In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth

Authors: L. Z. Adriano, J. C. Barasuol, M. Cardoso, M. Bolan

Published in: European Archives of Paediatric Dentistry | Issue 5/2019

Login to get access

Abstract

Aim

To compare the accuracy of three electronic apex locators in the measurement of the root length of primary teeth with the radiographic and direct visual techniques.

Methods

The sample of the in vitro study was 102 root canals of primary molars. For the direct visual technique, the root canal file was introduced into the canal until its tip passed the apical foramen, then it was moved back to the limit of the foramen or root resorption and the measurement was made with a millimetre ruler. For the radiographic measurement, a calibrated and blinded researcher (κ > 0.87) measured the root canals using the same rule. Measurements with apex locators were performed by inserting the file into the canal until the word “Apex” or audible warning to indicate the foramen or resorption. For data analysis, one-way ANOVA for repeated measurements and the Sidak post hoc tests were performed.

Results

The direct measurement (µ = 8.57) did not present statistical difference when compared to the Root ZX II measurement (µ = 8.45), as opposed to the radiographic measurements (µ = 8.12), the Endus (µ = 8.24) and iPex II (µ = 8.29).

Conclusion

The use of apex locators was superior to the radiographic method and the Root ZX II was the most effective.
Literature
go back to reference Abdullah A, Singh N, Rathore MS, Tandon S, Rajkumar B. Comparative evaluation of electronic apex locators and radiovisiography for working length determination in primary teeth in vivo. Int J Clin Pediatr Dent. 2016;2(9):118–23. Abdullah A, Singh N, Rathore MS, Tandon S, Rajkumar B. Comparative evaluation of electronic apex locators and radiovisiography for working length determination in primary teeth in vivo. Int J Clin Pediatr Dent. 2016;2(9):118–23.
go back to reference Ahmad IA, Pani SC. Accuracy of electronic apex locators in primary teeth: a meta-analysis. Int Endod J. 2015;48(3):298–307.PubMedCrossRef Ahmad IA, Pani SC. Accuracy of electronic apex locators in primary teeth: a meta-analysis. Int Endod J. 2015;48(3):298–307.PubMedCrossRef
go back to reference Ahmed HM. Anatomical challenges, electronic working length determination and current developments in root canal preparation of primary molar teeth. Int Endod J. 2013;46(11):1011–22.PubMed Ahmed HM. Anatomical challenges, electronic working length determination and current developments in root canal preparation of primary molar teeth. Int Endod J. 2013;46(11):1011–22.PubMed
go back to reference Angwaravong O, Panitvisai P. Accuracy of an electronic apex locator in primary teeth with root resorption. Int Endod J. 2009;42(2):115–21.PubMedCrossRef Angwaravong O, Panitvisai P. Accuracy of an electronic apex locator in primary teeth with root resorption. Int Endod J. 2009;42(2):115–21.PubMedCrossRef
go back to reference Beltrame AP, Triches TC, Sartori N, Bolan M. Electronic determination of root canal working length in primary molar teeth: an in vivo and ex vivo study. Int Endod J. 2011;44(5):402–6.PubMedCrossRef Beltrame AP, Triches TC, Sartori N, Bolan M. Electronic determination of root canal working length in primary molar teeth: an in vivo and ex vivo study. Int Endod J. 2011;44(5):402–6.PubMedCrossRef
go back to reference Bhat VK, Shetty P, Anandakrishna L. A Comparative evaluation of accuracy of new-generation electronic apex locator with conventional radiography to determine working length in primary teeth: an in vivo study. Int J Clin Pediatr Dent. 2017;10(1):34–6.PubMedPubMedCentralCrossRef Bhat VK, Shetty P, Anandakrishna L. A Comparative evaluation of accuracy of new-generation electronic apex locator with conventional radiography to determine working length in primary teeth: an in vivo study. Int J Clin Pediatr Dent. 2017;10(1):34–6.PubMedPubMedCentralCrossRef
