Published in: BMC Medical Physics 1/2009

Open Access 01-12-2009 | Research article

Repeatability of regional myocardial blood flow calculation in 82Rb PET imaging

Authors: Karin Knešaurek, Josef Machac, Zhuangyu Zhang

Published in: BMC Medical Physics | Issue 1/2009

Login to get access

Abstract

Background

We evaluated the repeatability of the calculation of myocardial blood flow (MBF) at rest and pharmacological stress, and calculated the coronary flow reserve (CFR) utilizing 82Rb PET imaging. The aim of the research was to prove high repeatability for global MBF and CFR values and good repeatability for regional MBF and CFR values. The results will have significant impact on cardiac PET imaging in terms of making it more affordable and increasing its use.

Methods

12 normal volunteers were imaged at rest and during pharmacological stress, with 2220 MBq of 82Rb each. A GE Advance PET system was used to acquire dynamic 50-frame studies. MBF was calculated with a 2-compartmental model using a modified PMOD program (PMOD; University Hospital Zurich, Zurich, Switzerland). Two differential equations, describing a 2-compartmental model, were solved by numerical integration and using Levenberg-Marquardt's method for fitting data. The PMOD program defines 16 standard segments and calculates myocardial flow for each segment, as well as average septal, anterior, lateral, inferior and global flow. Repeatability was evaluated according to the method of Bland and Altman.

Results

Global rest and stress MBF, as well as global CFR, showed very good repeatability. No significant differences were found between the paired resting global MBF (0.63 ± 0.13 vs. 0.64 ± 0.13 mL/min/g; mean difference, -1.0% ± 2.6%) and the stress global MBF (1.37 ± 0.23 vs. 1.37 ± 0.24; mean difference, 0.1% ± 2.3%). Global CFR was highly reproducible (2.25 ± 0.56 vs. 2.22 ± 0.54, P = not statistically significant; mean difference, 1.3% ± 14.3%). Repeatability coefficients for global rest MBF were 0.033 (5.2%) and stress MBF 0.062 (4.5%) mL/min/g. Regional rest and stress MBF and CFR have shown good reproducibility. The average per sector repeatability coefficients for rest MBF were 0.056 (8.5%) and stress MBF 0.089 (6.3%) mL/min/g, and average repeatability coefficient for CFR was 0.25 (10.6%).

