Skip to main content
Top
Published in: BMC Ear, Nose and Throat Disorders 1/2005

Open Access 01-12-2005 | Case report

Histopathological changes in the human larynx following expanded polytetrafluroethylene (Gore-Tex®) implantation

Authors: K Rajkumar, HS Khalil, M Elloy, E Sheffield, DL Baldwin

Published in: BMC Ear, Nose and Throat Disorders | Issue 1/2005

Login to get access

Abstract

Background

Expanded polytetrafluroethelyne (e PTFE, Gore-Tex®) has been advocated as an implant material for medialization of the vocal fold. Animal studies involving rabbits and a porcine model have demonstrated host tolerance of the implant. There have been no reports describing the histological changes in a human laryngectomy specimen with a Gore-Tex implant.

Case presentation

The histological findings in a laryngectomy specimen of a patient previously implanted with e PTFE for medialization of a paralyzed vocal fold following excision of a vagal neurofibroma were studied.
Histopathology revealed a mild foreign-body giant cell granulomatous reaction with some associated fibrosis. The granulomatous response was limited to the periphery of the Gore-Tex and although it closely followed the profile of the material it did not encroach into or significantly break up the material. There was no significant neutrophilic or lymphocytic inflammation.

Conclusions

Our findings are consistent with the animal models confirming that Gore-Tex implantation does not result in a significant granulomatous reaction in the human larynx over a 13-month period. Moreover, there is no evidence of resorption or infection. Further, the lack of lymphocytes in association with the granulomas indicates that there is no significant immunological hypersensitivity. Histologically, the slight permeation by connective tissue is similar to that seen in Gore-Tex vascular and cardiac implants. The degree of the slight giant cell response appears to be dependent on the profile of the material; a sharp edge incited more of a response than a flat surface.
Appendix
Available only for authorised users
Literature
1.
go back to reference McCulloch PM, Hoffman HT: Medialisation laryngoplasty with expanded polytetrafluroethylene. Surgical technique and preliminary results. Annals Otol Rhinol Laryngol. 1998, 107: 427-432.CrossRef McCulloch PM, Hoffman HT: Medialisation laryngoplasty with expanded polytetrafluroethylene. Surgical technique and preliminary results. Annals Otol Rhinol Laryngol. 1998, 107: 427-432.CrossRef
2.
go back to reference Cashman S, Simpson CB, McGuff HS: Soft tissue response of the rabbit larynx to Gortex implant. Annals Otol Rhinol Laryngol. 2002, 111: 977-982.CrossRef Cashman S, Simpson CB, McGuff HS: Soft tissue response of the rabbit larynx to Gortex implant. Annals Otol Rhinol Laryngol. 2002, 111: 977-982.CrossRef
3.
go back to reference Varvares MA, Montgomery WW, Hillman RE: Teflon Granuloma of the larynx: aetiology, pathophysiology and management. Annals Otol Rhinol Laryngol. 1995, 104: 511-515.CrossRef Varvares MA, Montgomery WW, Hillman RE: Teflon Granuloma of the larynx: aetiology, pathophysiology and management. Annals Otol Rhinol Laryngol. 1995, 104: 511-515.CrossRef
4.
go back to reference Bauer CA, Valentino J, Hoffman HT: Longterm results of vocal cord augmentation with autogenous fat. Annals Otol Rhinol Laryngol. 1995, 104: 871-874.CrossRef Bauer CA, Valentino J, Hoffman HT: Longterm results of vocal cord augmentation with autogenous fat. Annals Otol Rhinol Laryngol. 1995, 104: 871-874.CrossRef
5.
go back to reference Stein J, Eliachar I, Munoz-Ramirez H, Myles J, Strom EM: Histopathologic study of alternative substances for vocal fold medialization. Annals Otol Rhinol Laryngol. 2000, 109: 221-226.CrossRef Stein J, Eliachar I, Munoz-Ramirez H, Myles J, Strom EM: Histopathologic study of alternative substances for vocal fold medialization. Annals Otol Rhinol Laryngol. 2000, 109: 221-226.CrossRef
6.
go back to reference Giovanni A, Vallicioni J, Gras R, Zanaret M: Clinical experience with Gore-Tex for vocal fold medialization. Laryngoscope. 1999, 198: 284-288. 10.1097/00005537-199902000-00020.CrossRef Giovanni A, Vallicioni J, Gras R, Zanaret M: Clinical experience with Gore-Tex for vocal fold medialization. Laryngoscope. 1999, 198: 284-288. 10.1097/00005537-199902000-00020.CrossRef
8.
go back to reference Tamplenizza P, Acocella F, Addis F, Bolandrini E: Tubular e-PTFE implants in glottal and neoglottal insufficiency: tolerance and stability in an animal model. J Laryngol Otol. 2001, 115 (2): 106-111. 10.1258/0022215011907695.CrossRefPubMed Tamplenizza P, Acocella F, Addis F, Bolandrini E: Tubular e-PTFE implants in glottal and neoglottal insufficiency: tolerance and stability in an animal model. J Laryngol Otol. 2001, 115 (2): 106-111. 10.1258/0022215011907695.CrossRefPubMed
9.
go back to reference Hubmer MG, Hoffman C, Popper H, Scharnagl E: Expanded polytetrafluroethylene threads for lip augmentation induce foreign body granulomatous reaction. Plast Reconstr Surg. 1999, 103: 1277-1279.CrossRefPubMed Hubmer MG, Hoffman C, Popper H, Scharnagl E: Expanded polytetrafluroethylene threads for lip augmentation induce foreign body granulomatous reaction. Plast Reconstr Surg. 1999, 103: 1277-1279.CrossRefPubMed
10.
go back to reference Mole B: The use of Gore-Tex implants in aesthetic surgery of the face. Plast Reconstr Surg. 1992, 90: 200-206.CrossRefPubMed Mole B: The use of Gore-Tex implants in aesthetic surgery of the face. Plast Reconstr Surg. 1992, 90: 200-206.CrossRefPubMed
11.
go back to reference Conrad K, Gillman G: A six year experience with the use of expanded polytetrafluroethylene in rhinoplasty. Plast Reconstr Surg. 1998, 101: 1675-1684.CrossRefPubMed Conrad K, Gillman G: A six year experience with the use of expanded polytetrafluroethylene in rhinoplasty. Plast Reconstr Surg. 1998, 101: 1675-1684.CrossRefPubMed
12.
go back to reference Zeitels SM: New procedures for paralytic dysphonia: adduction arytenopexy, Gore-Tex medialization laryngoplasty and cricothyroid subluxation. Otolaryngol Clin North Am. 2000, 33: 841-854.CrossRefPubMed Zeitels SM: New procedures for paralytic dysphonia: adduction arytenopexy, Gore-Tex medialization laryngoplasty and cricothyroid subluxation. Otolaryngol Clin North Am. 2000, 33: 841-854.CrossRefPubMed
Metadata
Title
Histopathological changes in the human larynx following expanded polytetrafluroethylene (Gore-Tex®) implantation
Authors
K Rajkumar
HS Khalil
M Elloy
E Sheffield
DL Baldwin
Publication date
01-12-2005
Publisher
BioMed Central
Published in
BMC Ear, Nose and Throat Disorders / Issue 1/2005
Electronic ISSN: 1472-6815
DOI
https://doi.org/10.1186/1472-6815-5-1

Other articles of this Issue 1/2005

BMC Ear, Nose and Throat Disorders 1/2005 Go to the issue