Skip to main content
Top
Published in: BMC Ophthalmology 1/2017

Open Access 01-12-2017 | Research article

Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation

Authors: Dae-Young Park, Dong Hui Lim, Sungsoon Hwang, Joo Hyun, Tae-Young Chung

Published in: BMC Ophthalmology | Issue 1/2017

Login to get access

Abstract

Background

To evaluate and compare the astigmatism prediction errors taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens (IOL) implantation.

Methods

Phacoemulsification with toric Precizon IOL implantation was performed in 41 eyes with corneal astigmatism (range, 1 to 5 diopters (D)) determined by IOLMaster and SimK on Pentacam. Preoperative corneal astigmatism measurements were obtained from IOLMaster readings (IOLMaster, Baylor-IOLMaster, and Barrett-IOLMaster) and Pentacam readings (SimK, Baylor-SimK, Barrett-SimK, wavefront, true net power, total corneal refractive power, and vector derived by manual vector summation using corneal front and back astigmatism). Prediction error and intraclass correlation coefficient (ICC) between the measured (or calculated) astigmatism by IOLMaster and Pentacam and the estimated corneal astigmatism estimated by IOL toricity power and residual astigmatism were determined.

Results

The centroid errors in prediction error with IOLMaster, SimK, Baylor-IOLMaster, Baylor-SimK, Barrett-IOLMaster, Barrett-SimK, wavefront, true net power, total corneal refractive power, and vector were 0.59@103, 0.61 @103, 0.37@161, 0.41@162, 0.24@171, 0.36@162, 0.42@106, 0.04@8, 0.07@82, and 0.03@82, respectively, in with-the-rule (WTR) astigmatism eyes at postoperative 3-month. They were 0.22@87, 0.20@74, 0.16@21, 0.54@10, 0.43@3, 0.33@19, 0.51@25, 0.31@58, 0.29@50, and 0.14@50 in against-the-rule (ATR) astigmatism eyes. Of the ten modalities, vector showed the lowest WTR astigmatism prediction error and the highest ICC between the predicted and the estimated corneal astigmatism for both WTR and ATR eyes.

