Open Access 01-12-2007 | Research article

Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration

Authors: Christopher Prasser, Sylvia Bele, Cornelius Keyl, Stefan Schweiger, Benedikt Trabold, Matthias Amann, Julia Welnhofer, Christoph Wiesenack

Published in: BMC Anesthesiology | Issue 1/2007

Login to get access

Abstract

Background

Several techniques have been discussed as alternatives to the intermittent bolus thermodilution cardiac output (COPAC) measurement by the pulmonary artery catheter (PAC). However, these techniques usually require a central venous line, an additional catheter, or a special calibration procedure. A new arterial pressure-based cardiac output (COAP) device (FloTrac™, Vigileo™; Edwards Lifesciences, Irvine, CA, USA) only requires access to the radial or femoral artery using a standard arterial catheter and does not need an external calibration. We validated this technique in critically ill patients in the intensive care unit (ICU) using COPAC as the method of reference.

Methods

We studied 20 critically ill patients, aged 16 to 74 years (mean, 55.5 ± 18.8 years), who required both arterial and pulmonary artery pressure monitoring. COPAC measurements were performed at least every 4 hours and calculated as the average of 3 measurements, while COAP values were taken immediately at the end of bolus determinations. Accuracy of measurements was assessed by calculating the bias and limits of agreement using the method described by Bland and Altman.

Results

A total of 164 coupled measurements were obtained. Absolute values of COPAC ranged from 2.80 to 10.80 l/min (mean 5.93 ± 1.55 l/min). The bias and limits of agreement between COPAC and COAP for unequal numbers of replicates was 0.02 ± 2.92 l/min. The percentage error between COPAC and COAP was 49.3%. The bias between percentage changes in COPAC (ΔCOPAC) and percentage changes in COAP (ΔCOAP) for consecutive measurements was -0.70% ± 32.28%. COPAC and COAP showed a Pearson correlation coefficient of 0.58 (p < 0.01), while the correlation coefficient between ΔCOPAC and ΔCOAP was 0.46 (p < 0.01).

