Skip to main content
Top
Published in: Radiation Oncology 1/2013

Open Access 01-12-2013 | Research

Accurate IMRT fluence verification for prostate cancer patients using ‘in-vivo’ measured EPID images and in-room acquired kilovoltage cone-beam CT scans

Authors: Ali SAM Ali, Maarten LP Dirkx, Ruud M Cools, Ben JM Heijmen

Published in: Radiation Oncology | Issue 1/2013

Login to get access

Abstract

Background

To investigate for prostate cancer patients the comparison of ‘in-vivo’ measured portal dose images (PDIs) with predictions based on a kilovoltage cone-beam CT scan (CBCT), acquired during the same treatment fraction, as an alternative for pre-treatment verification. For evaluation purposes, predictions were also performed using the patients’ planning CTs (pCT).

Methods

To get reliable CBCT electron densities for PDI predictions, Hounsfield units from the pCT were mapped onto the CBCT, while accounting for non-rigidity in patient anatomy in an approximate way. PDI prediction accuracy was first validated for an anatomical phantom, using IMRT treatment plans of ten prostate cancer patients. Clinical performance was studied using data acquired for 50 prostate cancer patients. For each patient, 4–5 CBCTs were available, resulting in a total of 1413 evaluated images. Measured and predicted PDIs were compared using γ-analyses with 3% global dose difference and 3 mm distance to agreement as reference criteria. Moreover, the pass rate for automated PDI comparison was assessed. To quantify improvements in IMRT fluence verification accuracy results from multiple fractions were combined by generating a γ-image with values halfway the minimum and median γ values, pixel by pixel.

Results

For patients, CBCT-based PDI predictions showed a high agreement with measurements, with an average percentage of rejected pixels of 1.41% only. In spite of possible intra-fraction motion and anatomy changes, this was only slightly larger than for phantom measurements (0.86%). For pCT-based predictions, the agreement deteriorated (average percentage of rejected pixels 2.98%), due to an enhanced impact of anatomy variations. For predictions based on CBCT, combination of the first 2 fractions yielded gamma results in close agreement with pre-treatment analyses (average percentage of rejected pixels 0.63% versus 0.35%, percentage of rejected beams 0.6% versus 0%). For the pCT-based approach, only combination of the first 5 fractions resulted in acceptable agreement with pre-treatment results.

