Skip to main content
Top
Published in: Journal of Orthopaedic Surgery and Research 1/2012

Open Access 01-12-2012 | Research article

Biomechanical stability of a supra-acetabular pedicle screw Internal Fixation device (INFIX) vs External Fixation and plates for vertically unstable pelvic fractures

Authors: Jonathan M Vigdorchik, Amanda O Esquivel, Xin Jin, King H Yang, Ndidi A Onwudiwe, Rahul Vaidya

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2012

Login to get access

Abstract

Background

We have recently developed a subcutaneous anterior pelvic fixation technique (INFIX). This internal fixator permits patients to sit, roll over in bed and lie on their sides without the cumbersome external appliances or their complications. The purpose of this study was to evaluate the biomechanical stability of this novel supraacetabular pedicle screw internal fixation construct (INFIX) and compare it to standard internal fixation and external fixation techniques in a single stance pelvic fracture model.

Methods

Nine synthetic pelves with a simulated anterior posterior compression type III injury were placed into three groups (External Fixator, INFIX and Internal Fixation). Displacement, total axial stiffness, and the stiffness at the pubic symphysis and SI joint were calculated. Displacement and stiffness were compared by ANOVA with a Bonferroni adjustment for multiple comparisons

Results

The mean displacement at the pubic symphysis was 20, 9 and 0.8 mm for external fixation, INFIX and internal fixation, respectively. Plate fixation was significantly stiffer than the INFIX and external Fixator (P = 0.01) at the symphysis pubis. The INFIX device was significantly stiffer than external fixation (P = 0.017) at the symphysis pubis. There was no significant difference in SI joint displacement between any of the groups.

