Published in: Sports Medicine 3/2019

01-03-2019 | Electrocardiography | Systematic Review

The Accuracy of Acquiring Heart Rate Variability from Portable Devices: A Systematic Review and Meta-Analysis

Authors: Ward C. Dobbs, Michael V. Fedewa, Hayley V. MacDonald, Clifton J. Holmes, Zackary S. Cicone, Daniel J. Plews, Michael R. Esco

Published in: Sports Medicine | Issue 3/2019

Login to get access

Abstract

Background

Advancements in wearable technology have provided practitioners and researchers with the ability to conveniently measure various health and/or fitness indices. Specifically, portable devices have been devised for convenient recordings of heart rate variability (HRV). Yet, their accuracies remain questionable.

Objective

The aim was to quantify the accuracy of portable devices compared to electrocardiography (ECG) for measuring a multitude of HRV metrics and to identify potential moderators of this effect.

Methods

This meta-analysis was conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement. Articles published before July 29, 2017 were located via four electronic databases using a combination of the terms related to HRV and validity. Separate effect sizes (ESs), defined as the absolute standardized difference between the HRV value recorded using the portable device compared to ECG, were generated for each HRV metric (ten metrics analyzed in total). A multivariate, multi-level model, incorporating random-effects assumptions, was utilized to quantify the mean ES and 95% confidence interval (CI) and explore potential moderators.

Results

Twenty-three studies yielded 301 effects and revealed that HRV measurements acquired from portable devices differed from those obtained from ECG (ES = 0.23, 95% CI 0.05–0.42), although this effect was small and highly heterogeneous (I2 = 78.6%, 95% CI 76.2–80.7). Moderator analysis revealed that HRV metric (p <0.001), position (p = 0.033), and biological sex (β = 0.45, 95% CI 0.30–0.61; p <0.001), but not portable device, modulated the degree of absolute error. Within metric, absolute error was significantly higher when expressed as standard deviation of all normal–normal (R–R) intervals (SDNN) (ES = 0.44) compared to any other metric, but was no longer significantly different after a sensitivity analysis removed outliers. Likewise, the error associated with the tilt/recovery position was significantly higher than any other position and remained significantly different without outliers in the model.

