Published in: Journal of Orthopaedic Surgery and Research 1/2008

Open Access 01-12-2008 | Research article

Failure of dual radius hydroxyapatite-coated acetabular cups

Authors: Fabio D'Angelo, Mauro Molina, Giacomo Riva, Giovanni Zatti, Paolo Cherubino

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2008

Login to get access

Abstract

Introduction

Many kind of hydroxyapatite-coated cups were used, with favorable results in short term studies; it was supposed that its use could improve osteointegration of the cup, enhancing thus stability and survivorship. The purpose of this study is to analyze the long term behavior of the hemispheric HA coated, Dual Radius Osteonics cup and to discuss the way of failure through the exam of the revised components and of both periacetabular and osteolysis tissue.

Materials and Methods

Between 1994 and 1997, at the Department of Orthopedic Sciences of the Insubria University, using the posterolateral approach, were implanted 276 Dual Radius Osteonics® in 256 patients, with mean age of 63 years.

Results

At a mean follow-up of 10 years (range 8–12 years), 183 cups in 165 patients, were available for clinical and radiographical evaluation. 22 Cups among the 183 were revised (11%). The cause of revision was aseptic loosening in 17 cases, septic loosening in one case, periprosthetic fracture in another case, osteolysis and polyethylene wear in two cases and, finally, recurrent dislocations in the last one. In the remaining patients, mean HHS increased from a preoperative value of 50,15 to a postoperative value of 92,69. The mean polyethylene wear was 1,25 mm (min. 0,08, max. 3,9 mm), with a mean annual wear of 0,17 mm. The mean acetabular migration on the two axis was 1,6 mm and 1,8 mm. Peri-acetabular osteolysis were recorded in 89% of the implants (163 cases). The cumulative survivorship (revision as endpoint) at the time was 88,9%.

