Published in: BMC Pediatrics 1/2022

Open Access 01-12-2022 | Research

Variability of respiratory rate measurements in neonates- every minute counts

Authors: Catherine Muthoni Njeru, J. Mark Ansermino, William M. Macharia, Dustin T. Dunsmuir

Published in: BMC Pediatrics | Issue 1/2022

Login to get access

Abstract

Background

Respiratory rate is difficult to measure, especially in neonates who have an irregular breathing pattern. The World Health Organisation recommends a one-minute count, but there is limited data to support this length of observation. We sought to evaluate agreement between the respiratory rate (RR) derived from capnography in neonates, over 15 s, 30 s, 120 s and 300 s, against the recommended 60 s.

Methods

Neonates at two hospitals in Nairobi were recruited and had capnograph waveforms recorded using the Masimo Rad 97. A single high quality 5 min epoch was randomly chosen from each subject. For each selected epoch, the mean RR was calculated using a breath-detection algorithm applied to the waveform. The RR in the first 60 s was compared to the mean RR measured over the first 15 s, 30 s, 120 s, full 300 s, and last 60 s. We calculated bias and limits of agreement for each comparison and used Bland-Altman plots for visual comparisons.

Results

A total of 306 capnographs were analysed from individual subjects. The subjects had a median gestation age of 39 weeks with slightly more females (52.3%) than males (47.7%). The majority of the population were term neonates (70.1%) with 39 (12.8%) having a primary respiratory pathology. There was poor agreement between all the comparisons based on the limits of agreement [confidence interval], ranging between 11.9 [− 6.79 to 6.23] breaths per minute in the one versus 2 min comparison, and 34.7 [− 17.59 to 20.53] breaths per minute in the first versus last minute comparison. Worsening agreement was observed in plots with higher RRs.