Conclusions

Our results suggest that HRV measurements acquired using portable devices demonstrate a small amount of absolute error when compared to ECG. However, this small error is acceptable when considering the improved practicality and compliance of HRV measurements acquired through portable devices in the field setting. Practitioners and researchers should consider the cost–benefit along with the simplicity of the measurement when attempting to increase compliance in acquiring HRV measurements.
Appendix
Available only for authorised users
Literature
1.
go back to reference Thompson WR. Worldwide survey of fitness trends for 2017. ACSMs Health Fit J. 2016;20(6):8–17. Thompson WR. Worldwide survey of fitness trends for 2017. ACSMs Health Fit J. 2016;20(6):8–17.
2.
go back to reference Thompson WR. Worldwide survey of fitness trends for 2018: the CREP edition. ACSMs Health Fit J. 2017;21(6):10–9.CrossRef Thompson WR. Worldwide survey of fitness trends for 2018: the CREP edition. ACSMs Health Fit J. 2017;21(6):10–9.CrossRef
3.
go back to reference Plews DJ, Laursen PB, Stanley J, Kilding AE, Buchheit M. Training adaptation and heart rate variability in elite endurance athletes: opening the door to effective monitoring. Sports Med. 2013;43(9):773–81.PubMedCrossRef Plews DJ, Laursen PB, Stanley J, Kilding AE, Buchheit M. Training adaptation and heart rate variability in elite endurance athletes: opening the door to effective monitoring. Sports Med. 2013;43(9):773–81.PubMedCrossRef
4.
go back to reference Silva VP, Oliveira NA, Silveira H, Mello RGT, Deslandes AC. Heart rate variability indexes as a marker of chronic adaptation in athletes: a systematic review. Ann Noninvasive Electrocardiol. 2015;20(2):108–18.PubMedCrossRef Silva VP, Oliveira NA, Silveira H, Mello RGT, Deslandes AC. Heart rate variability indexes as a marker of chronic adaptation in athletes: a systematic review. Ann Noninvasive Electrocardiol. 2015;20(2):108–18.PubMedCrossRef
5.
go back to reference Malik M, Camm AJ, Bigger JT Jr, Breithardt G, Cerutti S, Cohen RJ, et al. Heart rate variability. Standards of measurement, physiological interpretation, and clinical use. Eur Heart J. 1996;17(3):354–81.CrossRef Malik M, Camm AJ, Bigger JT Jr, Breithardt G, Cerutti S, Cohen RJ, et al. Heart rate variability. Standards of measurement, physiological interpretation, and clinical use. Eur Heart J. 1996;17(3):354–81.CrossRef
6.
go back to reference Radespiel-Tröger M, Rauh R, Mahlke C, Gottschalk T, Muck-Weymann M. Agreement of two different methods for measurement of heart rate variability. Clin Auton Res. 2003;13(2):99–102.PubMedCrossRef Radespiel-Tröger M, Rauh R, Mahlke C, Gottschalk T, Muck-Weymann M. Agreement of two different methods for measurement of heart rate variability. Clin Auton Res. 2003;13(2):99–102.PubMedCrossRef
7.
go back to reference Flatt AA, Esco MR, Nakamura FY, Plews DJ. Interpreting daily heart rate variability changes in collegiate female soccer players. J Sports Med Phys Fitness. 2017;57(6):907–15.PubMed Flatt AA, Esco MR, Nakamura FY, Plews DJ. Interpreting daily heart rate variability changes in collegiate female soccer players. J Sports Med Phys Fitness. 2017;57(6):907–15.PubMed
8.
go back to reference Flatt AA, Esco MR. Evaluating individual training adaptation with smartphone-derived heart rate variability in a collegiate female soccer team. J Strength Cond Res. 2016;30(2):378–85.PubMedCrossRef Flatt AA, Esco MR. Evaluating individual training adaptation with smartphone-derived heart rate variability in a collegiate female soccer team. J Strength Cond Res. 2016;30(2):378–85.PubMedCrossRef
9.
go back to reference Esco MR, Flatt AA, Nakamura FY. Initial weekly HRV response is related to the prospective change in VO2max in female soccer players. Int J Sports Med. 2016;37(6):436–41.PubMedCrossRef Esco MR, Flatt AA, Nakamura FY. Initial weekly HRV response is related to the prospective change in VO2max in female soccer players. Int J Sports Med. 2016;37(6):436–41.PubMedCrossRef
10.
go back to reference Nakamura FY, Pereira LA, Esco MR, Flatt AA, Moraes JE, Abad CCC, et al. Intraday and interday reliability of ultra-short-term heart rate variability in rugby union players. J Strength Cond Res. 2017;31(2):548–51.PubMed Nakamura FY, Pereira LA, Esco MR, Flatt AA, Moraes JE, Abad CCC, et al. Intraday and interday reliability of ultra-short-term heart rate variability in rugby union players. J Strength Cond Res. 2017;31(2):548–51.PubMed
