Towards highly accelerated Cartesian time-resolved 3D flow cardiovascular magnetic resonance in the clinical setting

Authors: Daniel Giese, James Wong, Gerald F Greil, Martin Buehrer, Tobias Schaeffter, Sebastian Kozerke

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2014

Login to get access

Abstract

Background

The clinical applicability of time-resolved 3D flow cardiovascular magnetic resonance (CMR) remains compromised by the long scan times associated with phase-contrast imaging. The present work demonstrates the applicability of 8-fold acceleration of Cartesian time-resolved 3D flow CMR in 10 volunteers and in 9 patients with different congenital heart diseases (CHD). It is demonstrated that accelerated 3D flow CMR data acquisition and image reconstruction using k-t PCA (principal component analysis) can be implemented into clinical workflow and results are sufficiently accurate relative to conventional 2D flow CMR to permit for comprehensive flow quantification in CHD patients.

Methods

The fidelity of k-t PCA was first investigated on retrospectively undersampled data for different acceleration factors and compared to k-t SENSE and fully sampled reference data. Subsequently, k-t PCA with 8-fold nominal undersampling was applied on 10 healthy volunteers and 9 CHD patients on a clinical 1.5 T MR scanner. Quantitative flow validation was performed in vessels of interest on the 3D flow datasets and compared to 2D through-plane flow acquisitions. Particle trace analysis was used to qualitatively visualise flow patterns in patients.

Results

Accelerated time-resolved 3D flow data were successfully acquired in all subjects with 8-fold nominal scan acceleration. Nominal scan times excluding navigator efficiency were on the order of 6 min and 7 min in patients and volunteers. Mean differences in stroke volume in selected vessels of interest were 2.5 ± 8.4 ml and 1.63 ± 4.8 ml in volunteers and patients, respectively. Qualitative flow pattern analysis in the time-resolved 3D dataset revealed valuable insights into hemodynamics including circular and helical patterns as well as flow distributions and origin in the Fontan circulation.

