Published in: BMC Infectious Diseases 1/2010

Open Access 01-12-2010 | Research article

EVOTECH® endoscope cleaner and reprocessor (ECR) simulated-use and clinical-use evaluation of cleaning efficacy

Authors: Michelle J Alfa, Pat DeGagne, Nancy Olson, Iram Fatima

Published in: BMC Infectious Diseases | Issue 1/2010

Login to get access

Abstract

Background

The objective of this study was to perform simulated-use testing as well as a clinical study to assess the efficacy of the EVOTECH® Endoscope Cleaner and Reprocessor (ECR) cleaning for flexible colonoscopes, duodenoscopes, gastroscopes and bronchoscopes. The main aim was to determine if the cleaning achieved using the ECR was at least equivalent to that achieved using optimal manual cleaning.

Methods

Simulated-use testing consisted of inoculating all scope channels and two surface sites with Artificial Test Soil (ATS) containing 108 cfu/mL of Enterococcus faecalis, Pseudomonas aeruginosa and Candida albicans. Duodenoscopes, colonoscopes, and bronchoscopes (all Olympus endoscopes) were included in the simulated use testing. Each endoscope type was tested in triplicate and all channels and two surface sites were sampled for each scope. The clinical study evaluated patient-used duodenoscopes, bronchoscopes, colonoscopes, and gastroscopes (scopes used for emergency procedures were excluded) that had only a bedside flush prior to being processed in the ECR (i.e. no manual cleaning). There were 10 to 15 endoscopes evaluated post-cleaning and to ensure the entire ECR cycle was effective, 5 endoscopes were evaluated post-cleaning and post-high level disinfection. All channels and two external surface locations were sampled to evaluate the residual organic and microbial load. Effective cleaning of endoscope surfaces and channels was deemed to have been achieved if there was < 6.4 μg/cm2 of residual protein, < 1.8 μg/cm2 of residual hemoglobin and < 4 Log10 viable bacteria/cm2. Published data indicate that routine manual cleaning can achieve these endpoints so the ECR cleaning efficacy must meet or exceed these to establish that the ECR cleaning cycle could replace manual cleaning

Results

In the clinical study 75 patient-used scopes were evaluated post cleaning and 98.8% of surfaces and 99.7% of lumens met or surpassed the cleaning endpoints set for protein, hemoglobin and bioburden residuals. In the simulated-use study 100% of the Olympus colonoscopes, duodenoscopes and bronchoscopes evaluated met or surpassed the cleaning endpoints set for protein, and bioburden residuals (hemoglobin was not evaluated).

