Skip to main content
Top
Published in: Critical Care 1/2006

Open Access 01-02-2006 | Research

Imposed work of breathing during high-frequency oscillatory ventilation: a bench study

Authors: Marc van Heerde, Huib R van Genderingen, Tom Leenhoven, Karel Roubik, Frans B Plötz, Dick G Markhorst

Published in: Critical Care | Issue 1/2006

Login to get access

Abstract

Introduction

The ventilator and the endotracheal tube impose additional workload in mechanically ventilated patients breathing spontaneously. The total work of breathing (WOB) includes elastic and resistive work. In a bench test we assessed the imposed WOB using 3100 A/3100 B SensorMedics high-frequency oscillatory ventilators.

Methods

A computer-controlled piston-driven test lung was used to simulate a spontaneously breathing patient. The test lung was connected to a high-frequency oscillatory ventilation (HFOV) ventilator by an endotracheal tube. The inspiratory and expiratory airway flows and pressures at various places were sampled. The spontaneous breath rate and volume, tube size and ventilator settings were simulated as representative of the newborn to adult range. The fresh gas flow rate was set at a low and a high level. The imposed WOB was calculated using the Campbell diagram.

Results

In the simulations for newborns (assumed body weight 3.5 kg) and infants (assumed body weight 10 kg) the imposed WOB (mean ± standard deviation) was 0.22 ± 0.07 and 0.87 ± 0.25 J/l, respectively. Comparison of the imposed WOB in low and high fresh gas flow rate measurements yielded values of 1.63 ± 0.32 and 0.96 ± 0.24 J/l (P = 0.01) in small children (assumed body weight 25 kg), of 1.81 ± 0.30 and 1.10 ± 0.27 J/l (P < 0.001) in large children (assumed body weight 40 kg), and of 1.95 ± 0.31 and 1.12 ± 0.34 J/l (P < 0.01) in adults (assumed body weight 70 kg). High peak inspiratory flow and low fresh gas flow rate significantly increased the imposed WOB. Mean airway pressure in the breathing circuit decreased dramatically during spontaneous breathing, most markedly at the low fresh gas flow rate. This led to ventilator shut-off when the inspiratory flow exceeded the fresh gas flow.

