Open Access 01-10-2008 | Research

Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study

Authors: Matthias Gruenewald, Jochen Renner, Patrick Meybohm, Jan Höcker, Jens Scholz, Berthold Bein

Published in: Critical Care | Issue 5/2008

Login to get access

Abstract

Introduction

Monitoring cardiac output (CO) may allow early detection of haemodynamic instability, aiming to reduce morbidity and mortality in critically ill patients. Continuous cardiac output (CCO) monitoring is recommended in septic or postoperative patients with high incidences of intra-abdominal hypertension (IAH). The aim of the present study was to compare the agreement between three CCO methods and a bolus thermodilution CO technique during acute IAH and volume loading.

Methods

Ten pigs were anaesthetised and instrumented for haemodynamic measurements. Cardiac output was obtained using CCO by pulse power analysis (PulseCO; LiDCO monitor), using CCO by pulse contour analysis (PCCO; PiCCO monitor) and using CCO by pulmonary artery catheter thermodilution (CCOPAC), and was compared with bolus transcardiopulmonary thermodilution CO (COTCP) at baseline, after fluid loading, at IAH and after an additional fluid loading at IAH. Whereas PulseCO was only calibrated at baseline, PCCO was calibrated at each experimental step.

Results

PulseCO and PCCO underestimated CO, as the overall bias ± standard deviation was 1.0 ± 1.5 l/min and 1.0 ± 1.1 l/min compared with COTCP. A clinically accepted agreement between all of the CCO methods and COTCP was observed only at baseline. Whereas IAH did not influence the CO, increased CO following fluid loading at IAH was only reflected by CCOPAC and COTCP, not by uncalibrated PulseCO and PCCO. After recalibration, PCCO was comparable with COTCP.