go back to reference Borges MMB, Guimarães BM, Alves JD, et al. Evaluation of the accuracy of two apex locators in apical limit: an in vitro study. Rev Odontol Bras Central. 2016;25(74):126–9. Borges MMB, Guimarães BM, Alves JD, et al. Evaluation of the accuracy of two apex locators in apical limit: an in vitro study. Rev Odontol Bras Central. 2016;25(74):126–9.
go back to reference Carvalho ANL, Moura-Netto C, Moura AAM, Marques MM, Davidowicz H. Accuracy of three electronic apex locators in the presence of different irrigating solutions. Braz Oral Res. 2010;24(4):394–8.PubMedCrossRef Carvalho ANL, Moura-Netto C, Moura AAM, Marques MM, Davidowicz H. Accuracy of three electronic apex locators in the presence of different irrigating solutions. Braz Oral Res. 2010;24(4):394–8.PubMedCrossRef
go back to reference Chen X, Liu X, Zhong J. Clinical and radiographic evaluation of pulpectomy in primary teeth: a 18-months clinical randomized controlled trial. Head Face Med. 2017;13(1):12.PubMedPubMedCentralCrossRef Chen X, Liu X, Zhong J. Clinical and radiographic evaluation of pulpectomy in primary teeth: a 18-months clinical randomized controlled trial. Head Face Med. 2017;13(1):12.PubMedPubMedCentralCrossRef
go back to reference Dandempally A, Muppa R, Duddu MK, Bhupatiraju P, Nallanchakrava S. Formulating a regression equation for determination of working length in primary molars using apex locators: a clinical study. Eur Arch Paediatr Dent. 2013;14(6):369–74.PubMedCrossRef Dandempally A, Muppa R, Duddu MK, Bhupatiraju P, Nallanchakrava S. Formulating a regression equation for determination of working length in primary molars using apex locators: a clinical study. Eur Arch Paediatr Dent. 2013;14(6):369–74.PubMedCrossRef
go back to reference Govindaraju L, Jeevanandan G, Subramanian EMG. Comparison of quality of obturation and instrumentation time using hand files and two rotary file systems in primary molars: a single-blinded randomized controlled trial. Eur J Dent. 2017;11(3):376–9.PubMedPubMedCentralCrossRef Govindaraju L, Jeevanandan G, Subramanian EMG. Comparison of quality of obturation and instrumentation time using hand files and two rotary file systems in primary molars: a single-blinded randomized controlled trial. Eur J Dent. 2017;11(3):376–9.PubMedPubMedCentralCrossRef
go back to reference Jenkins JA, Walker WA, Schindler WG, Flores CM. An in vitro evaluation of the accuracy of the Root ZX in the presence of various irrigants. J Endod. 2001;27(3):209–11.PubMedCrossRef Jenkins JA, Walker WA, Schindler WG, Flores CM. An in vitro evaluation of the accuracy of the Root ZX in the presence of various irrigants. J Endod. 2001;27(3):209–11.PubMedCrossRef
go back to reference Kielbassa AM, Muller U, Munz I, Monting JS. Clinical evaluation of the measuring accuracy of ROOT ZX in primary teeth. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;95(1):94–100.PubMedCrossRef Kielbassa AM, Muller U, Munz I, Monting JS. Clinical evaluation of the measuring accuracy of ROOT ZX in primary teeth. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;95(1):94–100.PubMedCrossRef
go back to reference Kumar LV, Sreelakshmi N, Reddy ER, et al. Clinical evaluation of conventional radiography, radiovisiography, and an electronic apex locator in determining the working length in primary teeth. Pediatr Dent. 2016;38(1):37–41.PubMed Kumar LV, Sreelakshmi N, Reddy ER, et al. Clinical evaluation of conventional radiography, radiovisiography, and an electronic apex locator in determining the working length in primary teeth. Pediatr Dent. 2016;38(1):37–41.PubMed
go back to reference Leonardo MR, Silva LA, Nelson-Filho P, Silva RA, Raffaini MS. Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth. Int Endod J. 2008;41(4):317–21.PubMedCrossRef Leonardo MR, Silva LA, Nelson-Filho P, Silva RA, Raffaini MS. Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth. Int Endod J. 2008;41(4):317–21.PubMedCrossRef
go back to reference Mello-Moura AC, Moura-Netto C, Araki AT, Guedes-Pinto AC, Mendes FM. Ex vivo performance of five methods for root canal length determination in primary anterior teeth. Int Endod J. 2010;43(2):142–7.PubMedCrossRef Mello-Moura AC, Moura-Netto C, Araki AT, Guedes-Pinto AC, Mendes FM. Ex vivo performance of five methods for root canal length determination in primary anterior teeth. Int Endod J. 2010;43(2):142–7.PubMedCrossRef