Conclusion

Highly accelerated time-resolved 3D flow using k-t PCA is readily applicable in clinical routine protocols of CHD patients. Nominal scan times of 6 min are well tolerated and allow for quantitative and qualitative flow assessment in all great vessels.
Appendix
Available only for authorised users
Literature
1.
go back to reference Markl M, Kilner PJ, Ebbers T: Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance. J. Cardiovasc. Magn. Reson. 2011, 13: 1-22.CrossRef Markl M, Kilner PJ, Ebbers T: Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance. J. Cardiovasc. Magn. Reson. 2011, 13: 1-22.CrossRef
2.
go back to reference Nordmeyer S, Riesenkampff E, Crelier G, Khasheei A, Schnackenburg B, Berger F, Kuehne T: Flow‐sensitive four‐dimensional cine magnetic resonance imaging for offline blood flow quantification in multiple vessels: a validation study. J. Magn. Reson. Imaging. 2010, 32: 677-683.CrossRefPubMed Nordmeyer S, Riesenkampff E, Crelier G, Khasheei A, Schnackenburg B, Berger F, Kuehne T: Flow‐sensitive four‐dimensional cine magnetic resonance imaging for offline blood flow quantification in multiple vessels: a validation study. J. Magn. Reson. Imaging. 2010, 32: 677-683.CrossRefPubMed
3.
go back to reference Markl M, Wallis W, Brendecke S, Simon J, Frydrychowicz A, Harloff A: Estimation of global aortic pulse wave velocity by flow‐sensitive 4D MRI. Magn. Reson. Med. 2010, 63: 1575-1582.CrossRefPubMed Markl M, Wallis W, Brendecke S, Simon J, Frydrychowicz A, Harloff A: Estimation of global aortic pulse wave velocity by flow‐sensitive 4D MRI. Magn. Reson. Med. 2010, 63: 1575-1582.CrossRefPubMed
4.
go back to reference Ebbers T, Wigström L, Bolger AF, Engvall J, Karlsson M: Estimation of relative cardiovascular pressures using time‐resolved three‐dimensional phase contrast MRI. Magn. Reson. Med. 2001, 45: 872-879.CrossRefPubMed Ebbers T, Wigström L, Bolger AF, Engvall J, Karlsson M: Estimation of relative cardiovascular pressures using time‐resolved three‐dimensional phase contrast MRI. Magn. Reson. Med. 2001, 45: 872-879.CrossRefPubMed
5.
go back to reference Stalder A, Russe M, Frydrychowicz A, Bock J, Hennig J, Markl M: Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn. Reson. Med. 2008, 60: 1218-1231.CrossRefPubMed Stalder A, Russe M, Frydrychowicz A, Bock J, Hennig J, Markl M: Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn. Reson. Med. 2008, 60: 1218-1231.CrossRefPubMed
6.
go back to reference Petersson S, Dyverfeldt P, Ebbers T: Assessment of the accuracy of MRI wall shear stress estimation using numerical simulations. J. Magn. Reson. Imaging. 2012, 36: 128-138.CrossRefPubMed Petersson S, Dyverfeldt P, Ebbers T: Assessment of the accuracy of MRI wall shear stress estimation using numerical simulations. J. Magn. Reson. Imaging. 2012, 36: 128-138.CrossRefPubMed
7.
go back to reference Wu SP, Ringgaard S, Pedersen EM: Three-dimensional phase contrast velocity mapping acquisition improves wall shear stress estimation in vivo. Magn. Reson. Imaging. 2004, 22: 345-351.CrossRefPubMed Wu SP, Ringgaard S, Pedersen EM: Three-dimensional phase contrast velocity mapping acquisition improves wall shear stress estimation in vivo. Magn. Reson. Imaging. 2004, 22: 345-351.CrossRefPubMed
8.
go back to reference Dyverfeldt P, Sigfridsson A, Kvitting JPE, Ebbers T: Quantification of intravoxel velocity standard deviation and turbulence intensity by generalizing phase‐contrast MRI. Magn. Reson. Med. 2006, 56: 850-858.CrossRefPubMed Dyverfeldt P, Sigfridsson A, Kvitting JPE, Ebbers T: Quantification of intravoxel velocity standard deviation and turbulence intensity by generalizing phase‐contrast MRI. Magn. Reson. Med. 2006, 56: 850-858.CrossRefPubMed
9.
go back to reference Binter C, Knobloch V, Manka R, Sigfridsson A, Kozerke S: Bayesian multipoint velocity encoding for concurrent flow and turbulence mapping. Magn. Reson. Med. 2012, doi:10.1002/mrm.24370 Binter C, Knobloch V, Manka R, Sigfridsson A, Kozerke S: Bayesian multipoint velocity encoding for concurrent flow and turbulence mapping. Magn. Reson. Med. 2012, doi:10.1002/mrm.24370
10.
go back to reference Tsao J, Boesiger P, Pruessmann K: kt BLAST and kt SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn. Reson. Med. 2003, 50: 1031-1042.CrossRefPubMed Tsao J, Boesiger P, Pruessmann K: kt BLAST and kt SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn. Reson. Med. 2003, 50: 1031-1042.CrossRefPubMed