Conclusions

The CO obtained by uncalibrated PulseCO and PCCO failed to agree with COTCP during IAH and fluid loading. In the critically ill patient, recalibration of continuous arterial waveform CO methods should be performed after fluid loading or before a major change in therapy is initiated.
Appendix
Available only for authorised users
Literature
1.
go back to reference Eisenberg PR, Jaffe AS, Schuster DP: Clinical evaluation compared to pulmonary artery catheterization in the hemodynamic assessment of critically ill patients. Crit Care Med 1984, 12: 549-553. 10.1097/00003246-198407000-00001CrossRefPubMed Eisenberg PR, Jaffe AS, Schuster DP: Clinical evaluation compared to pulmonary artery catheterization in the hemodynamic assessment of critically ill patients. Crit Care Med 1984, 12: 549-553. 10.1097/00003246-198407000-00001CrossRefPubMed
2.
go back to reference Rodig G, Prasser C, Keyl C, Liebold A, Hobbhahn J: Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth 1999, 82: 525-530.CrossRefPubMed Rodig G, Prasser C, Keyl C, Liebold A, Hobbhahn J: Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth 1999, 82: 525-530.CrossRefPubMed
3.
go back to reference Bein B, Worthmann F, Tonner PH, Paris A, Steinfath M, Hedderich J, Scholz J: Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output. J Cardiothorac Vasc Anesth 2004, 18: 185-189. 10.1053/j.jvca.2004.01.025CrossRefPubMed Bein B, Worthmann F, Tonner PH, Paris A, Steinfath M, Hedderich J, Scholz J: Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output. J Cardiothorac Vasc Anesth 2004, 18: 185-189. 10.1053/j.jvca.2004.01.025CrossRefPubMed
4.
go back to reference Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P: Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002, 88: 350-356. 10.1093/bja/88.3.350CrossRefPubMed Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P: Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002, 88: 350-356. 10.1093/bja/88.3.350CrossRefPubMed
5.
go back to reference Buhre W, Weyland A, Kazmaier S, Hanekop GG, Baryalei MM, Sydow M, Sonntag H: Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting. J Cardiothorac Vasc Anesth 1999, 13: 437-440. 10.1016/S1053-0770(99)90216-1CrossRefPubMed Buhre W, Weyland A, Kazmaier S, Hanekop GG, Baryalei MM, Sydow M, Sonntag H: Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting. J Cardiothorac Vasc Anesth 1999, 13: 437-440. 10.1016/S1053-0770(99)90216-1CrossRefPubMed
6.
go back to reference Bottiger BW, Sinner B, Motsch J, Bach A, Bauer H, Martin E: Continuous versus intermittent thermodilution cardiac output measurement during orthotopic liver transplantation. Anaesthesia 1997, 52: 207-214. 10.1111/j.1365-2044.1997.079-az0077.xCrossRefPubMed Bottiger BW, Sinner B, Motsch J, Bach A, Bauer H, Martin E: Continuous versus intermittent thermodilution cardiac output measurement during orthotopic liver transplantation. Anaesthesia 1997, 52: 207-214. 10.1111/j.1365-2044.1997.079-az0077.xCrossRefPubMed
7.
go back to reference Pittman J, Bar-Yosef S, SumPing J, Sherwood M, Mark J: Continuous cardiac output monitoring with pulse contour analysis: a comparison with lithium indicator dilution cardiac output measurement. Crit Care Med 2005, 33: 2015-2021. 10.1097/01.CCM.0000179021.36805.1FCrossRefPubMed Pittman J, Bar-Yosef S, SumPing J, Sherwood M, Mark J: Continuous cardiac output monitoring with pulse contour analysis: a comparison with lithium indicator dilution cardiac output measurement. Crit Care Med 2005, 33: 2015-2021. 10.1097/01.CCM.0000179021.36805.1FCrossRefPubMed
8.
go back to reference Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg 2002, 74: S1408-S1412. 10.1016/S0003-4975(02)04059-6CrossRefPubMed Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg 2002, 74: S1408-S1412. 10.1016/S0003-4975(02)04059-6CrossRefPubMed
9.
go back to reference Costa MG, Della Rocca G, Chiarandini P, Mattelig S, Pompei L, Barriga MS, Reynolds T, Cecconi M, Pietropaoli P: Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med 2008, 34: 257-263. 10.1007/s00134-007-0878-6CrossRefPubMed Costa MG, Della Rocca G, Chiarandini P, Mattelig S, Pompei L, Barriga MS, Reynolds T, Cecconi M, Pietropaoli P: Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med 2008, 34: 257-263. 10.1007/s00134-007-0878-6CrossRefPubMed
