Published in: Molecular Cancer 1/2004

Open Access 01-12-2004 | Research

Sensitivity and reproducibility of standardized-competitive RT-PCR for transcript quantification and its comparison with real time RT-PCR

Authors: Vincenzo Pagliarulo, Ben George, Stephen J Beil, Susan Groshen, Peter W Laird, Jie Cai, James Willey, Richard J Cote, Ram H Datar

Published in: Molecular Cancer | Issue 1/2004

Login to get access

Abstract

Background

Probe based detection assays form the mainstay of transcript quantification. Problems with these assays include varying hybridization efficiencies of the probes used for transcript quantification and the expense involved. We examined the ability of a standardized competitive RT-PCR (StaRT PCR) assay to quantify transcripts of 4 cell cycle associated genes (RB, E2F1, CDKN2A and PCNA) in two cell lines (T24 & LD419) and compared its efficacy with the established Taqman real time quantitative RT-PCR assay. We also assessed the sensitivity, reproducibility and consistency of StaRT PCR. StaRT PCR assay is based on the incorporation of competitive templates (CT) in precisely standardized quantities along with the native template (NT) in a PCR reaction. This enables transcript quantification by comparing the NT and CT band intensities at the end of the PCR amplification. The CT serves as an ideal internal control. The transcript numbers are expressed as copies per million transcripts of a control gene such as β-actin (ACTB).

Results

The NT and CT were amplified at remarkably similar rates throughout the StaRT PCR amplification cycles, and the coefficient of variation was least (<3.8%) when the NT/CT ratio was kept as close to 1:1 as possible. The variability between the rates of amplification in different tubes subjected to the same StaRT PCR reaction was very low and within the range of experimental noise. Further, StaRT PCR was sensitive enough to detect variations as low as 10% in endogenous actin transcript quantity (p < 0.01 by the paired student's t-test). StaRT PCR correlated well with Taqman real time RT-PCR assay in terms of transcript quantification efficacy (p < 0.01 for all 4 genes by the Spearman Rank correlation method) and the ability to discriminate between cell types and confluence patterns.

