Published in: Journal of Orthopaedic Surgery and Research 1/2009

Open Access 01-12-2009 | Research article

Comparison of migration behavior between single and dual lag screw implants for intertrochanteric fracture fixation

Authors: George K Kouvidis, Mark B Sommers, Peter V Giannoudis, Pavlos G Katonis, Michael Bottlang

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2009

Login to get access

Abstract

Background

Lag screw cut-out failure following fixation of unstable intertrochanteric fractures in osteoporotic bone remains an unsolved challenge. This study tested if resistance to cut-out failure can be improved by using a dual lag screw implant in place of a single lag screw implant. Migration behavior and cut-out resistance of a single and a dual lag screw implant were comparatively evaluated in surrogate specimens using an established laboratory model of hip screw cut-out failure.

Methods

Five dual lag screw implants (Endovis, Citieffe) and five single lag screw implants (DHS, Synthes) were tested in the Hip Implant Performance Simulator (HIPS) of the Legacy Biomechanics Laboratory. This model simulated osteoporotic bone, an unstable fracture, and biaxial rocking motion representative of hip loading during normal gait. All constructs were loaded up to 20,000 cycles of 1.45 kN peak magnitude under biaxial rocking motion. The migration kinematics was continuously monitored with 6-degrees of freedom motion tracking system and the number of cycles to implant cut-out was recorded.

Results

The dual lag screw implant exhibited significantly less migration and sustained more loading cycles in comparison to the DHS single lag screw. All DHS constructs failed before 20,000 cycles, on average at 6,638 ± 2,837 cycles either by cut-out or permanent screw bending. At failure, DHS constructs exhibited 10.8 ± 2.3° varus collapse and 15.5 ± 9.5° rotation around the lag screw axis. Four out of five dual screws constructs sustained 20,000 loading cycles. One dual screw specimens sustained cut-out by medial migration of the distal screw after 10,054 cycles. At test end, varus collapse and neck rotation in dual screws implants advanced to 3.7 ± 1.7° and 1.6 ± 1.0°, respectively.