Conclusion

Our study confirms the bad behavior of this type of cup probably related to the design, to the method of HA fixation. The observations carried out on the revised cup confirm these hypotheses but did not clarify if the third body wear could be a further problem. Another interesting aspect is the high incidence of osteolysis, which are often asymptomatic becoming a problem for the surgeon as the patient refuses the possibility of a revision.
Appendix
Available only for authorised users
Literature
1.
go back to reference Geesink RG, de Groot K, Klein CP: Chemical implant fixation using hydroxyl-apatite coatings. The development of a human total hip prosthesis for chemical fixation to bone using hydroxyl-apatite coatings on titanium substrates. Clin Orthop Relat Res. 1987, 225: 147-170.PubMed Geesink RG, de Groot K, Klein CP: Chemical implant fixation using hydroxyl-apatite coatings. The development of a human total hip prosthesis for chemical fixation to bone using hydroxyl-apatite coatings on titanium substrates. Clin Orthop Relat Res. 1987, 225: 147-170.PubMed
2.
go back to reference Xenos JS, Callaghan JJ, Heekin RD, Hopkinson WJ, Savory CG, Moore MS: The porous-coated anatomic total hip prosthesis, inserted without cement. A prospective study with a minimum of ten years of follow-up. J Bone Joint Surg Am. 1999, 81: 74-82.PubMed Xenos JS, Callaghan JJ, Heekin RD, Hopkinson WJ, Savory CG, Moore MS: The porous-coated anatomic total hip prosthesis, inserted without cement. A prospective study with a minimum of ten years of follow-up. J Bone Joint Surg Am. 1999, 81: 74-82.PubMed
3.
go back to reference Harris WH, Maloney WJ: Hybrid total hip arthroplasty. Clin Orthop Relat Res. 1989, 249: 21-29.PubMed Harris WH, Maloney WJ: Hybrid total hip arthroplasty. Clin Orthop Relat Res. 1989, 249: 21-29.PubMed
4.
go back to reference Harris WH: Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969, 51 (4): 737-755.PubMed Harris WH: Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969, 51 (4): 737-755.PubMed
5.
go back to reference Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS, Chiesa R, McKellop HA, Dorr LD: Accuracy of measurament of polyethylene wear with use of radiographs of total hip replacements. J Bone Joint Surg Am. 2003, 85: 2378-84.PubMed Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS, Chiesa R, McKellop HA, Dorr LD: Accuracy of measurament of polyethylene wear with use of radiographs of total hip replacements. J Bone Joint Surg Am. 2003, 85: 2378-84.PubMed
6.
go back to reference Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CK: Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003, 85: 505-511.PubMed Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CK: Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003, 85: 505-511.PubMed
7.
go back to reference Lawrence JM, Engh CA, Macalino GE, Lauro GR: Outcome of revision hip arthroplasty done without cement. J Bone Joint Surg Am. 1994, 76: 965-973.PubMed Lawrence JM, Engh CA, Macalino GE, Lauro GR: Outcome of revision hip arthroplasty done without cement. J Bone Joint Surg Am. 1994, 76: 965-973.PubMed
8.
go back to reference DeLee JG, Charnley J: Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976, 121: 20-32.PubMed DeLee JG, Charnley J: Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976, 121: 20-32.PubMed
9.