Conclusion

The results of the study show that software calculation of MBF and CFR with 82Rb myocardial PET imaging is highly repeatable for global values and has good repeatability for regional values.
Appendix
Available only for authorised users
Literature
1.
go back to reference Jadvar H, Strauss HW, Segall GM: SPECT and PET in the Evaluation of Coronary Artery Disease. Radiographics. 1999, 19: 915-926.CrossRefPubMed Jadvar H, Strauss HW, Segall GM: SPECT and PET in the Evaluation of Coronary Artery Disease. Radiographics. 1999, 19: 915-926.CrossRefPubMed
2.
go back to reference Kaufmann PA, Camici PG: Myocardial Blood Flow Measurement by PET: Technical Aspects and Clinical Applications. J Nucl Med. 2005, 46: 75-88.PubMed Kaufmann PA, Camici PG: Myocardial Blood Flow Measurement by PET: Technical Aspects and Clinical Applications. J Nucl Med. 2005, 46: 75-88.PubMed
3.
go back to reference Mullani NA, Goldstein RA, Gould KL, et al: Myocardial perfusion with rubidium-82. I. Measurement of extraction fraction and flow with external detectors. J Nucl Med. 1983, 24: 898-906.PubMed Mullani NA, Goldstein RA, Gould KL, et al: Myocardial perfusion with rubidium-82. I. Measurement of extraction fraction and flow with external detectors. J Nucl Med. 1983, 24: 898-906.PubMed
4.
go back to reference Goldstein RA, Mullani NA, Marani SK, Fisher DJ, Gould KL, O'Brien HA: Myocardial perfusion with rubidium-82. II. Effects of metabolic and pharmacologic interventions. J Nucl Med. 1983, 24: 907-915.PubMed Goldstein RA, Mullani NA, Marani SK, Fisher DJ, Gould KL, O'Brien HA: Myocardial perfusion with rubidium-82. II. Effects of metabolic and pharmacologic interventions. J Nucl Med. 1983, 24: 907-915.PubMed
5.
go back to reference Goldstein RA, Mullani NA, Wong WH, et al: Positron imaging of myocardial infarction with rubidium-82. J Nucl Med. 1986, 27: 1824-1829.PubMed Goldstein RA, Mullani NA, Wong WH, et al: Positron imaging of myocardial infarction with rubidium-82. J Nucl Med. 1986, 27: 1824-1829.PubMed
6.
go back to reference Herrero P, Markham J, Shelton ME, Weinheimer CJ, Bergmann SR: Noninvasive quantification of regional myocardial perfusion with rubidium-82 and positron emission tomography: exploration of a mathematical model. Circulation. 1990, 82: 1377-1386.CrossRefPubMed Herrero P, Markham J, Shelton ME, Weinheimer CJ, Bergmann SR: Noninvasive quantification of regional myocardial perfusion with rubidium-82 and positron emission tomography: exploration of a mathematical model. Circulation. 1990, 82: 1377-1386.CrossRefPubMed
7.
go back to reference Herrero P, Markham J, Shelton ME, Bergmann SR: Implementation and evaluation of a two-compartment model for quantification of myocardial perfusion with rubidium-82 and positron emission tomography. Circ Res. 1992, 70: 496-507.CrossRefPubMed Herrero P, Markham J, Shelton ME, Bergmann SR: Implementation and evaluation of a two-compartment model for quantification of myocardial perfusion with rubidium-82 and positron emission tomography. Circ Res. 1992, 70: 496-507.CrossRefPubMed
8.
go back to reference Yoshida K, Mullani N, Gould KL: Coronary flow and flow reserve by PET simplified for clinical applications using rubidium-82 or nitrogen-13-ammonia. J Nucl Med. 1996, 37: 1701-1712.PubMed Yoshida K, Mullani N, Gould KL: Coronary flow and flow reserve by PET simplified for clinical applications using rubidium-82 or nitrogen-13-ammonia. J Nucl Med. 1996, 37: 1701-1712.PubMed
9.
go back to reference Lin JW, Robert R, Sciacca RR, Chou R-L, Laine AF, Bergmann SR: Quantification of myocardial perfusion in human subjects using 82Rb and wavelet-based noise reduction. J Nucl Med. 2001, 42: 201-208.PubMed Lin JW, Robert R, Sciacca RR, Chou R-L, Laine AF, Bergmann SR: Quantification of myocardial perfusion in human subjects using 82Rb and wavelet-based noise reduction. J Nucl Med. 2001, 42: 201-208.PubMed
10.
go back to reference Votaw JR, White M: Comparison of 2D and 3D cardiac 82Rb PET Studies. J Nucl Med. 2001, 42: 701-706.PubMed Votaw JR, White M: Comparison of 2D and 3D cardiac 82Rb PET Studies. J Nucl Med. 2001, 42: 701-706.PubMed
11.
go back to reference Knešaurek K, Machac J, Krynyckyi BR, Almeida OD: Comparison of 2D and 3D 82Rb myocardial perfusion PET imaging. J Nucl Med. 2003, 44: 1350-1356.PubMed Knešaurek K, Machac J, Krynyckyi BR, Almeida OD: Comparison of 2D and 3D 82Rb myocardial perfusion PET imaging. J Nucl Med. 2003, 44: 1350-1356.PubMed
12.
go back to reference Dimitrakopoulou-Strauss A, Strauss LG, Schwarzbach M, Burger C, Heichel T, Willeke F, Mechtersheimer G, Lehnert T: Dynamic PET 18F-FDG Studies in Patients with Primary and Recurrent Soft-Tissue Sarcomas: Impact on Diagnosis and Correlation with Grading. J Nucl Med. 2001, 42: 713-720.PubMed Dimitrakopoulou-Strauss A, Strauss LG, Schwarzbach M, Burger C, Heichel T, Willeke F, Mechtersheimer G, Lehnert T: Dynamic PET 18F-FDG Studies in Patients with Primary and Recurrent Soft-Tissue Sarcomas: Impact on Diagnosis and Correlation with Grading. J Nucl Med. 2001, 42: 713-720.PubMed