Conclusion

The single and double lag screw implants demonstrated a significantly different migration resistance in surrogate specimens under gait loading simulation with the HIPS model. In this model, the double screw construct provided significantly greater resistance against varus collapse and neck rotation in comparison to a standard DHS lag screw implant.
Appendix
Available only for authorised users
Literature
1.
go back to reference Massie WK: Fracture of the Hip. J Bone Joint Surg. 1964, 46-A: 658-90. Massie WK: Fracture of the Hip. J Bone Joint Surg. 1964, 46-A: 658-90.
2.
go back to reference Schumpelick W, Jantzen PM: A new principle in the operative treatment of trochanteric fracture of the hip. J Bone Joint Surg. 1988, 70-A: 1297-303. Schumpelick W, Jantzen PM: A new principle in the operative treatment of trochanteric fracture of the hip. J Bone Joint Surg. 1988, 70-A: 1297-303.
3.
go back to reference Pugh WL: A self adjusting nail plate for fractures about the hip joint. J Bone Joint Surg. 1955, 37-A: 1085-93.PubMed Pugh WL: A self adjusting nail plate for fractures about the hip joint. J Bone Joint Surg. 1955, 37-A: 1085-93.PubMed
4.
go back to reference Parker MJ, Handoll HH: Intramedullary nails for extracapsular hip fractures in adults. Cochrane Database Syst Rev. 2006, 3: CD004961PubMed Parker MJ, Handoll HH: Intramedullary nails for extracapsular hip fractures in adults. Cochrane Database Syst Rev. 2006, 3: CD004961PubMed
5.
go back to reference Saudan M, Lubbeke A, Sadowski C, Riand N, Stem R, Hoffmeyer P: Pertrochanteric fractures: is there an advantage to an intramedullary nail? A randomized, prospective study of 206 patients comparing the dynamic hip screw and proximal femoral nail. J Orthop Trauma. 2002, 16: 386-393. 10.1097/00005131-200207000-00004.CrossRefPubMed Saudan M, Lubbeke A, Sadowski C, Riand N, Stem R, Hoffmeyer P: Pertrochanteric fractures: is there an advantage to an intramedullary nail? A randomized, prospective study of 206 patients comparing the dynamic hip screw and proximal femoral nail. J Orthop Trauma. 2002, 16: 386-393. 10.1097/00005131-200207000-00004.CrossRefPubMed
6.
go back to reference Radford PJ, Needoff M, Webb JK: A prospective randomised comparison of the dynamic hip screw and the Gamma locking nail. J Bone Joint Surg Br. 1993, 75 (5): 789-793.PubMed Radford PJ, Needoff M, Webb JK: A prospective randomised comparison of the dynamic hip screw and the Gamma locking nail. J Bone Joint Surg Br. 1993, 75 (5): 789-793.PubMed
7.
go back to reference Davis TR, Sher JL, Horsman A, Simpson M, Porter BB, Checketts RG: Intertrochanteric femoral fractures: mechanical failures after internal fixation. J Bone Joint Surg Br. 1990, 72B: 26-31. Davis TR, Sher JL, Horsman A, Simpson M, Porter BB, Checketts RG: Intertrochanteric femoral fractures: mechanical failures after internal fixation. J Bone Joint Surg Br. 1990, 72B: 26-31.
8.
go back to reference Haynes RC, Poll RG, Miles AW, Weston RB: Failure of femoral head fixation: a cadaveric analysis of lag screw cut-out with the Gamma locking nail and AO dynamic hip screw. Injury. 1997, 28: 337-341. 10.1016/S0020-1383(97)00035-1.CrossRefPubMed Haynes RC, Poll RG, Miles AW, Weston RB: Failure of femoral head fixation: a cadaveric analysis of lag screw cut-out with the Gamma locking nail and AO dynamic hip screw. Injury. 1997, 28: 337-341. 10.1016/S0020-1383(97)00035-1.CrossRefPubMed
9.
go back to reference Leung KS, So WS, Shen WY, Hui PW: Gamma nails and dynamic hip screws for pertrochanteric fractures: a randomised prospective study in elderly patients. J Bone Joint Surg Br. 1992, 74B: 345-351. Leung KS, So WS, Shen WY, Hui PW: Gamma nails and dynamic hip screws for pertrochanteric fractures: a randomised prospective study in elderly patients. J Bone Joint Surg Br. 1992, 74B: 345-351.
10.