Conclusion

In-room acquired CBCT scans can be used for high accuracy IMRT fluence verification based on in-vivo measured EPID images. Combination of γ results for the first 2 fractions can largely compensate for small accuracy reductions, with respect to pre-treatment verification, related to intra-fraction motion and anatomy changes.
Appendix
Available only for authorised users
Literature
1.
go back to reference Greer PB, Popescu CC: Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy. Med Phys 2003, 30: 1618-1627. 10.1118/1.1582469CrossRefPubMed Greer PB, Popescu CC: Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy. Med Phys 2003, 30: 1618-1627. 10.1118/1.1582469CrossRefPubMed
2.
go back to reference van Zijtveld M, Dirkx MLP, de Boer HCJ, Heijmen BJM: Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience. Radiother Oncol 2006, 81: 168-175. 10.1016/j.radonc.2006.09.008CrossRefPubMed van Zijtveld M, Dirkx MLP, de Boer HCJ, Heijmen BJM: Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience. Radiother Oncol 2006, 81: 168-175. 10.1016/j.radonc.2006.09.008CrossRefPubMed
3.
go back to reference Mc Dermott LN, Wendling M, Van Asselen B, et al.: Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification. Med Phys 2006, 33: 3921-3930. 10.1118/1.2230810CrossRef Mc Dermott LN, Wendling M, Van Asselen B, et al.: Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification. Med Phys 2006, 33: 3921-3930. 10.1118/1.2230810CrossRef
4.
go back to reference Partridge M, Symonds-Tayler JR, Evans PM: IMRT verification with a camera-based electronic portal imaging system. Phys Med Biol 2000, 45: N183-N196. 10.1088/0031-9155/45/12/402CrossRefPubMed Partridge M, Symonds-Tayler JR, Evans PM: IMRT verification with a camera-based electronic portal imaging system. Phys Med Biol 2000, 45: N183-N196. 10.1088/0031-9155/45/12/402CrossRefPubMed
5.
go back to reference Change J, Ling CC: Using the frame averaging of aS500 EPID for IMRT verification. J Appl Clin Med Phys 2003, 4: 287-299. 10.1120/1.1615071CrossRef Change J, Ling CC: Using the frame averaging of aS500 EPID for IMRT verification. J Appl Clin Med Phys 2003, 4: 287-299. 10.1120/1.1615071CrossRef
6.
go back to reference Warkentin B, Steciw S, Rathee S, et al.: Dosimetric IMRT verification with a flat-panel EPID. Med Phys 2003, 30: 3143-3155. 10.1118/1.1625440CrossRefPubMed Warkentin B, Steciw S, Rathee S, et al.: Dosimetric IMRT verification with a flat-panel EPID. Med Phys 2003, 30: 3143-3155. 10.1118/1.1625440CrossRefPubMed
7.
go back to reference Wendling M, Louwe RJW, McDermott LN, et al.: Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method. Med Phys 2006, 33: 259-273. 10.1118/1.2147744CrossRefPubMed Wendling M, Louwe RJW, McDermott LN, et al.: Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method. Med Phys 2006, 33: 259-273. 10.1118/1.2147744CrossRefPubMed
8.
go back to reference Kroonwijk M, Pasma KL, Quint S, Koper PC, Visser AG, Heijmen BJM: In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID): Demonstration of internal organ motion. Radiother Oncol 1998, 49: 125-132. 10.1016/S0167-8140(98)00122-4CrossRefPubMed Kroonwijk M, Pasma KL, Quint S, Koper PC, Visser AG, Heijmen BJM: In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID): Demonstration of internal organ motion. Radiother Oncol 1998, 49: 125-132. 10.1016/S0167-8140(98)00122-4CrossRefPubMed
9.
go back to reference Pasma KL, Kroonwijk M, Quint S, Visser AG, Heijmen BJM: Transit dosimetry with an electronic portal imaging device _EPID_ for 115 prostate cancer patients. Int J Radiat Oncol Biol Phys 1999, 45: 1297-1303. 10.1016/S0360-3016(99)00328-4CrossRefPubMed Pasma KL, Kroonwijk M, Quint S, Visser AG, Heijmen BJM: Transit dosimetry with an electronic portal imaging device _EPID_ for 115 prostate cancer patients. Int J Radiat Oncol Biol Phys 1999, 45: 1297-1303. 10.1016/S0360-3016(99)00328-4CrossRefPubMed
10.
go back to reference Zijtveld M, Dirkx M, Breuers M, de Boer H, Heijmen B: Portal dose image prediction for in vivo treatment verification completely based on EPID measurements. Med Phys 2009, 36: 946-952. 10.1118/1.3070545CrossRefPubMed Zijtveld M, Dirkx M, Breuers M, de Boer H, Heijmen B: Portal dose image prediction for in vivo treatment verification completely based on EPID measurements. Med Phys 2009, 36: 946-952. 10.1118/1.3070545CrossRefPubMed
11.