13.
go back to reference Wyss CA, Koepfli P, Mikolajczyk K, Burger C, Schulthess GK, Kaufmann PA: Bicycle exercise stress in PET for assessment of coronary flow reserve: repeatability and comparison with adenosine stress. J Nucl Med. 2003, 44: 146-154.PubMed Wyss CA, Koepfli P, Mikolajczyk K, Burger C, Schulthess GK, Kaufmann PA: Bicycle exercise stress in PET for assessment of coronary flow reserve: repeatability and comparison with adenosine stress. J Nucl Med. 2003, 44: 146-154.PubMed
14.
go back to reference Daubechies I: Ten Lectures On Wavelets. SIAM. 1992 Daubechies I: Ten Lectures On Wavelets. SIAM. 1992
15.
go back to reference Schiller NB, Shah PM, Crawford M, et al: Recommendations for quantification of the left ventricle by two-dimensional echocardiography. J Am Soc Echocardiogr. 1989, 2: 358-367.CrossRefPubMed Schiller NB, Shah PM, Crawford M, et al: Recommendations for quantification of the left ventricle by two-dimensional echocardiography. J Am Soc Echocardiogr. 1989, 2: 358-367.CrossRefPubMed
16.
go back to reference Passing H, Bablok W: A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part I. J Clin Chem Clin Biochem. 1983, 21: 709-720.PubMed Passing H, Bablok W: A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part I. J Clin Chem Clin Biochem. 1983, 21: 709-720.PubMed
17.
go back to reference Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 327: 307-310. 10.1016/S0140-6736(86)91008-1.CrossRef Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 327: 307-310. 10.1016/S0140-6736(86)91008-1.CrossRef
18.
go back to reference Precision of Test Methods. I: Guide for the Determination and Reproducibility for a Standard Test Method. 1976, London, U.K.: British Standards Institution, 54-97. Precision of Test Methods. I: Guide for the Determination and Reproducibility for a Standard Test Method. 1976, London, U.K.: British Standards Institution, 54-97.
19.
go back to reference Phelps EM: PET – Molecular Imaging and Its Biological Applications. 2004, Springer, New York, 396-408. Phelps EM: PET – Molecular Imaging and Its Biological Applications. 2004, Springer, New York, 396-408.
20.
go back to reference Machac J: Cardiac Positron Emission Tomography Imaging. Semin Nucl Med. 2004, 35: 17-36. 10.1053/j.semnuclmed.2004.09.002.CrossRef Machac J: Cardiac Positron Emission Tomography Imaging. Semin Nucl Med. 2004, 35: 17-36. 10.1053/j.semnuclmed.2004.09.002.CrossRef
21.
go back to reference Lortie M, Mostert K, Kelly C, et al: Quantification of myocardial blood flow with 82Rb dynamic PET imaging. Eur J Nucl Med Mol Imagin. 2005, 34 (11): 1765-1774. 10.1007/s00259-007-0478-2.CrossRef Lortie M, Mostert K, Kelly C, et al: Quantification of myocardial blood flow with 82Rb dynamic PET imaging. Eur J Nucl Med Mol Imagin. 2005, 34 (11): 1765-1774. 10.1007/s00259-007-0478-2.CrossRef
22.
go back to reference Chareonthaitawee P, Kaufmann PA, Rimoldi O, Camici PG: Heterogeneity of resting and hyperemic myocardial blood flow in healthy humans. Cardiovascular Research. 2001, 50: 151-161. 10.1016/S0008-6363(01)00202-4.CrossRefPubMed Chareonthaitawee P, Kaufmann PA, Rimoldi O, Camici PG: Heterogeneity of resting and hyperemic myocardial blood flow in healthy humans. Cardiovascular Research. 2001, 50: 151-161. 10.1016/S0008-6363(01)00202-4.CrossRefPubMed
23.
go back to reference Nagamachi S, Czernin J, Kim AS, Sun KT, Bottcher M, Phelps ME, Schelbert HR: Reproducibility of measurements of regional resting and hyperemic myocardial blood flow assessed with PET. J Nucl Med. 1996, 37: 1626-1631.PubMed Nagamachi S, Czernin J, Kim AS, Sun KT, Bottcher M, Phelps ME, Schelbert HR: Reproducibility of measurements of regional resting and hyperemic myocardial blood flow assessed with PET. J Nucl Med. 1996, 37: 1626-1631.PubMed
24.
go back to reference El Fakhri G, Sitek A, Kijewski MF, Di Carli MF, Moore SC: Quantitative parametric 82Rb cardiac PET using generalized factor analysis of dynamic Sequences (GFADS) [abstract]. J Nuc Med. 2005, 46 (8): 1264-1271. El Fakhri G, Sitek A, Kijewski MF, Di Carli MF, Moore SC: Quantitative parametric 82Rb cardiac PET using generalized factor analysis of dynamic Sequences (GFADS) [abstract]. J Nuc Med. 2005, 46 (8): 1264-1271.
Metadata
Title
Repeatability of regional myocardial blood flow calculation in 82Rb PET imaging
Authors
Karin Knešaurek
Josef Machac
Zhuangyu Zhang
Publication date
01-12-2009
Publisher
BioMed Central
Published in
BMC Medical Physics / Issue 1/2009
Electronic ISSN: 1756-6649
DOI
https://doi.org/10.1186/1756-6649-9-2

Other articles of this Issue 1/2009

BMC Medical Physics 1/2009 Go to the issue