go back to reference Audige L, Hanson B, Swiontkowski MF: Implant-related complications in the treatment of unstable intertrochanteric fractures: meta-analysis of dynamic screw-plate versus dynamic screw-intramedullary nail devices. Int Orthop. 2003, 27: 197-203. 10.1007/s00264-003-0457-6.PubMedCentralCrossRefPubMed Audige L, Hanson B, Swiontkowski MF: Implant-related complications in the treatment of unstable intertrochanteric fractures: meta-analysis of dynamic screw-plate versus dynamic screw-intramedullary nail devices. Int Orthop. 2003, 27: 197-203. 10.1007/s00264-003-0457-6.PubMedCentralCrossRefPubMed
11.
go back to reference Ahrengart L, Tornkvist H, Fornander P, Thorngren KG, Pasanen L, Wahlstrom P, Honkonen S, Lindgren U: A randomized study of the compression hip screw and Gamma nail in 426 fractures. Clin Orthop. 2002, 401: 209-222. 10.1097/00003086-200208000-00024.CrossRefPubMed Ahrengart L, Tornkvist H, Fornander P, Thorngren KG, Pasanen L, Wahlstrom P, Honkonen S, Lindgren U: A randomized study of the compression hip screw and Gamma nail in 426 fractures. Clin Orthop. 2002, 401: 209-222. 10.1097/00003086-200208000-00024.CrossRefPubMed
12.
go back to reference Jensen JS, Tondevold E, Mossing N: Unstable trochanteric fractures treated with the sliding screw plate system: A biomechanical study of unstable trochanteric fractures III. Acta Orthop Scand. 1978, 49 (4): 392-397.CrossRefPubMed Jensen JS, Tondevold E, Mossing N: Unstable trochanteric fractures treated with the sliding screw plate system: A biomechanical study of unstable trochanteric fractures III. Acta Orthop Scand. 1978, 49 (4): 392-397.CrossRefPubMed
13.
go back to reference Yoshimine F, Latta LL, Milne EL: Sliding characteristics of compression hip screws in the intertrochanteric fracture. J Orthop Trauma. 1993, 7: 348-353. 10.1097/00005131-199308000-00011.CrossRefPubMed Yoshimine F, Latta LL, Milne EL: Sliding characteristics of compression hip screws in the intertrochanteric fracture. J Orthop Trauma. 1993, 7: 348-353. 10.1097/00005131-199308000-00011.CrossRefPubMed
14.
go back to reference Adams CI, Robinson CM, Court-Brown CM, McQueen MM: Prospective randomized controlled trial of an intramedullary nail versus dynamic screw and plate for intertrochanteric fractures of the femur. J Orthop Trauma. 2001, 15: 394-400. 10.1097/00005131-200108000-00003.CrossRefPubMed Adams CI, Robinson CM, Court-Brown CM, McQueen MM: Prospective randomized controlled trial of an intramedullary nail versus dynamic screw and plate for intertrochanteric fractures of the femur. J Orthop Trauma. 2001, 15: 394-400. 10.1097/00005131-200108000-00003.CrossRefPubMed
15.
go back to reference Bess RJ, Jolly SA: Comparison of compression hip screw and gamma nail for treatment of peritrochanteric fractures. J South Orthop Assoc. 1997, 6: 173-179.PubMed Bess RJ, Jolly SA: Comparison of compression hip screw and gamma nail for treatment of peritrochanteric fractures. J South Orthop Assoc. 1997, 6: 173-179.PubMed
16.
go back to reference Parker MJ, Handoll HH: Gamma and other cephalocondylic intramedullary nails versus extramedullary implants for extracapsular hip fractures. Cochrane Database Syst Rev. 2008, 3: CD000093PubMed Parker MJ, Handoll HH: Gamma and other cephalocondylic intramedullary nails versus extramedullary implants for extracapsular hip fractures. Cochrane Database Syst Rev. 2008, 3: CD000093PubMed
17.
go back to reference Valverde JA, Alonso MG, Porro JG, Rueda D, Larrauri PM, Soler JJ: Use of the Gamma nail in the treatment of fractures of the proximal femur. Clin Orthop. 1998, 350: 56-61.CrossRefPubMed Valverde JA, Alonso MG, Porro JG, Rueda D, Larrauri PM, Soler JJ: Use of the Gamma nail in the treatment of fractures of the proximal femur. Clin Orthop. 1998, 350: 56-61.CrossRefPubMed
18.
