Open Access 01-12-2017 | Advances in Monitoring for Anesthesia (LAH Critchley, Section Editor)

Stroke Volume Monitoring: Novel Continuous Wave Doppler Parameters, Algorithms and Advanced Noninvasive Haemodynamic Concepts

Authors: R. A. Phillips, B. E. Smith, V. M. Madigan

Published in: Current Anesthesiology Reports | Issue 4/2017

Login to get access

Abstract

Purpose of Review

Adequate oxygen delivery is essential for life, with hypoxia resulting in dysfunction, and ultimately death, of the cells, organs and organism. Blood flow delivers the oxygen bound in the blood, while haemodynamics is the science of blood flow. Stroke volume (SV) is the fundamental unit of blood flow, and reflects the interdependent performance of the heart, the vessels and the autonomic nervous system. However, haemodynamic management remains generally poor and predominantly guided by simple blood pressure observations alone.

Recent Findings

Doppler ultrasound measures SV with unequalled clinical precision when operated by trained personnel. Combining SV with BP measurements allows calculation of flow-pressure based measures which better reflect cardiovascular performance and allows personalised physiologic and pathophysiologic modelling consistent with Frank’s and Starling’s observations.

Summary

Doppler SV monitoring and novel flow-pressure parameters may improve our understanding of the cardiovascular system and lead to improved diagnosis and therapy. This review examines the physics and practice of Doppler SV monitoring and its application in advanced haemodynamics.
Literature
1.
go back to reference Guyton AC, Lindsey AW, Kaufmann BN. Effect of mean circulatory filling pressure and other peripheral circulatory factors on cardiac output. Am J Phys. 1955;180(3):463–8. Guyton AC, Lindsey AW, Kaufmann BN. Effect of mean circulatory filling pressure and other peripheral circulatory factors on cardiac output. Am J Phys. 1955;180(3):463–8.
3.
go back to reference • Meng L, Heerdt PM. Peri-operative goal-directed haemodynamic therapy based on flow parameters: a concept in evolution. Brit J Anesth. 2016;117(S3):iii3–iii17. https://doi.org/10.1093/bja/aew363. Although this extensive review has omitted several haemodynamic methods, there is still a wealth of knowledge to provide a good overview of the recent changes in haemodynamic monitoring used clinically. The authors also give good insights into why haemodynamic monitoring is required as well as the problems of determining what could and should be used. Unfortunately, there is little emphasis on oxygen delivery indices and rather more on resuscitation and short-term volume-augmented perfusion. CrossRef • Meng L, Heerdt PM. Peri-operative goal-directed haemodynamic therapy based on flow parameters: a concept in evolution. Brit J Anesth. 2016;117(S3):iii3–iii17. https://​doi.​org/​10.​1093/​bja/​aew363. Although this extensive review has omitted several haemodynamic methods, there is still a wealth of knowledge to provide a good overview of the recent changes in haemodynamic monitoring used clinically. The authors also give good insights into why haemodynamic monitoring is required as well as the problems of determining what could and should be used. Unfortunately, there is little emphasis on oxygen delivery indices and rather more on resuscitation and short-term volume-augmented perfusion. CrossRef
6.
go back to reference Arikan AA, Zappitelli M, Goldstein SL, Naipaul A, Jefferson LS, Loftus LL. Fluid overload is associated with impaired oxygenation and morbidity in critically ill children. Pediatr Care Med. 2012;13(3):253–8.CrossRef Arikan AA, Zappitelli M, Goldstein SL, Naipaul A, Jefferson LS, Loftus LL. Fluid overload is associated with impaired oxygenation and morbidity in critically ill children. Pediatr Care Med. 2012;13(3):253–8.CrossRef