Conclusion

StaRT PCR is thus a reliable and sensitive technique that can be applied to medium-high throughput quantitative transcript measurement. Further, it correlates well with Taqman real time PCR in terms of quantitative and discriminatory ability. This label-free, inexpensive technique may provide the ability to generate prognostically important molecular signatures unique to individual tumors and may enable identification of novel therapeutic targets.
Appendix
Available only for authorised users
Literature
1.
go back to reference Wu TD: Analyzing gene expression data from DNA microarrays to identify candidate genes. J Pathol. 2001, 195: 53-65.CrossRefPubMed Wu TD: Analyzing gene expression data from DNA microarrays to identify candidate genes. J Pathol. 2001, 195: 53-65.CrossRefPubMed
2.
go back to reference Willey JC, Crawford EL, Jackson CM, Weaver DA, Hoban JC, Khude SA, DeMuth JP: Expression measurement of many genes simultaneously by quantitative RT-PCR using standardized mixtures of competitive templates. AM J Respir Cell Mol Biol. 1998, 19: 6-17.CrossRefPubMed Willey JC, Crawford EL, Jackson CM, Weaver DA, Hoban JC, Khude SA, DeMuth JP: Expression measurement of many genes simultaneously by quantitative RT-PCR using standardized mixtures of competitive templates. AM J Respir Cell Mol Biol. 1998, 19: 6-17.CrossRefPubMed
3.
go back to reference Raeymaekers L: Quantitative PCR: Theoretical considerations with practical implications. Anal Biochem. 1993, 214: 582-5.CrossRefPubMed Raeymaekers L: Quantitative PCR: Theoretical considerations with practical implications. Anal Biochem. 1993, 214: 582-5.CrossRefPubMed
4.
go back to reference Pfaffl MW, Hageleit M: Validities of mRNA quantification using recombinant RNA and recombinant DNA external calibration curves in real-time RT-PCR. Biotechnol Lett. 2001, 23: 275-282. 10.1023/A:1005658330108.CrossRef Pfaffl MW, Hageleit M: Validities of mRNA quantification using recombinant RNA and recombinant DNA external calibration curves in real-time RT-PCR. Biotechnol Lett. 2001, 23: 275-282. 10.1023/A:1005658330108.CrossRef
5.
go back to reference Crawford EL, Godfridus JP, Noordhuis P, Rots MG, Vondracek M, Grafström RC, Lieuallen K, Lennon G, Zahorchak RJ, Georgeson MJ, Wali A, Lechner JF, Fan P-S, Kahaleh MB, Khuder SA, Warner KA, Weaver DA, Willey JC: Reproducible gene expression measurement among multiple laboratories obtained in a blinded study using standardized RT (StaRT)-PCR. Mol Diagnosis. 2001, 6: 217-225. 10.1054/modi.2001.29789.CrossRef Crawford EL, Godfridus JP, Noordhuis P, Rots MG, Vondracek M, Grafström RC, Lieuallen K, Lennon G, Zahorchak RJ, Georgeson MJ, Wali A, Lechner JF, Fan P-S, Kahaleh MB, Khuder SA, Warner KA, Weaver DA, Willey JC: Reproducible gene expression measurement among multiple laboratories obtained in a blinded study using standardized RT (StaRT)-PCR. Mol Diagnosis. 2001, 6: 217-225. 10.1054/modi.2001.29789.CrossRef
6.
go back to reference Yue H, Eastman PS, Wang BB, Minor J, Doctolero MH, Nuttall RL, Stack R, Becker JW, Montgomery JR, Vainer M, Johnston R: An evaluation of the performance of cDNA microarrays for detecting changes in global mRNA expression. Nucleic Acids Res. 2001, 29: e41-PubMedCentralCrossRefPubMed Yue H, Eastman PS, Wang BB, Minor J, Doctolero MH, Nuttall RL, Stack R, Becker JW, Montgomery JR, Vainer M, Johnston R: An evaluation of the performance of cDNA microarrays for detecting changes in global mRNA expression. Nucleic Acids Res. 2001, 29: e41-PubMedCentralCrossRefPubMed
7.
go back to reference Crawford EL, Warner KA, Khuder SA, Zahorchak RJ, Willey JC: Multiplex standardized RT-PCR for expression analysis of many genes in small samples. Biochem Biophys Res Commun. 2002, 293: 509-516.CrossRefPubMed Crawford EL, Warner KA, Khuder SA, Zahorchak RJ, Willey JC: Multiplex standardized RT-PCR for expression analysis of many genes in small samples. Biochem Biophys Res Commun. 2002, 293: 509-516.CrossRefPubMed
Metadata
Title
Sensitivity and reproducibility of standardized-competitive RT-PCR for transcript quantification and its comparison with real time RT-PCR
Authors
Vincenzo Pagliarulo
Ben George
Stephen J Beil
Susan Groshen
Peter W Laird
Jie Cai
James Willey
Richard J Cote
Ram H Datar
Publication date
01-12-2004
Publisher
BioMed Central
Published in
Molecular Cancer / Issue 1/2004
Electronic ISSN: 1476-4598
DOI
https://doi.org/10.1186/1476-4598-3-5

Other articles of this Issue 1/2004

Molecular Cancer 1/2004 Go to the issue
Webinar | 19-02-2024 | 17:30 (CET)

Keynote webinar | Spotlight on antibody–drug conjugates in cancer

Antibody–drug conjugates (ADCs) are novel agents that have shown promise across multiple tumor types. Explore the current landscape of ADCs in breast and lung cancer with our experts, and gain insights into the mechanism of action, key clinical trials data, existing challenges, and future directions.

Dr. Véronique Diéras
Prof. Fabrice Barlesi
Developed by: Springer Medicine