12.
go back to reference Flatt AA, Esco MR. Heart rate variability stabilization in athletes: towards more convenient data acquisition. Clin Physiol Funct Imaging. 2016;36(5):331–6.PubMedCrossRef Flatt AA, Esco MR. Heart rate variability stabilization in athletes: towards more convenient data acquisition. Clin Physiol Funct Imaging. 2016;36(5):331–6.PubMedCrossRef
13.
go back to reference Berntson GG, Bigger JT Jr, Eckberg DL, Grossman P, Kaufmann PG, Malik M, et al. Heart rate variability: origins, methods, and interpretive caveats. Psychophysiology. 1997;34(6):623–48.PubMedCrossRef Berntson GG, Bigger JT Jr, Eckberg DL, Grossman P, Kaufmann PG, Malik M, et al. Heart rate variability: origins, methods, and interpretive caveats. Psychophysiology. 1997;34(6):623–48.PubMedCrossRef
14.
go back to reference Board L, Ispoglou T, Ingle L. Validity of telemetric-derived measures of heart rate variability: a systematic review. J Exerc Physiol Online. 2016;19(6):64–84. Board L, Ispoglou T, Ingle L. Validity of telemetric-derived measures of heart rate variability: a systematic review. J Exerc Physiol Online. 2016;19(6):64–84.
15.
go back to reference Barbosa MP, da Silva NT, de Azevedo FM, Pastre CM, Vanderlei LC. Comparison of Polar(R) RS800G3 heart rate monitor with Polar(R) S810i and electrocardiogram to obtain the series of RR intervals and analysis of heart rate variability at rest. Clin Physiol Funct Imaging. 2016;36(2):112–7.PubMedCrossRef Barbosa MP, da Silva NT, de Azevedo FM, Pastre CM, Vanderlei LC. Comparison of Polar(R) RS800G3 heart rate monitor with Polar(R) S810i and electrocardiogram to obtain the series of RR intervals and analysis of heart rate variability at rest. Clin Physiol Funct Imaging. 2016;36(2):112–7.PubMedCrossRef
16.
go back to reference Bolkhovsky JB, Scully CG, Chon KH, editors. Statistical analysis of heart rate and heart rate variability monitoring through the use of smart phone cameras. In: Engineering in Medicine and Biology Society (EMBC), 2012 annual international conference of the IEEE; 2012: IEEE. Bolkhovsky JB, Scully CG, Chon KH, editors. Statistical analysis of heart rate and heart rate variability monitoring through the use of smart phone cameras. In: Engineering in Medicine and Biology Society (EMBC), 2012 annual international conference of the IEEE; 2012: IEEE.
17.
go back to reference Wallen MB, Hasson D, Theorell T, Canlon B, Osika W. Possibilities and limitations of the Polar RS800 in measuring heart rate variability at rest. Eur J Appl Physiol. 2012;112(3):1153–65.PubMedCrossRef Wallen MB, Hasson D, Theorell T, Canlon B, Osika W. Possibilities and limitations of the Polar RS800 in measuring heart rate variability at rest. Eur J Appl Physiol. 2012;112(3):1153–65.PubMedCrossRef
18.
go back to reference Hernando D, Garatachea N, Almeida R, Casajús JA, Bailón R. Validation of heart rate monitor Polar RS800 for heart rate variability analysis during exercise. J Strength Cond Res. 2018;32(3):716–25.PubMed Hernando D, Garatachea N, Almeida R, Casajús JA, Bailón R. Validation of heart rate monitor Polar RS800 for heart rate variability analysis during exercise. J Strength Cond Res. 2018;32(3):716–25.PubMed
19.
go back to reference Kingsley M, Lewis MJ, Marson RE. Comparison of Polar 810 s and an ambulatory ECG System for RR interval measures during progressive exercise. Int J Sports Med. 2005;26(1):39–44.PubMedCrossRef Kingsley M, Lewis MJ, Marson RE. Comparison of Polar 810 s and an ambulatory ECG System for RR interval measures during progressive exercise. Int J Sports Med. 2005;26(1):39–44.PubMedCrossRef
20.
go back to reference Romagnoli M, Alis R, Guillen J, Basterra J, Villacastin J, Guillen S. A novel device based on smart textile to control heart’s activity during exercise. Australas Phys Eng Sci Med. 2014;37(2):377–84.PubMedCrossRef Romagnoli M, Alis R, Guillen J, Basterra J, Villacastin J, Guillen S. A novel device based on smart textile to control heart’s activity during exercise. Australas Phys Eng Sci Med. 2014;37(2):377–84.PubMedCrossRef
21.