Conclusions

Neonates have high variability of RR, even over a short period of time. A slight degradation in the agreement is noted over periods shorter than 1 min. However, this is smaller than observations done 3 min apart in the same subject. Longer periods of observation also reduce agreement. For device developers, precise synchronization is needed when comparing devices to reduce the impact of RR variation. For clinicians, where possible, continuous or repeated monitoring of neonates would be preferable to one time RR measurements.
Literature
1.
go back to reference Lovett PB, Buchwald JM, Stürmann K, Bijur P. The vexatious vital: neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage. Ann Emerg Med. 2005;45(1):68–76.CrossRef Lovett PB, Buchwald JM, Stürmann K, Bijur P. The vexatious vital: neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage. Ann Emerg Med. 2005;45(1):68–76.CrossRef
2.
go back to reference Curzi-Dascalova L, Gaudebout C, Dreyfus-Brisac C. Respiratory frequencies of sleeping infants during the first months of life: correlations between values in different sleep states. Early Hum Dev. 1981;5(1):39–54.CrossRef Curzi-Dascalova L, Gaudebout C, Dreyfus-Brisac C. Respiratory frequencies of sleeping infants during the first months of life: correlations between values in different sleep states. Early Hum Dev. 1981;5(1):39–54.CrossRef
4.
go back to reference World Health Organization (WHO). Fourth Programme Report, 1988-1989: ARI Programme for Control of Acute Respiratory Infections 1990. World Health Organization (WHO). Fourth Programme Report, 1988-1989: ARI Programme for Control of Acute Respiratory Infections 1990.
5.
go back to reference Health MOF. Integrated management of newborn & childhood ILLNESSES ( IMNCI ) A guide for healthcare workers; 2018. Health MOF. Integrated management of newborn & childhood ILLNESSES ( IMNCI ) A guide for healthcare workers; 2018.
6.
go back to reference Ansermino JM, Dumont G, Ginsburg AS. How uncertain is our reference standard for respiratory rate measurement? Am J Respir Crit Care Med. 2019;199(8):1036–7.CrossRef Ansermino JM, Dumont G, Ginsburg AS. How uncertain is our reference standard for respiratory rate measurement? Am J Respir Crit Care Med. 2019;199(8):1036–7.CrossRef
8.
go back to reference Ermer S, Brewer L, Orr J, Egan TD, Johnson K. Comparison of 7 different sensors for detecting low respiratory rates using a single breath detection algorithm in nonintubated, Sedated Volunteers. Anesth Analg. 2019;129(2):399–408.CrossRef Ermer S, Brewer L, Orr J, Egan TD, Johnson K. Comparison of 7 different sensors for detecting low respiratory rates using a single breath detection algorithm in nonintubated, Sedated Volunteers. Anesth Analg. 2019;129(2):399–408.CrossRef
9.
go back to reference Ginsburg AS, Nkwopara E, MacHaria W, Ochieng R, Waiyego M, Zhou G, et al. Evaluation of non-invasive continuous physiological monitoring devices for neonates in Nairobi, Kenya: a research protocol. BMJ Open. 2020;10(4):1–7. Ginsburg AS, Nkwopara E, MacHaria W, Ochieng R, Waiyego M, Zhou G, et al. Evaluation of non-invasive continuous physiological monitoring devices for neonates in Nairobi, Kenya: a research protocol. BMJ Open. 2020;10(4):1–7.
10.
go back to reference Karlen W, Ansermino JMDG. Adaptive pulse segmentation and artifact detection in photoplethysmography for mobile applications. Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:3131–4.PubMed Karlen W, Ansermino JMDG. Adaptive pulse segmentation and artifact detection in photoplethysmography for mobile applications. Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:3131–4.PubMed
12.
go back to reference Bland U, Giavarina D. Lessons in biostatistics. Past Present Futur Stat Sci. 2014;25(2):359–72. Bland U, Giavarina D. Lessons in biostatistics. Past Present Futur Stat Sci. 2014;25(2):359–72.
13.
go back to reference Simoes EAF, Roark R, Berman S, Esler LL, Murphy J. Respiratory rate: measurement of variability over time and accuracy at different counting periods. Arch Dis Child. 1991;66(10):1199–203.CrossRef Simoes EAF, Roark R, Berman S, Esler LL, Murphy J. Respiratory rate: measurement of variability over time and accuracy at different counting periods. Arch Dis Child. 1991;66(10):1199–203.CrossRef
14.
go back to reference MPH MSNSM, MD RGB,DLSM, MD, Mark IN, et al. Does this child have pneumonia? The rational clinical examination systematic review. JAMA. 2017;318(5):462–71.CrossRef MPH MSNSM, MD RGB,DLSM, MD, Mark IN, et al. Does this child have pneumonia? The rational clinical examination systematic review. JAMA. 2017;318(5):462–71.CrossRef
16.
go back to reference Ross KR, Rosen CL. Sleep and respiratory physiology in children. Clin Chest Med. 2014;35(3):457–67.CrossRef Ross KR, Rosen CL. Sleep and respiratory physiology in children. Clin Chest Med. 2014;35(3):457–67.CrossRef
17.
go back to reference Horne RSC. Cardio-respiratory control during sleep in infancy. Paediatr Respir Rev. 2014;15(2):163–9.PubMed Horne RSC. Cardio-respiratory control during sleep in infancy. Paediatr Respir Rev. 2014;15(2):163–9.PubMed
Metadata
Title
Variability of respiratory rate measurements in neonates- every minute counts
Authors
Catherine Muthoni Njeru
J. Mark Ansermino
William M. Macharia
Dustin T. Dunsmuir
Publication date
01-12-2022
Publisher
BioMed Central
Published in
BMC Pediatrics / Issue 1/2022
Electronic ISSN: 1471-2431
DOI
https://doi.org/10.1186/s12887-021-03087-z

Other articles of this Issue 1/2022

BMC Pediatrics 1/2022 Go to the issue