go back to reference Shin-Yoon Kim, Do-Heon Kim, Yong-Goo Kim, Chang-Wug Oh, Joo-Chul Ihn: Early Failure of Hemispheric Hydroxyapatite-coated Acetabular Cups. Clin Orthop Relat Res. 2006, 446: 233-238. 10.1097/01.blo.0000195923.28480.d4.CrossRef Shin-Yoon Kim, Do-Heon Kim, Yong-Goo Kim, Chang-Wug Oh, Joo-Chul Ihn: Early Failure of Hemispheric Hydroxyapatite-coated Acetabular Cups. Clin Orthop Relat Res. 2006, 446: 233-238. 10.1097/01.blo.0000195923.28480.d4.CrossRef
10.
go back to reference Capello WN, D'Antonio JA, Manley MT, Feinberg JR: Hydroxyapatite in total hip arthoplasty: clinical results and critical issues. Clin Orthop Relat Res. 1998, 355: 200-211. 10.1097/00003086-199810000-00021.CrossRefPubMed Capello WN, D'Antonio JA, Manley MT, Feinberg JR: Hydroxyapatite in total hip arthoplasty: clinical results and critical issues. Clin Orthop Relat Res. 1998, 355: 200-211. 10.1097/00003086-199810000-00021.CrossRefPubMed
11.
go back to reference Kawamura H, Dunbar MJ, Murray P, Bourne RB, Rorabeck CH: The porous coated anatomic total hip replacement: a ten to fourteen year follow-up study of cementless total hip arthroplasty. J Bone Joint Surg Am. 2001, 83: 1333-1338.PubMed Kawamura H, Dunbar MJ, Murray P, Bourne RB, Rorabeck CH: The porous coated anatomic total hip replacement: a ten to fourteen year follow-up study of cementless total hip arthroplasty. J Bone Joint Surg Am. 2001, 83: 1333-1338.PubMed
12.
go back to reference Engh CA, Bobyn JD, Glassman AH: Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br. 1987, 69: 45-49.PubMed Engh CA, Bobyn JD, Glassman AH: Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br. 1987, 69: 45-49.PubMed
13.
go back to reference Overgaard S, Knudsen HM, Hansen LN, Mossing N: Hip arthroplasty in Jutland, Denmark: age and sex-specific incidences of primary operation. Acta Orthop Scand. 1992, 63: 536-542.CrossRefPubMed Overgaard S, Knudsen HM, Hansen LN, Mossing N: Hip arthroplasty in Jutland, Denmark: age and sex-specific incidences of primary operation. Acta Orthop Scand. 1992, 63: 536-542.CrossRefPubMed
14.
go back to reference deGroot K, Geesink R, Klein CP, Serekian P: Plasma spray coatings of hydroxyapatite. J Biomed Mater Res. 1987, 21: 1375-1381. 10.1002/jbm.820211203.CrossRef deGroot K, Geesink R, Klein CP, Serekian P: Plasma spray coatings of hydroxyapatite. J Biomed Mater Res. 1987, 21: 1375-1381. 10.1002/jbm.820211203.CrossRef
15.
go back to reference Soballe K, Hansen ES, BRockstedt-Rasmussen H, Hjortdal VE, Juhl GI, Pedersen CM, Hvid I, Bunger C: Gap healing enhanced by hydroxyapatite coating in dogs. Clin Orthop Relat Res. 1991, 272: 300-307.PubMed Soballe K, Hansen ES, BRockstedt-Rasmussen H, Hjortdal VE, Juhl GI, Pedersen CM, Hvid I, Bunger C: Gap healing enhanced by hydroxyapatite coating in dogs. Clin Orthop Relat Res. 1991, 272: 300-307.PubMed
16.
go back to reference Lintner F, Bohm G, Huber M, Scholz R: Histology of tissue adjacent to an HAC-coated femoral prosthesis: a case report. J Bone Joint Surg Br. 1994, 76: 824-830.PubMed Lintner F, Bohm G, Huber M, Scholz R: Histology of tissue adjacent to an HAC-coated femoral prosthesis: a case report. J Bone Joint Surg Br. 1994, 76: 824-830.PubMed
17.
go back to reference D'Lima DD, Walker RH, Colwell CW: Omnifit-HA stem in total hip arthroplasty: a 2- to 5-year followup. Clin Orthop Relat Res. 1999, 363: 163-169. 10.1097/00003086-199906000-00021.PubMed D'Lima DD, Walker RH, Colwell CW: Omnifit-HA stem in total hip arthroplasty: a 2- to 5-year followup. Clin Orthop Relat Res. 1999, 363: 163-169. 10.1097/00003086-199906000-00021.PubMed
18.
go back to reference Furlong RJ, Osborn JF: Fixation of hip prostheses by hydroxyapatite ceramic coatings. J Bone Joint Surg Br. 1991, 73: 741-74.PubMed Furlong RJ, Osborn JF: Fixation of hip prostheses by hydroxyapatite ceramic coatings. J Bone Joint Surg Br. 1991, 73: 741-74.PubMed
