Published in: BMC Ophthalmology 1/2013

Open Access 01-12-2013 | Research article

Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions

Authors: Caterina Schweier, James VM Hanson, Jens Funk, Marc Töteberg-Harms

Published in: BMC Ophthalmology | Issue 1/2013

Login to get access

Abstract

Background

Icare PRO (ICP) is a new Rebound tonometer that is able to measure intraocular pressure (IOP) in both sitting and reclining positions. In this study, the gold standard Goldmann tonometer (GAT) was compared to ICP and Tono-Pen AVIA (TPA). Hypothesis was that repeatability of GAT is superior to ICP and TPA.

Methods

36 eyes of 36 healthy caucasian individuals, 13 male and 26 females, 17 right and 19 left eyes have been included in this prospective, randomized, cross-sectional study. The study was conducted at a single site (Dept. of Ophthalmology, UniversityHospital Zurich, Switzerland). Primary outcome measures were Intraclass correlation coefficients (ICC) and coefficients of variation (COV) and test-retest repeatability as visualized by Bland-Altman analysis. Secondary outcome measures were IOP in sitting (GAT, ICP and TPA) and in reclining (ICP and TPA) position.

Results

Mean IOP measured by GAT was 14.9±3.5 mmHg. Mean IOP measured by ICP was 15.6±3.1 mmHg (with TPA 14.8±2.7 mmHg) in sitting and 16.5±3.5 mmHg (with TPA 17.0±3.0 mmHg) in reclining positions. COVs ranged from 2.9% (GAT) to 6.9% (ICP reclining) and ICCs from 0.819 (ICP reclining) to 0.972 (GAT).

Conclusions

Repeatability is good with all three devices. GAT has higher repeatability compared to the two tested hand-held devices with lowest COVs and highest ICCs. IOP was higher in the reclining compared to the sitting position.