Conclusion

Spontaneous breathing during HFOV resulted in considerable imposed WOB in pediatric and adult simulations, explaining the discomfort seen in those patients breathing spontaneously during HFOV. The level of imposed WOB was lower in the newborn and infant simulations, explaining why these patients tolerate spontaneous breathing during HFOV well. A high fresh gas flow rate reduced the imposed WOB. These findings suggest the need for a demand flow system based on patient need allowing spontaneous breathing during HFOV.
Appendix
Available only for authorised users
Literature
1.
go back to reference Cereda M, Foti G, Marcora B, Gili M, Giacomini M, Sparacino ME, Pesenti A: Pressure support ventilation in patients with acute lung injury. Crit Care Med 2000, 28: 1269-1275. 10.1097/00003246-200005000-00002CrossRefPubMed Cereda M, Foti G, Marcora B, Gili M, Giacomini M, Sparacino ME, Pesenti A: Pressure support ventilation in patients with acute lung injury. Crit Care Med 2000, 28: 1269-1275. 10.1097/00003246-200005000-00002CrossRefPubMed
2.
go back to reference Hering R, Peters D, Zinserling J, Wrigge H, von Spiegel T, Putensen C: Effects of spontaneous breathing during airway pressure release ventilation on renal perfusion and function in patients with acute lung injury. Intensive Care Med 2002, 28: 1426-1433. 10.1007/s00134-002-1442-zCrossRefPubMed Hering R, Peters D, Zinserling J, Wrigge H, von Spiegel T, Putensen C: Effects of spontaneous breathing during airway pressure release ventilation on renal perfusion and function in patients with acute lung injury. Intensive Care Med 2002, 28: 1426-1433. 10.1007/s00134-002-1442-zCrossRefPubMed
3.
go back to reference Neumann P, Wrigge H, Zinserling J, Hinz J, Maripuu E, Andersson LG, Putensen C, Hedenstierna G: Spontaneous breathing affects the spatial ventilation and perfusion distribution during mechanical ventilatory support. Crit Care Med 2005, 33: 1090-1095. 10.1097/01.CCM.0000163226.34868.0ACrossRefPubMed Neumann P, Wrigge H, Zinserling J, Hinz J, Maripuu E, Andersson LG, Putensen C, Hedenstierna G: Spontaneous breathing affects the spatial ventilation and perfusion distribution during mechanical ventilatory support. Crit Care Med 2005, 33: 1090-1095. 10.1097/01.CCM.0000163226.34868.0ACrossRefPubMed
4.
go back to reference Putensen C, Mutz NJ, Putensen-Himmer G, Zinserling J: Spontaneous breathing during ventilatory support improves ventilation–perfusion distributions in patients with acute respiratory distress syndrome. Am J Respir Crit Care Med 1999, 159: 1241-1248.CrossRefPubMed Putensen C, Mutz NJ, Putensen-Himmer G, Zinserling J: Spontaneous breathing during ventilatory support improves ventilation–perfusion distributions in patients with acute respiratory distress syndrome. Am J Respir Crit Care Med 1999, 159: 1241-1248.CrossRefPubMed
5.
go back to reference Putensen C, Zech S, Wrigge H, Zinserling J, Stuber F, Von Spiegel T, Mutz N: Long-term effects of spontaneous breathing during ventilatory support in patients with acute lung injury. Am J Respir Crit Care Med 2001, 164: 43-49.CrossRefPubMed Putensen C, Zech S, Wrigge H, Zinserling J, Stuber F, Von Spiegel T, Mutz N: Long-term effects of spontaneous breathing during ventilatory support in patients with acute lung injury. Am J Respir Crit Care Med 2001, 164: 43-49.CrossRefPubMed
6.
go back to reference Sydow M, Burchardi H, Ephraim E, Zielmann S, Crozier TA: Long-term effects of two different ventilatory modes on oxygenation in acute lung injury. Comparison of airway pressure release ventilation and volume-controlled inverse ratio ventilation. Am J Respir Crit Care Med 1994, 149: 1550-1556.CrossRefPubMed Sydow M, Burchardi H, Ephraim E, Zielmann S, Crozier TA: Long-term effects of two different ventilatory modes on oxygenation in acute lung injury. Comparison of airway pressure release ventilation and volume-controlled inverse ratio ventilation. Am J Respir Crit Care Med 1994, 149: 1550-1556.CrossRefPubMed
7.
go back to reference Bollen CW, Uiterwaal CS, van Vught AJ: Cumulative metaanalysis of high-frequency versus conventional ventilation in premature neonates. Am J Respir Crit Care Med 2003, 168: 1150-1155. 10.1164/rccm.200306-721CPCrossRefPubMed Bollen CW, Uiterwaal CS, van Vught AJ: Cumulative metaanalysis of high-frequency versus conventional ventilation in premature neonates. Am J Respir Crit Care Med 2003, 168: 1150-1155. 10.1164/rccm.200306-721CPCrossRefPubMed
8.
go back to reference Froese AB, Kinsella JP: High-frequency oscillatory ventilation: lessons from the neonatal/pediatric experience. Crit Care Med 2005, 33: 115-121. 10.1097/01.CCM.0000155923.97849.6DCrossRef Froese AB, Kinsella JP: High-frequency oscillatory ventilation: lessons from the neonatal/pediatric experience. Crit Care Med 2005, 33: 115-121. 10.1097/01.CCM.0000155923.97849.6DCrossRef