Conclusion

Although the COAP algorithm shows a minimal bias with COPAC over a wide range of values in an inhomogeneous group of critically ill patients, the scattering of the data remains relative wide. Therefore, the used algorithm (V 1.03) failed to demonstrate an acceptable accuracy in comparison to the clinical standard of cardiac output determination.
Appendix
Available only for authorised users
Literature
1.
go back to reference Connors AF, Speroff T, Dawson NV, Thomas C, Harrell FE, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart catheterization in the initial care of critically ill patients. JAMA. 1996, 276: 889-897.CrossRefPubMed Connors AF, Speroff T, Dawson NV, Thomas C, Harrell FE, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart catheterization in the initial care of critically ill patients. JAMA. 1996, 276: 889-897.CrossRefPubMed
2.
go back to reference Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M, Canadian Critical Care Clinical Trials Group: A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med. 2003, 348: 5-14.CrossRefPubMed Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M, Canadian Critical Care Clinical Trials Group: A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med. 2003, 348: 5-14.CrossRefPubMed
3.
go back to reference Cholley BP, Payen D: Noninvasive techniques for measurements of cardiac output. Curr Opin Crit Care. 2005, 11: 424-429.CrossRefPubMed Cholley BP, Payen D: Noninvasive techniques for measurements of cardiac output. Curr Opin Crit Care. 2005, 11: 424-429.CrossRefPubMed
4.
go back to reference Bein B, Worthmann F, Tonner PH, Paris A, Steinfath M, Hedderich J, Scholz J: Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output. J Cardiothorac Vasc Anesth. 2004, 18: 185-189.CrossRefPubMed Bein B, Worthmann F, Tonner PH, Paris A, Steinfath M, Hedderich J, Scholz J: Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output. J Cardiothorac Vasc Anesth. 2004, 18: 185-189.CrossRefPubMed
5.
go back to reference Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P: Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth. 2003, 50: 707-711.CrossRefPubMed Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P: Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth. 2003, 50: 707-711.CrossRefPubMed
6.
go back to reference Felbinger TW, Reuter DA, Eltzschig HK, Moerstedt K, Goedje O, Goetz AE: Comparison of pulmonary arterial thermodilution and arterial pulse contour analysis: evaluation of a new algorithm. J Clin Anesth. 2002, 14: 296-301.CrossRefPubMed Felbinger TW, Reuter DA, Eltzschig HK, Moerstedt K, Goedje O, Goetz AE: Comparison of pulmonary arterial thermodilution and arterial pulse contour analysis: evaluation of a new algorithm. J Clin Anesth. 2002, 14: 296-301.CrossRefPubMed
7.
go back to reference Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg. 2002, 74: 1408-1412.CrossRef Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg. 2002, 74: 1408-1412.CrossRef
8.
go back to reference Jonas MM, Tanser SJ: Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. Curr Opin Crit Care. 2002, 8: 257-261.CrossRefPubMed Jonas MM, Tanser SJ: Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. Curr Opin Crit Care. 2002, 8: 257-261.CrossRefPubMed
9.
go back to reference Pittman J, Bar Yosef S, SumPing J, Sherwood M, Mark J: Continuous cardiac output monitoring with pulse contour analysis: A comparison with lithium indicator dilution cardiac output measurement. Crit Care Med. 2005, 33: 2015-2021.CrossRefPubMed Pittman J, Bar Yosef S, SumPing J, Sherwood M, Mark J: Continuous cardiac output monitoring with pulse contour analysis: A comparison with lithium indicator dilution cardiac output measurement. Crit Care Med. 2005, 33: 2015-2021.CrossRefPubMed
10.
go back to reference Rauch H, Muller M, Fleischer F, Bauer H, Martin E, Bottiger BW: Pulse contour analysis versus thermodilution in cardiac surgery patients. Acta Anaesthesiol Scand. 2002, 46: 424-429.CrossRefPubMed Rauch H, Muller M, Fleischer F, Bauer H, Martin E, Bottiger BW: Pulse contour analysis versus thermodilution in cardiac surgery patients. Acta Anaesthesiol Scand. 2002, 46: 424-429.CrossRefPubMed
11.
go back to reference Rödig G, Prasser C, Keyl C, Liebold A, Hobbhahn J: Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth. 1999, 82: 525-530.CrossRefPubMed Rödig G, Prasser C, Keyl C, Liebold A, Hobbhahn J: Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth. 1999, 82: 525-530.CrossRefPubMed
12.
go back to reference Wiesenack C, Fiegl C, Keyser A, Prasser C, Keyl C: Assessment of fluid responsiveness in mechanically ventilated cardiac surgical patients. Eur J Anaesth. 2005, 22: 658-665.CrossRef Wiesenack C, Fiegl C, Keyser A, Prasser C, Keyl C: Assessment of fluid responsiveness in mechanically ventilated cardiac surgical patients. Eur J Anaesth. 2005, 22: 658-665.CrossRef