Trial registration

The study was registered to the Clinical Trials Register of the US National Institute of Health, NCT01325324.
Appendix
Available only for authorised users
Literature
1.
go back to reference Goldmann H, Schmidt T: Applanation tonometry. Ophthalmologica. 1957, 134 (4): 221-242. 10.1159/000303213.CrossRefPubMed Goldmann H, Schmidt T: Applanation tonometry. Ophthalmologica. 1957, 134 (4): 221-242. 10.1159/000303213.CrossRefPubMed
2.
go back to reference Punjabi OS, Ho HK, Kniestedt C, Bostrom AG, Stamper RL, Lin SC: Intraocular pressure and ocular pulse amplitude comparisons in different types of glaucoma using dynamic contour tonometry. Curr Eye Res. 2006, 31 (10): 851-862. 10.1080/02713680600899887.CrossRefPubMed Punjabi OS, Ho HK, Kniestedt C, Bostrom AG, Stamper RL, Lin SC: Intraocular pressure and ocular pulse amplitude comparisons in different types of glaucoma using dynamic contour tonometry. Curr Eye Res. 2006, 31 (10): 851-862. 10.1080/02713680600899887.CrossRefPubMed
3.
go back to reference Kaufmann C, Bachmann LM, Thiel MA: Comparison of dynamic contour tonometry with goldmann applanation tonometry. Invest Ophthalmol Vis Sci. 2004, 45 (9): 3118-3121. 10.1167/iovs.04-0018.CrossRefPubMed Kaufmann C, Bachmann LM, Thiel MA: Comparison of dynamic contour tonometry with goldmann applanation tonometry. Invest Ophthalmol Vis Sci. 2004, 45 (9): 3118-3121. 10.1167/iovs.04-0018.CrossRefPubMed
4.
go back to reference Medeiros FA, Sample PA, Weinreb RN: Comparison of dynamic contour tonometry and goldmann applanation tonometry in African American subjects. Ophthalmology. 2007, 114 (4): 658-665. 10.1016/j.ophtha.2006.06.065.CrossRefPubMed Medeiros FA, Sample PA, Weinreb RN: Comparison of dynamic contour tonometry and goldmann applanation tonometry in African American subjects. Ophthalmology. 2007, 114 (4): 658-665. 10.1016/j.ophtha.2006.06.065.CrossRefPubMed
5.
go back to reference Huang Y, Tham CC, Zhang M: Central corneal thickness and applanation tonometry. J Cataract Refract Surg. 2008, 34 (3): 347-CrossRefPubMed Huang Y, Tham CC, Zhang M: Central corneal thickness and applanation tonometry. J Cataract Refract Surg. 2008, 34 (3): 347-CrossRefPubMed
6.
go back to reference Holladay JT, Allison ME, Prager TC: Goldmann applanation tonometry in patients with regular corneal astigmatism. Am J Ophthalmol. 1983, 96 (1): 90-93.CrossRefPubMed Holladay JT, Allison ME, Prager TC: Goldmann applanation tonometry in patients with regular corneal astigmatism. Am J Ophthalmol. 1983, 96 (1): 90-93.CrossRefPubMed
7.
go back to reference Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ: Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry. Am J Ophthalmol. 2007, 143 (1): 39-47. 10.1016/j.ajo.2006.09.036.CrossRefPubMed Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ: Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry. Am J Ophthalmol. 2007, 143 (1): 39-47. 10.1016/j.ajo.2006.09.036.CrossRefPubMed
8.
go back to reference Broman AT, Congdon NG, Bandeen-Roche K, Quigley HA: Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure. J Glaucoma. 2007, 16 (7): 581-588. 10.1097/IJG.0b013e3180640f40.CrossRefPubMed Broman AT, Congdon NG, Bandeen-Roche K, Quigley HA: Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure. J Glaucoma. 2007, 16 (7): 581-588. 10.1097/IJG.0b013e3180640f40.CrossRefPubMed
9.
go back to reference Iester M, Mermoud A, Achache F, Roy S: New Tonopen XL: comparison with the Goldmann tonometer. Eye (Lond). 2001, 15 (Pt 1): 52-58.CrossRef Iester M, Mermoud A, Achache F, Roy S: New Tonopen XL: comparison with the Goldmann tonometer. Eye (Lond). 2001, 15 (Pt 1): 52-58.CrossRef
10.
go back to reference Bhartiya S, Bali SJ, Sharma R, Chaturvedi N, Dada T: Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry. Int Ophthalmol. 2011, 31 (4): 297-302. 10.1007/s10792-011-9458-4.CrossRefPubMed Bhartiya S, Bali SJ, Sharma R, Chaturvedi N, Dada T: Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry. Int Ophthalmol. 2011, 31 (4): 297-302. 10.1007/s10792-011-9458-4.CrossRefPubMed
11.
go back to reference Mackay RS, Marg E: Fast, automatic, electronic tonometers based on an exact theroy. Acta Ophthalmol (Copenh). 1959, 37: 495-507.CrossRef Mackay RS, Marg E: Fast, automatic, electronic tonometers based on an exact theroy. Acta Ophthalmol (Copenh). 1959, 37: 495-507.CrossRef
12.
go back to reference Mackay RS, Marg E, Oechsli R: Automatic tonometer with exact theory: various biological applications. Science. 1960, 131: 1668-1669. 10.1126/science.131.3414.1668.CrossRefPubMed Mackay RS, Marg E, Oechsli R: Automatic tonometer with exact theory: various biological applications. Science. 1960, 131: 1668-1669. 10.1126/science.131.3414.1668.CrossRefPubMed