Conclusions

Anterior plate fixation is stiffer than both the INFIX and external fixation in single stance pelvic fracture model. The INFIX was stiffer than external fixation for both overall axial stiffness, and stiffness at the pubic symphysis. Combined with the presumed benefit of minimizing the complications associated with external fixation, the INFIX may be a more preferable option for temporary anterior pelvic fixation in situations where external fixation may have otherwise been used.
Appendix
Available only for authorised users
Literature
1.
go back to reference Gardner MJ, Kendoff D, Ostermeier S, Citak M, Hüfner T, Krettek C, Nork SE: Sacroiliac joint compression using an anterior pelvic compressor: a mechanical study in synthetic bone. J Orthop Trauma. 2007, 21 (7): 435-441. 10.1097/BOT.0b013e318126bb8e.CrossRef Gardner MJ, Kendoff D, Ostermeier S, Citak M, Hüfner T, Krettek C, Nork SE: Sacroiliac joint compression using an anterior pelvic compressor: a mechanical study in synthetic bone. J Orthop Trauma. 2007, 21 (7): 435-441. 10.1097/BOT.0b013e318126bb8e.CrossRef
2.
go back to reference Gardner MJ, Nork SE: Stabilization of unstable pelvic fractures with supraacetabular compression external fixation. J Orthop Trauma. 2007, 21 (4): 269273-CrossRef Gardner MJ, Nork SE: Stabilization of unstable pelvic fractures with supraacetabular compression external fixation. J Orthop Trauma. 2007, 21 (4): 269273-CrossRef
3.
go back to reference Grimm MR, Vrahas MS, Thomas KA: Pressure-volume characteristics of the intact and disrupted pelvic retroperitoneum. J Trauma. 1998, 44 (3): 454-459. 10.1097/00005373-199803000-00006.CrossRef Grimm MR, Vrahas MS, Thomas KA: Pressure-volume characteristics of the intact and disrupted pelvic retroperitoneum. J Trauma. 1998, 44 (3): 454-459. 10.1097/00005373-199803000-00006.CrossRef
4.
go back to reference Kregor PJ, Routt ML: Unstable pelvic ring disruptions in unstable patients. Injury. 1999, 30 (Suppl 2): B19-B28.CrossRef Kregor PJ, Routt ML: Unstable pelvic ring disruptions in unstable patients. Injury. 1999, 30 (Suppl 2): B19-B28.CrossRef
5.
go back to reference Riemer BL, Butterfield SL, Diamond DL, Young JC, Raves JJ, Cottington E, Kislan K: Acute mortality associated with injuries to the pelvic ring: the role of early patient mobilization and external fixation. J Trauma. 1993, 35 (5): 671-675. 10.1097/00005373-199311000-00003. discussion 676–677CrossRef Riemer BL, Butterfield SL, Diamond DL, Young JC, Raves JJ, Cottington E, Kislan K: Acute mortality associated with injuries to the pelvic ring: the role of early patient mobilization and external fixation. J Trauma. 1993, 35 (5): 671-675. 10.1097/00005373-199311000-00003. discussion 676–677CrossRef
6.
go back to reference Hupel TM, McKee MD, Waddell JP, Schemitsch EH: Primary external fixation of rotationally unstable pelvic fractures in obese patients. J Trauma. 1998, 45 (1): 111-115. 10.1097/00005373-199807000-00024.CrossRef Hupel TM, McKee MD, Waddell JP, Schemitsch EH: Primary external fixation of rotationally unstable pelvic fractures in obese patients. J Trauma. 1998, 45 (1): 111-115. 10.1097/00005373-199807000-00024.CrossRef
7.
go back to reference Vaidya R, Colen R, Vigdorchik J, Tonnos F, Sethi A: Treatment of unstable pelvic ring injuries with an internal anterior fixator and posterior fixation: initial clinical series. J Orthop Trauma. 2012, 26 (1): 1-8. 10.1097/BOT.0b013e318233b8a7.CrossRef Vaidya R, Colen R, Vigdorchik J, Tonnos F, Sethi A: Treatment of unstable pelvic ring injuries with an internal anterior fixator and posterior fixation: initial clinical series. J Orthop Trauma. 2012, 26 (1): 1-8. 10.1097/BOT.0b013e318233b8a7.CrossRef
8.
go back to reference Vaidya R, Kubiak EN, Bergin PF, Dombroski DG, Critchlow RJ, Sethi A, Starr AJ: Complications of Anterior Subcutaneous Internal Fixation for Unstable Pelvis Fractures: a Multicenter Study. Clin Orthop Relat Res. 2012, Epub ahead of print Vaidya R, Kubiak EN, Bergin PF, Dombroski DG, Critchlow RJ, Sethi A, Starr AJ: Complications of Anterior Subcutaneous Internal Fixation for Unstable Pelvis Fractures: a Multicenter Study. Clin Orthop Relat Res. 2012, Epub ahead of print
9.
go back to reference Varga E, Hearn T, Powell J, Tile M: Effects of method of internal fixation of symphyseal disruptions on stability of the pelvic ring. Injury. 1995, 26: 75-80. 10.1016/0020-1383(95)92180-I.CrossRef Varga E, Hearn T, Powell J, Tile M: Effects of method of internal fixation of symphyseal disruptions on stability of the pelvic ring. Injury. 1995, 26: 75-80. 10.1016/0020-1383(95)92180-I.CrossRef
10.
go back to reference Yinger K, Scalise J, Olson SA, Bay BK, Finkemeier CG: Biomechanical comparison of posterior pelvic ring fixation. J Orthop Trauma. 2003, 17 (7): 481-487. 10.1097/00005131-200308000-00002.CrossRef Yinger K, Scalise J, Olson SA, Bay BK, Finkemeier CG: Biomechanical comparison of posterior pelvic ring fixation. J Orthop Trauma. 2003, 17 (7): 481-487. 10.1097/00005131-200308000-00002.CrossRef
11.
go back to reference Ponsen KJ, Joosse P, Van Dijke GA, Snijders CJ: External fixation of the pelvic ring: an experimental study on the role of pin diameter, pin position, and parasymphyseal fixator pins. Acta Orthop. 2007, 78 (5): 648-653. 10.1080/17453670710014347.CrossRef Ponsen KJ, Joosse P, Van Dijke GA, Snijders CJ: External fixation of the pelvic ring: an experimental study on the role of pin diameter, pin position, and parasymphyseal fixator pins. Acta Orthop. 2007, 78 (5): 648-653. 10.1080/17453670710014347.CrossRef
12.
go back to reference Pohlemann T, Angst M, Schneider E, Ganz R, Tscherne H: Fixation of transforaminal sacrum fractures: a biomechanical study. J Orthop Trauma. 1993, 7 (2): 107-117. 10.1097/00005131-199304000-00002.CrossRef Pohlemann T, Angst M, Schneider E, Ganz R, Tscherne H: Fixation of transforaminal sacrum fractures: a biomechanical study. J Orthop Trauma. 1993, 7 (2): 107-117. 10.1097/00005131-199304000-00002.CrossRef