9.
go back to reference Bollen CW, van Well GT, Sherr T, Beale RJ, Shah S, Findlay G, Monchi M, Chiche JD, Weiler N, Uiterwaal CS, et al.: High frequency oscillatory ventilation compared with conventional mechanical ventilation in adult respiratory distress syndrome: a randomized controlled trial [ISRCTN24242669]. Crit Care 2005, 9: R430-R439. 10.1186/cc3737PubMedCentralCrossRefPubMed Bollen CW, van Well GT, Sherr T, Beale RJ, Shah S, Findlay G, Monchi M, Chiche JD, Weiler N, Uiterwaal CS, et al.: High frequency oscillatory ventilation compared with conventional mechanical ventilation in adult respiratory distress syndrome: a randomized controlled trial [ISRCTN24242669]. Crit Care 2005, 9: R430-R439. 10.1186/cc3737PubMedCentralCrossRefPubMed
10.
go back to reference Derdak S, Mehta S, Stewart TE, Smith T, Rogers M, Buchman TG, Carlin B, Lowson S, Granton J: High-frequency oscillatory ventilation for acute respiratory distress syndrome in adults: a randomized, controlled trial. Am J Respir Crit Care Med 2002, 166: 801-808. 10.1164/rccm.2108052CrossRefPubMed Derdak S, Mehta S, Stewart TE, Smith T, Rogers M, Buchman TG, Carlin B, Lowson S, Granton J: High-frequency oscillatory ventilation for acute respiratory distress syndrome in adults: a randomized, controlled trial. Am J Respir Crit Care Med 2002, 166: 801-808. 10.1164/rccm.2108052CrossRefPubMed
11.
go back to reference Singh JM, Ferguson ND: Is it time to increase the frequency of use of high-frequency oscillatory ventilation? Crit Care 2005, 9: 339-340. 10.1186/cc3761PubMedCentralCrossRefPubMed Singh JM, Ferguson ND: Is it time to increase the frequency of use of high-frequency oscillatory ventilation? Crit Care 2005, 9: 339-340. 10.1186/cc3761PubMedCentralCrossRefPubMed
12.
go back to reference Slee-Wijffels FY, van der Vaart KR, Twisk JW, Markhorst DG, Plötz FB: High-frequency oscillatory ventilation in children: a single-center experience of 53 cases. Crit Care 2005, 9: R274-R279. 10.1186/cc3520PubMedCentralCrossRefPubMed Slee-Wijffels FY, van der Vaart KR, Twisk JW, Markhorst DG, Plötz FB: High-frequency oscillatory ventilation in children: a single-center experience of 53 cases. Crit Care 2005, 9: R274-R279. 10.1186/cc3520PubMedCentralCrossRefPubMed
13.
go back to reference Sessler CN: Sedation, analgesia, and neuromuscular blockade for high-frequency oscillatory ventilation. Crit Care Med 2005, 33: 209-216. 10.1097/01.CCM.0000156794.96880.DFCrossRef Sessler CN: Sedation, analgesia, and neuromuscular blockade for high-frequency oscillatory ventilation. Crit Care Med 2005, 33: 209-216. 10.1097/01.CCM.0000156794.96880.DFCrossRef
14.
go back to reference Banner MJ, Jaeger MJ, Kirby RR: Components of the work of breathing and implications for monitoring ventilator-dependent patients. Crit Care Med 1994, 22: 515-523.CrossRefPubMed Banner MJ, Jaeger MJ, Kirby RR: Components of the work of breathing and implications for monitoring ventilator-dependent patients. Crit Care Med 1994, 22: 515-523.CrossRefPubMed
15.
go back to reference Kirton OC, DeHaven CB, Morgan JP, Windsor J, Civetta JM: Elevated imposed work of breathing masquerading as ventilator weaning intolerance. Chest 1995, 108: 1021-1025.CrossRefPubMed Kirton OC, DeHaven CB, Morgan JP, Windsor J, Civetta JM: Elevated imposed work of breathing masquerading as ventilator weaning intolerance. Chest 1995, 108: 1021-1025.CrossRefPubMed
16.
go back to reference Kacmarek RM, Hess D: Basic principles of ventilator machinery. In Principles and practice of mechanical ventilation. Edited by: Tobin MJ. New York: McGraw-Hill, Inc.; 1994:65-110. Kacmarek RM, Hess D: Basic principles of ventilator machinery. In Principles and practice of mechanical ventilation. Edited by: Tobin MJ. New York: McGraw-Hill, Inc.; 1994:65-110.
17.
go back to reference SensorMedics: 3100 A High Frequency Oscillatory Ventilator Operator Manual. Yorba Linda, CA: SensorMedics; 1993. SensorMedics: 3100 A High Frequency Oscillatory Ventilator Operator Manual. Yorba Linda, CA: SensorMedics; 1993.
18.
go back to reference SensorMedics: 3100 B High Frequency Oscillatory Ventilator Operator's Manual. Yorba Linda, CA: SensorMedics; 1999. SensorMedics: 3100 B High Frequency Oscillatory Ventilator Operator's Manual. Yorba Linda, CA: SensorMedics; 1999.
19.
go back to reference Campbell EJ: The Respiratory Muscles and the Mechanics of Breathing. Chicago, IL: Year Book Medical; 1958. Campbell EJ: The Respiratory Muscles and the Mechanics of Breathing. Chicago, IL: Year Book Medical; 1958.
