies. Am J Infect Control. 2016;44:e53–62.CrossRef Rutala WA, Weber DJ. Reprocessing semicritical items: current issues and new technologies. Am J Infect Control. 2016;44:e53–62.CrossRef
23.
go back to reference Smith PN, Palenik CJ, Blanchard SB. Microbial contamination and the sterilization/disinfection of surgical guides used in the placement of endosteal implant. Int J Oral Maxillofac Implants. 2011;26(2):274–81.PubMed Smith PN, Palenik CJ, Blanchard SB. Microbial contamination and the sterilization/disinfection of surgical guides used in the placement of endosteal implant. Int J Oral Maxillofac Implants. 2011;26(2):274–81.PubMed
24.
go back to reference Marei HF, Alshaia A, Alarifi S, Almasoud N, Abdelhady A. Effect of steam heat sterilization on the accuracy of 3D printed surgical guides. Implant Dent. 2019;28(4):372–7.CrossRef Marei HF, Alshaia A, Alarifi S, Almasoud N, Abdelhady A. Effect of steam heat sterilization on the accuracy of 3D printed surgical guides. Implant Dent. 2019;28(4):372–7.CrossRef
25.
go back to reference International Organization for Standardization, Sterilization of health care products — Low temperature vaporized hydrogen peroxide — Requirements for the development, validation and routine control of a sterilization process for medical devices (ISO/NP 22441). Retrieved from https://www.iso.org/standard/73214.html International Organization for Standardization, Sterilization of health care products — Low temperature vaporized hydrogen peroxide — Requirements for the development, validation and routine control of a sterilization process for medical devices (ISO/NP 22441). Retrieved from https://​www.​iso.​org/​standard/​73214.​html
26.
go back to reference Wallace CA. New developments in disinfection and sterilization. Am J Infect Control. 2016;44:23–7.CrossRef Wallace CA. New developments in disinfection and sterilization. Am J Infect Control. 2016;44:23–7.CrossRef
28.
go back to reference Rutala WA, Gergen MF, Weber DJ. Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid. Am J Infect Control. 1998;26:393–8.CrossRef Rutala WA, Gergen MF, Weber DJ. Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid. Am J Infect Control. 1998;26:393–8.CrossRef
29.
go back to reference Shaheen E, Alhelwani A, Van De Casteele E, Politis C, Jacobs R. Evaluation of Dimensional Changes of 3D Printed Models After Sterilization: A Pilot Study. Open Dent J. 2018;12(Suppl-1, M3):72–9.CrossRef Shaheen E, Alhelwani A, Van De Casteele E, Politis C, Jacobs R. Evaluation of Dimensional Changes of 3D Printed Models After Sterilization: A Pilot Study. Open Dent J. 2018;12(Suppl-1, M3):72–9.CrossRef
30.
go back to reference Figl M, Weber C, Assadan O, Toma CD, Traxler H, Seemann R, Guevara-Rojas G, Pöschl WP, Ewers R, Schicho K. Splint sterilization-a potencial registration hazard in computer-assisted surgery. J Oral Maxillofac Surg. 2012;70:966–71.CrossRef Figl M, Weber C, Assadan O, Toma CD, Traxler H, Seemann R, Guevara-Rojas G, Pöschl WP, Ewers R, Schicho K. Splint sterilization-a potencial registration hazard in computer-assisted surgery. J Oral Maxillofac Surg. 2012;70:966–71.CrossRef
Metadata
Title
Effects of disinfection and sterilization on the dimensional changes and mechanical properties of 3D printed surgical guides for implant therapy – pilot study
Authors
Gréta Török
Péter Gombocz
Eszter Bognár
Péter Nagy
Elek Dinya
Barbara Kispélyi
Péter Hermann
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Oral Health / Issue 1/2020
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-020-1005-0

Other articles of this Issue 1/2020

BMC Oral Health 1/2020 Go to the issue