Skip to main content
Top
Published in: BMC Ophthalmology 1/2015

Open Access 01-12-2015 | Research article

Comparison of corneal endothelial cell measurements by two non-contact specular microscopes

Authors: Laura Gasser, Thomas Reinhard, Daniel Böhringer

Published in: BMC Ophthalmology | Issue 1/2015

Login to get access

Abstract

Background

Measurement of corneal endothelial cell density is important both for clinical diagnosis as well as clinical studies. Since endothelial cell loss is considered irreversible in humans, even small changes in endothelial cell density are relevant. Therefore it is important to know whether different instruments for endothelial cell density measurements give the same results and can thus be used interchangeably. In this study we compare corneal endothelial cell density and morphometry measurements from two widely used non-contact specular microscopes, the Topcon SP3000P and Konan Noncon Robo SP8000.

Methods

Endothelial cell measurements were performed with both the Topcon SP3000P and Konan Noncon Robo SP8000 on 34 eyes of 18 consecutive patients of our cornea clinics with poor image quality being the only exclusion criterion. Images were obtained using the auto-focussing method and manual cell selection. Endothelial cell density (ECD), hexagonal cell ratio (HEX) and coefficient of value (CV) of the endothelial cell layer were calculated by the instruments’ built-in software.

Results

ECD values calculated by the Konan were systematically higher than Topcon values: in 94 % of eyes Konan gave a higher value than Topcon, leading to a mean difference in ECD between the instruments of 187 cells/mm2 (P < 0.001 in paired Wilcoxon test). HEX showed a broad range of values and differed greatly with only weak correlation between the two instruments. CV values for Konan mostly exceeded Topcon values, and only showed a weak correlation between the two instruments as well.