11.
go back to reference Baltes C, Kozerke S, Hansen MS, Pruessmann KP, Tsao J, Boesiger P: Accelerating cine phase-contrast flow measurements using k-t BLAST and k-t SENSE. Magn. Reson. Med. 2005, 54: 1430-1438.CrossRefPubMed Baltes C, Kozerke S, Hansen MS, Pruessmann KP, Tsao J, Boesiger P: Accelerating cine phase-contrast flow measurements using k-t BLAST and k-t SENSE. Magn. Reson. Med. 2005, 54: 1430-1438.CrossRefPubMed
12.
go back to reference Jung B, Ullmann P, Honal M, Bauer S, Hennig J, Markl M: Parallel MRI with extended and averaged GRAPPA kernels (PEAK-GRAPPA): optimized spatiotemporal dynamic imaging. J. Magn. Reson. Imaging. 2008, 28: 1226-1232.CrossRefPubMed Jung B, Ullmann P, Honal M, Bauer S, Hennig J, Markl M: Parallel MRI with extended and averaged GRAPPA kernels (PEAK-GRAPPA): optimized spatiotemporal dynamic imaging. J. Magn. Reson. Imaging. 2008, 28: 1226-1232.CrossRefPubMed
13.
go back to reference Stadlbauer A, Van Der Riet W, Crelier G, Salomonowitz E: Accelerated time-resolved three-dimensional MR velocity mapping of blood flow patterns in the aorta using SENSE and kt BLAST. Eur. J. Radiol. 2010, 75: 15-21.CrossRef Stadlbauer A, Van Der Riet W, Crelier G, Salomonowitz E: Accelerated time-resolved three-dimensional MR velocity mapping of blood flow patterns in the aorta using SENSE and kt BLAST. Eur. J. Radiol. 2010, 75: 15-21.CrossRef
14.
go back to reference Carlsson M, Töger J, Kanski M, Bloch KM, Ståhlberg F, Heiberg E, Arheden H: Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or kt BLAST: head to head comparison and validation at 1.5 T and 3 T. J. Cardiovasc. Magn. Reson. 2011, 13: 1-7.CrossRef Carlsson M, Töger J, Kanski M, Bloch KM, Ståhlberg F, Heiberg E, Arheden H: Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or kt BLAST: head to head comparison and validation at 1.5 T and 3 T. J. Cardiovasc. Magn. Reson. 2011, 13: 1-7.CrossRef
15.
go back to reference Schnell S, Markl M, Entezari P, Mahadewia RJ, Semaan E, Stankovic Z, Collins J, Carr J, Jung B: k‐t GRAPPA accelerated four‐dimensional flow MRI in the aorta: effect on scan time, image quality, and quantification of flow and wall shear stress. Magn. Reson. Med. 2013, doi:10.1002/mrm.24925 Schnell S, Markl M, Entezari P, Mahadewia RJ, Semaan E, Stankovic Z, Collins J, Carr J, Jung B: k‐t GRAPPA accelerated four‐dimensional flow MRI in the aorta: effect on scan time, image quality, and quantification of flow and wall shear stress. Magn. Reson. Med. 2013, doi:10.1002/mrm.24925
16.
go back to reference Stadlbauer A, van der Riet W, Globits S, Crelier G, Salomonowitz E: Accelerated phase‐contrast MR imaging: comparison of k‐t BLAST with SENSE and doppler ultrasound for velocity and flow measurements in the aorta. J. Magn. Reson. Imaging. 2009, 29: 817-824.CrossRefPubMed Stadlbauer A, van der Riet W, Globits S, Crelier G, Salomonowitz E: Accelerated phase‐contrast MR imaging: comparison of k‐t BLAST with SENSE and doppler ultrasound for velocity and flow measurements in the aorta. J. Magn. Reson. Imaging. 2009, 29: 817-824.CrossRefPubMed
17.
go back to reference Knobloch V, Boesiger P, Kozerke S: Sparsity transform k‐t principal component analysis for accelerating cine three‐dimensional flow measurements. Magn. Reson. Med. 2012, 70: 53-63.CrossRefPubMed Knobloch V, Boesiger P, Kozerke S: Sparsity transform k‐t principal component analysis for accelerating cine three‐dimensional flow measurements. Magn. Reson. Med. 2012, 70: 53-63.CrossRefPubMed
18.
go back to reference Hsiao A, Lustig M, Alley MT, Murphy MJ, Vasanawala SS: Evaluation of valvular insufficiency and shunts with parallel-imaging compressed-sensing 4D phase-contrast MR imaging with stereoscopic 3D velocity-fusion volume-rendered visualization. Radiology. 2012, 265: 87-95.PubMedCentralCrossRefPubMed Hsiao A, Lustig M, Alley MT, Murphy MJ, Vasanawala SS: Evaluation of valvular insufficiency and shunts with parallel-imaging compressed-sensing 4D phase-contrast MR imaging with stereoscopic 3D velocity-fusion volume-rendered visualization. Radiology. 2012, 265: 87-95.PubMedCentralCrossRefPubMed
19.
go back to reference Frydrychowicz A, Wieben O, Niespodsany E, Reeder S, Johnson K, Francois C, et al: Quantification of thoracic blood flow using volumetric magnetic resonance imaging with radial velocity encoding. Invest Radiol. 2013, 48: 819-825.CrossRefPubMed Frydrychowicz A, Wieben O, Niespodsany E, Reeder S, Johnson K, Francois C, et al: Quantification of thoracic blood flow using volumetric magnetic resonance imaging with radial velocity encoding. Invest Radiol. 2013, 48: 819-825.CrossRefPubMed