go back to reference Rosenblum SF, Zuckerman JD, Kummer FJ, Tam BS: A biomechanical evaluation of the Gamma nail. J Bone Joint Surg Br. 1992, 74: 352-357.PubMed Rosenblum SF, Zuckerman JD, Kummer FJ, Tam BS: A biomechanical evaluation of the Gamma nail. J Bone Joint Surg Br. 1992, 74: 352-357.PubMed
19.
go back to reference Bridle SH, Patel AD, Bircher M, Calvert PT: Fixation of intertrochanteric fractures of the femur: a randomised prospective comparison of the gamma nail and the dynamic hip screw. J Bone Joint Surg Br. 1991, 73: 330-334.PubMed Bridle SH, Patel AD, Bircher M, Calvert PT: Fixation of intertrochanteric fractures of the femur: a randomised prospective comparison of the gamma nail and the dynamic hip screw. J Bone Joint Surg Br. 1991, 73: 330-334.PubMed
20.
go back to reference Wang CJ, Brown CJ, Yettram AL, Procter P: Intramedullary femoral nails: one or two lag screws? A preliminary study. Med Eng Phys. 2000, 22: 613-624. 10.1016/S1350-4533(00)00081-3.CrossRefPubMed Wang CJ, Brown CJ, Yettram AL, Procter P: Intramedullary femoral nails: one or two lag screws? A preliminary study. Med Eng Phys. 2000, 22: 613-624. 10.1016/S1350-4533(00)00081-3.CrossRefPubMed
21.
go back to reference Ingman AM: Percutaneous intramedullary fixation of trochanteric fractures of the femur: clinical trial of a new hip nail. Injury. 2000, 31: 483-487. 10.1016/S0020-1383(00)00027-9.CrossRefPubMed Ingman AM: Percutaneous intramedullary fixation of trochanteric fractures of the femur: clinical trial of a new hip nail. Injury. 2000, 31: 483-487. 10.1016/S0020-1383(00)00027-9.CrossRefPubMed
22.
go back to reference Simmermacher RK, Bosch AM, Werken Van der C: The AO/ASIF proximal femoral nail (PFN): a new device for the treatment of unstable proximal femoral fractures. Injury. 1999, 30: 327-332. 10.1016/S0020-1383(99)00091-1.CrossRefPubMed Simmermacher RK, Bosch AM, Werken Van der C: The AO/ASIF proximal femoral nail (PFN): a new device for the treatment of unstable proximal femoral fractures. Injury. 1999, 30: 327-332. 10.1016/S0020-1383(99)00091-1.CrossRefPubMed
23.
go back to reference Kubiak EN, Bong M, Park S, Kummer F, Egol K, Koval KJ: Intramedullary fixation of unstable intertrochanteric hip fractures. One or two lag screws. J Orthop Trauma. 2004, 18: 12-17. 10.1097/00005131-200401000-00003.CrossRefPubMed Kubiak EN, Bong M, Park S, Kummer F, Egol K, Koval KJ: Intramedullary fixation of unstable intertrochanteric hip fractures. One or two lag screws. J Orthop Trauma. 2004, 18: 12-17. 10.1097/00005131-200401000-00003.CrossRefPubMed
24.
go back to reference Ropars M, Mitton D, Skalli W: Minimally invasive screw plates for surgery of unstable intertrochanteric femoral fractures: A biomechanical study. Clin Biomech (Bristol, Avon). 2008, 23 (8): 1012-7. 10.1016/j.clinbiomech.2008.04.018.CrossRef Ropars M, Mitton D, Skalli W: Minimally invasive screw plates for surgery of unstable intertrochanteric femoral fractures: A biomechanical study. Clin Biomech (Bristol, Avon). 2008, 23 (8): 1012-7. 10.1016/j.clinbiomech.2008.04.018.CrossRef
25.
go back to reference Saikko V, Calonius O, Keränen J: Effect of extent of motion and type of load on the wear of polyethylene in a biaxial hip simulator. J Biomed Mater Res B Appl Biomater. 2003, 65 (1): 186-192. 10.1002/jbm.b.10556.CrossRefPubMed Saikko V, Calonius O, Keränen J: Effect of extent of motion and type of load on the wear of polyethylene in a biaxial hip simulator. J Biomed Mater Res B Appl Biomater. 2003, 65 (1): 186-192. 10.1002/jbm.b.10556.CrossRefPubMed
26.
go back to reference Pedersen DR, Brown TD, Maxian TA, Callaghan JJ: Temporal and spatial distributions of directional counterface motion at the acetabular bearing surface in total hip arthroplasty. Iowa Orthop J. 1998, 18: 43-53.PubMedCentralPubMed Pedersen DR, Brown TD, Maxian TA, Callaghan JJ: Temporal and spatial distributions of directional counterface motion at the acetabular bearing surface in total hip arthroplasty. Iowa Orthop J. 1998, 18: 43-53.PubMedCentralPubMed