go back to reference Weippert M, Kumar M, Kreuzfeld S, Arndt D, Rieger A, Stoll R. Comparison of three mobile devices for measuring R-R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system. Eur J Appl Physiol. 2010;109(4):779–86.PubMedCrossRef Weippert M, Kumar M, Kreuzfeld S, Arndt D, Rieger A, Stoll R. Comparison of three mobile devices for measuring R-R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system. Eur J Appl Physiol. 2010;109(4):779–86.PubMedCrossRef
22.
go back to reference Chellakumar PJ, Brumfield A, Kunderu K, Schopper AW. Heart rate variability: comparison among devices with different temporal resolutions. Physiol Meas. 2005;26(6):979–86.PubMedCrossRef Chellakumar PJ, Brumfield A, Kunderu K, Schopper AW. Heart rate variability: comparison among devices with different temporal resolutions. Physiol Meas. 2005;26(6):979–86.PubMedCrossRef
23.
go back to reference Russoniello CV, Zhirnov YN, Pougatchev VI, Gribkov EN. Heart rate variability and biological age: implications for health and gaming. Cyberpsychol Behav Soc Netw. 2013;16(4):302–8.PubMedCrossRef Russoniello CV, Zhirnov YN, Pougatchev VI, Gribkov EN. Heart rate variability and biological age: implications for health and gaming. Cyberpsychol Behav Soc Netw. 2013;16(4):302–8.PubMedCrossRef
24.
go back to reference Esco MR, Flatt AA, Nakamura FY. Agreement between a smartphone pulse sensor application and electrocardiography for determining lnRMSSD. J Strength Cond Res. 2017;31(2):380–5.PubMed Esco MR, Flatt AA, Nakamura FY. Agreement between a smartphone pulse sensor application and electrocardiography for determining lnRMSSD. J Strength Cond Res. 2017;31(2):380–5.PubMed
25.
go back to reference Flatt AA, Esco MR. Validity of the ithlete™ smart phone application for determining ultra-short-term heart rate variability. J Hum Kinet. 2013;39:85–92.PubMedPubMedCentralCrossRef Flatt AA, Esco MR. Validity of the ithlete™ smart phone application for determining ultra-short-term heart rate variability. J Hum Kinet. 2013;39:85–92.PubMedPubMedCentralCrossRef
26.
go back to reference Vanderlei LCM, Silva RA, Pastre CM, Azevedo FMD, Godoy MF. Comparison of the Polar S810i monitor and the ECG for the analysis of heart rate variability in the time and frequency domains. Braz J Med Biol Res. 2008;41(10):854–9.PubMedCrossRef Vanderlei LCM, Silva RA, Pastre CM, Azevedo FMD, Godoy MF. Comparison of the Polar S810i monitor and the ECG for the analysis of heart rate variability in the time and frequency domains. Braz J Med Biol Res. 2008;41(10):854–9.PubMedCrossRef
27.
go back to reference Gamelin FX, Berthoin S, Bosquet L. Validity of the Polar S810 heart rate monitor to measure R-R intervals at rest. Med Sci Sports Exerc. 2006;38(5):887–93.PubMedCrossRef Gamelin FX, Berthoin S, Bosquet L. Validity of the Polar S810 heart rate monitor to measure R-R intervals at rest. Med Sci Sports Exerc. 2006;38(5):887–93.PubMedCrossRef
28.
go back to reference Giles D, Draper N, Neil W. Validity of the Polar V800 heart rate monitor to measure RR intervals at rest. Eur J Appl Physiol. 2016;116(3):563–71.PubMedCrossRef Giles D, Draper N, Neil W. Validity of the Polar V800 heart rate monitor to measure RR intervals at rest. Eur J Appl Physiol. 2016;116(3):563–71.PubMedCrossRef
29.
go back to reference Montaño A, Brown F, Credeur DP, Williams MA, Stoner L. Telemetry-derived heart rate variability responses to a physical stressor. Clin Physiol Funct Imaging. 2017;37(4):421–7.PubMedCrossRef Montaño A, Brown F, Credeur DP, Williams MA, Stoner L. Telemetry-derived heart rate variability responses to a physical stressor. Clin Physiol Funct Imaging. 2017;37(4):421–7.PubMedCrossRef
30.
go back to reference Vasconcellos FV, Seabra A, Cunha FA, Montenegro RA, Bouskela E, Farinatti P. Heart rate variability assessment with fingertip photoplethysmography and Polar RS800cx as compared with electrocardiography in obese adolescents. Blood Press Monit. 2015;20(6):351–60.PubMedCrossRef Vasconcellos FV, Seabra A, Cunha FA, Montenegro RA, Bouskela E, Farinatti P. Heart rate variability assessment with fingertip photoplethysmography and Polar RS800cx as compared with electrocardiography in obese adolescents. Blood Press Monit. 2015;20(6):351–60.PubMedCrossRef
31.