19.
go back to reference Geesink RG, de Groot K, Klein CP: Chemical implant fixation using hydroxyl-apatite coatings: the development of a human total hip prosthesis for chemical fixation to bone using hydroxyl-apatite coatings on titanium substrates. Clin Orthop Relat Res. 1987, 225: 147-170.PubMed Geesink RG, de Groot K, Klein CP: Chemical implant fixation using hydroxyl-apatite coatings: the development of a human total hip prosthesis for chemical fixation to bone using hydroxyl-apatite coatings on titanium substrates. Clin Orthop Relat Res. 1987, 225: 147-170.PubMed
20.
go back to reference Capello WN, D'Antonio JA, Feinberg JR, Manley MT: Hydroxyapatite-coated total hip femoral components in patients less than fifty years old: clinical and radiographic results after five to eight years of follow-up. J Bone Joint Surg Am. 1997, 79: 1023-1029.PubMed Capello WN, D'Antonio JA, Feinberg JR, Manley MT: Hydroxyapatite-coated total hip femoral components in patients less than fifty years old: clinical and radiographic results after five to eight years of follow-up. J Bone Joint Surg Am. 1997, 79: 1023-1029.PubMed
21.
go back to reference Jaffe WL, Scott DF: Total hip arthroplasty with hydroxyapatitecoated prostheses. J Bone Joint Surg M. 1996, 78: 1918-1934. Jaffe WL, Scott DF: Total hip arthroplasty with hydroxyapatitecoated prostheses. J Bone Joint Surg M. 1996, 78: 1918-1934.
22.
go back to reference Kavanagh BF, Dewitz MA, Ilstrup DM, Stauffer RN, Coventry MB: Charnley total hip arthroplasty with cement: fifteen-year results. J Bone Joint Surg Am. 1989, 71: 1496-1503.PubMed Kavanagh BF, Dewitz MA, Ilstrup DM, Stauffer RN, Coventry MB: Charnley total hip arthroplasty with cement: fifteen-year results. J Bone Joint Surg Am. 1989, 71: 1496-1503.PubMed
23.
go back to reference Manley MT, Capello WN, D'Antonio JA, Geesink RGT: Fixation of acetabular cups without cement in total hip arthroplasty. A comparison of three different implant surface at minimum duration of follow-up of five years. J Bone Joint Surg Am. 1998, 80: 1175-1185.PubMed Manley MT, Capello WN, D'Antonio JA, Geesink RGT: Fixation of acetabular cups without cement in total hip arthroplasty. A comparison of three different implant surface at minimum duration of follow-up of five years. J Bone Joint Surg Am. 1998, 80: 1175-1185.PubMed
24.
go back to reference Røkkum M, Brandt M, Bye K, Hetland KR, Waage S, Reigstad A: Polythylene wear, osteolysis and adetabular loosening with an HA-coated hip prosthesis. A follow-up of 94 consecutive arthroplasties. J Bone Joint Surg Br. 1999, 81: 582-589. 10.1302/0301-620X.81B4.8715.CrossRefPubMed Røkkum M, Brandt M, Bye K, Hetland KR, Waage S, Reigstad A: Polythylene wear, osteolysis and adetabular loosening with an HA-coated hip prosthesis. A follow-up of 94 consecutive arthroplasties. J Bone Joint Surg Br. 1999, 81: 582-589. 10.1302/0301-620X.81B4.8715.CrossRefPubMed
25.
go back to reference Geesink RGT: Hydroxyapatite-coated total hip prostesis: two year clinical and roentgenographic results of 100 cases. Clin Orthop Relat Res. 1990, 261: 39-58.PubMed Geesink RGT: Hydroxyapatite-coated total hip prostesis: two year clinical and roentgenographic results of 100 cases. Clin Orthop Relat Res. 1990, 261: 39-58.PubMed
26.
go back to reference D'Antonio JA, Capello WN, Jaffe WL: Hydroxyapatite-coated hip implants: multicenter three-year clinical and roentgenographic results. Clin Orthop Relat Res. 1992, 285: 102-115.PubMed D'Antonio JA, Capello WN, Jaffe WL: Hydroxyapatite-coated hip implants: multicenter three-year clinical and roentgenographic results. Clin Orthop Relat Res. 1992, 285: 102-115.PubMed