10.
go back to reference Linton NW, Linton RA: Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb. Br J Anaesth 2001, 86: 486-496. 10.1093/bja/86.4.486CrossRefPubMed Linton NW, Linton RA: Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb. Br J Anaesth 2001, 86: 486-496. 10.1093/bja/86.4.486CrossRefPubMed
11.
go back to reference Bein B, Meybohm P, Cavus E, Renner J, Tonner PH, Steinfath M, Scholz J, Doerges V: The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg 2007, 105: 107-113. 10.1213/01.ane.0000268140.02147.edCrossRefPubMed Bein B, Meybohm P, Cavus E, Renner J, Tonner PH, Steinfath M, Scholz J, Doerges V: The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg 2007, 105: 107-113. 10.1213/01.ane.0000268140.02147.edCrossRefPubMed
12.
go back to reference Cooper ES, Muir WW: Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs. Crit Care Med 2007, 35: 1724-1729. 10.1097/01.CCM.0000266590.25109.F2CrossRefPubMed Cooper ES, Muir WW: Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs. Crit Care Med 2007, 35: 1724-1729. 10.1097/01.CCM.0000266590.25109.F2CrossRefPubMed
13.
go back to reference Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL: Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med 2008, 36: 434-440. 10.1097/CCM.0b013e31818a88b6CrossRefPubMed Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL: Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med 2008, 36: 434-440. 10.1097/CCM.0b013e31818a88b6CrossRefPubMed
14.
go back to reference Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M: Cardiac output by PulseCO is not interchangeable with thermodilution in patients undergoing OPCAB. Can J Anaesth 2005, 52: 530-534.CrossRefPubMed Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M: Cardiac output by PulseCO is not interchangeable with thermodilution in patients undergoing OPCAB. Can J Anaesth 2005, 52: 530-534.CrossRefPubMed
15.
go back to reference Michard F: Pulse contour analysis: fairy tale or new reality? Crit Care Med 2007, 35: 1791-1792. 10.1097/01.CCM.0000269351.38762.B9CrossRefPubMed Michard F: Pulse contour analysis: fairy tale or new reality? Crit Care Med 2007, 35: 1791-1792. 10.1097/01.CCM.0000269351.38762.B9CrossRefPubMed
16.
go back to reference Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, Bihari D, Innes R, Cohen J, Singer P, Japiassu A, Kurtop E, De Keulenaer BL, Daelemans R, Del Turco M, Cosimini P, Ranieri M, Jacquet L, Laterre PF, Gattinoni L: Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study. Intensive Care Med 2004, 30: 822-829. 10.1007/s00134-004-2169-9CrossRefPubMed Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, Bihari D, Innes R, Cohen J, Singer P, Japiassu A, Kurtop E, De Keulenaer BL, Daelemans R, Del Turco M, Cosimini P, Ranieri M, Jacquet L, Laterre PF, Gattinoni L: Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study. Intensive Care Med 2004, 30: 822-829. 10.1007/s00134-004-2169-9CrossRefPubMed
17.
go back to reference Malbrain M, De Iaet I, Viaene D, Vermeiren G, Schoonheydt K, Dits H: Validation of four different continuous minimal invasive cardiac output methods in patients with abdominal hypertension. Acta Clin Belg Suppl 2007, 62: 256. Malbrain M, De Iaet I, Viaene D, Vermeiren G, Schoonheydt K, Dits H: Validation of four different continuous minimal invasive cardiac output methods in patients with abdominal hypertension. Acta Clin Belg Suppl 2007, 62: 256.
18.
go back to reference Wesseling KH, Jansen JR, Settels JJ, Schreuder JJ: Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol 1993, 74: 2566-2573.PubMed Wesseling KH, Jansen JR, Settels JJ, Schreuder JJ: Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol 1993, 74: 2566-2573.PubMed
19.
go back to reference Cheatham ML, Malbrain ML, Kirkpatrick A, Sugrue M, Parr M, De Waele J, Balogh Z, Leppaniemi A, Olvera C, Ivatury R, D'Amours S, Wendon J, Hillman K, Wilmer A: Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. II. Recommendations. Intensive Care Med 2007, 33: 951-962. 10.1007/s00134-007-0592-4CrossRefPubMed Cheatham ML, Malbrain ML, Kirkpatrick A, Sugrue M, Parr M, De Waele J, Balogh Z, Leppaniemi A, Olvera C, Ivatury R, D'Amours S, Wendon J, Hillman K, Wilmer A: Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. II. Recommendations. Intensive Care Med 2007, 33: 951-962. 10.1007/s00134-007-0592-4CrossRefPubMed