27.
go back to reference Mejia LC, Brierley TJ: A hip wear simulator for the evaluation of biomaterials in hip arthroplasty components. Biomed Mater Eng. 1994, 4 (4): 259-71.PubMed Mejia LC, Brierley TJ: A hip wear simulator for the evaluation of biomaterials in hip arthroplasty components. Biomed Mater Eng. 1994, 4 (4): 259-71.PubMed
29.
go back to reference Ehmke LW, Fitzpatrick DC, Krieg JC, Madey SM, Bottlang M: Lag screws for hip fracture fixation: Evaluation of migration resistance under simulated walking. J Orthop Res. 2005, 23: 1329-35.CrossRefPubMed Ehmke LW, Fitzpatrick DC, Krieg JC, Madey SM, Bottlang M: Lag screws for hip fracture fixation: Evaluation of migration resistance under simulated walking. J Orthop Res. 2005, 23: 1329-35.CrossRefPubMed
30.
go back to reference Sommers MB, Roth C, Hall H, Kam BCC, Ehmke LW, Krieg JC, Madey SM, Bottlang M: A laboratory model to evaluate cutout resistance of implants for pertrochanteric fracture fixation. J Orthop Trauma. 2004, 18: 361-8. 10.1097/00005131-200407000-00006.CrossRefPubMed Sommers MB, Roth C, Hall H, Kam BCC, Ehmke LW, Krieg JC, Madey SM, Bottlang M: A laboratory model to evaluate cutout resistance of implants for pertrochanteric fracture fixation. J Orthop Trauma. 2004, 18: 361-8. 10.1097/00005131-200407000-00006.CrossRefPubMed
31.
32.
go back to reference Linde F, Gothgen CB, Hvid I, Pongsoipetch B: Mechanical properties of trabecular bone by a non-destructive compression testing approach. Eng Med. 1988, 17: 23-9. 10.1243/EMED_JOUR_1988_017_008_02.CrossRefPubMed Linde F, Gothgen CB, Hvid I, Pongsoipetch B: Mechanical properties of trabecular bone by a non-destructive compression testing approach. Eng Med. 1988, 17: 23-9. 10.1243/EMED_JOUR_1988_017_008_02.CrossRefPubMed
33.
go back to reference Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77: 1058-1064.PubMed Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77: 1058-1064.PubMed
34.
go back to reference Chang WS, Zuckerman JD, Kummer FJ, Frankel VH: Biomechanical evaluation of anatomic reduction versus medial displacement osteotomy in unstable intertrochanteric fractures. Clin Orthop. 1987, 225: 141-146.PubMed Chang WS, Zuckerman JD, Kummer FJ, Frankel VH: Biomechanical evaluation of anatomic reduction versus medial displacement osteotomy in unstable intertrochanteric fractures. Clin Orthop. 1987, 225: 141-146.PubMed
35.
go back to reference Baumgart F, Cordey J, Morikawa K, Perren SM, Rahn BA, Schavan R, Snyder S: AO/ASIF Self-tapping screws (STS). Injury. 1993, 24 (Suppl 1): S1-17. 10.1016/0020-1383(93)90312-T.CrossRefPubMed Baumgart F, Cordey J, Morikawa K, Perren SM, Rahn BA, Schavan R, Snyder S: AO/ASIF Self-tapping screws (STS). Injury. 1993, 24 (Suppl 1): S1-17. 10.1016/0020-1383(93)90312-T.CrossRefPubMed
36.
go back to reference Windolf Markus, Muths Raphael, Braunstein Volker, Gueorguiev Boyko, Hänni Markus, Schwieger Karsten: Quantification of cancellous bone-compaction due to DHS® Blade insertion and influence upon cut-out resistance. J Clin Biomech. 2009, 24: 53-58. 10.1016/j.clinbiomech.2008.09.005.CrossRef Windolf Markus, Muths Raphael, Braunstein Volker, Gueorguiev Boyko, Hänni Markus, Schwieger Karsten: Quantification of cancellous bone-compaction due to DHS® Blade insertion and influence upon cut-out resistance. J Clin Biomech. 2009, 24: 53-58. 10.1016/j.clinbiomech.2008.09.005.CrossRef
37.
go back to reference Giannoudis PV, Schneider E: Principles of fixation of osteoporotic fractures. J Bone Joint Surg [Br]. 2006, 88: 1272-78. 10.2106/JBJS.E.00414.CrossRef Giannoudis PV, Schneider E: Principles of fixation of osteoporotic fractures. J Bone Joint Surg [Br]. 2006, 88: 1272-78. 10.2106/JBJS.E.00414.CrossRef
38.