13.
go back to reference Dekking HM, Coster HD: Dynamic tonometry. Ophthalmologica. 1967, 154 (1): 59-74. 10.1159/000305149.CrossRefPubMed Dekking HM, Coster HD: Dynamic tonometry. Ophthalmologica. 1967, 154 (1): 59-74. 10.1159/000305149.CrossRefPubMed
14.
go back to reference Kontiola A: A new electromechanical method for measuring intraocular pressure. Doc Ophthalmol. 1996, 93 (3): 265-276.CrossRefPubMed Kontiola A: A new electromechanical method for measuring intraocular pressure. Doc Ophthalmol. 1996, 93 (3): 265-276.CrossRefPubMed
15.
go back to reference Kontiola AI: A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand. 2000, 78 (2): 142-145. 10.1034/j.1600-0420.2000.078002142.x.CrossRefPubMed Kontiola AI: A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand. 2000, 78 (2): 142-145. 10.1034/j.1600-0420.2000.078002142.x.CrossRefPubMed
16.
go back to reference Kontiola A, Puska P: Measuring intraocular pressure with the pulsair 3000 and rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol. 2004, 242 (1): 3-7. 10.1007/s00417-003-0671-3.CrossRefPubMed Kontiola A, Puska P: Measuring intraocular pressure with the pulsair 3000 and rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol. 2004, 242 (1): 3-7. 10.1007/s00417-003-0671-3.CrossRefPubMed
17.
go back to reference Rabe-Hesketh S, Skrondal A: Estimation using xtmixed. Multilevel and Longitudinal Modeling Using STATA. Edited by: Rabe-Hesketh S, Skrondal A. 2008, College Station, TX: STATA Press, 433-436. 2 Rabe-Hesketh S, Skrondal A: Estimation using xtmixed. Multilevel and Longitudinal Modeling Using STATA. Edited by: Rabe-Hesketh S, Skrondal A. 2008, College Station, TX: STATA Press, 433-436. 2
18.
go back to reference Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8 (2): 135-160. 10.1191/096228099673819272.CrossRefPubMed Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8 (2): 135-160. 10.1191/096228099673819272.CrossRefPubMed
19.
go back to reference Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1 (8476): 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1 (8476): 307-310.CrossRefPubMed
20.
go back to reference Buchanan RA, Williams TD: Intraocular pressure, ocular pulse pressure, and body position. Am J Optom Physiol Opt. 1985, 62 (1): 59-62. 10.1097/00006324-198501000-00008.CrossRefPubMed Buchanan RA, Williams TD: Intraocular pressure, ocular pulse pressure, and body position. Am J Optom Physiol Opt. 1985, 62 (1): 59-62. 10.1097/00006324-198501000-00008.CrossRefPubMed
21.
go back to reference Jorge J, Ramoa-Marques R, Lourenco A, Silva S, Nascimento S, Queiros A, Gonzalez-Meijome JM: IOP variations in the sitting and supine positions. J Glaucoma. 2010, 19 (9): 609-612. 10.1097/IJG.0b013e3181ca7ca5.CrossRefPubMed Jorge J, Ramoa-Marques R, Lourenco A, Silva S, Nascimento S, Queiros A, Gonzalez-Meijome JM: IOP variations in the sitting and supine positions. J Glaucoma. 2010, 19 (9): 609-612. 10.1097/IJG.0b013e3181ca7ca5.CrossRefPubMed
22.
go back to reference Prata TS, De-Moraes CG, Kanadani FN, Ritch R, Paranhos A: Posture-induced intraocular pressure changes: considerations regarding body position in glaucoma patients. Surv Ophthalmol. 2010, 55 (5): 445-453. 10.1016/j.survophthal.2009.12.002.CrossRefPubMed Prata TS, De-Moraes CG, Kanadani FN, Ritch R, Paranhos A: Posture-induced intraocular pressure changes: considerations regarding body position in glaucoma patients. Surv Ophthalmol. 2010, 55 (5): 445-453. 10.1016/j.survophthal.2009.12.002.CrossRefPubMed
23.
go back to reference Chui WS, Lam A, Chen D, Chiu R: The influence of corneal properties on rebound tonometry. Ophthalmology. 2008, 115 (1): 80-84. 10.1016/j.ophtha.2007.03.061.CrossRefPubMed Chui WS, Lam A, Chen D, Chiu R: The influence of corneal properties on rebound tonometry. Ophthalmology. 2008, 115 (1): 80-84. 10.1016/j.ophtha.2007.03.061.CrossRefPubMed
24.
go back to reference Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B: Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol. 2006, 90 (7): 833-835. 10.1136/bjo.2005.089870.CrossRefPubMedPubMedCentral Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B: Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol. 2006, 90 (7): 833-835. 10.1136/bjo.2005.089870.CrossRefPubMedPubMedCentral
25.
go back to reference Martinez-de-la-Casa JM, Garcia-Feijoo J, Castillo A, Garcia-Sanchez J: Reproducibility and clinical evaluation of rebound tonometry. Invest Ophthalmol Vis Sci. 2005, 46 (12): 4578-4580. 10.1167/iovs.05-0586.CrossRefPubMed Martinez-de-la-Casa JM, Garcia-Feijoo J, Castillo A, Garcia-Sanchez J: Reproducibility and clinical evaluation of rebound tonometry. Invest Ophthalmol Vis Sci. 2005, 46 (12): 4578-4580. 10.1167/iovs.05-0586.CrossRefPubMed