go back to reference Nunan D, Jakovljevic DG, Donovan G, Hodges LD, Sandercock GR, Brodie DA. Levels of agreement for RR intervals and short-term heart rate variability obtained from the Polar S810 and an alternative system. Eur J Appl Physiol. 2008;103(5):529–37.PubMedCrossRef Nunan D, Jakovljevic DG, Donovan G, Hodges LD, Sandercock GR, Brodie DA. Levels of agreement for RR intervals and short-term heart rate variability obtained from the Polar S810 and an alternative system. Eur J Appl Physiol. 2008;103(5):529–37.PubMedCrossRef
32.
go back to reference Moher D, Liberati A, Tetzlaff J, Altman D. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. BMJ. 2009;339:332–6.CrossRef Moher D, Liberati A, Tetzlaff J, Altman D. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. BMJ. 2009;339:332–6.CrossRef
33.
go back to reference Cohen JF, Korevaar DA, Altman DG, Bruns DE, Gatsonis CA, Hooft L, et al. STARD 2015 guidelines for reporting diagnostic accuracy studies: explanation and elaboration. BMJ Open. 2016;6(11):e012799.PubMedPubMedCentralCrossRef Cohen JF, Korevaar DA, Altman DG, Bruns DE, Gatsonis CA, Hooft L, et al. STARD 2015 guidelines for reporting diagnostic accuracy studies: explanation and elaboration. BMJ Open. 2016;6(11):e012799.PubMedPubMedCentralCrossRef
34.
go back to reference Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig LM, et al. Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD initiative. Standards for reporting of diagnostic accuracy. Clin Chem. 2003;49(1):1–6.PubMedCrossRef Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig LM, et al. Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD initiative. Standards for reporting of diagnostic accuracy. Clin Chem. 2003;49(1):1–6.PubMedCrossRef
35.
go back to reference Quintana DS, Alvares GA, Heathers JA. Guidelines for reporting articles on psychiatry and heart rate variability (GRAPH): recommendations to advance research communication. Transl Psychiatry. 2016;6:e803.PubMedPubMedCentralCrossRef Quintana DS, Alvares GA, Heathers JA. Guidelines for reporting articles on psychiatry and heart rate variability (GRAPH): recommendations to advance research communication. Transl Psychiatry. 2016;6:e803.PubMedPubMedCentralCrossRef
36.
go back to reference Laborde S, Mosley E, Thayer JF. Heart rate variability and cardiac vagal tone in psychophysiological research—recommendations for experiment planning, data analysis, and data reporting. Front Psychol. 2017;8:213.PubMedPubMedCentralCrossRef Laborde S, Mosley E, Thayer JF. Heart rate variability and cardiac vagal tone in psychophysiological research—recommendations for experiment planning, data analysis, and data reporting. Front Psychol. 2017;8:213.PubMedPubMedCentralCrossRef
38.
go back to reference Hedges L, Olkin I. Statistical methods for meta-analysis. 6th ed. San Diego: Academic Press; 1985. p. 79–201. Hedges L, Olkin I. Statistical methods for meta-analysis. 6th ed. San Diego: Academic Press; 1985. p. 79–201.
39.
go back to reference Crossingham IR, Nethercott DR, Columb MO. Comparing cardiac output monitors and defining agreement: a systematic review and meta-analysis. J Intensive Care Soc. 2016;17(4):302–13.PubMedPubMedCentralCrossRef Crossingham IR, Nethercott DR, Columb MO. Comparing cardiac output monitors and defining agreement: a systematic review and meta-analysis. J Intensive Care Soc. 2016;17(4):302–13.PubMedPubMedCentralCrossRef
40.
go back to reference Zaki R, Bulgiba A, Ismail R, Ismail NA. Statistical methods used to test for agreement of medical instruments measuring continuous variables in method comparison studies: a systematic review. PLoS One. 2012;7(5):e37908.PubMedPubMedCentralCrossRef Zaki R, Bulgiba A, Ismail R, Ismail NA. Statistical methods used to test for agreement of medical instruments measuring continuous variables in method comparison studies: a systematic review. PLoS One. 2012;7(5):e37908.PubMedPubMedCentralCrossRef
41.
go back to reference Plews DJ, Scott B, Altini M, Wood M, Kilding AE, Laursen PB. Comparison of heart rate variability recording with smart phone photoplethysmographic, Polar H7 chest strap and electrocardiogram methods. Int J Sports Physiol Perform. 2017;12(10):1324–8.PubMedCrossRef Plews DJ, Scott B, Altini M, Wood M, Kilding AE, Laursen PB. Comparison of heart rate variability recording with smart phone photoplethysmographic, Polar H7 chest strap and electrocardiogram methods. Int J Sports Physiol Perform. 2017;12(10):1324–8.PubMedCrossRef