go back to reference Mente J, Seidel J, Buchalla W, Koch MJ. Electronic determination of root canal length in primary teeth with and without root resorption. Int Endod J. 2002;35(5):447–52.PubMedCrossRef Mente J, Seidel J, Buchalla W, Koch MJ. Electronic determination of root canal length in primary teeth with and without root resorption. Int Endod J. 2002;35(5):447–52.PubMedCrossRef
go back to reference Navit S, Jaiswal N, Khan SA, et al. Antimicrobial efficacy of contemporary obturating materials used in primary teeth—an in-vitro study. journal of clinical and diagnostic research. J Clin Diagn Res. 2016;10(9):ZC09–12.PubMedPubMedCentral Navit S, Jaiswal N, Khan SA, et al. Antimicrobial efficacy of contemporary obturating materials used in primary teeth—an in-vitro study. journal of clinical and diagnostic research. J Clin Diagn Res. 2016;10(9):ZC09–12.PubMedPubMedCentral
go back to reference Nelson-Filho P, Romualdo PC, Bonifácio KC, et al. Accuracy of the iPex multi-frequency electronic apex locator in primary molars: an ex vivo study. Int Endod J. 2011;44(4):303–6.PubMedCrossRef Nelson-Filho P, Romualdo PC, Bonifácio KC, et al. Accuracy of the iPex multi-frequency electronic apex locator in primary molars: an ex vivo study. Int Endod J. 2011;44(4):303–6.PubMedCrossRef
go back to reference Oznurhan F, Ünal M, Kapdan A, Ozturk C, Aksoy S. Clinical evaluation of apex locator and radiography in primary teeth. Int J Paediatr Dent. 2015;25(3):199–203.PubMedCrossRef Oznurhan F, Ünal M, Kapdan A, Ozturk C, Aksoy S. Clinical evaluation of apex locator and radiography in primary teeth. Int J Paediatr Dent. 2015;25(3):199–203.PubMedCrossRef
go back to reference Parisay I, Ghoddusi J, Forghanib M. A review on vital pulp therapy in primary teeth. Iran Endod J. 2015;10(1):6–15.PubMed Parisay I, Ghoddusi J, Forghanib M. A review on vital pulp therapy in primary teeth. Iran Endod J. 2015;10(1):6–15.PubMed
go back to reference Patiño-Marín N, Zavala-Alonso NV, Martínez-Castañón GA, et al. Clinical evaluation of the accuracy of conventional radiography and apex locators in primary teeth. Pediatr Dent. 2011;33(1):19–22.PubMed Patiño-Marín N, Zavala-Alonso NV, Martínez-Castañón GA, et al. Clinical evaluation of the accuracy of conventional radiography and apex locators in primary teeth. Pediatr Dent. 2011;33(1):19–22.PubMed
go back to reference Saritha S, Uloopi KS, Vinay C, Chandra SR, Rao VV. Clinical evaluation of Root ZX II electronic apex locator in primary teeth. Eur Arch Paediatr Dent. 2012;13(1):32–5.PubMedCrossRef Saritha S, Uloopi KS, Vinay C, Chandra SR, Rao VV. Clinical evaluation of Root ZX II electronic apex locator in primary teeth. Eur Arch Paediatr Dent. 2012;13(1):32–5.PubMedCrossRef
go back to reference Tosun G, Erdemir A, Eldeniz AU, Sermet U, Sener Y. Accuracy of two electronic apex locators in primary teeth with and without apical resorption: a laboratory study. Int Endod J. 2008;41(5):436–41.PubMedCrossRef Tosun G, Erdemir A, Eldeniz AU, Sermet U, Sener Y. Accuracy of two electronic apex locators in primary teeth with and without apical resorption: a laboratory study. Int Endod J. 2008;41(5):436–41.PubMedCrossRef
go back to reference Vasconcelos BC, Bueno MM, Luna-Cruz SM, Duarte MA, Fernandes CA. Accuracy of five electronic foramen locators with different operating systems: an ex vivo study. J Appl Oral Sci. 2013;21(2):132–7.PubMedPubMedCentralCrossRef Vasconcelos BC, Bueno MM, Luna-Cruz SM, Duarte MA, Fernandes CA. Accuracy of five electronic foramen locators with different operating systems: an ex vivo study. J Appl Oral Sci. 2013;21(2):132–7.PubMedPubMedCentralCrossRef
Metadata
Title
In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth
Authors
L. Z. Adriano
J. C. Barasuol
M. Cardoso
M. Bolan
Publication date
01-10-2019
Publisher
Springer Berlin Heidelberg
Published in
European Archives of Paediatric Dentistry / Issue 5/2019
Print ISSN: 1818-6300
Electronic ISSN: 1996-9805
DOI
https://doi.org/10.1007/s40368-018-00413-5

Other articles of this Issue 5/2019

European Archives of Paediatric Dentistry 5/2019 Go to the issue