20.
go back to reference Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1: 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1: 307-310.CrossRefPubMed
21.
go back to reference Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999, 15: 85-91. 10.1023/A:1009982611386CrossRefPubMed Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999, 15: 85-91. 10.1023/A:1009982611386CrossRefPubMed
22.
go back to reference Spohr F, Hettrich P, Bauer H, Haas U, Martin E, Bottiger BW: Comparison of two methods for enhanced continuous circulatory monitoring in patients with septic shock. Intensive Care Med 2007, 33: 1805-1810. 10.1007/s00134-007-0703-2CrossRefPubMed Spohr F, Hettrich P, Bauer H, Haas U, Martin E, Bottiger BW: Comparison of two methods for enhanced continuous circulatory monitoring in patients with septic shock. Intensive Care Med 2007, 33: 1805-1810. 10.1007/s00134-007-0703-2CrossRefPubMed
23.
go back to reference Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P: Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 2003, 50: 707-711.CrossRefPubMed Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P: Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 2003, 50: 707-711.CrossRefPubMed
24.
go back to reference Cheatham ML, Safcsak K, Block EF, Nelson LD: Preload assessment in patients with an open abdomen. J Trauma 1999, 46: 16-22.CrossRefPubMed Cheatham ML, Safcsak K, Block EF, Nelson LD: Preload assessment in patients with an open abdomen. J Trauma 1999, 46: 16-22.CrossRefPubMed
25.
go back to reference Sumpelmann R, Schuerholz T, Marx G, Jesch NK, Osthaus WA, Ure BM: Hemodynamic changes during acute elevation of intra-abdominal pressure in rabbits. Paediatr Anaesth 2006, 16: 1262-1267. 10.1111/j.1460-9592.2006.01987.xCrossRefPubMed Sumpelmann R, Schuerholz T, Marx G, Jesch NK, Osthaus WA, Ure BM: Hemodynamic changes during acute elevation of intra-abdominal pressure in rabbits. Paediatr Anaesth 2006, 16: 1262-1267. 10.1111/j.1460-9592.2006.01987.xCrossRefPubMed
26.
go back to reference Piehl MD, Manning JE, McCurdy SL, Rhue TS, Kocis KC, Cairns CB, Cairns BA: Pulse contour cardiac output analysis in a piglet model of severe hemorrhagic shock. Crit Care Med 2008, 36: 1189-1195. 10.1097/CCM.0B013E31816592A3CrossRefPubMed Piehl MD, Manning JE, McCurdy SL, Rhue TS, Kocis KC, Cairns CB, Cairns BA: Pulse contour cardiac output analysis in a piglet model of severe hemorrhagic shock. Crit Care Med 2008, 36: 1189-1195. 10.1097/CCM.0B013E31816592A3CrossRefPubMed
27.
go back to reference Sander M, von Heymann C, Foer A, von Dossow V, Grosse J, Dushe S, Konertz WF, Spies CD: Pulse contour analysis after normothermic cardiopulmonary bypass in cardiac surgery patients. Crit Care 2005, 9: R729-R734. 10.1186/cc3903PubMedCentralCrossRefPubMed Sander M, von Heymann C, Foer A, von Dossow V, Grosse J, Dushe S, Konertz WF, Spies CD: Pulse contour analysis after normothermic cardiopulmonary bypass in cardiac surgery patients. Crit Care 2005, 9: R729-R734. 10.1186/cc3903PubMedCentralCrossRefPubMed
28.
go back to reference Sagawa K, Lie RK, Schaefer J: Translation of Otto Frank's paper ¸Die Grundform des Arteriellen Pulses' Zeitschrift fur Biologie 37: 483–526 (1899). J Mol Cell Cardiol 1990, 22: 253-277. 10.1016/0022-2828(90)91459-KCrossRefPubMed Sagawa K, Lie RK, Schaefer J: Translation of Otto Frank's paper ¸Die Grundform des Arteriellen Pulses' Zeitschrift fur Biologie 37: 483–526 (1899). J Mol Cell Cardiol 1990, 22: 253-277. 10.1016/0022-2828(90)91459-KCrossRefPubMed
Metadata
Title
Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study
Authors
Matthias Gruenewald
Jochen Renner
Patrick Meybohm
Jan Höcker
Jens Scholz
Berthold Bein
Publication date
01-10-2008
Publisher
BioMed Central
Published in
Critical Care / Issue 5/2008
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc7102

Other articles of this Issue 5/2008

Critical Care 5/2008 Go to the issue