42.
go back to reference Boos CJ, Bakker-Dyos J, Watchorn J, Woods DR, O’Hara JP, Macconnachie L, et al. A comparison of two methods of heart rate variability assessment at high altitude. Clin Physiol Funct Imaging. 2017;37(6):582–7.PubMedCrossRef Boos CJ, Bakker-Dyos J, Watchorn J, Woods DR, O’Hara JP, Macconnachie L, et al. A comparison of two methods of heart rate variability assessment at high altitude. Clin Physiol Funct Imaging. 2017;37(6):582–7.PubMedCrossRef
43.
go back to reference Hox J. Multilevel analysis: techniques and applications. 2nd edn. New York, NY: Taylor & Francis; 2010. p. 205–32.CrossRef Hox J. Multilevel analysis: techniques and applications. 2nd edn. New York, NY: Taylor & Francis; 2010. p. 205–32.CrossRef
44.
go back to reference Singer JD, Using SAS. PROC MIXED to fit multilevel models, hierarchical models, and individual growth models. J Educ Behav Stat. 1998;23(4):323–55.CrossRef Singer JD, Using SAS. PROC MIXED to fit multilevel models, hierarchical models, and individual growth models. J Educ Behav Stat. 1998;23(4):323–55.CrossRef
45.
go back to reference Higgins J, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med. 2002;21(11):1539–58.PubMedCrossRef Higgins J, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med. 2002;21(11):1539–58.PubMedCrossRef
47.
go back to reference Schäfer A, Vagedes J. How accurate is pulse rate variability as an estimate of heart rate variability? A review on studies comparing photoplethysmographic technology with an electrocardiogram. Int J Cardiol. 2013;166(1):15–29.PubMedCrossRef Schäfer A, Vagedes J. How accurate is pulse rate variability as an estimate of heart rate variability? A review on studies comparing photoplethysmographic technology with an electrocardiogram. Int J Cardiol. 2013;166(1):15–29.PubMedCrossRef
49.
go back to reference Rosenberg M. The file-drawer problem revisited: a general weighted method for calculating fail-safe numbers in meta-analysis. Evolution. 2005;59(2):464–8.PubMedCrossRef Rosenberg M. The file-drawer problem revisited: a general weighted method for calculating fail-safe numbers in meta-analysis. Evolution. 2005;59(2):464–8.PubMedCrossRef
50.
go back to reference Viechtbauer W. Conducting meta-analyses in R with the metafor package. J Stat Softw. 2010;36(3):1–48.CrossRef Viechtbauer W. Conducting meta-analyses in R with the metafor package. J Stat Softw. 2010;36(3):1–48.CrossRef
51.
go back to reference R Development Core Team. R: a language and environment for statistical computing. Vienna, Austria: R Foundation for Statistical Computing. 2017. R Development Core Team. R: a language and environment for statistical computing. Vienna, Austria: R Foundation for Statistical Computing. 2017.
52.
go back to reference Gamelin FX, Baquet G, Berthoin S, Bosquet L. Validity of the Polar S810 to measure R-R intervals in children. Int J Sports Med. 2008;29(2):134–8.PubMedCrossRef Gamelin FX, Baquet G, Berthoin S, Bosquet L. Validity of the Polar S810 to measure R-R intervals in children. Int J Sports Med. 2008;29(2):134–8.PubMedCrossRef
53.
go back to reference Heathers JAJ. Smartphone-enabled pulse rate variability: an alternative methodology for the collection of heart rate variability in psychophysiological research. Int J Psychophysiol. 2013;89:297–304.PubMedCrossRef Heathers JAJ. Smartphone-enabled pulse rate variability: an alternative methodology for the collection of heart rate variability in psychophysiological research. Int J Psychophysiol. 2013;89:297–304.PubMedCrossRef
54.
go back to reference Nunan D, Donovan G, Jakovljevic DG, Hodges LD, Sandercock GR, Brodie DA. Validity and reliability of short-term heart-rate variability from the Polar S810. Med Sci Sports Exerc. 2009;41(1):243–50.PubMedCrossRef Nunan D, Donovan G, Jakovljevic DG, Hodges LD, Sandercock GR, Brodie DA. Validity and reliability of short-term heart-rate variability from the Polar S810. Med Sci Sports Exerc. 2009;41(1):243–50.PubMedCrossRef
55.