13.
go back to reference Langewouters GJ, Wesseling KH, Goedhard WJ: The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component model. J Biomech. 1985, 18: 613-620.CrossRefPubMed Langewouters GJ, Wesseling KH, Goedhard WJ: The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component model. J Biomech. 1985, 18: 613-620.CrossRefPubMed
14.
go back to reference Mayer J, Boldt J, Schöllhorn T, Röhm KD, Mengistu AM, Suttner S: Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth. 2007, 98: 176-182.CrossRefPubMed Mayer J, Boldt J, Schöllhorn T, Röhm KD, Mengistu AM, Suttner S: Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth. 2007, 98: 176-182.CrossRefPubMed
15.
go back to reference Opdam HI, Wan L, Bellomo R: A pilot assessment of the FloTrac™ cardiac output monitoring system. Intensive Care Med. 2006 Oct 25. Opdam HI, Wan L, Bellomo R: A pilot assessment of the FloTrac™ cardiac output monitoring system. Intensive Care Med. 2006 Oct 25.
16.
go back to reference Sander M, Spies CD, Grubitzsch H, Foer A, Müller M, von Heymann C: Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care. 2006, 10: R164-CrossRefPubMedPubMedCentral Sander M, Spies CD, Grubitzsch H, Foer A, Müller M, von Heymann C: Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care. 2006, 10: R164-CrossRefPubMedPubMedCentral
17.
go back to reference Manecke GR: Edwards FloTrac™sensor and Vigileo™monitor: easy, accurate, reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices. 2005, 2: 523-527.CrossRefPubMed Manecke GR: Edwards FloTrac™sensor and Vigileo™monitor: easy, accurate, reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices. 2005, 2: 523-527.CrossRefPubMed
18.
go back to reference Boulain T, Achard J-M, Teboul J-L, Richard C, Perrotin D, Ginies G: Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest. 2002, 121: 1245-1252.CrossRefPubMed Boulain T, Achard J-M, Teboul J-L, Richard C, Perrotin D, Ginies G: Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest. 2002, 121: 1245-1252.CrossRefPubMed
19.
go back to reference Guyton AC, Hall JE: Vascular distensibility and functions of the arterial and venous systems. Textbook of medical physiology. Edited by: Guyton AC, Hall JE. 2000, Philadelphia: Saunders, 152-161. Guyton AC, Hall JE: Vascular distensibility and functions of the arterial and venous systems. Textbook of medical physiology. Edited by: Guyton AC, Hall JE. 2000, Philadelphia: Saunders, 152-161.
20.
go back to reference Headley JM: Arterial pressure-based technologies: a new trend in cardiac output monitoring. Crit Care Nurs Clin N Am. 2006, 18: 179-187.CrossRef Headley JM: Arterial pressure-based technologies: a new trend in cardiac output monitoring. Crit Care Nurs Clin N Am. 2006, 18: 179-187.CrossRef
21.
go back to reference Cecconi M, Wilson J, Rhodes A: Pulse Pressure Analysis. I Yearbook of intensive care and emergency medicine. Edited by: Vincent JL. 2006, Berlin, Heidelberg, New York: Springer, 176-183.CrossRef Cecconi M, Wilson J, Rhodes A: Pulse Pressure Analysis. I Yearbook of intensive care and emergency medicine. Edited by: Vincent JL. 2006, Berlin, Heidelberg, New York: Springer, 176-183.CrossRef
22.
go back to reference Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8: 135-160.CrossRefPubMed Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8: 135-160.CrossRefPubMed
23.
go back to reference Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999, 15: 85-91.CrossRefPubMed Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999, 15: 85-91.CrossRefPubMed
24.
go back to reference Taylor SH, Silke B: Is the measurement of cardiac output useful in the clinical practice?. Br J Anaesth. 1988, 60: 90-98.CrossRef Taylor SH, Silke B: Is the measurement of cardiac output useful in the clinical practice?. Br J Anaesth. 1988, 60: 90-98.CrossRef
25.
go back to reference Nishikawa T, Dohi S: Errors in the measurement of cardiac output by thermodilution. Can J Anaesth. 1993, 40: 142-153.CrossRefPubMed Nishikawa T, Dohi S: Errors in the measurement of cardiac output by thermodilution. Can J Anaesth. 1993, 40: 142-153.CrossRefPubMed
26.
go back to reference Stetz CW, Miller RG, Kelly GE, Raffin TA: Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Resp Dis. 1982, 126: 1001-1004.PubMed Stetz CW, Miller RG, Kelly GE, Raffin TA: Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Resp Dis. 1982, 126: 1001-1004.PubMed
Metadata
Title
Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration
Authors
Christopher Prasser
Sylvia Bele
Cornelius Keyl
Stefan Schweiger
Benedikt Trabold
Matthias Amann
Julia Welnhofer
Christoph Wiesenack
Publication date
01-12-2007
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2007
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/1471-2253-7-9

Other articles of this Issue 1/2007

BMC Anesthesiology 1/2007 Go to the issue