8.
go back to reference • Le Manach Y, Hofer CK, Lehot JJ, Vallet B, Goarin JP, Tavernier B, et al. Can changes in arterial pressure be used to detect changes in cardiac output during volume expansion in the perioperative period? Anesthesiology. 2012;117(6):1165–74. https://doi.org/10.1097/ALN.0b013e318275561d. In short, the answer to the question the authors ask is no, yet only around one third of anesthesiologists do anything to improve on the answer by way of measuring anything but basic haemodynamic parameters. Using a grey zone approach, the authors show that arterial pressure cannot be used as an index of cardiac output or to guide volume expansion and fluid therapy. Very few patients fulfilled the required criteria for use of PPV as an indicator of fluid status while cardiac arrythmia was a further major confounder. CrossRefPubMed • Le Manach Y, Hofer CK, Lehot JJ, Vallet B, Goarin JP, Tavernier B, et al. Can changes in arterial pressure be used to detect changes in cardiac output during volume expansion in the perioperative period? Anesthesiology. 2012;117(6):1165–74. https://​doi.​org/​10.​1097/​ALN.​0b013e318275561d​. In short, the answer to the question the authors ask is no, yet only around one third of anesthesiologists do anything to improve on the answer by way of measuring anything but basic haemodynamic parameters. Using a grey zone approach, the authors show that arterial pressure cannot be used as an index of cardiac output or to guide volume expansion and fluid therapy. Very few patients fulfilled the required criteria for use of PPV as an indicator of fluid status while cardiac arrythmia was a further major confounder. CrossRefPubMed
9.
go back to reference Rowell LB. Human Cardiovascular Control. Oxford University Press, 1993, p206. Rowell LB. Human Cardiovascular Control. Oxford University Press, 1993, p206.
10.
go back to reference Wade OL, Bishop JM. Cardiac output and regional blood flow. Oxford: Blackwell Scientific; 1962. p. p187. Wade OL, Bishop JM. Cardiac output and regional blood flow. Oxford: Blackwell Scientific; 1962. p. p187.
11.
go back to reference Frank O. Die Grundform des Arteriellen Pulses. Z Biol. 1899;37:483–526. (also - Sagawa K, Lie RK, Schaefer J. Translation of Otto Frank’s paper “Die Grundform des Arteriellen Pulses” Zeitschrift für Biologie 37: 483-526 1899). J Mol Cell Cardiol. 1990 22(3):253–4. Frank O. Die Grundform des Arteriellen Pulses. Z Biol. 1899;37:483–526. (also - Sagawa K, Lie RK, Schaefer J. Translation of Otto Frank’s paper “Die Grundform des Arteriellen Pulses” Zeitschrift für Biologie 37: 483-526 1899). J Mol Cell Cardiol. 1990 22(3):253–4.
13.
go back to reference Laketta EG. Starling’s law of the heart is explained by an intimate interaction of muscle length and myofilament calcium activation. J Am Coll Cardiol. 1987;10(5):1157–64.CrossRef Laketta EG. Starling’s law of the heart is explained by an intimate interaction of muscle length and myofilament calcium activation. J Am Coll Cardiol. 1987;10(5):1157–64.CrossRef
14.
go back to reference Riva-Rocci S, Zanchetti A, Mancia G. A new sphygmomanometer. Sphygmomanometric technique. J Hypertens. 1996;14(1):1–12.PubMed Riva-Rocci S, Zanchetti A, Mancia G. A new sphygmomanometer. Sphygmomanometric technique. J Hypertens. 1996;14(1):1–12.PubMed
15.
go back to reference Van Montfrans GA. Oscillometric blood pressure measurement: progress and problems. Blood Press Monit. 2001;6(6):287–90.CrossRefPubMed Van Montfrans GA. Oscillometric blood pressure measurement: progress and problems. Blood Press Monit. 2001;6(6):287–90.CrossRefPubMed
19.
