Objective

An endotracheal tube (ETT) imposes work of breathing on mechanically ventilated patients. Using a bellows-in-a-box model lung, we compared the tube compensation (TC) performances of the Nellcor Puritan-Bennett 840 ventilator and of the Dräger Evita 4 ventilator.

Measurements and results

Each ventilator was connected to the model lung. The respiratory rate of the model lung was set at 10 breaths/min with 1 s inspiratory time. Inspiratory flows were 30 or 60 l/min. A full-length 8 mm bore ETT was inserted between the ventilator circuit and the model lung. The TC was set at 0%, 10%, 50%, and 100% for both ventilators. Pressure was monitored at the airway, the trachea, and the pleura, and the data were recorded on a computer for later analysis of the delay time, of the inspiratory trigger pressure, and of the pressure–time product (PTP). The delay time was calculated as the time between the start of inspiration and minimum airway pressure, and the inspiratory trigger pressure was defined as the most negative pressure level. The same measurements were performed under pressure support ventilation of 4 and 8 cmH2O.
The PTP increased according to the magnitude of inspiratory flow. Even with 100% TC, neither ventilator could completely compensate for the PTP imposed by the ETT. At 0% TC the PTP tended to be less with the Nellcor Puritan-Bennett 840 ventilator, while at 100% TC the PTP tended to be less with the Dräger Evita 4 ventilator. A small amount of pressure support can be equally effective to reduce the inspiratory effort compared with the TC.