go back to reference Porto LGG, Junqueira J. Comparison of time-domain short-term heart interval variability analysis using a wrist-worn heart rate monitor and the conventional electrocardiogram. Pacing Clin Electrophysiol. 2009;32(1):43–51.PubMedCrossRef Porto LGG, Junqueira J. Comparison of time-domain short-term heart interval variability analysis using a wrist-worn heart rate monitor and the conventional electrocardiogram. Pacing Clin Electrophysiol. 2009;32(1):43–51.PubMedCrossRef
56.
go back to reference Guzik P, Piekos C, Pierog O, Fenech N, Krauze T, Piskorski J, et al. Classic electrocardiogram-based and mobile technology derived approaches to heart rate variability are not equivalent. Int J Cardiol. 2018;258:154–6.PubMedCrossRef Guzik P, Piekos C, Pierog O, Fenech N, Krauze T, Piskorski J, et al. Classic electrocardiogram-based and mobile technology derived approaches to heart rate variability are not equivalent. Int J Cardiol. 2018;258:154–6.PubMedCrossRef
57.
go back to reference Sterne JAC, Sutton AJ, Ioannidis JPA, Terrin N, Jones DR, Lau J, et al. Recommendations for examining and interpreting funnel plot asymmetry in meta-analyses of randomised controlled trials. BMJ. 2011;343:d4002.PubMedCrossRef Sterne JAC, Sutton AJ, Ioannidis JPA, Terrin N, Jones DR, Lau J, et al. Recommendations for examining and interpreting funnel plot asymmetry in meta-analyses of randomised controlled trials. BMJ. 2011;343:d4002.PubMedCrossRef
58.
go back to reference Begg CB, Berlin JA. Publication bias: a problem in interpreting medical data. J R Stat Soc Ser A. 1988;151(3):419–63.CrossRef Begg CB, Berlin JA. Publication bias: a problem in interpreting medical data. J R Stat Soc Ser A. 1988;151(3):419–63.CrossRef
59.
go back to reference Sterne JAC, Egger M. Funnel plots for detecting bias in meta-analysis: guidelines on choice of axis. J Clin Epidemiol. 2001;54(10):1046–55.PubMedCrossRef Sterne JAC, Egger M. Funnel plots for detecting bias in meta-analysis: guidelines on choice of axis. J Clin Epidemiol. 2001;54(10):1046–55.PubMedCrossRef
60.
go back to reference Rosenthal R. Meta-analytic procedures for social research. 2nd edn. Newbury Park, CA: SAGE Publications; 1991.CrossRef Rosenthal R. Meta-analytic procedures for social research. 2nd edn. Newbury Park, CA: SAGE Publications; 1991.CrossRef
61.
go back to reference Sammito S, Böckelmann I. Options and limitations of heart rate measurement and analysis of heart rate variability by mobile devices: a systematic review. Herzschrittmacherther Elektrophysiol. 2016;27(1):38–45.PubMedCrossRef Sammito S, Böckelmann I. Options and limitations of heart rate measurement and analysis of heart rate variability by mobile devices: a systematic review. Herzschrittmacherther Elektrophysiol. 2016;27(1):38–45.PubMedCrossRef
62.
go back to reference Buchheit M, Mendez-Villanueva A. Improbable effect of carbohydrate diet on cardiac autonomic modulation during exercise. Eur J Appl Physiol. 2010;109(3):571–4.PubMedCrossRef Buchheit M, Mendez-Villanueva A. Improbable effect of carbohydrate diet on cardiac autonomic modulation during exercise. Eur J Appl Physiol. 2010;109(3):571–4.PubMedCrossRef
63.
64.
go back to reference Penttila J, Helminen A, Jartti T, Kuusela T, Huikuri HV, Tulppo MP, et al. Time domain, geometrical and frequency domain analysis of cardiac vagal outflow: effects of various respiratory patterns. Clin Physiol. 2001;21(3):365–76.PubMedCrossRef Penttila J, Helminen A, Jartti T, Kuusela T, Huikuri HV, Tulppo MP, et al. Time domain, geometrical and frequency domain analysis of cardiac vagal outflow: effects of various respiratory patterns. Clin Physiol. 2001;21(3):365–76.PubMedCrossRef
66.
go back to reference Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19(1):231–40.PubMed Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19(1):231–40.PubMed
67.
go back to reference Esco MR, Flatt AA. Ultra-short-term heart rate variability indexes at rest and post-exercise in athletes: evaluating the agreement with accepted recommendations. J Sports Sci Med. 2014;13(3):535–41.PubMedPubMedCentral Esco MR, Flatt AA. Ultra-short-term heart rate variability indexes at rest and post-exercise in athletes: evaluating the agreement with accepted recommendations. J Sports Sci Med. 2014;13(3):535–41.PubMedPubMedCentral
68.