Conclusions

The ECR cleaning cycle provides an effective automated approach that ensures surfaces and channels of flexible endoscopes are adequately cleaned after having only a bedside flush but no manual cleaning. It is crucial to note that endoscopes used for emergency procedures or where reprocessing is delayed for more than one hour MUST still be manually cleaned prior to placing them in the ECR.
Appendix
Available only for authorised users
Literature
1.
go back to reference (AORN) AORN: Perioperative Standards and Recommended Practices. American Operating Room Nurses (AORN). 2008, 421-445. (AORN) AORN: Perioperative Standards and Recommended Practices. American Operating Room Nurses (AORN). 2008, 421-445.
2.
go back to reference Rutala WA, Weber DJ: Reprocessing endoscopes: United States perspective. J Hosp Infect. 2004, 56 (Suppl 2): S27-39. 10.1016/j.jhin.2003.12.035.CrossRefPubMed Rutala WA, Weber DJ: Reprocessing endoscopes: United States perspective. J Hosp Infect. 2004, 56 (Suppl 2): S27-39. 10.1016/j.jhin.2003.12.035.CrossRefPubMed
3.
go back to reference Canadian Standards Association Inc. Publishers M, ON: CSA Z314.8-08 Decontamination of Reusable Medical Devices. 2008 Canadian Standards Association Inc. Publishers M, ON: CSA Z314.8-08 Decontamination of Reusable Medical Devices. 2008
4.
go back to reference Bisset L, Cossart YE, Selby W, West R, Catterson D, O'Hara K, Vickery K: A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure. Am J Infect Control. 2006, 34 (5): 274-280. 10.1016/j.ajic.2005.08.007.CrossRefPubMed Bisset L, Cossart YE, Selby W, West R, Catterson D, O'Hara K, Vickery K: A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure. Am J Infect Control. 2006, 34 (5): 274-280. 10.1016/j.ajic.2005.08.007.CrossRefPubMed
5.
go back to reference Mehta AC, Prakash UB, Garland R, Haponik E, Moses L, Schaffner W, Silvestri G: American College of Chest Physicians and American Association for Bronchology [corrected] consensus statement: prevention of flexible bronchoscopy-associated infection. Chest. 2005, 128 (3): 1742-1755. 10.1378/chest.128.3.1742.CrossRefPubMed Mehta AC, Prakash UB, Garland R, Haponik E, Moses L, Schaffner W, Silvestri G: American College of Chest Physicians and American Association for Bronchology [corrected] consensus statement: prevention of flexible bronchoscopy-associated infection. Chest. 2005, 128 (3): 1742-1755. 10.1378/chest.128.3.1742.CrossRefPubMed
6.
go back to reference Moses FM, Lee JS: Current GI endoscope disinfection and QA practices. Dig Dis Sci. 2004, 49 (11-12): 1791-1797. 10.1007/s10620-004-9572-5.CrossRefPubMed Moses FM, Lee JS: Current GI endoscope disinfection and QA practices. Dig Dis Sci. 2004, 49 (11-12): 1791-1797. 10.1007/s10620-004-9572-5.CrossRefPubMed
7.
go back to reference Nelson DB, Muscarella LF: Current issues in endoscope reprocessing and infection control during gastrointestinal endoscopy. World J Gastroenterol. 2006, 12 (25): 3953-3964.CrossRefPubMedPubMedCentral Nelson DB, Muscarella LF: Current issues in endoscope reprocessing and infection control during gastrointestinal endoscopy. World J Gastroenterol. 2006, 12 (25): 3953-3964.CrossRefPubMedPubMedCentral
8.
go back to reference Association for the Advancement of Medical Instrumentation: ANSI/AAMI ST35: Safe handling and biolgoical decontamination of medical devices in health care facilities and in nonclinical settings. 1996 Association for the Advancement of Medical Instrumentation: ANSI/AAMI ST35: Safe handling and biolgoical decontamination of medical devices in health care facilities and in nonclinical settings. 1996
9.
go back to reference Pajkos A, Vickery K, Cossart Y: Is biofilm accumulation on endoscope tubing a contributor to the failure of cleaning and decontamination?. J Hosp Infect. 2004, 58 (3): 224-229. 10.1016/j.jhin.2004.06.023.CrossRefPubMed Pajkos A, Vickery K, Cossart Y: Is biofilm accumulation on endoscope tubing a contributor to the failure of cleaning and decontamination?. J Hosp Infect. 2004, 58 (3): 224-229. 10.1016/j.jhin.2004.06.023.CrossRefPubMed
10.
go back to reference Rutala WW, DJ; and the Healthcare Infection Control Practices Advisory Committee (HICPAC): Guideline for Disinfection and Sterilization in Healthcare facilities. Centre for Disease Control. 2008 Rutala WW, DJ; and the Healthcare Infection Control Practices Advisory Committee (HICPAC): Guideline for Disinfection and Sterilization in Healthcare facilities. Centre for Disease Control. 2008
11.
go back to reference Spach DH, Silverstein FE, Stamm WE: Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993, 118 (2): 117-128.CrossRefPubMed Spach DH, Silverstein FE, Stamm WE: Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993, 118 (2): 117-128.CrossRefPubMed
12.