go back to reference Windolf Markus, Braunstein Volker, Dutoit Christof, Schwieger Karsten: Is a helical shaped implant a superior alternative to the Dynamic Hip Screw for unstable femoral neck fractures? A biomechanical investigation. Clin Biomech (Bristol, Avon). 2009, 24 (1): 59-64. 10.1016/j.clinbiomech.2008.07.004.CrossRef Windolf Markus, Braunstein Volker, Dutoit Christof, Schwieger Karsten: Is a helical shaped implant a superior alternative to the Dynamic Hip Screw for unstable femoral neck fractures? A biomechanical investigation. Clin Biomech (Bristol, Avon). 2009, 24 (1): 59-64. 10.1016/j.clinbiomech.2008.07.004.CrossRef
39.
go back to reference Parker MJ, Blundell C: Choice of implant for internal fixation of femoral neck fractures. Meta-analysis of 25 randomized trials including 4925 patients. Acta Orthop Scand. 1998, 69: 138-43.CrossRefPubMed Parker MJ, Blundell C: Choice of implant for internal fixation of femoral neck fractures. Meta-analysis of 25 randomized trials including 4925 patients. Acta Orthop Scand. 1998, 69: 138-43.CrossRefPubMed
40.
go back to reference Goldhagen PR, O_Connor DR, Schwarze D, Schwartz E: A prospective comparative study of the compression hip screw and the gamma nail. J Orthop Trauma. 1994, 8: 367-72. 10.1097/00005131-199410000-00001.CrossRefPubMed Goldhagen PR, O_Connor DR, Schwarze D, Schwartz E: A prospective comparative study of the compression hip screw and the gamma nail. J Orthop Trauma. 1994, 8: 367-72. 10.1097/00005131-199410000-00001.CrossRefPubMed
41.
go back to reference Werner-Tutschku W, Lajtai G, Schmiedhuber G, Lang T, Pirkl C, Orthner E: Intra – and perioperative complications in the stabilization of per – and subtrochanteric femoral fractures by means of PFN. Unfallchirurg. 2002, 105 (10): 881-885. 10.1007/s00113-002-0416-5.CrossRefPubMed Werner-Tutschku W, Lajtai G, Schmiedhuber G, Lang T, Pirkl C, Orthner E: Intra – and perioperative complications in the stabilization of per – and subtrochanteric femoral fractures by means of PFN. Unfallchirurg. 2002, 105 (10): 881-885. 10.1007/s00113-002-0416-5.CrossRefPubMed
42.
go back to reference Weil YA, Gardner MJ, Mikhail G, Pierson G, Helfet DL, Lorich DG: Medial migration of intramedullary hip fixation devices. Arch Orthop Trauma Surg. 2008, 128: 227-234. 10.1007/s00402-007-0497-2.CrossRefPubMed Weil YA, Gardner MJ, Mikhail G, Pierson G, Helfet DL, Lorich DG: Medial migration of intramedullary hip fixation devices. Arch Orthop Trauma Surg. 2008, 128: 227-234. 10.1007/s00402-007-0497-2.CrossRefPubMed
43.
go back to reference Jinn Lin MD: Encouraging Results of Treating Femoral Trochanteric Fractures With Specially Designed Double-Screw Nails. J Trauma. 2007, 63: 866-874. 10.1097/TA.0b013e3180342087.CrossRefPubMed Jinn Lin MD: Encouraging Results of Treating Femoral Trochanteric Fractures With Specially Designed Double-Screw Nails. J Trauma. 2007, 63: 866-874. 10.1097/TA.0b013e3180342087.CrossRefPubMed
44.
go back to reference Friedl W, Clausen J: Experimental examination for optimized stabilisation of trochanteric femur fractures, intra- or extramedullary implant localisation and influence of femur neck component profile on cut-out risk. Chirurg. 2001, 72: 1344-52. 10.1007/s001040170041.CrossRefPubMed Friedl W, Clausen J: Experimental examination for optimized stabilisation of trochanteric femur fractures, intra- or extramedullary implant localisation and influence of femur neck component profile on cut-out risk. Chirurg. 2001, 72: 1344-52. 10.1007/s001040170041.CrossRefPubMed
Metadata
Title
Comparison of migration behavior between single and dual lag screw implants for intertrochanteric fracture fixation
Authors
George K Kouvidis
Mark B Sommers
Peter V Giannoudis
Pavlos G Katonis
Michael Bottlang
Publication date
01-12-2009
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2009
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/1749-799X-4-16

Other articles of this Issue 1/2009

Journal of Orthopaedic Surgery and Research 1/2009 Go to the issue