20.
go back to reference Gatehouse PD, Keegan J, Crowe LA, Masood S, Mohiaddin RH, Kreitner KF, Firmin DN: Applications of phase-contrast flow and velocity imaging in cardiovascular MRI. Eur. Radiol. 2005, 15: 2172-2184.CrossRefPubMed Gatehouse PD, Keegan J, Crowe LA, Masood S, Mohiaddin RH, Kreitner KF, Firmin DN: Applications of phase-contrast flow and velocity imaging in cardiovascular MRI. Eur. Radiol. 2005, 15: 2172-2184.CrossRefPubMed
21.
go back to reference Frydrychowicz A, François CJ, Turski PA: Four-dimensional phase contrast magnetic resonance angiography: potential clinical applications. Eur. J. Radiol. 2011, 80: 24-35.PubMedCentralCrossRefPubMed Frydrychowicz A, François CJ, Turski PA: Four-dimensional phase contrast magnetic resonance angiography: potential clinical applications. Eur. J. Radiol. 2011, 80: 24-35.PubMedCentralCrossRefPubMed
22.
go back to reference Westenberg JJ, Roes SD, Marsan NA, Binnendijk NM, Doornbos J, Bax JJ, Reiber JH, De Roos A, van der Geest RJ: Mitral valve and tricuspid valve blood flow: accurate quantification with 3D velocity-encoded MR imaging with retrospective valve tracking1. Radiology. 2008, 249: 792-800.CrossRefPubMed Westenberg JJ, Roes SD, Marsan NA, Binnendijk NM, Doornbos J, Bax JJ, Reiber JH, De Roos A, van der Geest RJ: Mitral valve and tricuspid valve blood flow: accurate quantification with 3D velocity-encoded MR imaging with retrospective valve tracking1. Radiology. 2008, 249: 792-800.CrossRefPubMed
23.
go back to reference Pedersen H, Kozerke S, Ringgaard S, Nehrke K, Kim WY: k-t PCA: temporally constrained k-t BLAST reconstruction using principal component analysis. Magn. Reson. Med. 2009, 62: 706-716.CrossRefPubMed Pedersen H, Kozerke S, Ringgaard S, Nehrke K, Kim WY: k-t PCA: temporally constrained k-t BLAST reconstruction using principal component analysis. Magn. Reson. Med. 2009, 62: 706-716.CrossRefPubMed
24.
go back to reference Pelc N, Bernstein M, Shimakawa A, Glover G: Encoding strategies for three-direction phase-contrast MR imaging of flow. J. Magn. Reson. Imaging. 1991, 1: 405-413.CrossRefPubMed Pelc N, Bernstein M, Shimakawa A, Glover G: Encoding strategies for three-direction phase-contrast MR imaging of flow. J. Magn. Reson. Imaging. 1991, 1: 405-413.CrossRefPubMed
25.
go back to reference Beerbaum P, Körperich H, Gieseke J, Barth P, Peuster M, Meyer H, et al: Blood flow quantification in adults by phase-contrast MRI combined with SENSE-a validation study. J. Cardiovasc. Magn. Reson. 2005, 7: 361-369.CrossRefPubMed Beerbaum P, Körperich H, Gieseke J, Barth P, Peuster M, Meyer H, et al: Blood flow quantification in adults by phase-contrast MRI combined with SENSE-a validation study. J. Cardiovasc. Magn. Reson. 2005, 7: 361-369.CrossRefPubMed
26.
go back to reference Bernstein M, Zhou X, Polzin J, King K, Ganin A, Pelc N, Glover G: Concomitant gradient terms in phase contrast MR: analysis and correction. Magn. Reson. Med. 1998, 39: 300-308.CrossRefPubMed Bernstein M, Zhou X, Polzin J, King K, Ganin A, Pelc N, Glover G: Concomitant gradient terms in phase contrast MR: analysis and correction. Magn. Reson. Med. 1998, 39: 300-308.CrossRefPubMed
27.
go back to reference Walker PG, Cranney GB, Scheidegger MB, Waseleski G, Pohost GM, Yoganathan AP: Semiautomated method for noise reduction and background phase error correction in MR phase velocity data. J. Magn. Reson. Imaging. 1993, 3: 521-530.CrossRefPubMed Walker PG, Cranney GB, Scheidegger MB, Waseleski G, Pohost GM, Yoganathan AP: Semiautomated method for noise reduction and background phase error correction in MR phase velocity data. J. Magn. Reson. Imaging. 1993, 3: 521-530.CrossRefPubMed
28.
go back to reference Lankhaar JW, Hofman MB, Marcus JT, Zwanenburg JJ, Faes TJ, Vonk-Noordegraaf A: Correction of phase offset errors in main pulmonary artery flow quantification. J. Magn. Reson. Imaging. 2005, 22: 73-79.CrossRefPubMed Lankhaar JW, Hofman MB, Marcus JT, Zwanenburg JJ, Faes TJ, Vonk-Noordegraaf A: Correction of phase offset errors in main pulmonary artery flow quantification. J. Magn. Reson. Imaging. 2005, 22: 73-79.CrossRefPubMed
29.
