Impact of different leaf velocities and dose rates on the number of monitor units and the dose-volume-histograms using intensity modulated radiotherapy with sliding-window technique

Authors: Hilke Vorwerk, Daniela Wagner, Clemens F Hess

Published in: Radiation Oncology | Issue 1/2008

Login to get access

Abstract

Background

Intensity modulated radiotherapy (IMRT) using sliding window technique utilises a leaf sequencing algorithm, which takes some control system limitations like dose rates (DR) and velocity of the leafs (LV) into account. The effect of altering these limitations on the number of monitor units and radiation dose to the organs at risk (OAR) were analysed.

Methods

IMRT plans for different LVs from 1.0 cm/sec to 10.0 cm/sec and different DRs from 100 MU/min to 600 MU/min for two patients with prostate cancer and two patients with squamous cell cancer of the scalp (SCCscalp) were calculated using the same "optimal fluence map". For each field the number of monitor units, the dose volume histograms and the differences in the "actual fluence maps" of the fields were analysed.

Results

With increase of the DR and decrease of the LV the number of monitor units increased and consequentially the radiation dose given to the OAR. In particular the serial OARs of patients with SCCscalp, which are located outside the end position of the leafs and inside the open field, received an additional dose of a higher DR and lower LV is used.

Conclusion

For best protection of organs at risk, a low DR and high LV should be applied. But the consequence of a low DR is both a long treatment time and also that a LV of higher than 3.0 cm/sec is mechanically not applicable. Our recommendation for an optimisation of the discussed parameters is a leaf velocity of 2.5 cm/sec and a dose rate of 300–400 MU/min (prostate cancer) and 100–200 MU/min (SCCscalp) for best protection of organs at risk, short treatment time and number of monitor units.
Appendix
Available only for authorised users
Literature
1.
go back to reference Litzenberg DW, Moran JM, Fraass BA: Incorporation of realistic delivery limitations into dynamic MLC treatment delivery. Med Phys 2002, 29: 810-820. 10.1118/1.1470499CrossRefPubMed Litzenberg DW, Moran JM, Fraass BA: Incorporation of realistic delivery limitations into dynamic MLC treatment delivery. Med Phys 2002, 29: 810-820. 10.1118/1.1470499CrossRefPubMed
2.
go back to reference Low DA, Sohn JW, Klein EE, Markman J, Mutic S, Dempsey JF: Characterization of a commercial multileaf collimator used for intensity modulated radiation therapy. Med Phys 2001, 28: 752-6. 10.1118/1.1367863CrossRefPubMed Low DA, Sohn JW, Klein EE, Markman J, Mutic S, Dempsey JF: Characterization of a commercial multileaf collimator used for intensity modulated radiation therapy. Med Phys 2001, 28: 752-6. 10.1118/1.1367863CrossRefPubMed
3.
go back to reference Emami B, Lyman J, Brown A, Coia L, Goitein M, Munzenrider JE, Shank B, Solin LJ, Wesson M: Tolerance of normal tissue to therapeutic irradiation. Int J Radiat Oncol Biol Phys 1991, 21: 109-122.CrossRefPubMed Emami B, Lyman J, Brown A, Coia L, Goitein M, Munzenrider JE, Shank B, Solin LJ, Wesson M: Tolerance of normal tissue to therapeutic irradiation. Int J Radiat Oncol Biol Phys 1991, 21: 109-122.CrossRefPubMed
4.
go back to reference De Martin E, Fiorino C, Broggi S, Longobardi B, Pierelli A, Perna L, Cattaneo GM, Calandrino R: Agreement criteria between expected and measured field fluences in IMRT of head and neck cancer: The importance and use of the gamma histograms statistical analysis. Radiother Oncol 2007, 85: 399-406. 10.1016/j.radonc.2007.10.045CrossRefPubMed De Martin E, Fiorino C, Broggi S, Longobardi B, Pierelli A, Perna L, Cattaneo GM, Calandrino R: Agreement criteria between expected and measured field fluences in IMRT of head and neck cancer: The importance and use of the gamma histograms statistical analysis. Radiother Oncol 2007, 85: 399-406. 10.1016/j.radonc.2007.10.045CrossRefPubMed
5.
go back to reference Wendling M, Zijp LJ, McDermott LN, Smit EJ, Sonke JJ, Mijnheer BJ, van Herk M: A fast algorithm for gamma evaluation in 3D. Med Phys 2007, 34: 1647-1654. 10.1118/1.2721657CrossRefPubMed Wendling M, Zijp LJ, McDermott LN, Smit EJ, Sonke JJ, Mijnheer BJ, van Herk M: A fast algorithm for gamma evaluation in 3D. Med Phys 2007, 34: 1647-1654. 10.1118/1.2721657CrossRefPubMed
6.
go back to reference Spezi E, Lewis DG: Gamma histograms for radiotherapy plan evaluation. Radiother Oncol 2006, 79: 224-230. 10.1016/j.radonc.2006.03.020CrossRefPubMed Spezi E, Lewis DG: Gamma histograms for radiotherapy plan evaluation. Radiother Oncol 2006, 79: 224-230. 10.1016/j.radonc.2006.03.020CrossRefPubMed
7.
go back to reference Schmidhalter D, Fix MK, Niederer P, Mini R, Manser P: Leaf transmission reduction using moving jaws for dynamic MLC IMRT. Med Phys 2007, 34: 3674-3687. 10.1118/1.2768864CrossRefPubMed Schmidhalter D, Fix MK, Niederer P, Mini R, Manser P: Leaf transmission reduction using moving jaws for dynamic MLC IMRT. Med Phys 2007, 34: 3674-3687. 10.1118/1.2768864CrossRefPubMed
8.
go back to reference Kry SF, Salehpour M, Followill DS, Stovall M, Kuban DA, White RA, Rosen II: Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy. Int J Radiat Oncol Biol Phys 2005, 62: 1204-1216.CrossRefPubMed Kry SF, Salehpour M, Followill DS, Stovall M, Kuban DA, White RA, Rosen II: Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy. Int J Radiat Oncol Biol Phys 2005, 62: 1204-1216.CrossRefPubMed
9.
go back to reference Giorgia N, Antonella F, Eugenio V, Alessandro C, Filippo A, Luca C: What is an acceptably smoothed fluence? Dosimetric and delivery considerations for dynamic sliding window IMRT. Radiat Oncol 2007, 2: 42. 10.1186/1748-717X-2-42PubMedCentralCrossRefPubMed Giorgia N, Antonella F, Eugenio V, Alessandro C, Filippo A, Luca C: What is an acceptably smoothed fluence? Dosimetric and delivery considerations for dynamic sliding window IMRT. Radiat Oncol 2007, 2: 42. 10.1186/1748-717X-2-42PubMedCentralCrossRefPubMed
Metadata
Title
Impact of different leaf velocities and dose rates on the number of monitor units and the dose-volume-histograms using intensity modulated radiotherapy with sliding-window technique
Authors
Hilke Vorwerk
Daniela Wagner
Clemens F Hess
Publication date
01-12-2008
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2008
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-3-31

Other articles of this Issue 1/2008

Radiation Oncology 1/2008 Go to the issue