Conclusion

Although both ventilators provided effective TC, even when set to 100% TC they could not entirely compensate for a ventilator and ETT-imposed work of breathing. The effect of TC is less than that of pressure support ventilation. Physicians should be aware of this when using TC in weaning trials.
Literature
1.
go back to reference Pelosi P, Cereda M, Foti G, Giacomini M, Pesenti A: Alterations of lung and chest wall mechanics in patients with acute lung injury: effects of positive end-expiratory pressure. Am J Respir Crit Care Med 1995, 152: 531-537.CrossRefPubMed Pelosi P, Cereda M, Foti G, Giacomini M, Pesenti A: Alterations of lung and chest wall mechanics in patients with acute lung injury: effects of positive end-expiratory pressure. Am J Respir Crit Care Med 1995, 152: 531-537.CrossRefPubMed
2.
go back to reference Pesenti A, Pelosi P, Rossi N, Virtuani A, Brazzi L, Rossi A: The effects of positive end-expiratory pressure on respiratory resistance in patients with the adult respiratory distress syndrome and in normal anesthetized subjects. Am Rev Respir Dis 1991, 144: 101-107.CrossRefPubMed Pesenti A, Pelosi P, Rossi N, Virtuani A, Brazzi L, Rossi A: The effects of positive end-expiratory pressure on respiratory resistance in patients with the adult respiratory distress syndrome and in normal anesthetized subjects. Am Rev Respir Dis 1991, 144: 101-107.CrossRefPubMed
3.
go back to reference Mols G, Kessler V, Benzing A, Lichtwarck-Aschoff M, Geiger K, Guttmann J: Is pulmonary resistance constant, within the range of tidal volume ventilation, in patients with ARDS? Br J Anaesth 2001, 86: 176-182. 10.1093/bja/86.2.176CrossRefPubMed Mols G, Kessler V, Benzing A, Lichtwarck-Aschoff M, Geiger K, Guttmann J: Is pulmonary resistance constant, within the range of tidal volume ventilation, in patients with ARDS? Br J Anaesth 2001, 86: 176-182. 10.1093/bja/86.2.176CrossRefPubMed
4.
go back to reference Straus C, Louis B, Isabey D, Lemaire F, Harf A, Brochard L: Comparison of the endotracheal tube and the upper airway to breathing workload. Am J Respir Crit Care Med 1998, 157: 23-30.CrossRefPubMed Straus C, Louis B, Isabey D, Lemaire F, Harf A, Brochard L: Comparison of the endotracheal tube and the upper airway to breathing workload. Am J Respir Crit Care Med 1998, 157: 23-30.CrossRefPubMed
5.
go back to reference Fabry B, Haberthür C, Zappe D, Guttmann J, Kuhlen R, Stocker R: Breathing pattern and additional work of breathing in spontaneously breathing patients with different ventilatory demand during inspiratory pressure support and automatic tube compensation. Intensive Care Med 1997, 23: 545-552. 10.1007/s001340050371CrossRefPubMed Fabry B, Haberthür C, Zappe D, Guttmann J, Kuhlen R, Stocker R: Breathing pattern and additional work of breathing in spontaneously breathing patients with different ventilatory demand during inspiratory pressure support and automatic tube compensation. Intensive Care Med 1997, 23: 545-552. 10.1007/s001340050371CrossRefPubMed
6.
go back to reference Haberthür C, Elsasser S, Eberhard L, Stocker R, Guttmann J: Total versus tube-related additional work of breathing in ventilator-dependent patients. Acta Anaesthesiol Scand 2000, 44: 749-757. 10.1034/j.1399-6576.2000.440615.xCrossRefPubMed Haberthür C, Elsasser S, Eberhard L, Stocker R, Guttmann J: Total versus tube-related additional work of breathing in ventilator-dependent patients. Acta Anaesthesiol Scand 2000, 44: 749-757. 10.1034/j.1399-6576.2000.440615.xCrossRefPubMed
7.
go back to reference Guttmann J, Bernhard H, Mols G, Benzing A, Hofmann P, Haberthür C, Zappe D, Fabry B, Geiger K: Respiratory comfort of automatic tube compensation and inspiratory pressure support in conscious human. Intensive Care Med 1997, 23: 1119-1124. 10.1007/s001340050467CrossRefPubMed Guttmann J, Bernhard H, Mols G, Benzing A, Hofmann P, Haberthür C, Zappe D, Fabry B, Geiger K: Respiratory comfort of automatic tube compensation and inspiratory pressure support in conscious human. Intensive Care Med 1997, 23: 1119-1124. 10.1007/s001340050467CrossRefPubMed
8.
go back to reference Fujino Y, Uchiyama A, Mashimo T, Nishimura M: Spontaneously breathing lung model comparison of work of breathing between automatic tube compensation and pressure support. Respir Care 2003, 48: 38-45.PubMed Fujino Y, Uchiyama A, Mashimo T, Nishimura M: Spontaneously breathing lung model comparison of work of breathing between automatic tube compensation and pressure support. Respir Care 2003, 48: 38-45.PubMed
9.
go back to reference Miyoshi E, Fujino Y, Mashimo T, Nishimura M: Performance of transport ventilator with patient-triggered ventilation. Chest 2000, 118: 1109-1115. 10.1378/chest.118.4.1109CrossRefPubMed Miyoshi E, Fujino Y, Mashimo T, Nishimura M: Performance of transport ventilator with patient-triggered ventilation. Chest 2000, 118: 1109-1115. 10.1378/chest.118.4.1109CrossRefPubMed
10.
go back to reference Bolder PM, Healy TEJ, Bolder AR, Beatty PCW, Kay B: The extra work of breathing through adult endotracheal tubes. Anesth Analg 1986, 65: 853-859.CrossRefPubMed Bolder PM, Healy TEJ, Bolder AR, Beatty PCW, Kay B: The extra work of breathing through adult endotracheal tubes. Anesth Analg 1986, 65: 853-859.CrossRefPubMed
11.
go back to reference Shapiro M, Wilson K, Casar G, Bloom K, Teague RB: Work of breathing through different sized endotracheal tubes. Crit Care Med 1986, 14: 1028-1031.CrossRefPubMed Shapiro M, Wilson K, Casar G, Bloom K, Teague RB: Work of breathing through different sized endotracheal tubes. Crit Care Med 1986, 14: 1028-1031.CrossRefPubMed
12.
go back to reference Brochard L, Rua F, Lorino H, Lemaire F, Harf A: Inspiratory pressure support compensates for the additional work of breathing caused by the endotracheal tube. Anesthesiology 1991, 75: 739-745.CrossRefPubMed Brochard L, Rua F, Lorino H, Lemaire F, Harf A: Inspiratory pressure support compensates for the additional work of breathing caused by the endotracheal tube. Anesthesiology 1991, 75: 739-745.CrossRefPubMed
13.
go back to reference Takeuchi M, Imanaka H, Miyano H, Kumon K, Nishimura M: Effect of patient-triggered ventilation on respiratory workload in infants after cardiac surgery. Anesthesiology 2000, 93: 1238-1244. 10.1097/00000542-200011000-00017CrossRefPubMed Takeuchi M, Imanaka H, Miyano H, Kumon K, Nishimura M: Effect of patient-triggered ventilation on respiratory workload in infants after cardiac surgery. Anesthesiology 2000, 93: 1238-1244. 10.1097/00000542-200011000-00017CrossRefPubMed
14.
go back to reference Sassoon CS, Lodia R, Rheeman CH, Kuei JH, Light RW, Mahutte CK: Inspiratory muscle work of breathing during flow-by, demand-flow, and continuous-flow systems in patients with chronic obstructive pulmonary disease. Am Rev Respir Dis 1992, 145: 1219-1222.CrossRefPubMed Sassoon CS, Lodia R, Rheeman CH, Kuei JH, Light RW, Mahutte CK: Inspiratory muscle work of breathing during flow-by, demand-flow, and continuous-flow systems in patients with chronic obstructive pulmonary disease. Am Rev Respir Dis 1992, 145: 1219-1222.CrossRefPubMed
15.
go back to reference Sassoon CS, Del Rosario N, Fei R, Rheeman CH, Gruer SE, Mahutte CK: Influence of pressure- and flow-triggered synchronous intermittent mandatory ventilation on inspiratory muscle work. Crit Care Med 1994, 22: 1933-1941.CrossRefPubMed Sassoon CS, Del Rosario N, Fei R, Rheeman CH, Gruer SE, Mahutte CK: Influence of pressure- and flow-triggered synchronous intermittent mandatory ventilation on inspiratory muscle work. Crit Care Med 1994, 22: 1933-1941.CrossRefPubMed
Metadata
Title
Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
Authors
Yoshiko Maeda
Yuji Fujino
Akinori Uchiyama
Nobuyuki Taenaka
Takashi Mashimo
Masaji Nishimura
Publication date
01-10-2003
Publisher
BioMed Central
Published in
Critical Care / Issue 5/2003
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc2343

Other articles of this Issue 5/2003

Critical Care 5/2003 Go to the issue