go back to reference Valverde I, Nordmeyer S, Uribe S, Greil G, Berger F, Kuehne T, Beerbaum P: Systemic-to-pulmonary collateral flow in patients with palliated univentricular heart physiology: measurement using cardiovascular magnetic resonance 4D velocity acquisition. J. Cardiovasc. Magn. Reson. 2012, 14: 25-PubMedCentralCrossRefPubMed Valverde I, Nordmeyer S, Uribe S, Greil G, Berger F, Kuehne T, Beerbaum P: Systemic-to-pulmonary collateral flow in patients with palliated univentricular heart physiology: measurement using cardiovascular magnetic resonance 4D velocity acquisition. J. Cardiovasc. Magn. Reson. 2012, 14: 25-PubMedCentralCrossRefPubMed
30.
go back to reference Uribe S, Beerbaum P, Soerensen T, Rasmusson A, Razavi R, Schaeffter T: Four-dimensional (4D) flow of the whole heart and great vessels using real-time respiratory self-gating. Magn. Reson. Med. 2009, 62: 984-992.CrossRefPubMed Uribe S, Beerbaum P, Soerensen T, Rasmusson A, Razavi R, Schaeffter T: Four-dimensional (4D) flow of the whole heart and great vessels using real-time respiratory self-gating. Magn. Reson. Med. 2009, 62: 984-992.CrossRefPubMed
31.
go back to reference Gatehouse P, Rolf M, Graves M, Hofman M, Totman J, Werner B, Quest R, Liu Y, von Spiczak J, Dieringer M: Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. J. Cardiovasc. Magn. Reson. 2010, 12: 5-PubMedCentralCrossRefPubMed Gatehouse P, Rolf M, Graves M, Hofman M, Totman J, Werner B, Quest R, Liu Y, von Spiczak J, Dieringer M: Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. J. Cardiovasc. Magn. Reson. 2010, 12: 5-PubMedCentralCrossRefPubMed
32.
go back to reference Giese D, Haeberlin M, Barmet C, Pruessmann KP, Schaeffter T, Kozerke S: Analysis and correction of background velocity offsets in phase contrast flow measurements using magnetic field monitoring. Magn. Reson. Med. 2012, 67: 1294-1302.CrossRefPubMed Giese D, Haeberlin M, Barmet C, Pruessmann KP, Schaeffter T, Kozerke S: Analysis and correction of background velocity offsets in phase contrast flow measurements using magnetic field monitoring. Magn. Reson. Med. 2012, 67: 1294-1302.CrossRefPubMed
33.
go back to reference Wolf RL, Ehman RL, Riederer SJ, Rossman PJ: Analysis of systematic and random error in MR volumetric flow measurements. Magn. Reson. Med. 1993, 30: 82-91.CrossRefPubMed Wolf RL, Ehman RL, Riederer SJ, Rossman PJ: Analysis of systematic and random error in MR volumetric flow measurements. Magn. Reson. Med. 1993, 30: 82-91.CrossRefPubMed
34.
go back to reference Burkart DJ, Felmlee JP, Johnson CD, Wolf RL, Weaver AL, Ehman RL: Cine phase-contrast MR flow measurements: improved precision using an automated method of vessel detection. J. Comput. Assist. Tomogr. 1994, 18: 469-475.CrossRefPubMed Burkart DJ, Felmlee JP, Johnson CD, Wolf RL, Weaver AL, Ehman RL: Cine phase-contrast MR flow measurements: improved precision using an automated method of vessel detection. J. Comput. Assist. Tomogr. 1994, 18: 469-475.CrossRefPubMed
35.
go back to reference Conturo TE, Smith GD: Signal‐to‐noise in phase angle reconstruction: dynamic range extension using phase reference offsets. Magn. Reson. Med. 1990, 15: 420-437.CrossRefPubMed Conturo TE, Smith GD: Signal‐to‐noise in phase angle reconstruction: dynamic range extension using phase reference offsets. Magn. Reson. Med. 1990, 15: 420-437.CrossRefPubMed
36.
go back to reference Lee A, Pike G, Pelc N: Three-point phase-contrast velocity measurements with increased velocity-to-noise ratio. Magn. Reson. Med. 1995, 33: 122-2.CrossRefPubMed Lee A, Pike G, Pelc N: Three-point phase-contrast velocity measurements with increased velocity-to-noise ratio. Magn. Reson. Med. 1995, 33: 122-2.CrossRefPubMed
37.
go back to reference Lamothe M, Rutt BK: Multistep phase difference phase contrast imaging. J. Magn. Reson. Imaging. 1997, 7: 838-842.CrossRefPubMed Lamothe M, Rutt BK: Multistep phase difference phase contrast imaging. J. Magn. Reson. Imaging. 1997, 7: 838-842.CrossRefPubMed
Metadata
Title
Towards highly accelerated Cartesian time-resolved 3D flow cardiovascular magnetic resonance in the clinical setting
Authors
Daniel Giese
James Wong
Gerald F Greil
Martin Buehrer
Tobias Schaeffter
Sebastian Kozerke
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2014
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-16-42

Other articles of this Issue 1/2014

Journal of Cardiovascular Magnetic Resonance 1/2014 Go to the issue