Conclusion

Vector summation using anterior and posterior corneal surface power taken with the Pentacam yields the least astigmatism prediction error and is a promising tool for determining toric IOL cylinder power.
Literature
1.
go back to reference Chang S-W, Su T-Y, Chen Y-L. Influence of ocular features and incision width on surgically induced astigmatism after cataract surgery. J Refract Surg. 2015;31:82–8.CrossRefPubMed Chang S-W, Su T-Y, Chen Y-L. Influence of ocular features and incision width on surgically induced astigmatism after cataract surgery. J Refract Surg. 2015;31:82–8.CrossRefPubMed
2.
go back to reference Dunne MC, Royston JM, Barnes DA. Posterior corneal surface toricity and total corneal astigmatism. Optom Vis Sci. 1991;68:708–10.CrossRefPubMed Dunne MC, Royston JM, Barnes DA. Posterior corneal surface toricity and total corneal astigmatism. Optom Vis Sci. 1991;68:708–10.CrossRefPubMed
3.
go back to reference Prisant O, Hoang-Xuan T, Proano C, Hernandez E, Awad S, Azar DT. Vector summation of anterior and posterior corneal topographical astigmatism. J Cataract Refract Surg. 2002;28:1636–43.CrossRefPubMed Prisant O, Hoang-Xuan T, Proano C, Hernandez E, Awad S, Azar DT. Vector summation of anterior and posterior corneal topographical astigmatism. J Cataract Refract Surg. 2002;28:1636–43.CrossRefPubMed
4.
go back to reference Koch DD, Jenkins RB, Weikert MP, Yeu E, Wang L. Correcting astigmatism with toric intraocular lenses: effect of posterior corneal astigmatism. J Cataract Refract Surg. 2013;39:1803–9.CrossRefPubMed Koch DD, Jenkins RB, Weikert MP, Yeu E, Wang L. Correcting astigmatism with toric intraocular lenses: effect of posterior corneal astigmatism. J Cataract Refract Surg. 2013;39:1803–9.CrossRefPubMed
5.
go back to reference Abulafia A, Hill WE, Franchina M, Barrett GD. Comparison of methods to predict residual astigmatism after intraocular lens implantation. J Refract Surg. 2015;31:699–707.CrossRefPubMed Abulafia A, Hill WE, Franchina M, Barrett GD. Comparison of methods to predict residual astigmatism after intraocular lens implantation. J Refract Surg. 2015;31:699–707.CrossRefPubMed
6.
go back to reference Alpins NA. A new method of analyzing vectors for changes in astigmatism. J Cataract Refract Surg. 1993;19:524–33.CrossRefPubMed Alpins NA. A new method of analyzing vectors for changes in astigmatism. J Cataract Refract Surg. 1993;19:524–33.CrossRefPubMed
7.
go back to reference Abulafia A, Barrett GD, Kleinmann G, Ofir S, Levy A, Marcovich AL, et al. Prediction of refractive outcomes with toric intraocular lens implantation. J Cataract Refract Surg. 2015;41:936–44.CrossRefPubMed Abulafia A, Barrett GD, Kleinmann G, Ofir S, Levy A, Marcovich AL, et al. Prediction of refractive outcomes with toric intraocular lens implantation. J Cataract Refract Surg. 2015;41:936–44.CrossRefPubMed
8.
go back to reference Stavridis E, Eppig T, Szentmáry N, Seitz B, Langenbucher A. Reproducibility of Scheimpflug tomography measurements regarding corneal front and back surface power. Klin Monatsbl Augenheilkd. 2015;232:1297–303.CrossRefPubMed Stavridis E, Eppig T, Szentmáry N, Seitz B, Langenbucher A. Reproducibility of Scheimpflug tomography measurements regarding corneal front and back surface power. Klin Monatsbl Augenheilkd. 2015;232:1297–303.CrossRefPubMed
9.
go back to reference Holladay JT, Moran JR, Kezirian GM. Analysis of aggregate surgically induced refractive change, prediction error, and intraocular astigmatism. J Cataract Refract Surg. 2001;27:61–79.CrossRefPubMed Holladay JT, Moran JR, Kezirian GM. Analysis of aggregate surgically induced refractive change, prediction error, and intraocular astigmatism. J Cataract Refract Surg. 2001;27:61–79.CrossRefPubMed
10.
go back to reference Eom Y, Song JS, Kim YY, Kim HM. Comparison of SRK/T and Haigis formulas for predicting corneal astigmatism correction with toric intraocular lenses. J Cataract Refract Surg. 2015;41:1650–7.CrossRefPubMed Eom Y, Song JS, Kim YY, Kim HM. Comparison of SRK/T and Haigis formulas for predicting corneal astigmatism correction with toric intraocular lenses. J Cataract Refract Surg. 2015;41:1650–7.CrossRefPubMed
11.
go back to reference Viera AJ, Garrett JM. Understanding interobserver agreement: the kappa statistic. Fam Med. 2005;37:360–3.PubMed Viera AJ, Garrett JM. Understanding interobserver agreement: the kappa statistic. Fam Med. 2005;37:360–3.PubMed
12.
go back to reference Villegas EA, Alcón E, Artal. minimum Amount of astigmatism that should be corrected. J Cataract Refract Surg. 2014;40:13–9.CrossRefPubMed Villegas EA, Alcón E, Artal. minimum Amount of astigmatism that should be corrected. J Cataract Refract Surg. 2014;40:13–9.CrossRefPubMed
13.
go back to reference Tonn B, Klaproth OK, Kohnen T. Anterior surface-based keratometry compared with Scheimpflug tomography-based total corneal astigmatism. Invest Ophthalmol Vis Sci. 2015;56:291–8.CrossRef Tonn B, Klaproth OK, Kohnen T. Anterior surface-based keratometry compared with Scheimpflug tomography-based total corneal astigmatism. Invest Ophthalmol Vis Sci. 2015;56:291–8.CrossRef