go back to reference Altini M, Amft O. HRV4Training: large-scale longitudinal training load analysis in unconstrained free-living settings using a smartphone application. Conf Proc IEEE Eng Med Biol Soc. 2016;2016:2610–3.PubMed Altini M, Amft O. HRV4Training: large-scale longitudinal training load analysis in unconstrained free-living settings using a smartphone application. Conf Proc IEEE Eng Med Biol Soc. 2016;2016:2610–3.PubMed
69.
go back to reference Plews DJ, Laursen PB, Buchheit M. Day-to-day heart-rate variability recordings in world-champion rowers: appreciating unique athlete characteristics. Int J Sports Physiol Perform. 2017;12(5):697–703.PubMedCrossRef Plews DJ, Laursen PB, Buchheit M. Day-to-day heart-rate variability recordings in world-champion rowers: appreciating unique athlete characteristics. Int J Sports Physiol Perform. 2017;12(5):697–703.PubMedCrossRef
70.
go back to reference Nuuttila OP, Nikander A, Polomoshnov D, Laukkanen JA, Hakkinen K. Effects of HRV-guided vs. predetermined block training on performance, HRV and serum hormones. Int J Sports Med. 2017;38(12):909–20.PubMedCrossRef Nuuttila OP, Nikander A, Polomoshnov D, Laukkanen JA, Hakkinen K. Effects of HRV-guided vs. predetermined block training on performance, HRV and serum hormones. Int J Sports Med. 2017;38(12):909–20.PubMedCrossRef
71.
go back to reference Plews DJ, Laursen PB, Kilding AE, Buchheit M. Evaluating training adaptation with heart-rate measures: a methodological comparison. Int J Sports Physiol Perform. 2013;8(6):688–91.PubMedCrossRef Plews DJ, Laursen PB, Kilding AE, Buchheit M. Evaluating training adaptation with heart-rate measures: a methodological comparison. Int J Sports Physiol Perform. 2013;8(6):688–91.PubMedCrossRef
72.
go back to reference Lemeshow AR, Blum RE, Berlin JA, Stoto MA, Colditz GA. Searching one or two databases was insufficient for meta-analysis of observational studies. J Clin Epidemiol. 2005;58(9):867–73.PubMedCrossRef Lemeshow AR, Blum RE, Berlin JA, Stoto MA, Colditz GA. Searching one or two databases was insufficient for meta-analysis of observational studies. J Clin Epidemiol. 2005;58(9):867–73.PubMedCrossRef
73.
go back to reference Whiting P, Westwood M, Burke M, Sterne J, Glanville J. Systematic reviews of test accuracy should search a range of databases to identify primary studies. J Clin Epidemiol. 2008;61(4):357–64.PubMedCrossRef Whiting P, Westwood M, Burke M, Sterne J, Glanville J. Systematic reviews of test accuracy should search a range of databases to identify primary studies. J Clin Epidemiol. 2008;61(4):357–64.PubMedCrossRef
74.
go back to reference Papaioannou D, Sutton A, Carroll C, Booth A, Wong R. Literature searching for social science systematic reviews: consideration of a range of search techniques. Health Inf Libr J. 2010;27(2):114–22.CrossRef Papaioannou D, Sutton A, Carroll C, Booth A, Wong R. Literature searching for social science systematic reviews: consideration of a range of search techniques. Health Inf Libr J. 2010;27(2):114–22.CrossRef
75.
go back to reference Linder SK, Kamath GR, Pratt GF, Saraykar SS, Volk RJ. Citation searches are more sensitive than keyword searches to identify studies using specific measurement instruments. J Clin Epidemiol. 2015;68(4):412–7.PubMedCrossRef Linder SK, Kamath GR, Pratt GF, Saraykar SS, Volk RJ. Citation searches are more sensitive than keyword searches to identify studies using specific measurement instruments. J Clin Epidemiol. 2015;68(4):412–7.PubMedCrossRef
76.
go back to reference Allen IE, Olkin I. Estimating time to conduct a meta-analysis from number of citations retrieved. JAMA. 1999;282(7):634–5.PubMedCrossRef Allen IE, Olkin I. Estimating time to conduct a meta-analysis from number of citations retrieved. JAMA. 1999;282(7):634–5.PubMedCrossRef
Metadata
Title
The Accuracy of Acquiring Heart Rate Variability from Portable Devices: A Systematic Review and Meta-Analysis
Authors
Ward C. Dobbs
Michael V. Fedewa
Hayley V. MacDonald
Clifton J. Holmes
Zackary S. Cicone
Daniel J. Plews
Michael R. Esco
Publication date
01-03-2019
Publisher
Springer International Publishing
Published in
Sports Medicine / Issue 3/2019
Print ISSN: 0112-1642
Electronic ISSN: 1179-2035
DOI
https://doi.org/10.1007/s40279-019-01061-5

Other articles of this Issue 3/2019

Sports Medicine 3/2019 Go to the issue