go back to reference Puszko GB: Manual cleaning of endoscopes: a comparison study of syringe versus suction methods using the endo-suction cleaning system. Gastroenterol Nurs. 2001, 24 (2): 69-74. 10.1097/00001610-200103000-00005.CrossRefPubMed Puszko GB: Manual cleaning of endoscopes: a comparison study of syringe versus suction methods using the endo-suction cleaning system. Gastroenterol Nurs. 2001, 24 (2): 69-74. 10.1097/00001610-200103000-00005.CrossRefPubMed
13.
go back to reference Agerton T, Valway S, Gore B, Pozsik C, Plikaytis B, Woodley C, Onorato I: Transmission of a highly drug-resistant strain (strain W1) of Mycobacterium tuberculosis. Community outbreak and nosocomial transmission via a contaminated bronchoscope. JAMA. 1997, 278 (13): 1073-1077. 10.1001/jama.278.13.1073.CrossRefPubMed Agerton T, Valway S, Gore B, Pozsik C, Plikaytis B, Woodley C, Onorato I: Transmission of a highly drug-resistant strain (strain W1) of Mycobacterium tuberculosis. Community outbreak and nosocomial transmission via a contaminated bronchoscope. JAMA. 1997, 278 (13): 1073-1077. 10.1001/jama.278.13.1073.CrossRefPubMed
14.
go back to reference Bronowicki JP, Venard V, Botte C, Monhoven N, Gastin I, Chone L, Hudziak H, Rihn B, Delanoe C, LeFaou A, et al: Patient-to-patient transmission of hepatitis C virus during colonoscopy. N Engl J Med. 1997, 337 (4): 237-240. 10.1056/NEJM199707243370404.CrossRefPubMed Bronowicki JP, Venard V, Botte C, Monhoven N, Gastin I, Chone L, Hudziak H, Rihn B, Delanoe C, LeFaou A, et al: Patient-to-patient transmission of hepatitis C virus during colonoscopy. N Engl J Med. 1997, 337 (4): 237-240. 10.1056/NEJM199707243370404.CrossRefPubMed
15.
go back to reference Ahishali E U-BO, Dalopcioglu D: Chemical colistis due to glutaraldehyde: Case series and review of the literature. Dig Dis Sci. 2008, Ahishali E U-BO, Dalopcioglu D: Chemical colistis due to glutaraldehyde: Case series and review of the literature. Dig Dis Sci. 2008,
16.
go back to reference Nelson DB: Recent advances in epidemiology and prevention of gastrointestinal endoscopy related infections. Curr Opin Infect Dis. 2005, 18 (4): 326-330. 10.1097/01.qco.0000171925.47452.8f.CrossRefPubMed Nelson DB: Recent advances in epidemiology and prevention of gastrointestinal endoscopy related infections. Curr Opin Infect Dis. 2005, 18 (4): 326-330. 10.1097/01.qco.0000171925.47452.8f.CrossRefPubMed
17.
go back to reference Vickery K, Ngo QD, Zou J, Cossart YE: The effect of multiple cycles of contamination, detergent washing, and disinfection on the development of biofilm in endoscope tubing. Am J Infect Control. 2009, 37 (6): 470-475. 10.1016/j.ajic.2008.09.016.CrossRefPubMed Vickery K, Ngo QD, Zou J, Cossart YE: The effect of multiple cycles of contamination, detergent washing, and disinfection on the development of biofilm in endoscope tubing. Am J Infect Control. 2009, 37 (6): 470-475. 10.1016/j.ajic.2008.09.016.CrossRefPubMed
18.
go back to reference Zuhlsdorf B, Emmrich M, Floss H, Martiny H: Cleaning efficacy of nine different cleaners in a washer-disinfector designed for flexible endoscopes. J Hosp Infect. 2002, 52 (3): 206-211. 10.1053/jhin.2002.1284.CrossRefPubMed Zuhlsdorf B, Emmrich M, Floss H, Martiny H: Cleaning efficacy of nine different cleaners in a washer-disinfector designed for flexible endoscopes. J Hosp Infect. 2002, 52 (3): 206-211. 10.1053/jhin.2002.1284.CrossRefPubMed
19.
go back to reference Zuhlsdorf B, Martiny H: Intralaboratory reproducibility of the German test method of prEN ISO 15883-1 for determination of the cleaning efficacy of washer-disinfectors for flexible endoscopes. J Hosp Infect. 2005, 59 (4): 286-291. 10.1016/j.jhin.2004.09.033.CrossRefPubMed Zuhlsdorf B, Martiny H: Intralaboratory reproducibility of the German test method of prEN ISO 15883-1 for determination of the cleaning efficacy of washer-disinfectors for flexible endoscopes. J Hosp Infect. 2005, 59 (4): 286-291. 10.1016/j.jhin.2004.09.033.CrossRefPubMed
20.
go back to reference Zuhlsdorf B, Winkler A, Dietze B, Floss H, Martiny H: Gastroscope processing in washer-disinfectors at three different temperatures. J Hosp Infect. 2003, 55 (4): 276-282. 10.1016/j.jhin.2003.08.014.CrossRefPubMed Zuhlsdorf B, Winkler A, Dietze B, Floss H, Martiny H: Gastroscope processing in washer-disinfectors at three different temperatures. J Hosp Infect. 2003, 55 (4): 276-282. 10.1016/j.jhin.2003.08.014.CrossRefPubMed
21.
go back to reference Mean M, Mallaret MR, Bichard P, Shum J, Zarski JP: Gastrointestinal endoscopes cleaned without detergent substance following an automated endoscope washer/disinfector dysfunction. Gastroenterol Clin Biol. 2006, 30 (5): 665-668. 10.1016/S0399-8320(06)73258-4.CrossRefPubMed Mean M, Mallaret MR, Bichard P, Shum J, Zarski JP: Gastrointestinal endoscopes cleaned without detergent substance following an automated endoscope washer/disinfector dysfunction. Gastroenterol Clin Biol. 2006, 30 (5): 665-668. 10.1016/S0399-8320(06)73258-4.CrossRefPubMed
22.