26.
go back to reference Ismail AR, Lamont M, Perera S, Khan-Lim D, Mehta R, Macleod JD, Anderson DF: Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplasties. Br J Ophthalmol. 2007, 91 (7): 980-981. 10.1136/bjo.2006.099564.CrossRefPubMedPubMedCentral Ismail AR, Lamont M, Perera S, Khan-Lim D, Mehta R, Macleod JD, Anderson DF: Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplasties. Br J Ophthalmol. 2007, 91 (7): 980-981. 10.1136/bjo.2006.099564.CrossRefPubMedPubMedCentral
27.
go back to reference Moreno-Montanes J, Olmo N, Zarranz-Ventura J, Heras-Mulero H: Dynamic contour tonometry in eyes after penetrating keratoplasty. Cornea. 2009, 28 (7): 836-837.CrossRefPubMed Moreno-Montanes J, Olmo N, Zarranz-Ventura J, Heras-Mulero H: Dynamic contour tonometry in eyes after penetrating keratoplasty. Cornea. 2009, 28 (7): 836-837.CrossRefPubMed
28.
go back to reference Rootman DS, Insler MS, Thompson HW, Parelman J, Poland D, Unterman SR: Accuracy and precision of the Tono-Pen in measuring intraocular pressure after keratoplasty and epikeratophakia and in scarred corneas. Arch Ophthalmol. 1988, 106 (12): 1697-1700. 10.1001/archopht.1988.01060140869030.CrossRefPubMed Rootman DS, Insler MS, Thompson HW, Parelman J, Poland D, Unterman SR: Accuracy and precision of the Tono-Pen in measuring intraocular pressure after keratoplasty and epikeratophakia and in scarred corneas. Arch Ophthalmol. 1988, 106 (12): 1697-1700. 10.1001/archopht.1988.01060140869030.CrossRefPubMed
29.
go back to reference Ceruti P, Morbio R, Marraffa M, Marchini G: Comparison of dynamic contour tonometry and goldmann applanation tonometry in deep lamellar and penetrating keratoplasties. Am J Ophthalmol. 2008, 145 (2): 215-221. 10.1016/j.ajo.2007.10.003.CrossRefPubMed Ceruti P, Morbio R, Marraffa M, Marchini G: Comparison of dynamic contour tonometry and goldmann applanation tonometry in deep lamellar and penetrating keratoplasties. Am J Ophthalmol. 2008, 145 (2): 215-221. 10.1016/j.ajo.2007.10.003.CrossRefPubMed
30.
go back to reference Kniestedt C, Lin S, Choe J, Bostrom A, Nee M, Stamper RL: Clinical comparison of contour and applanation tonometry and their relationship to pachymetry. Arch Ophthalmol. 2005, 123 (11): 1532-1537. 10.1001/archopht.123.11.1532.CrossRefPubMed Kniestedt C, Lin S, Choe J, Bostrom A, Nee M, Stamper RL: Clinical comparison of contour and applanation tonometry and their relationship to pachymetry. Arch Ophthalmol. 2005, 123 (11): 1532-1537. 10.1001/archopht.123.11.1532.CrossRefPubMed
31.
go back to reference Kniestedt C, Nee M, Stamper RL: Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol. 2005, 243 (4): 359-366. 10.1007/s00417-004-1024-6.CrossRefPubMed Kniestedt C, Nee M, Stamper RL: Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol. 2005, 243 (4): 359-366. 10.1007/s00417-004-1024-6.CrossRefPubMed
32.
go back to reference Frenkel RE, Hong YJ, Shin DH: Comparison of the Tono-Pen to the Goldmann applanation tonometer. Arch Ophthalmol. 1988, 106 (6): 750-753. 10.1001/archopht.1988.01060130820030.CrossRefPubMed Frenkel RE, Hong YJ, Shin DH: Comparison of the Tono-Pen to the Goldmann applanation tonometer. Arch Ophthalmol. 1988, 106 (6): 750-753. 10.1001/archopht.1988.01060130820030.CrossRefPubMed
33.
go back to reference Kao SF, Lichter PR, Bergstrom TJ, Rowe S, Musch DC: Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer. Ophthalmology. 1987, 94 (12): 1541-1544.CrossRefPubMed Kao SF, Lichter PR, Bergstrom TJ, Rowe S, Musch DC: Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer. Ophthalmology. 1987, 94 (12): 1541-1544.CrossRefPubMed
34.
go back to reference Tonnu PA, Ho T, Sharma K, White E, Bunce C, Garway-Heath D: A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol. 2005, 89 (7): 847-850. 10.1136/bjo.2004.056614.CrossRefPubMedPubMedCentral Tonnu PA, Ho T, Sharma K, White E, Bunce C, Garway-Heath D: A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol. 2005, 89 (7): 847-850. 10.1136/bjo.2004.056614.CrossRefPubMedPubMedCentral
Metadata
Title
Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions
Authors
Caterina Schweier
James VM Hanson
Jens Funk
Marc Töteberg-Harms
Publication date
01-12-2013
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2013
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/1471-2415-13-44

Other articles of this Issue 1/2013

BMC Ophthalmology 1/2013 Go to the issue