Published in: BMC Anesthesiology 1/2008

Open Access 01-12-2008 | Research article

Desflurane consumption during automated closed-circuit delivery is higher than when a conventional anesthesia machine is used with a simple vaporizer-O2-N2O fresh gas flow sequence

Authors: Sofie De Cooman, Nathalie De Mey, Bram BC Dewulf, Rik Carette, Thierry Deloof, Maurice Sosnowski, Andre M De Wolf, Jan FA Hendrickx

Published in: BMC Anesthesiology | Issue 1/2008

Login to get access

Abstract

Background

The Zeus® (Dräger, Lübeck, Germany), an automated closed-circuit anesthesia machine, uses high fresh gas flows (FGF) to wash-in the circuit and the lungs, and intermittently flushes the system to remove unwanted N2. We hypothesized this could increase desflurane consumption to such an extent that agent consumption might become higher than with a conventional anesthesia machine (Anesthesia Delivery Unit [ADU®], GE, Helsinki, Finland) used with a previously derived desflurane-O2-N2O administration schedule that allows early FGF reduction.

Methods

Thirty-four ASA PS I or II patients undergoing plastic, urologic, or gynecologic surgery received desflurane in O2/N2O. In the ADU group (n = 24), an initial 3 min high FGF of O2 and N2O (2 and 4 L.min-1, respectively) was used, followed by 0.3 L.min-1 O2 + 0.4 L.min-1 N2O. The desflurane vaporizer setting (FD) was 6.5% for the first 15 min, and 5.5% during the next 25 min. In the Zeus group (n = 10), the Zeus® was used in automated closed circuit anesthesia mode with a selected end-expired (FA) desflurane target of 4.6%, and O2/N2O as the carrier gases with a target inspired O2% of 30%. Desflurane FA and consumption during the first 40 min were compared using repeated measures one-way ANOVA.

Results

Age and weight did not differ between the groups (P > 0.05), but patients in the Zeus group were taller (P = 0.04). In the Zeus group, the desflurane FA was lower during the first 3 min (P < 0.05), identical at 4 min (P > 0.05), and slightly higher after 4 min (P < 0.05). Desflurane consumption was higher in the Zeus group at all times, a difference that persisted after correcting for the small difference in FA between the two groups.

Conclusion

Agent consumption with an automated closed-circuit anesthesia machine is higher than with a conventional anesthesia machine when the latter is used with a specific vaporizer-FGF sequence. Agent consumption during automated delivery might be further reduced by optimizing the algorithm(s) that manages the initial FGF or by tolerating some N2 in the circuit to minimize the need for intermittent flushing.
Appendix
Available only for authorised users
Literature
1.
go back to reference Lowe J, Ernst E: The quantitative practice of anesthesia – use of a closed circuit. 1981, Baltimore: Williams & Wilkins Lowe J, Ernst E: The quantitative practice of anesthesia – use of a closed circuit. 1981, Baltimore: Williams & Wilkins
2.
go back to reference Struys MMRF, Kalmar AF, De Baerdemaeker LEC, Mortier EP, Rolly G, Manigel J, Buschke W: Time course of inhaled anaesthetic drug delivery using multifunctional closed-circuit anaesthesia ventilator. In vitro comparison with a classical anaesthesia machine. Br J Anaesth. 2005, 94: 306-317. 10.1093/bja/aei051.CrossRefPubMed Struys MMRF, Kalmar AF, De Baerdemaeker LEC, Mortier EP, Rolly G, Manigel J, Buschke W: Time course of inhaled anaesthetic drug delivery using multifunctional closed-circuit anaesthesia ventilator. In vitro comparison with a classical anaesthesia machine. Br J Anaesth. 2005, 94: 306-317. 10.1093/bja/aei051.CrossRefPubMed
3.
go back to reference Hendrickx JFA, Cardinael S, Carette R, Lemmens HJ, De Wolf AM: The Ideal O2/N2O Fresh Gas Flow (FGF) Sequence with the Anesthesia Delivery Unit Machine. J Clin Anesth. 2007, 19: 274-279. 10.1016/j.jclinane.2007.01.003.CrossRefPubMed Hendrickx JFA, Cardinael S, Carette R, Lemmens HJ, De Wolf AM: The Ideal O2/N2O Fresh Gas Flow (FGF) Sequence with the Anesthesia Delivery Unit Machine. J Clin Anesth. 2007, 19: 274-279. 10.1016/j.jclinane.2007.01.003.CrossRefPubMed
4.
go back to reference Hendrickx JFA, Dewulf BBC, De Mey N, Carette R, Deloof T, De Cooman S, De Wolf AM: Development and performance of a two step desflurane – O2/N2O fresh gas flow sequence. J Clin Anesth. Hendrickx JFA, Dewulf BBC, De Mey N, Carette R, Deloof T, De Cooman S, De Wolf AM: Development and performance of a two step desflurane – O2/N2O fresh gas flow sequence. J Clin Anesth.
5.
go back to reference Hendrickx JFA, Van Zundert AAJ, De Wolf AM: Influence of the reference gas of paramagnetic oxygen analyzers on nitrogen concentrations during closed-circuit anesthesia. J Clin Monit Comp. 1998, 14: 381-84. 10.1023/A:1009969219569.CrossRef Hendrickx JFA, Van Zundert AAJ, De Wolf AM: Influence of the reference gas of paramagnetic oxygen analyzers on nitrogen concentrations during closed-circuit anesthesia. J Clin Monit Comp. 1998, 14: 381-84. 10.1023/A:1009969219569.CrossRef
6.
go back to reference Lerou JG, Verheijen R, Booij LH: Model-based administration of inhalation anaesthesia. 4. Applying the system model. Br J Anaesth. 2002, 88: 175-83. 10.1093/bja/88.2.175.CrossRefPubMed Lerou JG, Verheijen R, Booij LH: Model-based administration of inhalation anaesthesia. 4. Applying the system model. Br J Anaesth. 2002, 88: 175-83. 10.1093/bja/88.2.175.CrossRefPubMed
Metadata
Title
Desflurane consumption during automated closed-circuit delivery is higher than when a conventional anesthesia machine is used with a simple vaporizer-O2-N2O fresh gas flow sequence
Authors
Sofie De Cooman
Nathalie De Mey
Bram BC Dewulf
Rik Carette
Thierry Deloof
Maurice Sosnowski
Andre M De Wolf
Jan FA Hendrickx
Publication date
01-12-2008
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2008
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/1471-2253-8-4

Other articles of this Issue 1/2008

BMC Anesthesiology 1/2008 Go to the issue