go back to reference Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al. Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on high blood pressure research. Circ. 2005;111(5):697–716. https://doi.org/10.1161/01.CIR.0000154900.76284.F6.CrossRef Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al. Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on high blood pressure research. Circ. 2005;111(5):697–716. https://​doi.​org/​10.​1161/​01.​CIR.​0000154900.​76284.​F6.CrossRef
20.
go back to reference Picone DS, Schultz MG, Otahal P, Aakhus S, Al-Jumaily AM, Black JA, et al. Accuracy of cuff-measured pressure systematic reviews and meta-analyses. J Am Coll Cardiol. 2017;70(5):572–86.CrossRefPubMed Picone DS, Schultz MG, Otahal P, Aakhus S, Al-Jumaily AM, Black JA, et al. Accuracy of cuff-measured pressure systematic reviews and meta-analyses. J Am Coll Cardiol. 2017;70(5):572–86.CrossRefPubMed
21.
go back to reference Wo CC, Shoemaker WC, Appel PL, Bishop MH, Kram HB, Hardin E. Unreliability of blood pressure and heart rate to evaluate cardiac output in emergency resuscitation and critical illness. Crit Care Med. 1993 Feb;21(2):218–23.CrossRefPubMed Wo CC, Shoemaker WC, Appel PL, Bishop MH, Kram HB, Hardin E. Unreliability of blood pressure and heart rate to evaluate cardiac output in emergency resuscitation and critical illness. Crit Care Med. 1993 Feb;21(2):218–23.CrossRefPubMed
22.
go back to reference Tomlin PJ, Duck F, McNulty M, Green CD. A comparison of methods of evaluating myocardial contractility. Can Anaesth Soc J. 1975;22:436–8.CrossRefPubMed Tomlin PJ, Duck F, McNulty M, Green CD. A comparison of methods of evaluating myocardial contractility. Can Anaesth Soc J. 1975;22:436–8.CrossRefPubMed
23.
go back to reference • Hall JB. Searching for evidence to support pulmonary artery catheter use in critically ill patients. JAMA. 2005;294(13):1693–4. Although a little dated, this editorial from 2005, some 35 years after the introduction of the pulmonary artery catheter (PAC), discusses the evidence in terms of patient outcomes and the impact, or more accurately, lack of impact, of the PAC. Although often considered or referred to as the ‘gold standard’ of haemodynamic monitoring and cardiac output measurement, in fact there is little evidence to support this view. The ESCAPE trial which was terminated early due to excessive adverse effects in the PAC group is also discussed as is the case for using non-PAC methods of evaluating haemodynamics. The final admonishment of ‘Don’t just do something, stand there! And then think about it…’ is as true today as 12 years ago. CrossRefPubMed • Hall JB. Searching for evidence to support pulmonary artery catheter use in critically ill patients. JAMA. 2005;294(13):1693–4. Although a little dated, this editorial from 2005, some 35 years after the introduction of the pulmonary artery catheter (PAC), discusses the evidence in terms of patient outcomes and the impact, or more accurately, lack of impact, of the PAC. Although often considered or referred to as the ‘gold standard’ of haemodynamic monitoring and cardiac output measurement, in fact there is little evidence to support this view. The ESCAPE trial which was terminated early due to excessive adverse effects in the PAC group is also discussed as is the case for using non-PAC methods of evaluating haemodynamics. The final admonishment of ‘Don’t just do something, stand there! And then think about it…’ is as true today as 12 years ago. CrossRefPubMed
24.
go back to reference Connors AF, McCaffree DR, Gray BA. Evaluation of right heart catheterization in the critically ill patient without acute myocardial infarction. N Engl J Med. 1983;308:263–7.CrossRefPubMed Connors AF, McCaffree DR, Gray BA. Evaluation of right heart catheterization in the critically ill patient without acute myocardial infarction. N Engl J Med. 1983;308:263–7.CrossRefPubMed
26.
go back to reference Phillips RA, Hood SG, Jacobson BM, West MJ, Wan L, May CN. Pulmonary artery catheter (PAC) accuracy and efficacy compared with flow probe and transcutaneous Doppler (USCOM): an ovine validation. Crit Care Res Prac. 2012; https://doi.org/10.1155/2012/621496. Phillips RA, Hood SG, Jacobson BM, West MJ, Wan L, May CN. Pulmonary artery catheter (PAC) accuracy and efficacy compared with flow probe and transcutaneous Doppler (USCOM): an ovine validation. Crit Care Res Prac. 2012; https://​doi.​org/​10.​1155/​2012/​621496.
29.
go back to reference Fick A. Ueber Diffusion. Annalen der Physik und Chemie von J C Pogendorff. 1855;94:59–86.CrossRef Fick A. Ueber Diffusion. Annalen der Physik und Chemie von J C Pogendorff. 1855;94:59–86.CrossRef
30.
33.
go back to reference Wilde D, de Wilde RBP, Schreuder JJ, van den Berg PCM, Jansen JRC. An evaluation of five pulse arterial contour techniques during cardiac surgery. Anesthesia. 2007;62:760–8.CrossRef Wilde D, de Wilde RBP, Schreuder JJ, van den Berg PCM, Jansen JRC. An evaluation of five pulse arterial contour techniques during cardiac surgery. Anesthesia. 2007;62:760–8.CrossRef
39.
go back to reference Magliocca A, Rezoagli E, Anderson TA, Burns SM, Ichinose F, Chitilian HV. Cardiac output measurements based on the pulse wave transit time and thoracic impedance exhibit limited agreement with Thermodilution method during orthotopic liver transplantation. Anesth Analg. 2017; https://doi.org/10.1213/ANE.0000000000002171 Magliocca A, Rezoagli E, Anderson TA, Burns SM, Ichinose F, Chitilian HV. Cardiac output measurements based on the pulse wave transit time and thoracic impedance exhibit limited agreement with Thermodilution method during orthotopic liver transplantation. Anesth Analg. 2017; https://​doi.​org/​10.​1213/​ANE.​0000000000002171​
46.
go back to reference Franklin DL, Schlegel W, Rushmer RF. Blood flow measured by Doppler frequency shift of back scattered ultrasound. Science. 1961;134:564–5.CrossRefPubMed Franklin DL, Schlegel W, Rushmer RF. Blood flow measured by Doppler frequency shift of back scattered ultrasound. Science. 1961;134:564–5.CrossRefPubMed
47.
go back to reference Huntsman LL, Stewart DK, Barnes SR, Franklin SB, Colocousis JS, Hessel EA. Noninvasive Doppler determination of cardiac output in man. Clinical validation Circulation. 1983;67(3):593–602.CrossRefPubMed Huntsman LL, Stewart DK, Barnes SR, Franklin SB, Colocousis JS, Hessel EA. Noninvasive Doppler determination of cardiac output in man. Clinical validation Circulation. 1983;67(3):593–602.CrossRefPubMed
48.
go back to reference •• Hatle L, Angelsen B. Doppler ultrasound in cardiology. Physical principles and applications. Philadelphia: Lea and Febiger; 1982. p. 192. A seminal work that should be compulsory reading for anybody intending to use echocardiography or Doppler techniques to measure volume flows in the heart. The scientific basis of the methods and more importantly, the limitations and pitfalls of the techniques are well explained and are both enlightening and sobering. This should be viewed along with reference 89 for a more complete understanding of the issues involved in using echocardiographic techniques at the point of care. •• Hatle L, Angelsen B. Doppler ultrasound in cardiology. Physical principles and applications. Philadelphia: Lea and Febiger; 1982. p. 192. A seminal work that should be compulsory reading for anybody intending to use echocardiography or Doppler techniques to measure volume flows in the heart. The scientific basis of the methods and more importantly, the limitations and pitfalls of the techniques are well explained and are both enlightening and sobering. This should be viewed along with reference 89 for a more complete understanding of the issues involved in using echocardiographic techniques at the point of care.
50.
go back to reference Gillam LD, Kritzer GL, Ascah KJ. Which cardiac valve provides the best Doppler estimate of cardiac output in humans. Circulation. 1985;72(Suppl III):99. Gillam LD, Kritzer GL, Ascah KJ. Which cardiac valve provides the best Doppler estimate of cardiac output in humans. Circulation. 1985;72(Suppl III):99.
51.
go back to reference McLennan FM, Haites NE, Mackenzie JD, Daniel MK, Rawles JM. Reproducibility of linear cardiac output measurement of Doppler ultrasound alone. Br Heart J. 1986;55:25–31.CrossRefPubMedPubMedCentral McLennan FM, Haites NE, Mackenzie JD, Daniel MK, Rawles JM. Reproducibility of linear cardiac output measurement of Doppler ultrasound alone. Br Heart J. 1986;55:25–31.CrossRefPubMedPubMedCentral
52.
go back to reference Ihlen H, Endresen K, Myreng Y, Myhre E. Reproducibility of cardiac stroke volume estimated by Doppler echocardiography. Am J Cardiol. 1987;59(9):975–8.CrossRefPubMed Ihlen H, Endresen K, Myreng Y, Myhre E. Reproducibility of cardiac stroke volume estimated by Doppler echocardiography. Am J Cardiol. 1987;59(9):975–8.CrossRefPubMed
53.
go back to reference Moulinier L, Venet T, Schiller NB, Kurtz TW, Morris RC, Sebastian A. Measurement of aortic blood flow by Doppler echocardiography: day to day variability in normal subjects and applicability in clinical research. J Am Coll Cardiol. 1991;17:1326–33.CrossRefPubMed Moulinier L, Venet T, Schiller NB, Kurtz TW, Morris RC, Sebastian A. Measurement of aortic blood flow by Doppler echocardiography: day to day variability in normal subjects and applicability in clinical research. J Am Coll Cardiol. 1991;17:1326–33.CrossRefPubMed
54.
go back to reference Kusumoto FM, Venet T, Schiller NB, Sebastian A, Foster E. Measurement of aortic blood flow by Doppler echocardiography: temporal, technician and reader variability in normal subjects and the application of generalizability theory in clinical research. J Am Soc Echocardiogr. 1995;8:647–53.CrossRefPubMed Kusumoto FM, Venet T, Schiller NB, Sebastian A, Foster E. Measurement of aortic blood flow by Doppler echocardiography: temporal, technician and reader variability in normal subjects and the application of generalizability theory in clinical research. J Am Soc Echocardiogr. 1995;8:647–53.CrossRefPubMed
55.
go back to reference Phillips RA, Dadd MJ, Gill RW, West MJ, Burstow DJ. Transcutaneous continuous wave Doppler cardiac output monitoring is feasible producing reliable and reproducible signals. J Am Coll Cardiol. 2002;39(supplB):283B. Phillips RA, Dadd MJ, Gill RW, West MJ, Burstow DJ. Transcutaneous continuous wave Doppler cardiac output monitoring is feasible producing reliable and reproducible signals. J Am Coll Cardiol. 2002;39(supplB):283B.
56.
go back to reference Phillips RA, Paradisis M, Evans NJ, Southwell DL, Burstow DJ, West MJ. Validation of USCOM CO measurements in preterm neonates by comparison with echocardiography. Crit Care. 2006;10(Supl1):144. Phillips RA, Paradisis M, Evans NJ, Southwell DL, Burstow DJ, West MJ. Validation of USCOM CO measurements in preterm neonates by comparison with echocardiography. Crit Care. 2006;10(Supl1):144.
57.
go back to reference Su BC, Yu HP, Yang MW, Lin CC, Kao MC, Chang CH, et al. Reliability of a new ultrasonic cardiac output monitor in recipients of living donor liver transplantation. Liver Transpl. 2008;14:1029–37.CrossRefPubMed Su BC, Yu HP, Yang MW, Lin CC, Kao MC, Chang CH, et al. Reliability of a new ultrasonic cardiac output monitor in recipients of living donor liver transplantation. Liver Transpl. 2008;14:1029–37.CrossRefPubMed
60.
go back to reference Walker A, Olsson E, Wranne B, Ringqvist I, Ask P. Accuracy of spectral Doppler flow and tissue velocity measurements in ultrasound systems. Ultrasound Med Biol. 2004;30(1):127–32.CrossRefPubMed Walker A, Olsson E, Wranne B, Ringqvist I, Ask P. Accuracy of spectral Doppler flow and tissue velocity measurements in ultrasound systems. Ultrasound Med Biol. 2004;30(1):127–32.CrossRefPubMed
61.
go back to reference Gent R. Applied Physics and Technology of Diagnostic Ultrasound. Milner Publishing 1997. Gent R. Applied Physics and Technology of Diagnostic Ultrasound. Milner Publishing 1997.
62.
go back to reference Quinones MA, et al. Recommendations for quantification of Doppler echocardiography: a report from the Doppler Quantification Task Force of the Nomenclature and Standards Committee of the American Society of Echocardiography. J Am Soc Echocardiogr. 2002;15(2):167–84.CrossRefPubMed Quinones MA, et al. Recommendations for quantification of Doppler echocardiography: a report from the Doppler Quantification Task Force of the Nomenclature and Standards Committee of the American Society of Echocardiography. J Am Soc Echocardiogr. 2002;15(2):167–84.CrossRefPubMed
63.
go back to reference Mark JB, Steinbrook RA, Gugino LD, Maddi R, Hartwell B, Shemin R, et al. Continuous noninvasive monitoring of cardiac output with esophageal Doppler ultrasound during cardiac surgery. Anesth Analg. 1986;65(10):1013–20.CrossRefPubMed Mark JB, Steinbrook RA, Gugino LD, Maddi R, Hartwell B, Shemin R, et al. Continuous noninvasive monitoring of cardiac output with esophageal Doppler ultrasound during cardiac surgery. Anesth Analg. 1986;65(10):1013–20.CrossRefPubMed
64.
go back to reference Boyd E. The growth of the surface area of the human body. Minneapolis: University of Minnesota Press; 1935. Boyd E. The growth of the surface area of the human body. Minneapolis: University of Minnesota Press; 1935.
65.
go back to reference Tanner JM. Fallacy of per weight and per surface area standards and their relation to spurious correlation. J Apllied Physiol. 1949;2(1):1–15. Tanner JM. Fallacy of per weight and per surface area standards and their relation to spurious correlation. J Apllied Physiol. 1949;2(1):1–15.
66.
go back to reference Gutgesell HP, Rembold CM. Growth of the human heart relative to body surface area. Am J Cardiol. 1990;65(9):662–8.CrossRefPubMed Gutgesell HP, Rembold CM. Growth of the human heart relative to body surface area. Am J Cardiol. 1990;65(9):662–8.CrossRefPubMed
67.
go back to reference Nidorf SM, Picard MH, Triulzi MO, Thomas JD, Newell J, King ME, et al. New perspectives in the assessment of cardiac chamber dimensions during development and adulthood. J Am Coll Cardiol. 1992;19(5):983–8.CrossRefPubMed Nidorf SM, Picard MH, Triulzi MO, Thomas JD, Newell J, King ME, et al. New perspectives in the assessment of cardiac chamber dimensions during development and adulthood. J Am Coll Cardiol. 1992;19(5):983–8.CrossRefPubMed
68.
go back to reference Sheil ML, Jenkins O, Sholler GF. Echocardiographic assessment of aortic root dimensions in normal children based on measurement of a new ratio of aortic size independent of growth. Am J Cardiol. 1995;75(10):711–5.CrossRefPubMed Sheil ML, Jenkins O, Sholler GF. Echocardiographic assessment of aortic root dimensions in normal children based on measurement of a new ratio of aortic size independent of growth. Am J Cardiol. 1995;75(10):711–5.CrossRefPubMed
69.
go back to reference Capps SB, Elkins RC, Fronk DM. Body surface area as a predictor of aortic and pulmonary valve diameter. J Thorac Cardiovasc Surg. 2000;119(5):975–82.CrossRefPubMed Capps SB, Elkins RC, Fronk DM. Body surface area as a predictor of aortic and pulmonary valve diameter. J Thorac Cardiovasc Surg. 2000;119(5):975–82.CrossRefPubMed
71.
go back to reference Weber E. Wagner’s Handwörterbuch der Physiologie mit Rücksicht auf physiologische Pathologie. Leipzig Vieweg. 1846;3(pt II):110. Weber E. Wagner’s Handwörterbuch der Physiologie mit Rücksicht auf physiologische Pathologie. Leipzig Vieweg. 1846;3(pt II):110.
72.
go back to reference Suga H. Left ventricular time-varying pressure-volume ratio in systole as an index of inotropism. Jpn Heart J. 1971;12(2):153–60.CrossRefPubMed Suga H. Left ventricular time-varying pressure-volume ratio in systole as an index of inotropism. Jpn Heart J. 1971;12(2):153–60.CrossRefPubMed
73.
go back to reference Suga H, Sagawa K. Instantaneous pressure-volume relationships and their ratio in the excised, supported canine left ventricle. Circ Res. 1974;35(1):117–26.CrossRefPubMed Suga H, Sagawa K. Instantaneous pressure-volume relationships and their ratio in the excised, supported canine left ventricle. Circ Res. 1974;35(1):117–26.CrossRefPubMed
74.
go back to reference Sagawa K, Suga H, Shoukas AA, Bakalar KM. End-systolic pressure/volume ratio: a new index of ventricular contractility. Am J Cardiol. 1977;40(5):748–53.CrossRefPubMed Sagawa K, Suga H, Shoukas AA, Bakalar KM. End-systolic pressure/volume ratio: a new index of ventricular contractility. Am J Cardiol. 1977;40(5):748–53.CrossRefPubMed
75.
go back to reference Sagawa K, Maughan L, Suga H, Sunagawa K. Cardiac contraction and the pressure-volume relationship. New-York: Oxford University Press; 1988. p. 42–106. Sagawa K, Maughan L, Suga H, Sunagawa K. Cardiac contraction and the pressure-volume relationship. New-York: Oxford University Press; 1988. p. 42–106.
76.
go back to reference Williams SG, Cooke GA, Wright DJ, Parsons WJ, Riley RL, Marshall P, et al. Peak exercise cardiac output power; a direct indicator of cardiac function strongly predictive of prognosis in chronic heart failure. Eur Heart J. 2001;22(16):1496–503.CrossRefPubMed Williams SG, Cooke GA, Wright DJ, Parsons WJ, Riley RL, Marshall P, et al. Peak exercise cardiac output power; a direct indicator of cardiac function strongly predictive of prognosis in chronic heart failure. Eur Heart J. 2001;22(16):1496–503.CrossRefPubMed
77.
78.
go back to reference Fincke R, Hochman JS, Lowe AM, Menon V, Slater JN, Webb JG, et al. Cardiac power is the strongest hemodynamic correlate of mortality in cardiogenic shock: a report from the SCOTT trial register. J Am Coll Cardiol. 2004;44(2):340–8.CrossRefPubMed Fincke R, Hochman JS, Lowe AM, Menon V, Slater JN, Webb JG, et al. Cardiac power is the strongest hemodynamic correlate of mortality in cardiogenic shock: a report from the SCOTT trial register. J Am Coll Cardiol. 2004;44(2):340–8.CrossRefPubMed
79.
go back to reference Glower DD, Spratt JA, Snow ND, Kabas JS, Davis JW, Olsen CO, et al. Linearity of the Frank-Starling relationship in the intact heart: the concept of preload recruitable stroke work. Circulation. 1985;71(5):994–1009.CrossRefPubMed Glower DD, Spratt JA, Snow ND, Kabas JS, Davis JW, Olsen CO, et al. Linearity of the Frank-Starling relationship in the intact heart: the concept of preload recruitable stroke work. Circulation. 1985;71(5):994–1009.CrossRefPubMed
83.
go back to reference Oommen B, Karamanoglu M, Kovács SJ. Modeling time varying elastance: the meaning of “load-independence”. Cardiovasc Eng. 2003;3(4):123–30.CrossRef Oommen B, Karamanoglu M, Kovács SJ. Modeling time varying elastance: the meaning of “load-independence”. Cardiovasc Eng. 2003;3(4):123–30.CrossRef
91.
go back to reference • Porter TR, Shillcutt SK, Adams MS, et al. Guidelines for the use of echocardiography as a monitor for therapeutic intervention in adults: a report from the American Society of Echocardiography. J Am Soc Echocardiogr. 2015;28(1):40–56. https://doi.org/10.1016/j.echo.2014.09.009. With the huge growth in the use of bedside echocardiography in critical care, it is salutary to understand just what echocardiography can do, and more importantly, what it cannot do reliably or easily. The strict techniques and considerable technical challenges that must be overcome for the accurate measurement of volume flows are daunting for the non-expert echocardiographer. Although a fairly technical document, understanding of the limitations of bedside echocardiography is essential for any practitioner intending to use such data in the therapeutic situation. CrossRefPubMed • Porter TR, Shillcutt SK, Adams MS, et al. Guidelines for the use of echocardiography as a monitor for therapeutic intervention in adults: a report from the American Society of Echocardiography. J Am Soc Echocardiogr. 2015;28(1):40–56. https://​doi.​org/​10.​1016/​j.​echo.​2014.​09.​009. With the huge growth in the use of bedside echocardiography in critical care, it is salutary to understand just what echocardiography can do, and more importantly, what it cannot do reliably or easily. The strict techniques and considerable technical challenges that must be overcome for the accurate measurement of volume flows are daunting for the non-expert echocardiographer. Although a fairly technical document, understanding of the limitations of bedside echocardiography is essential for any practitioner intending to use such data in the therapeutic situation. CrossRefPubMed
92.
go back to reference Singer M, Clarke J, Bennett ED. Continuous hemodynamic monitoring by esophageal Doppler. Crit Care Med. 1989;17(5):447–52.CrossRefPubMed Singer M, Clarke J, Bennett ED. Continuous hemodynamic monitoring by esophageal Doppler. Crit Care Med. 1989;17(5):447–52.CrossRefPubMed
93.
go back to reference Critchley LA, Peng ZY, Fok BS, Flach J, Wong SC, Lee A, et al. Testing the reliability of a new ultrasonic cardiac output monitor, the USCOM, using aortic flow probes in anaesthetized dogs. Anesth Analg. 2005;100:748–53.CrossRefPubMed Critchley LA, Peng ZY, Fok BS, Flach J, Wong SC, Lee A, et al. Testing the reliability of a new ultrasonic cardiac output monitor, the USCOM, using aortic flow probes in anaesthetized dogs. Anesth Analg. 2005;100:748–53.CrossRefPubMed
94.
go back to reference Critchley LA. Differences between Cardio-Q and USCOM Doppler cardiac output readings in high risk surgical patients. Br J Anaesth. 2012;108(S2):ii113. Critchley LA. Differences between Cardio-Q and USCOM Doppler cardiac output readings in high risk surgical patients. Br J Anaesth. 2012;108(S2):ii113.
Metadata
Title
Stroke Volume Monitoring: Novel Continuous Wave Doppler Parameters, Algorithms and Advanced Noninvasive Haemodynamic Concepts
Authors
R. A. Phillips
B. E. Smith
V. M. Madigan
Publication date
01-12-2017
Publisher
Springer US
Published in
Current Anesthesiology Reports / Issue 4/2017
Electronic ISSN: 2167-6275
DOI
https://doi.org/10.1007/s40140-017-0235-4

Other articles of this Issue 4/2017

Current Anesthesiology Reports 4/2017 Go to the issue

Patient Safety in Anesthesia (A Barbeito, Section Editor)

Transitions of Care in the Perioperative Period

Patient Safety in Anesthesia (A Barbeito, Section Editor)

Teamwork, Safety, and Non-Technical Skills

Geriatric Anesthesia (S Akhtar, Section Editor)

Regional Versus General Anesthesia in the Elderly: New Insights

Patient Safety in Anesthesia (A Barbeito, Section Editor)

Improvement Science in Anaesthesia

Advances in Monitoring for Anesthesia (LAH Critchley, Section Editor)

Cardiac Output Monitoring: Validation Studies–how Results Should be Presented