27.
go back to reference Önsten I, Carlsson AS, Sanzén L, Besjakov J: Migration and wear of a hydroxyapatite-coated hip prosthesis a controlled roentgen stereophotogrammetric study. J Bone Joint Surg Br. 1996, 78: 85-91.PubMed Önsten I, Carlsson AS, Sanzén L, Besjakov J: Migration and wear of a hydroxyapatite-coated hip prosthesis a controlled roentgen stereophotogrammetric study. J Bone Joint Surg Br. 1996, 78: 85-91.PubMed
28.
go back to reference Lai KA, Shen WJ, Chen CH, Yang CY, Hu WP, Chang GL: Failure of hydroxyapatite-coated acetabular cups: ten-year follow-up of 85 Landos Atoll arthroplasties. J Bone Joint Surg Br. 2002, 84: 641-646. 10.1302/0301-620X.84B5.12384.CrossRefPubMed Lai KA, Shen WJ, Chen CH, Yang CY, Hu WP, Chang GL: Failure of hydroxyapatite-coated acetabular cups: ten-year follow-up of 85 Landos Atoll arthroplasties. J Bone Joint Surg Br. 2002, 84: 641-646. 10.1302/0301-620X.84B5.12384.CrossRefPubMed
29.
go back to reference Capello WN, D'Antonio JA, Feinberg JR: Hydroxyapatite-coated stems in patients under 50 years old: clinical and radiographic results at five-year minimum follow-up. Orthop Trans. 1995, 19: 399-404. Capello WN, D'Antonio JA, Feinberg JR: Hydroxyapatite-coated stems in patients under 50 years old: clinical and radiographic results at five-year minimum follow-up. Orthop Trans. 1995, 19: 399-404.
30.
go back to reference Giannikas KA, Din R, Sadiq S, Dunningham TH: Medium-term results of the ABG total hip arthroplasty in young patients. J Arthroplasty. 2002, 17: 184-188. 10.1054/arth.2002.29394.CrossRefPubMed Giannikas KA, Din R, Sadiq S, Dunningham TH: Medium-term results of the ABG total hip arthroplasty in young patients. J Arthroplasty. 2002, 17: 184-188. 10.1054/arth.2002.29394.CrossRefPubMed
31.
go back to reference Tompkins GS, Jacobs JJ, Kull LR, Rosenberg AG, Galante JO: Primary total hip arthroplasty with a porous-coated acetabular component. Seven-to-ten year results. J Bone Joint Surg Am. 1997, 79: 169-176.PubMed Tompkins GS, Jacobs JJ, Kull LR, Rosenberg AG, Galante JO: Primary total hip arthroplasty with a porous-coated acetabular component. Seven-to-ten year results. J Bone Joint Surg Am. 1997, 79: 169-176.PubMed
32.
go back to reference Latimer H, Lachiewicz PF, Chapel H: Porous-coated acetabular components with screw fixartion. Five to ten-year results. J Bone Joint Surg Am. 1996, 78: 975-981.PubMed Latimer H, Lachiewicz PF, Chapel H: Porous-coated acetabular components with screw fixartion. Five to ten-year results. J Bone Joint Surg Am. 1996, 78: 975-981.PubMed
33.
go back to reference Bloebaum RD, Beeks D, orr LD, Savory CG, DuPont J, Hofmann AA: Complication with hydroxyapatite particulate separation in total hip arthroplasty. Clin Orthop Relat Res. 1994, 298: 19-26.PubMed Bloebaum RD, Beeks D, orr LD, Savory CG, DuPont J, Hofmann AA: Complication with hydroxyapatite particulate separation in total hip arthroplasty. Clin Orthop Relat Res. 1994, 298: 19-26.PubMed
34.
go back to reference Morscher EW, Hefti A, Aebi U: Severe osteolysis after third-body wear due to hydroxyapatite particles from acetabular cup coating. J Bone Joint Surg Br. 1998, 80: 267-272. 10.1302/0301-620X.80B2.8316.CrossRefPubMed Morscher EW, Hefti A, Aebi U: Severe osteolysis after third-body wear due to hydroxyapatite particles from acetabular cup coating. J Bone Joint Surg Br. 1998, 80: 267-272. 10.1302/0301-620X.80B2.8316.CrossRefPubMed
35.
go back to reference Clohisy JC, Harris WH: The Harris-Galante porous coated acetabular component with screw fixation: an average ten year follow-up study. J Bone Joint Surg Am. 1999, 81: 66-73.PubMed Clohisy JC, Harris WH: The Harris-Galante porous coated acetabular component with screw fixation: an average ten year follow-up study. J Bone Joint Surg Am. 1999, 81: 66-73.PubMed
36.
go back to reference Pilliar RM, Lee JM, Maniatopolous C: Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res. 1986, 108-113. 208PubMed Pilliar RM, Lee JM, Maniatopolous C: Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res. 1986, 108-113. 208PubMed
37.
go back to reference Sun Limin, Berndt Christopher, Gross Karlis, Kucuk Ahmet: Material Fundamentals and Clinical Performance of Plasma-Sprayed Hydroxyapatite Coatings: A Review. J Biomed Mater Res (Appl Biomater). 2001, 58: 570-592. 10.1002/jbm.1056.CrossRef Sun Limin, Berndt Christopher, Gross Karlis, Kucuk Ahmet: Material Fundamentals and Clinical Performance of Plasma-Sprayed Hydroxyapatite Coatings: A Review. J Biomed Mater Res (Appl Biomater). 2001, 58: 570-592. 10.1002/jbm.1056.CrossRef
38.
go back to reference Rapperport DJ, Carter DR, Schurman DJ: Contact finite element stress analysis of porous ingrowth acetabular cup implantation, ingrowth, and loosening. J Orthop Res. 1987, 5: 548-561. 10.1002/jor.1100050410.CrossRefPubMed Rapperport DJ, Carter DR, Schurman DJ: Contact finite element stress analysis of porous ingrowth acetabular cup implantation, ingrowth, and loosening. J Orthop Res. 1987, 5: 548-561. 10.1002/jor.1100050410.CrossRefPubMed
39.
go back to reference Geesink RG, Hoefnagels NH: Six-year results of hydroxyapatite-coated total hip replacement. J Bone Joint Surg Br. 1995, 77: 534-547.PubMed Geesink RG, Hoefnagels NH: Six-year results of hydroxyapatite-coated total hip replacement. J Bone Joint Surg Br. 1995, 77: 534-547.PubMed
40.
go back to reference Rothman RH, Hozack WJ, Ranawat A, Moriarty L: Hydroxyapatite-coated femoral stems. A matched-pair analysis of coated and uncoated implants. J Bone Joint Surg Am. 1996, 78: 319-324.PubMed Rothman RH, Hozack WJ, Ranawat A, Moriarty L: Hydroxyapatite-coated femoral stems. A matched-pair analysis of coated and uncoated implants. J Bone Joint Surg Am. 1996, 78: 319-324.PubMed
41.
go back to reference Bauer TW, Stulberg BN, Jiang M, Geesink RGT: Uncemented acetabular components, histologic analysis of retrieved hydroxyapatite coated and porous implants. J Arthroplasty. 1993, 8: 167-177.CrossRefPubMed Bauer TW, Stulberg BN, Jiang M, Geesink RGT: Uncemented acetabular components, histologic analysis of retrieved hydroxyapatite coated and porous implants. J Arthroplasty. 1993, 8: 167-177.CrossRefPubMed
42.
go back to reference Collier JP, Surprenant VA, Mayor MB, Wrona M, Jensen RE, Surprenant HP: Loss of hydroxyapatite coating on retrieved, total hip components. J Arthroplasty. 1993, 8: 389-393. 10.1016/S0883-5403(06)80037-9.CrossRefPubMed Collier JP, Surprenant VA, Mayor MB, Wrona M, Jensen RE, Surprenant HP: Loss of hydroxyapatite coating on retrieved, total hip components. J Arthroplasty. 1993, 8: 389-393. 10.1016/S0883-5403(06)80037-9.CrossRefPubMed
43.
go back to reference Hardy DC, Frayssinet P, Bonel G, Authom T, Le Naelou SA, Delincé PE: Two-year outcome of hydroxyapatitecoated prostheses: two femoral prostheses retrieved at autopsy. Acta Orthop Scand. 1994, 65: 253-257.CrossRefPubMed Hardy DC, Frayssinet P, Bonel G, Authom T, Le Naelou SA, Delincé PE: Two-year outcome of hydroxyapatitecoated prostheses: two femoral prostheses retrieved at autopsy. Acta Orthop Scand. 1994, 65: 253-257.CrossRefPubMed