go back to reference Alfa MJ, Olson N, DeGagne P: Automated washing with the Reliance Endoscope Processing System and its equivalence to optimal manual cleaning. Am J Infect Control. 2006, 34 (9): 561-570. 10.1016/j.ajic.2006.01.010.CrossRefPubMed Alfa MJ, Olson N, DeGagne P: Automated washing with the Reliance Endoscope Processing System and its equivalence to optimal manual cleaning. Am J Infect Control. 2006, 34 (9): 561-570. 10.1016/j.ajic.2006.01.010.CrossRefPubMed
23.
go back to reference Society of Gastrointestinal and Associates IP: SGNA Standards and Guidelines: Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes. 2009 Society of Gastrointestinal and Associates IP: SGNA Standards and Guidelines: Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes. 2009
24.
go back to reference Society of Gastrointestinal and Associates IP: SGNA Standards and Guidelines: Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes. 2007 Society of Gastrointestinal and Associates IP: SGNA Standards and Guidelines: Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes. 2007
25.
go back to reference Association for the Advancement of Medical Instrumentation: AAMI TIR30: A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices. 2003 Association for the Advancement of Medical Instrumentation: AAMI TIR30: A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices. 2003
26.
go back to reference Alfa M, DeGagne P, Olson N: Validation of ATS as an appropriate test soil. Zentr Steril (Central Service). 1999, 7: 68-382. Alfa M, DeGagne P, Olson N: Validation of ATS as an appropriate test soil. Zentr Steril (Central Service). 1999, 7: 68-382.
27.
go back to reference Alfa MJ, Degagne P, Olson N: Worst-case soiling levels for patient-used flexible endoscopes before and after cleaning. Am J Infect Control. 1999, 27 (5): 392-401. 10.1016/S0196-6553(99)70004-0.CrossRefPubMed Alfa MJ, Degagne P, Olson N: Worst-case soiling levels for patient-used flexible endoscopes before and after cleaning. Am J Infect Control. 1999, 27 (5): 392-401. 10.1016/S0196-6553(99)70004-0.CrossRefPubMed
28.
go back to reference US Department of Health and Human Services PHS, Food and Drug Administration Center for Devices and Radiologic Health: FDA Reviewer Guidance on the content and format of premarket notification 510(k) submissions for washers and washer-disinfectors. 1998 US Department of Health and Human Services PHS, Food and Drug Administration Center for Devices and Radiologic Health: FDA Reviewer Guidance on the content and format of premarket notification 510(k) submissions for washers and washer-disinfectors. 1998
29.
go back to reference Gillespie EE, Kotsanas D, Stuart RL: Microbiological monitoring of endoscopes: 5-year review. J Gastroenterol Hepatol. 2008, 23 (7 Pt 1): 1069-1074. 10.1111/j.1440-1746.2007.05264.x.CrossRefPubMed Gillespie EE, Kotsanas D, Stuart RL: Microbiological monitoring of endoscopes: 5-year review. J Gastroenterol Hepatol. 2008, 23 (7 Pt 1): 1069-1074. 10.1111/j.1440-1746.2007.05264.x.CrossRefPubMed
30.
31.
go back to reference Vickery K, Pajkos A, Cossart Y: Removal of biofilm from endoscopes: evaluation of detergent efficiency. Am J Infect Control. 2004, 32 (3): 170-176. 10.1016/j.ajic.2003.10.009.CrossRefPubMed Vickery K, Pajkos A, Cossart Y: Removal of biofilm from endoscopes: evaluation of detergent efficiency. Am J Infect Control. 2004, 32 (3): 170-176. 10.1016/j.ajic.2003.10.009.CrossRefPubMed
32.
go back to reference Marion K, Freney J, James G, Bergeron E, Renaud FN, Costerton JW: Using an efficient biofilm detaching agent: an essential step for the improvement of endoscope reprocessing protocols. J Hosp Infect. 2006, 64 (2): 136-142. 10.1016/j.jhin.2006.06.011.CrossRefPubMed Marion K, Freney J, James G, Bergeron E, Renaud FN, Costerton JW: Using an efficient biofilm detaching agent: an essential step for the improvement of endoscope reprocessing protocols. J Hosp Infect. 2006, 64 (2): 136-142. 10.1016/j.jhin.2006.06.011.CrossRefPubMed
33.
go back to reference Kircheis U, Martiny H: Comparison of the cleaning and disinfecting efficacy of four washer-disinfectors for flexible endoscopes. J Hosp Infect. 2007, 66 (3): 255-261. 10.1016/j.jhin.2007.03.020.CrossRefPubMed Kircheis U, Martiny H: Comparison of the cleaning and disinfecting efficacy of four washer-disinfectors for flexible endoscopes. J Hosp Infect. 2007, 66 (3): 255-261. 10.1016/j.jhin.2007.03.020.CrossRefPubMed
Metadata
Title
EVOTECH® endoscope cleaner and reprocessor (ECR) simulated-use and clinical-use evaluation of cleaning efficacy
Authors
Michelle J Alfa
Pat DeGagne
Nancy Olson
Iram Fatima
Publication date
01-12-2010
Publisher
BioMed Central
Published in
BMC Infectious Diseases / Issue 1/2010
Electronic ISSN: 1471-2334
DOI
https://doi.org/10.1186/1471-2334-10-200

Other articles of this Issue 1/2010

BMC Infectious Diseases 1/2010 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine