Published in: Drug Safety 12/2014

01-12-2014 | Systematic Review

Benefits and Risks of Using Smart Pumps to Reduce Medication Error Rates: A Systematic Review

Authors: Kumiko Ohashi, Olivia Dalleur, Patricia C. Dykes, David W. Bates

Published in: Drug Safety | Issue 12/2014

Login to get access

Abstract

Background

Smart infusion pumps have been introduced to prevent medication errors and have been widely adopted nationally in the USA, though they are not always used in Europe or other regions. Despite widespread usage of smart pumps, intravenous medication errors have not been fully eliminated.

Objective

Through a systematic review of recent studies and reports regarding smart pump implementation and use, we aimed to identify the impact of smart pumps on error reduction and on the complex process of medication administration, and strategies to maximize the benefits of smart pumps.

Methods

The medical literature related to the effects of smart pumps for improving patient safety was searched in PUBMED, EMBASE, and the Cochrane Central Register of Controlled Trials (CENTRAL) (2000–2014) and relevant papers were selected by two researchers.

Results

After the literature search, 231 papers were identified and the full texts of 138 articles were assessed for eligibility. Of these, 22 were included after removal of papers that did not meet the inclusion criteria. We assessed both the benefits and negative effects of smart pumps from these studies. One of the benefits of using smart pumps was intercepting errors such as the wrong rate, wrong dose, and pump setting errors. Other benefits include reduction of adverse drug event rates, practice improvements, and cost effectiveness. Meanwhile, the current issues or negative effects related to using smart pumps were lower compliance rates of using smart pumps, the overriding of soft alerts, non-intercepted errors, or the possibility of using the wrong drug library.

Conclusion

The literature suggests that smart pumps reduce but do not eliminate programming errors. Although the hard limits of a drug library play a main role in intercepting medication errors, soft limits were still not as effective as hard limits because of high override rates. Compliance in using smart pumps is key towards effectively preventing errors. Opportunities for improvement include upgrading drug libraries, developing standardized drug libraries, decreasing the number of unnecessary warnings, and developing stronger approaches to minimize workarounds. Also, as with other clinical information systems, smart pumps should be implemented with the idea of using continuous quality improvement processes to iteratively improve their use.
Appendix
Available only for authorised users
Literature
1.
go back to reference Pang RKY, Kong DCM, deClifford JM, Lam SS, Leung BK. Smart infusion pumps reduce intravenous medication administration errors at an Australian teaching hospital. J Pharm Pract Res. 2011;41(3):192–5. Pang RKY, Kong DCM, deClifford JM, Lam SS, Leung BK. Smart infusion pumps reduce intravenous medication administration errors at an Australian teaching hospital. J Pharm Pract Res. 2011;41(3):192–5.
2.
go back to reference Snodgrass RD. Smart pump technology. Biomed Instrum Technol. 2005;39(6):444–6.PubMed Snodgrass RD. Smart pump technology. Biomed Instrum Technol. 2005;39(6):444–6.PubMed
5.
go back to reference Murdoch LJ, Cameron VL. Smart infusion technology: a minimum safety standard for intensive care? Br J Nurs. 2008;17(10):630–6.PubMedCrossRef Murdoch LJ, Cameron VL. Smart infusion technology: a minimum safety standard for intensive care? Br J Nurs. 2008;17(10):630–6.PubMedCrossRef
6.
go back to reference Mansfield J, Jarrett S. Using smart pumps to understand and evaluate clinician practice patterns to ensure patient safety. Hosp Pharm. 2013;48(11):942–50.PubMedCentralPubMedCrossRef Mansfield J, Jarrett S. Using smart pumps to understand and evaluate clinician practice patterns to ensure patient safety. Hosp Pharm. 2013;48(11):942–50.PubMedCentralPubMedCrossRef
7.
go back to reference Pedersen CA, Schneider PJ, Scheckelhoff DJ. ASHP national survey of pharmacy practice in hospital settings: monitoring and patient education—2012. Am J Health Syst Pharm. 2013;70(9):787–803.PubMedCrossRef Pedersen CA, Schneider PJ, Scheckelhoff DJ. ASHP national survey of pharmacy practice in hospital settings: monitoring and patient education—2012. Am J Health Syst Pharm. 2013;70(9):787–803.PubMedCrossRef
8.
go back to reference Rothschild JM, Keohane CA, Cook EF, et al. A controlled trial of smart infusion pumps to improve medication safety in critically ill patients. Crit Care Med. 2005;33(3):533–40.PubMedCrossRef Rothschild JM, Keohane CA, Cook EF, et al. A controlled trial of smart infusion pumps to improve medication safety in critically ill patients. Crit Care Med. 2005;33(3):533–40.PubMedCrossRef
9.
go back to reference Manrique-Rodriguez S, Sanchez-Galindo A, Fernandez-Llamazares CM, et al. Smart pump alerts: all that glitters is not gold. Int J Med Inform. 2012;81(5):344–50.PubMedCrossRef Manrique-Rodriguez S, Sanchez-Galindo A, Fernandez-Llamazares CM, et al. Smart pump alerts: all that glitters is not gold. Int J Med Inform. 2012;81(5):344–50.PubMedCrossRef
10.
go back to reference Williams CK, Maddox RR. Implementation of an i.v. medication safety system. Am J Health Syst Pharm. 2005;62(5):530–6.PubMed Williams CK, Maddox RR. Implementation of an i.v. medication safety system. Am J Health Syst Pharm. 2005;62(5):530–6.PubMed
11.
go back to reference Trbovich PL, Pinkney S, Cafazzo JA, Easty AC. The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit. Qual Saf Health Care. 2010;19(5):430–4.PubMedCentralPubMed Trbovich PL, Pinkney S, Cafazzo JA, Easty AC. The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit. Qual Saf Health Care. 2010;19(5):430–4.PubMedCentralPubMed
12.
go back to reference Hertzel C, Sousa VD. The use of smart pumps for preventing medication errors. J Infus Nurs. 2009;32(5):257–67.PubMedCrossRef Hertzel C, Sousa VD. The use of smart pumps for preventing medication errors. J Infus Nurs. 2009;32(5):257–67.PubMedCrossRef
13.
go back to reference Manrique-Rodriguez S, Sanchez-Galindo AC, Lopez-Herce J, et al. Impact of implementing smart infusion pumps in a pediatric intensive care unit. Am J Health Syst Pharm. 2013;70(21):1897–906.PubMedCrossRef Manrique-Rodriguez S, Sanchez-Galindo AC, Lopez-Herce J, et al. Impact of implementing smart infusion pumps in a pediatric intensive care unit. Am J Health Syst Pharm. 2013;70(21):1897–906.PubMedCrossRef
14.
go back to reference Malashock CM, Shull SS, Gould DA. Effect of smart infusion pumps on medication errors related to infusion device programming. Hosp Pharm. 2004;39(5):460–9. Malashock CM, Shull SS, Gould DA. Effect of smart infusion pumps on medication errors related to infusion device programming. Hosp Pharm. 2004;39(5):460–9.
15.
go back to reference Moher D, Liberati A, Tetzlaff J, Altman DG, Group P. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med. 2009;6(7):e1000097.PubMedCentralPubMedCrossRef Moher D, Liberati A, Tetzlaff J, Altman DG, Group P. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med. 2009;6(7):e1000097.PubMedCentralPubMedCrossRef
16.
go back to reference Fields M, Peterman J. Intravenous medication safety system averts high-risk medication errors and provides actionable data. Nurs Adm Q. 2005;29(1):78–87.PubMedCrossRef Fields M, Peterman J. Intravenous medication safety system averts high-risk medication errors and provides actionable data. Nurs Adm Q. 2005;29(1):78–87.PubMedCrossRef
17.
go back to reference Wilson K, Sullivan M. Preventing medication errors with smart infusion technology. Am J Health Syst Pharm. 2004;61(2):177–83.PubMed Wilson K, Sullivan M. Preventing medication errors with smart infusion technology. Am J Health Syst Pharm. 2004;61(2):177–83.PubMed
18.
go back to reference Ohashi K, Dykes P, McIntosh K, Buckley E, Wien M, Bates DW. Evaluation of intravenous medication errors with smart infusion pumps in an academic medical center. AMIA Annu Symp Proc. 2013;2013:1089–98.PubMedCentralPubMed Ohashi K, Dykes P, McIntosh K, Buckley E, Wien M, Bates DW. Evaluation of intravenous medication errors with smart infusion pumps in an academic medical center. AMIA Annu Symp Proc. 2013;2013:1089–98.PubMedCentralPubMed
19.
go back to reference Russell RA, Murkowski K, Scanlon MC. Discrepancies between medication orders and infusion pump programming in a paediatric intensive care unit. Qual Saf Health Care. 2010;19(Suppl 3):i31–5.PubMedCrossRef Russell RA, Murkowski K, Scanlon MC. Discrepancies between medication orders and infusion pump programming in a paediatric intensive care unit. Qual Saf Health Care. 2010;19(Suppl 3):i31–5.PubMedCrossRef
20.
go back to reference Evans RS, Carlson R, Johnson KV, Palmer BK, Lloyd JF. Enhanced notification of infusion pump programming errors. Stud Health Technol Inform. 2010;160(Pt 1):734–8.PubMed Evans RS, Carlson R, Johnson KV, Palmer BK, Lloyd JF. Enhanced notification of infusion pump programming errors. Stud Health Technol Inform. 2010;160(Pt 1):734–8.PubMed
21.
go back to reference Husch M, Sullivan C, Rooney D, et al. Insights from the sharp end of intravenous medication errors: implications for infusion pump technology. Qual Saf Health Care. 2005;14(2):80–6.PubMedCentralPubMedCrossRef Husch M, Sullivan C, Rooney D, et al. Insights from the sharp end of intravenous medication errors: implications for infusion pump technology. Qual Saf Health Care. 2005;14(2):80–6.PubMedCentralPubMedCrossRef
22.
go back to reference Skledar SJ, Niccolai CS, Schilling D, et al. Quality-improvement analytics for intravenous infusion pumps. Am J Health Syst Pharm. 2013;70(8):680–6.PubMedCrossRef Skledar SJ, Niccolai CS, Schilling D, et al. Quality-improvement analytics for intravenous infusion pumps. Am J Health Syst Pharm. 2013;70(8):680–6.PubMedCrossRef
23.
go back to reference Fanikos J, Fiumara K, Baroletti S, et al. Impact of smart infusion technology on administration of anticoagulants (unfractionated heparin, argatroban, lepirudin, and bivalirudin). Am J Cardiol. 2007;99(7):1002–5.PubMedCrossRef Fanikos J, Fiumara K, Baroletti S, et al. Impact of smart infusion technology on administration of anticoagulants (unfractionated heparin, argatroban, lepirudin, and bivalirudin). Am J Cardiol. 2007;99(7):1002–5.PubMedCrossRef
24.
go back to reference Larsen GY, Parker HB, Cash J, O’Connell M, Grant MC. Standard drug concentrations and smart-pump technology reduce continuous-medication-infusion errors in pediatric patients. Pediatrics. 2005;116(1):e21–5.PubMedCrossRef Larsen GY, Parker HB, Cash J, O’Connell M, Grant MC. Standard drug concentrations and smart-pump technology reduce continuous-medication-infusion errors in pediatric patients. Pediatrics. 2005;116(1):e21–5.PubMedCrossRef
25.
go back to reference Prewitt J, Schneider S, Horvath M, Hammond J, Jackson J, Ginsberg B. PCA safety data review after clinical decision support and smart pump technology implementation. J Patient Saf. 2013;9(2):103–9.PubMedCrossRef Prewitt J, Schneider S, Horvath M, Hammond J, Jackson J, Ginsberg B. PCA safety data review after clinical decision support and smart pump technology implementation. J Patient Saf. 2013;9(2):103–9.PubMedCrossRef
26.