go back to reference Dirkx M, Heijmen B: Testing of the stability of intensity modulated beams generated with dynamic multileaf collimation, applied to the MM50 Racetrack Microtron. Med Phys 2000, 27: 2701-2707. 10.1118/1.1326450CrossRefPubMed Dirkx M, Heijmen B: Testing of the stability of intensity modulated beams generated with dynamic multileaf collimation, applied to the MM50 Racetrack Microtron. Med Phys 2000, 27: 2701-2707. 10.1118/1.1326450CrossRefPubMed
12.
go back to reference Vieira SC, Dirkx MLP, Heijmen BJM, de Boer HCJ: SIFT: a method to verify the IMRT fluence delivered during patient treatment using an electronic portal imaging device. Int J Radiat Oncol Biol Phys 2004, 60: 981-993.CrossRefPubMed Vieira SC, Dirkx MLP, Heijmen BJM, de Boer HCJ: SIFT: a method to verify the IMRT fluence delivered during patient treatment using an electronic portal imaging device. Int J Radiat Oncol Biol Phys 2004, 60: 981-993.CrossRefPubMed
13.
go back to reference Franken EM, de Boer JCJ, Barnhoorn JC, Heijmen BJM: Characteristics relevant to portal dosimetry of a cooled CCD camera-based EPID. Med Phys 2004, 31: 2549-2551. 10.1118/1.1785893CrossRefPubMed Franken EM, de Boer JCJ, Barnhoorn JC, Heijmen BJM: Characteristics relevant to portal dosimetry of a cooled CCD camera-based EPID. Med Phys 2004, 31: 2549-2551. 10.1118/1.1785893CrossRefPubMed
14.
go back to reference Franken EM, de Boer JCJ, Heijmen BJM: A novel approach to accurate portal dosimetry using CCD-camera based EPIDs. Med Phys 2006, 33: 888-903. 10.1118/1.2172746CrossRefPubMed Franken EM, de Boer JCJ, Heijmen BJM: A novel approach to accurate portal dosimetry using CCD-camera based EPIDs. Med Phys 2006, 33: 888-903. 10.1118/1.2172746CrossRefPubMed
15.
go back to reference Pasma KL, Heijmen BJM, Kroonwijk M, Visser AG: Portal dose image prediction for dosimetric treatment verification in radiotherapy I: an algorithm for open beams. Med Phys 1998, 25: 830-840. 10.1118/1.598293CrossRefPubMed Pasma KL, Heijmen BJM, Kroonwijk M, Visser AG: Portal dose image prediction for dosimetric treatment verification in radiotherapy I: an algorithm for open beams. Med Phys 1998, 25: 830-840. 10.1118/1.598293CrossRefPubMed
16.
go back to reference Ali ASAM, Dirkx MLP, Breuers MGJ, Heijmen BJM: Inclusion of the treatment couch in Portal dose image prediction for high precision EPID dosimetry. Med Phys 2011, 38: 377-381. 10.1118/1.3523615CrossRefPubMed Ali ASAM, Dirkx MLP, Breuers MGJ, Heijmen BJM: Inclusion of the treatment couch in Portal dose image prediction for high precision EPID dosimetry. Med Phys 2011, 38: 377-381. 10.1118/1.3523615CrossRefPubMed
17.
go back to reference Siewerdsen JH, Jaffray DA: Cone-beam computed tomography with a flat-panel imager: magnitude and effects of X-ray scatter. Med Phys 2001, 28: 220-231. 10.1118/1.1339879CrossRefPubMed Siewerdsen JH, Jaffray DA: Cone-beam computed tomography with a flat-panel imager: magnitude and effects of X-ray scatter. Med Phys 2001, 28: 220-231. 10.1118/1.1339879CrossRefPubMed
18.
go back to reference van Zijtveld M, Dirkx MLP, Heijmen BJM: Correction of Cone beam CT values using a planning CT for derivation of the “dose of the day. Radiother Oncol 2007, 85: 195-200. 10.1016/j.radonc.2007.08.010CrossRefPubMed van Zijtveld M, Dirkx MLP, Heijmen BJM: Correction of Cone beam CT values using a planning CT for derivation of the “dose of the day. Radiother Oncol 2007, 85: 195-200. 10.1016/j.radonc.2007.08.010CrossRefPubMed
19.
go back to reference Mutanga TF, de Boer HCJ, Van der Wielen GJ, et al.: Stereotactic targeting in prostate radiotherapy: speed and precision by daily automatic positioning corrections using kilovoltage/megavoltage images pairs. Int J Radiat Oncol Biol Phys 2008, 71: 1074-1083. 10.1016/j.ijrobp.2007.11.022CrossRefPubMed Mutanga TF, de Boer HCJ, Van der Wielen GJ, et al.: Stereotactic targeting in prostate radiotherapy: speed and precision by daily automatic positioning corrections using kilovoltage/megavoltage images pairs. Int J Radiat Oncol Biol Phys 2008, 71: 1074-1083. 10.1016/j.ijrobp.2007.11.022CrossRefPubMed
20.
go back to reference Mutanga TF, de Boer HC, Rajan V, Dirkx ML, van Os MJ, Incrocci L, Heijmen BJ: Software-controlled, highly automated intrafraction prostate motion correction with intrafraction stereographic targeting: System description and clinical results. Med Phys 2012, 39: 1314-1321. 10.1118/1.3684953CrossRefPubMed Mutanga TF, de Boer HC, Rajan V, Dirkx ML, van Os MJ, Incrocci L, Heijmen BJ: Software-controlled, highly automated intrafraction prostate motion correction with intrafraction stereographic targeting: System description and clinical results. Med Phys 2012, 39: 1314-1321. 10.1118/1.3684953CrossRefPubMed
21.