14.
go back to reference Savini G, Naeser K. An analysis of the factors influencing the residual refractive astigmatism after cataract surgery with Toric intraocular lenses. Invest Ophthalmol Vis Sci. 2015;56:827–35.CrossRefPubMed Savini G, Naeser K. An analysis of the factors influencing the residual refractive astigmatism after cataract surgery with Toric intraocular lenses. Invest Ophthalmol Vis Sci. 2015;56:827–35.CrossRefPubMed
15.
go back to reference Crawford AZ, Patel DV, Mcghee CNJ. Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems. Am J Ophthalmol. 2013;156:53–60.CrossRefPubMed Crawford AZ, Patel DV, Mcghee CNJ. Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems. Am J Ophthalmol. 2013;156:53–60.CrossRefPubMed
16.
go back to reference Reitblat O, Levy A, Kleinmann G, Abulafia A, Assia EI. Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: comparison of methodologies. J Cataract Refract Surg. 2016;42:217–25.CrossRefPubMed Reitblat O, Levy A, Kleinmann G, Abulafia A, Assia EI. Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: comparison of methodologies. J Cataract Refract Surg. 2016;42:217–25.CrossRefPubMed
17.
go back to reference Kim H-Y. Statistical notes for clinical researchers: evaluation of measurement error 2: Dahlberg’s error, bland-Altman method, and kappa coefficient. Restor Dent Endod. 2013;38:182–5.CrossRefPubMedPubMedCentral Kim H-Y. Statistical notes for clinical researchers: evaluation of measurement error 2: Dahlberg’s error, bland-Altman method, and kappa coefficient. Restor Dent Endod. 2013;38:182–5.CrossRefPubMedPubMedCentral
18.
go back to reference Zheng T, Chen Z, Lu Y. Influence factors of estimation errors for total corneal astigmatism using keratometric astigmatism in patients before cataract surgery. J Cataract Refract Surg. 2016;42:84–94.CrossRefPubMed Zheng T, Chen Z, Lu Y. Influence factors of estimation errors for total corneal astigmatism using keratometric astigmatism in patients before cataract surgery. J Cataract Refract Surg. 2016;42:84–94.CrossRefPubMed
19.
go back to reference Tadros A, Habib M, Tejwani D, Von Lany H, Thomas P. Opposite clear corneal incisions on the steep meridian in phacoemulsification: early effects on the cornea. J Cataract Refract Surg. 2004;30:414–7.CrossRefPubMed Tadros A, Habib M, Tejwani D, Von Lany H, Thomas P. Opposite clear corneal incisions on the steep meridian in phacoemulsification: early effects on the cornea. J Cataract Refract Surg. 2004;30:414–7.CrossRefPubMed
20.
go back to reference Denoyer A, Ricaud X, Van Went C, Labbé A, Baudouin C. Influence of corneal biomechanical properties on surgically induced astigmatism in cataract surgery. J Cataract Refract Surg. 2013;39:1204–10.CrossRefPubMed Denoyer A, Ricaud X, Van Went C, Labbé A, Baudouin C. Influence of corneal biomechanical properties on surgically induced astigmatism in cataract surgery. J Cataract Refract Surg. 2013;39:1204–10.CrossRefPubMed
21.
go back to reference Yong Park C, Do JR, Chuck RS. Predicting postoperative astigmatism using Scheimpflug keratometry (Pentacam) and automated keratometry (IOLMaster). Curr Eye Res. 2012;37:1091–8.CrossRefPubMed Yong Park C, Do JR, Chuck RS. Predicting postoperative astigmatism using Scheimpflug keratometry (Pentacam) and automated keratometry (IOLMaster). Curr Eye Res. 2012;37:1091–8.CrossRefPubMed
22.
23.
go back to reference Toto L, Vecchiarino L, D’Ugo E, Cardone D, Mastropasqua A, Mastropasqua R, et al. Astigmatism correction with toric IOL: analysis of visual performance, position, and wavefront error. J Refract Surg. 2013;29:476–83.CrossRefPubMed Toto L, Vecchiarino L, D’Ugo E, Cardone D, Mastropasqua A, Mastropasqua R, et al. Astigmatism correction with toric IOL: analysis of visual performance, position, and wavefront error. J Refract Surg. 2013;29:476–83.CrossRefPubMed
24.
go back to reference Olsen T, Corydon L, Gimbel H. Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg. 1995;21:313–9.CrossRefPubMed Olsen T, Corydon L, Gimbel H. Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg. 1995;21:313–9.CrossRefPubMed
25.
go back to reference Eom Y, Kang SY, Song JS, Kim YY, Kim HM. Effect of effective lens position on cylinder power of toric intraocular lenses. Can J Ophthalmol. 2015;50:26–32.CrossRefPubMed Eom Y, Kang SY, Song JS, Kim YY, Kim HM. Effect of effective lens position on cylinder power of toric intraocular lenses. Can J Ophthalmol. 2015;50:26–32.CrossRefPubMed
Metadata
Title
Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation
Authors
Dae-Young Park
Dong Hui Lim
Sungsoon Hwang
Joo Hyun
Tae-Young Chung
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2017
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0550-z

Other articles of this Issue 1/2017

BMC Ophthalmology 1/2017 Go to the issue