13.
go back to reference Shaw JA, Mino DE, Werner FW, Murray DG: Posterior stabilization of pelvic fractures by use of threaded compression rods. Case reports and mechanical testing. Clin Orthop Relat Res. 1985, 192: 240-254. Shaw JA, Mino DE, Werner FW, Murray DG: Posterior stabilization of pelvic fractures by use of threaded compression rods. Case reports and mechanical testing. Clin Orthop Relat Res. 1985, 192: 240-254.
14.
go back to reference Comstock CP, van der Meulen MC, Goodman SB: Biomechanical comparison of posterior internal fixation techniques for unstable pelvic fractures. J Orthop Trauma. 1996, 10: 517-522. 10.1097/00005131-199611000-00001.CrossRef Comstock CP, van der Meulen MC, Goodman SB: Biomechanical comparison of posterior internal fixation techniques for unstable pelvic fractures. J Orthop Trauma. 1996, 10: 517-522. 10.1097/00005131-199611000-00001.CrossRef
15.
go back to reference Leighton RK, Waddell JP, Bray TJ, Bray TJ, Hapman MW, Simpson L, Martin RB, Sharkey NA: Biomechanical testing of new and old fixation devices for vertical shear fractures of the pelvis. J Orthop Trauma. 1991, 5: 313-317. 10.1097/00005131-199109000-00010.CrossRef Leighton RK, Waddell JP, Bray TJ, Bray TJ, Hapman MW, Simpson L, Martin RB, Sharkey NA: Biomechanical testing of new and old fixation devices for vertical shear fractures of the pelvis. J Orthop Trauma. 1991, 5: 313-317. 10.1097/00005131-199109000-00010.CrossRef
16.
go back to reference Simonian PT, Routt C, Harrington RM, Tencer AF: Internal fixation for the transforaminal sacral fracture. Clin Orthop Relat Res. 1996, 323: 202-209.CrossRef Simonian PT, Routt C, Harrington RM, Tencer AF: Internal fixation for the transforaminal sacral fracture. Clin Orthop Relat Res. 1996, 323: 202-209.CrossRef
17.
go back to reference Simonian PT, Routt ML, Harrington RM, Mayo KA, Tencer AF: Biomechanical simulation of the anteroposterior compression injury of the pelvis. An understanding of instability and fixation. Clin Orthop Relat Res. 1994, 309: 245-256. Simonian PT, Routt ML, Harrington RM, Mayo KA, Tencer AF: Biomechanical simulation of the anteroposterior compression injury of the pelvis. An understanding of instability and fixation. Clin Orthop Relat Res. 1994, 309: 245-256.
18.
go back to reference Simonian PT, Routt ML, Harrington RM, Tencer AF: Internal fixation of the unstable anterior pelvic ring: a biomechanical comparison of standard plating techniques and the retrograde medullary superior pubic ramus screw. J Orthop Trauma. 1994, 8 (6): 476-482.CrossRef Simonian PT, Routt ML, Harrington RM, Tencer AF: Internal fixation of the unstable anterior pelvic ring: a biomechanical comparison of standard plating techniques and the retrograde medullary superior pubic ramus screw. J Orthop Trauma. 1994, 8 (6): 476-482.CrossRef
19.
go back to reference Van Zwienen CM, van den Bosch EW, Snijders CJ, Kleinrensink GJ, van Vugt AB: Biomechanical comparison of sacroiliac screw techniques for unstable pelvic ring fractures. J Orthop Trauma. 2004, 18: 589-595. 10.1097/00005131-200410000-00002.CrossRef Van Zwienen CM, van den Bosch EW, Snijders CJ, Kleinrensink GJ, van Vugt AB: Biomechanical comparison of sacroiliac screw techniques for unstable pelvic ring fractures. J Orthop Trauma. 2004, 18: 589-595. 10.1097/00005131-200410000-00002.CrossRef
20.
go back to reference Dujardin FH, Roussignol X, Hossenbaccus M, Thomine JM: Experimental study of the sacroiliac joint micromotion in pelvic disruption. J Orthop Trauma. 2002, 16: 99-103. 10.1097/00005131-200202000-00005.CrossRef Dujardin FH, Roussignol X, Hossenbaccus M, Thomine JM: Experimental study of the sacroiliac joint micromotion in pelvic disruption. J Orthop Trauma. 2002, 16: 99-103. 10.1097/00005131-200202000-00005.CrossRef
21.
go back to reference MacAvoy MC, McClellan RT, Goodman SB, Chien CR, Allen WA, van der Meulen MC: Stability of open-book pelvic fractures using a new biomechanical model of single-limb stance. J Orthop Trauma. 1997, 11: 590-593. 10.1097/00005131-199711000-00008.CrossRef MacAvoy MC, McClellan RT, Goodman SB, Chien CR, Allen WA, van der Meulen MC: Stability of open-book pelvic fractures using a new biomechanical model of single-limb stance. J Orthop Trauma. 1997, 11: 590-593. 10.1097/00005131-199711000-00008.CrossRef
22.
go back to reference Pohlemann T, Culemann U, Tscherne H: Comparative biomechanical studies of internal stabilization of trans-foraminal sacrum fractures. Orthopade. 1992, 21 (6): 413-421. Pohlemann T, Culemann U, Tscherne H: Comparative biomechanical studies of internal stabilization of trans-foraminal sacrum fractures. Orthopade. 1992, 21 (6): 413-421.
23.
go back to reference Rubash HE, Brown TD, Nelson DD, Mears DC: Comparative mechanical performances of some new devices for fixation of unstable pelvic ring fractures. Med Biol Eng Comput. 1983, 21: 657-663. 10.1007/BF02464027.CrossRef Rubash HE, Brown TD, Nelson DD, Mears DC: Comparative mechanical performances of some new devices for fixation of unstable pelvic ring fractures. Med Biol Eng Comput. 1983, 21: 657-663. 10.1007/BF02464027.CrossRef
Metadata
Title
Biomechanical stability of a supra-acetabular pedicle screw Internal Fixation device (INFIX) vs External Fixation and plates for vertically unstable pelvic fractures
Authors
Jonathan M Vigdorchik
Amanda O Esquivel
Xin Jin
King H Yang
Ndidi A Onwudiwe
Rahul Vaidya
Publication date
01-12-2012
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2012
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/1749-799X-7-31

Other articles of this Issue 1/2012

Journal of Orthopaedic Surgery and Research 1/2012 Go to the issue