go back to reference McDermott LN, Wending W, Sonke JJ, Van Herk M, Mijnheer BJ: Replacing pretreatment verification with an in vivo EPID dosimetry for prostate IMRT. Int J Radiat Oncol Biol Phys 2007, 67: 1568-1577. 10.1016/j.ijrobp.2006.11.047CrossRefPubMed McDermott LN, Wending W, Sonke JJ, Van Herk M, Mijnheer BJ: Replacing pretreatment verification with an in vivo EPID dosimetry for prostate IMRT. Int J Radiat Oncol Biol Phys 2007, 67: 1568-1577. 10.1016/j.ijrobp.2006.11.047CrossRefPubMed
22.
go back to reference Dirkx MLP, Akhiat A, et al.: Plan100: a computer-controlled system to achieve “100%” certainty in correct daily plan delivery. In Proceedings of the XVI th International Conference on the Use of Computers in Radiation Therapy (ICCR2010): 31 May - June 3 2010. Edited by: Jan-Jakob S. Antoni van Leeuwenhoek Ziekenhuis: Het Nederlands Kanker Instituut; 2010. Dirkx MLP, Akhiat A, et al.: Plan100: a computer-controlled system to achieve “100%” certainty in correct daily plan delivery. In Proceedings of the XVI th International Conference on the Use of Computers in Radiation Therapy (ICCR2010): 31 May - June 3 2010. Edited by: Jan-Jakob S. Antoni van Leeuwenhoek Ziekenhuis: Het Nederlands Kanker Instituut; 2010.
23.
go back to reference Vieira SC, Dirkx ML, Pasma KL, Heijmen BJ: Fast and accurate leaf verification for dynamic multileaf collimation using an electronic portal imaging device. Med Phys 2002, 29: 2034-2040. 10.1118/1.1501141CrossRefPubMed Vieira SC, Dirkx ML, Pasma KL, Heijmen BJ: Fast and accurate leaf verification for dynamic multileaf collimation using an electronic portal imaging device. Med Phys 2002, 29: 2034-2040. 10.1118/1.1501141CrossRefPubMed
24.
go back to reference Vieira SC, Bolt RA, Dirkx MLP, Visser AG, Heijmen BJM: Fast, daily linac verification for segmented IMRT using electronic portal imaging. Radiother Oncol 2006, 80: 86-92. 10.1016/j.radonc.2006.06.010CrossRefPubMed Vieira SC, Bolt RA, Dirkx MLP, Visser AG, Heijmen BJM: Fast, daily linac verification for segmented IMRT using electronic portal imaging. Radiother Oncol 2006, 80: 86-92. 10.1016/j.radonc.2006.06.010CrossRefPubMed
25.
go back to reference McDermott LN, Wendling M, Nijkamp J, Mans A, Sonke JJ, Mijnheer BJ, van Herk M: 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT. Radiother Oncol 2008, 86: 35-42. 10.1016/j.radonc.2007.11.010CrossRefPubMed McDermott LN, Wendling M, Nijkamp J, Mans A, Sonke JJ, Mijnheer BJ, van Herk M: 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT. Radiother Oncol 2008, 86: 35-42. 10.1016/j.radonc.2007.11.010CrossRefPubMed
26.
go back to reference van Elmpt W, Nijsten S, Petit S, Mijnheer B, Lambin P, Dekker A: 3D in vivo dosimetry using megavoltage cone-beam CT and EPID dosimetry. Int J Radiat Oncol Biol Phys 2009, 73: 1580-1587. 10.1016/j.ijrobp.2008.11.051CrossRefPubMed van Elmpt W, Nijsten S, Petit S, Mijnheer B, Lambin P, Dekker A: 3D in vivo dosimetry using megavoltage cone-beam CT and EPID dosimetry. Int J Radiat Oncol Biol Phys 2009, 73: 1580-1587. 10.1016/j.ijrobp.2008.11.051CrossRefPubMed
27.
go back to reference Van Elmpt WJC, Petit SF, Dirk DR, Lambin P, Dekker ALAJ: 3D dose delivery verification using repeated cone-beam imaging and EPID dosimetry for stereotactic body radiotherapy of non-small cell lung cancer. Radiother Oncol 2010, 94: 188-194. 10.1016/j.radonc.2009.12.024CrossRefPubMed Van Elmpt WJC, Petit SF, Dirk DR, Lambin P, Dekker ALAJ: 3D dose delivery verification using repeated cone-beam imaging and EPID dosimetry for stereotactic body radiotherapy of non-small cell lung cancer. Radiother Oncol 2010, 94: 188-194. 10.1016/j.radonc.2009.12.024CrossRefPubMed
28.
go back to reference Zijtveld M, Dirkx MLP, Breuers MGJ, Kuipers R, Heijmen BJM: Evaluation of the ‘dose of the day’ for IMRT prostate cancer patients derived from portal dose measurements and cone-beam CT. Radiother Oncol 2010, 96: 172-177. 10.1016/j.radonc.2010.05.015CrossRefPubMed Zijtveld M, Dirkx MLP, Breuers MGJ, Kuipers R, Heijmen BJM: Evaluation of the ‘dose of the day’ for IMRT prostate cancer patients derived from portal dose measurements and cone-beam CT. Radiother Oncol 2010, 96: 172-177. 10.1016/j.radonc.2010.05.015CrossRefPubMed
Metadata
Title
Accurate IMRT fluence verification for prostate cancer patients using ‘in-vivo’ measured EPID images and in-room acquired kilovoltage cone-beam CT scans
Authors
Ali SAM Ali
Maarten LP Dirkx
Ruud M Cools
Ben JM Heijmen
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2013
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-8-211

Other articles of this Issue 1/2013

Radiation Oncology 1/2013 Go to the issue