go back to reference Nuckols TK, Bower AG, Paddock SM, et al. Programmable infusion pumps in ICUs: an analysis of corresponding adverse drug events. J Gen Intern Med. 2008;23(Suppl 1):41–5.PubMedCentralPubMedCrossRef Nuckols TK, Bower AG, Paddock SM, et al. Programmable infusion pumps in ICUs: an analysis of corresponding adverse drug events. J Gen Intern Med. 2008;23(Suppl 1):41–5.PubMedCentralPubMedCrossRef
27.
go back to reference Paul JE, Bertram B, Antoni K, et al. Impact of a comprehensive safety initiative on patient-controlled analgesia errors. Anesthesiology. 2010;113(6):1427–32.PubMedCrossRef Paul JE, Bertram B, Antoni K, et al. Impact of a comprehensive safety initiative on patient-controlled analgesia errors. Anesthesiology. 2010;113(6):1427–32.PubMedCrossRef
28.
go back to reference Tran M, Ciarkowski S, Wagner D, Stevenson JG. A case study on the safety impact of implementing smart patient-controlled analgesic pumps at a tertiary care academic medical center. Jt Comm J Qual Patient Saf. 2012;38(3):112–9.PubMed Tran M, Ciarkowski S, Wagner D, Stevenson JG. A case study on the safety impact of implementing smart patient-controlled analgesic pumps at a tertiary care academic medical center. Jt Comm J Qual Patient Saf. 2012;38(3):112–9.PubMed
29.
go back to reference Gerhart D, O’Shea K, Muller S. Advancing medication infusion safety through the clinical integration of technology. Hosp Pract (1995). 2013;41(4):7–14.CrossRef Gerhart D, O’Shea K, Muller S. Advancing medication infusion safety through the clinical integration of technology. Hosp Pract (1995). 2013;41(4):7–14.CrossRef
30.
go back to reference Kastrup M, Balzer F, Volk T, Spies C. Analysis of event logs from syringe pumps: a retrospective pilot study to assess possible effects of syringe pumps on safety in a university hospital critical care unit in Germany. Drug Saf. 2012;35(7):563–74.PubMedCrossRef Kastrup M, Balzer F, Volk T, Spies C. Analysis of event logs from syringe pumps: a retrospective pilot study to assess possible effects of syringe pumps on safety in a university hospital critical care unit in Germany. Drug Saf. 2012;35(7):563–74.PubMedCrossRef
31.
go back to reference Bates DW, Leape LL, Cullen DJ, et al. Effect of computerized physician order entry and a team intervention on prevention of serious medication errors. JAMA. 1998;280:1311–6.PubMedCrossRef Bates DW, Leape LL, Cullen DJ, et al. Effect of computerized physician order entry and a team intervention on prevention of serious medication errors. JAMA. 1998;280:1311–6.PubMedCrossRef
32.
go back to reference Poon EG, Keohane CA, Yoon CS, et al. Effect of bar-code technology on the safety of medication administration. N Engl J Med. 2010;362(18):1698–707.PubMedCrossRef Poon EG, Keohane CA, Yoon CS, et al. Effect of bar-code technology on the safety of medication administration. N Engl J Med. 2010;362(18):1698–707.PubMedCrossRef
33.
go back to reference Poon EG, Cina JL, Churchill W, et al. Medication dispensing errors and potential adverse drug events before and after implementing bar code technology in the pharmacy. Ann Intern Med. 2006;145:426–34.PubMedCrossRef Poon EG, Cina JL, Churchill W, et al. Medication dispensing errors and potential adverse drug events before and after implementing bar code technology in the pharmacy. Ann Intern Med. 2006;145:426–34.PubMedCrossRef
34.
go back to reference Manrique-Rodriguez S, Sanchez-Galindo A, Fernandez-Llamazares CM, et al. Developing a drug library for smart pumps in a pediatric intensive care unit. Artif Intell Med. 2012;54(3):155–61.PubMedCrossRef Manrique-Rodriguez S, Sanchez-Galindo A, Fernandez-Llamazares CM, et al. Developing a drug library for smart pumps in a pediatric intensive care unit. Artif Intell Med. 2012;54(3):155–61.PubMedCrossRef
35.
go back to reference Eastham J, Rizos A, Gama J, et al. Reduction in variation of intravenous drug administration in seventeen San Diego hospitals with standardized drug concentrations and dosage units. Hosp Pharm. 2009;44(2):150–8.CrossRef Eastham J, Rizos A, Gama J, et al. Reduction in variation of intravenous drug administration in seventeen San Diego hospitals with standardized drug concentrations and dosage units. Hosp Pharm. 2009;44(2):150–8.CrossRef
Metadata
Title
Benefits and Risks of Using Smart Pumps to Reduce Medication Error Rates: A Systematic Review
Authors
Kumiko Ohashi
Olivia Dalleur
Patricia C. Dykes
David W. Bates
Publication date
01-12-2014
Publisher
Springer International Publishing
Published in
Drug Safety / Issue 12/2014
Print ISSN: 0114-5916
Electronic ISSN: 1179-1942
DOI
https://doi.org/10.1007/s40264-014-0232-1

Other articles of this Issue 12/2014

Drug Safety 12/2014 Go to the issue

Acknowledgement to Referees

Acknowledgement to Referees