Conclusions

Values for ECD between the Konan and the Topcon do correlate well, but the ECDs calculated by the Konan are systematically higher than Topcon values. Both HEX and CV vary greatly and do not correlate sufficiently. Thus we recommend not to use the Konan and the Topcon instrument interchangeably.
Literature
1.
go back to reference Olsen T. Non-contact specular microscopy of human corneal endothelium. Acta Ophthalmol (Copenh). 1979;57:986–98.CrossRef Olsen T. Non-contact specular microscopy of human corneal endothelium. Acta Ophthalmol (Copenh). 1979;57:986–98.CrossRef
3.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed
4.
go back to reference R Development Core Team: R. A language and environment for statistical computing. Vienna, Austria: R Foundation for Statistical Computing; 2008. R Development Core Team: R. A language and environment for statistical computing. Vienna, Austria: R Foundation for Statistical Computing; 2008.
5.
go back to reference Cheung SW, Cho P. Endothelial cells analysis with the TOPCON specular microscope SP-2000P and IMAGEnet system. Curr Eye Res. 2000;21:788–98.CrossRefPubMed Cheung SW, Cho P. Endothelial cells analysis with the TOPCON specular microscope SP-2000P and IMAGEnet system. Curr Eye Res. 2000;21:788–98.CrossRefPubMed
6.
go back to reference Abib FC, Holzchuh R, Schaefer A, Schaefer T, Godois R. The endothelial sample size analysis in corneal specular microscopy clinical examinations. Cornea. 2012;31:546–50.CrossRefPubMed Abib FC, Holzchuh R, Schaefer A, Schaefer T, Godois R. The endothelial sample size analysis in corneal specular microscopy clinical examinations. Cornea. 2012;31:546–50.CrossRefPubMed
7.
go back to reference Böhringer D, Hettich L, Maier PC, Reinhard T. Data quality of unsupervised endothelial cell counting vs. reading centre analysis in multicentric clinical trials. Klin Monbl Augenheilkd. 2012;229:628–31.CrossRefPubMed Böhringer D, Hettich L, Maier PC, Reinhard T. Data quality of unsupervised endothelial cell counting vs. reading centre analysis in multicentric clinical trials. Klin Monbl Augenheilkd. 2012;229:628–31.CrossRefPubMed
8.
go back to reference Doughty MJ, Müller A, Zaman ML. Assessment of the reliability of human corneal endothelial cell-density estimates using a noncontact specular microscope. Cornea. 2000;19:148–58.CrossRefPubMed Doughty MJ, Müller A, Zaman ML. Assessment of the reliability of human corneal endothelial cell-density estimates using a noncontact specular microscope. Cornea. 2000;19:148–58.CrossRefPubMed
9.
go back to reference McCarey BE, Edelhauser HF, Lynn MJ. Review of corneal endothelial specular microscopy for FDA clinical trials of refractive procedures, surgical devices and New intraocular drugs and solutions. Cornea. 2008;27:1–16.CrossRefPubMedPubMedCentral McCarey BE, Edelhauser HF, Lynn MJ. Review of corneal endothelial specular microscopy for FDA clinical trials of refractive procedures, surgical devices and New intraocular drugs and solutions. Cornea. 2008;27:1–16.CrossRefPubMedPubMedCentral
10.
go back to reference Isager P, Hjortdal JO, Guo S, Ehlers N. Comparison of endothelial cell density estimated by contact and non-contact specular microscopy. Acta Ophthalmol Scand. 2000;78:42–4.CrossRefPubMed Isager P, Hjortdal JO, Guo S, Ehlers N. Comparison of endothelial cell density estimated by contact and non-contact specular microscopy. Acta Ophthalmol Scand. 2000;78:42–4.CrossRefPubMed
11.
go back to reference De Sanctis U, Machetta F, Razzano L, Dalmasso P, Grignolo FM. Corneal endothelium evaluation with 2 noncontact specular microscopes and their semiautomated methods of analysis. Cornea. 2006;25:501–6.CrossRefPubMed De Sanctis U, Machetta F, Razzano L, Dalmasso P, Grignolo FM. Corneal endothelium evaluation with 2 noncontact specular microscopes and their semiautomated methods of analysis. Cornea. 2006;25:501–6.CrossRefPubMed
12.
go back to reference Thuret G, Deb-Joardar N, Zhao M, Gain P, Gavet Y, Nguyen F. Agreement between two non-contact specular microscopes: Topcon SP2000P versus Rhine-Tec. Br J Ophthalmol. 2007;91:979–80.CrossRefPubMedPubMedCentral Thuret G, Deb-Joardar N, Zhao M, Gain P, Gavet Y, Nguyen F. Agreement between two non-contact specular microscopes: Topcon SP2000P versus Rhine-Tec. Br J Ophthalmol. 2007;91:979–80.CrossRefPubMedPubMedCentral
13.
go back to reference Landesz M, Siertsema JV, Van Rij G. Comparative study of three semiautomated specular microscopes. J Cataract Refract Surg. 1995;21:409–16.CrossRefPubMed Landesz M, Siertsema JV, Van Rij G. Comparative study of three semiautomated specular microscopes. J Cataract Refract Surg. 1995;21:409–16.CrossRefPubMed
14.
go back to reference Kitzmann AS, Winter EJ, Nau CB, McLaren JW, Hodge DO, Bourne WM. Comparison of corneal endothelial cell images from a noncontact specular microscope and a scanning confocal microscope. Cornea. 2005;24:980–4.CrossRefPubMed Kitzmann AS, Winter EJ, Nau CB, McLaren JW, Hodge DO, Bourne WM. Comparison of corneal endothelial cell images from a noncontact specular microscope and a scanning confocal microscope. Cornea. 2005;24:980–4.CrossRefPubMed
15.
go back to reference Luft N, Hirnschall N, Schuschitz S, Draschl P, Findl O. Comparison of 4 specular microscopes in healthy eyes and eyes with cornea guttata or corneal grafts. Cornea. 2015;34:381–6.CrossRefPubMed Luft N, Hirnschall N, Schuschitz S, Draschl P, Findl O. Comparison of 4 specular microscopes in healthy eyes and eyes with cornea guttata or corneal grafts. Cornea. 2015;34:381–6.CrossRefPubMed
Metadata
Title
Comparison of corneal endothelial cell measurements by two non-contact specular microscopes
Authors
Laura Gasser
Thomas Reinhard
Daniel Böhringer
Publication date
01-12-2015
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2015
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-015-0068-1

Other articles of this Issue 1/2015

BMC Ophthalmology 1/2015 Go to the issue