44.
go back to reference Buma P, Gardeniers JW: Tissue reactions around a hydroxyapatite-coated hip prostheses: case report of a retrieved specimen. J Arthroplasty. 1995, 10: 389-395. 10.1016/S0883-5403(05)80190-1.CrossRefPubMed Buma P, Gardeniers JW: Tissue reactions around a hydroxyapatite-coated hip prostheses: case report of a retrieved specimen. J Arthroplasty. 1995, 10: 389-395. 10.1016/S0883-5403(05)80190-1.CrossRefPubMed
45.
go back to reference Friedman RJ: Advances in biomaterials and factors affecting implant fixation. Instr Course Lect. 1992, 41: 127-136.PubMed Friedman RJ: Advances in biomaterials and factors affecting implant fixation. Instr Course Lect. 1992, 41: 127-136.PubMed
46.
go back to reference Mallory TH, Head WC, Lombardi AV: Tapered design for the cementless total hip arthroplasty femoral component. Clin Orthop Relat Res. 1997, 344: 172-178. 10.1097/00003086-199711000-00018.CrossRefPubMed Mallory TH, Head WC, Lombardi AV: Tapered design for the cementless total hip arthroplasty femoral component. Clin Orthop Relat Res. 1997, 344: 172-178. 10.1097/00003086-199711000-00018.CrossRefPubMed
47.
go back to reference Kim YH, Kim VE: Results of the Harris-Galante cementless hip prosthesis. J Bone Joint Surg Br. 1992, 74: 83-87.PubMed Kim YH, Kim VE: Results of the Harris-Galante cementless hip prosthesis. J Bone Joint Surg Br. 1992, 74: 83-87.PubMed
48.
go back to reference Bauer TW, Geesink RC, Zimmerman R, McMahon JT: Hydroxyapatite-coated femoral stems. Histological analysis of components retrieved at autopsy. J Bone Joint Surg Am. 1991, 73: 1439-1452.PubMed Bauer TW, Geesink RC, Zimmerman R, McMahon JT: Hydroxyapatite-coated femoral stems. Histological analysis of components retrieved at autopsy. J Bone Joint Surg Am. 1991, 73: 1439-1452.PubMed
49.
go back to reference Yee AJ, Kreder HK, Bookman I, Davey JR: A randomized trial of hydroxyapatite coated prostheses in total hip arthroplasty. Clin Orthop Relat Res. 1999, 366: 120-132. 10.1097/00003086-199909000-00016.CrossRefPubMed Yee AJ, Kreder HK, Bookman I, Davey JR: A randomized trial of hydroxyapatite coated prostheses in total hip arthroplasty. Clin Orthop Relat Res. 1999, 366: 120-132. 10.1097/00003086-199909000-00016.CrossRefPubMed
50.
go back to reference Huk OL, Bansal M, Betts F, Rimnac CM, Lieberman JR, Huo MH, Salvati EA: Polyethylene and metal debris generated by nonarticulating surfaces of modular acetabular components. J Bone Joint Surg Br. 1994, 76: 568-574.PubMed Huk OL, Bansal M, Betts F, Rimnac CM, Lieberman JR, Huo MH, Salvati EA: Polyethylene and metal debris generated by nonarticulating surfaces of modular acetabular components. J Bone Joint Surg Br. 1994, 76: 568-574.PubMed
51.
go back to reference Bloebaum RD, Mihalopoulus NL, Jensen JW, Dorr LD: Post-mortem analysis of bone growth into porous-coated acetabular components. J Bone Joint Surg Am. 1997, 79: 1013-1022.PubMed Bloebaum RD, Mihalopoulus NL, Jensen JW, Dorr LD: Post-mortem analysis of bone growth into porous-coated acetabular components. J Bone Joint Surg Am. 1997, 79: 1013-1022.PubMed
Metadata
Title
Failure of dual radius hydroxyapatite-coated acetabular cups
Authors
Fabio D'Angelo
Mauro Molina
Giacomo Riva
Giovanni Zatti
Paolo Cherubino
Publication date
01-12-2008
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2008
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/1749-799X-3-35

Other articles of this Issue 1/2008

Journal of Orthopaedic Surgery and Research 1/2008 Go to the issue