20.
go back to reference Agostini E, Campbell EJ, Freedman S: Energetics. In The Respiratory Muscles. Edited by: Campbell EJ, Agostini E, Newsom Davis J. Philadelphia, PA: Saunders; 1970:115-137. Agostini E, Campbell EJ, Freedman S: Energetics. In The Respiratory Muscles. Edited by: Campbell EJ, Agostini E, Newsom Davis J. Philadelphia, PA: Saunders; 1970:115-137.
21.
go back to reference Chiumello D, Pelosi P, Croci M, Bigatello LM, Gattinoni L: The effects of pressurization rate on breathing pattern, work of breathing, gas exchange and patient comfort in pressure support ventilation. Eur Respir J 2001, 18: 107-114. 10.1183/09031936.01.00083901CrossRefPubMed Chiumello D, Pelosi P, Croci M, Bigatello LM, Gattinoni L: The effects of pressurization rate on breathing pattern, work of breathing, gas exchange and patient comfort in pressure support ventilation. Eur Respir J 2001, 18: 107-114. 10.1183/09031936.01.00083901CrossRefPubMed
22.
go back to reference Vitacca M, Bianchi L, Zanotti E, Vianello A, Barbano L, Porta R, Clini E: Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilation. Chest 2004, 126: 851-859. 10.1378/chest.126.3.851CrossRefPubMed Vitacca M, Bianchi L, Zanotti E, Vianello A, Barbano L, Porta R, Clini E: Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilation. Chest 2004, 126: 851-859. 10.1378/chest.126.3.851CrossRefPubMed
23.
go back to reference French CJ: Work of breathing measurement in the critically ill patient. Anaesth Intensive Care 1999, 27: 561-573.PubMed French CJ: Work of breathing measurement in the critically ill patient. Anaesth Intensive Care 1999, 27: 561-573.PubMed
24.
go back to reference Girault C, Breton L, Richard JC, Tamion F, Vandelet P, Aboab J, Leroy J, Bonmarchand G: Mechanical effects of airway humidification devices in difficult to wean patients. Crit Care Med 2003, 31: 1306-1311. 10.1097/01.CCM.0000063284.92122.0ECrossRefPubMed Girault C, Breton L, Richard JC, Tamion F, Vandelet P, Aboab J, Leroy J, Bonmarchand G: Mechanical effects of airway humidification devices in difficult to wean patients. Crit Care Med 2003, 31: 1306-1311. 10.1097/01.CCM.0000063284.92122.0ECrossRefPubMed
25.
go back to reference MacIntyre NR: Respiratory mechanics in the patient who is weaning from the ventilator. Respir Care 2005, 50: 275-286.PubMed MacIntyre NR: Respiratory mechanics in the patient who is weaning from the ventilator. Respir Care 2005, 50: 275-286.PubMed
26.
go back to reference MacIntyre NR, Cook DJ, Ely EW, Epstein SK, Fink JB, Heffner JE, Hess D, Hubmayer RD, Scheinhorn DJ: Evidence-based guidelines for weaning and discontinuing ventilatory support: a collective task force facilitated by the American College of Chest Physicians; the American Association for Respiratory Care; and the American College of Critical Care Medicine. Chest 2001, 120: S375-S395. 10.1378/chest.120.6_suppl.375SCrossRef MacIntyre NR, Cook DJ, Ely EW, Epstein SK, Fink JB, Heffner JE, Hess D, Hubmayer RD, Scheinhorn DJ: Evidence-based guidelines for weaning and discontinuing ventilatory support: a collective task force facilitated by the American College of Chest Physicians; the American Association for Respiratory Care; and the American College of Critical Care Medicine. Chest 2001, 120: S375-S395. 10.1378/chest.120.6_suppl.375SCrossRef
27.
go back to reference Putensen C, Hering R, Muders T, Wrigge H: Assisted breathing is better in acute respiratory failure. Curr Opin Crit Care 2005, 11: 63-68. 10.1097/00075198-200502000-00010CrossRefPubMed Putensen C, Hering R, Muders T, Wrigge H: Assisted breathing is better in acute respiratory failure. Curr Opin Crit Care 2005, 11: 63-68. 10.1097/00075198-200502000-00010CrossRefPubMed
28.
go back to reference Zapletal A, Samanek M, Paul T: Lung Function in Children and Adolescents. Basel: Karger; 1987. Zapletal A, Samanek M, Paul T: Lung Function in Children and Adolescents. Basel: Karger; 1987.
29.
go back to reference Thibeault DW, Clutario B, Awld PA: The oxygen cost of breathing in the premature infant. Pediatrics 1966, 37: 954-959.PubMed Thibeault DW, Clutario B, Awld PA: The oxygen cost of breathing in the premature infant. Pediatrics 1966, 37: 954-959.PubMed
Metadata
Title
Imposed work of breathing during high-frequency oscillatory ventilation: a bench study
Authors
Marc van Heerde
Huib R van Genderingen
Tom Leenhoven
Karel Roubik
Frans B Plötz
Dick G Markhorst
Publication date
01-02-2006
Publisher
BioMed Central
Published in
Critical Care / Issue 1/2006
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc3988

Other articles of this Issue 1/2006

Critical Care 1/2006 Go to the issue