Open Access 01-08-2017 | Review

Cardiovascular implantable electronic devices: a review of the dangers and difficulties in MR scanning and attempts to improve safety

Authors: Pei Ghim Poh, Charlene Liew, Colin Yeo, Le Roy Chong, Andrew Tan, Angeline Poh

Published in: Insights into Imaging | Issue 4/2017

Login to get access

Abstract

An increasing number of patients are being treated with cardiovascular implantable electronic devices (CIEDs), many of which are MR conditional. There is a lack of literature on the safe scanning of MR conditional CIEDs. This review article discusses MR imaging safety in patients with implanted CIEDs. Guidelines on safe use and indications of imaging patients with MR conditional CIEDs are described, followed by a pictorial essay of the radiographic features of these devices. We also discuss the challenges of monitoring the patient in the MR environment, advances in MRI conditional imaging of devices, availability, limitations and workflow including vendor-specific and other collaborative efforts to simplify the scanning process. Radiologists must be able to facilitate the safe utilization of MR imaging in patients who have CIEDs. A thorough knowledge of the hazards of imaging non-MR compatible devices is required as well as knowing how to correctly identify and manage the imaging of patients with MR conditional CIEDs. Finally, we propose steps required to facilitate the safe scanning of patients with MR conditional CIEDs adopted in our institution and a contingency plan in the event that an inadvertent MR scan of a patient with a MRI unsafe CIED should occur.

Main Messages

Risks of MR imaging in patients who have CIEDs have been worked around.
There are many technical limitations in enabling safe MR scanning of CIEDs.
Radiological identification of MRI-conditional status of CIEDs is useful.
Standardizing conditions for safe MRI scanning is important.
We offer example algorithms for facilitating safe MRI scanning of CIEDs.
Literature
1.
go back to reference Götte MJW, Rüssel IK, de Roest GJ et al (2010) Magnetic resonance imaging, pacemakers and implantable cardioverter-defibrillators: current situation and clinical perspective. Neth Hear J 18(1):31–37 Götte MJW, Rüssel IK, de Roest GJ et al (2010) Magnetic resonance imaging, pacemakers and implantable cardioverter-defibrillators: current situation and clinical perspective. Neth Hear J 18(1):31–37
3.
go back to reference Shellock FG, Spinazzi A (2008) MRI safety update 2008: part 2, screening patients for MRI. AJR Am J Roentgenol 191(4):1140–1149CrossRefPubMed Shellock FG, Spinazzi A (2008) MRI safety update 2008: part 2, screening patients for MRI. AJR Am J Roentgenol 191(4):1140–1149CrossRefPubMed
4.
go back to reference Shinbane J, Colletti P, Shellock FG (2007) MR in patients with pacemakers and ICDs: defining the issues. J Cardiovasc Magn Reson 9(1):5–13CrossRefPubMed Shinbane J, Colletti P, Shellock FG (2007) MR in patients with pacemakers and ICDs: defining the issues. J Cardiovasc Magn Reson 9(1):5–13CrossRefPubMed
5.
go back to reference Nazarian S, Hansford R, Roguin A et al (2011) A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med 155(7):415–424CrossRefPubMedPubMedCentral Nazarian S, Hansford R, Roguin A et al (2011) A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med 155(7):415–424CrossRefPubMedPubMedCentral
6.
go back to reference Shinbane JS et al (2011) Magnetic resonance imaging in patients with cardiac pacemakers: era of "MR conditional" designs. J Cardiovasc Magn Reson 13:63CrossRefPubMedPubMedCentral Shinbane JS et al (2011) Magnetic resonance imaging in patients with cardiac pacemakers: era of "MR conditional" designs. J Cardiovasc Magn Reson 13:63CrossRefPubMedPubMedCentral
7.
go back to reference Del Ojo JL, Moya F, Villalba J, Sanz O, Pavon R, Garcia D, Pastor L (2005) Is magnetic resonance imaging safe in cardiac pacemaker recipients? Pacing Clin Electrophysiol 28(4):274–278CrossRefPubMed Del Ojo JL, Moya F, Villalba J, Sanz O, Pavon R, Garcia D, Pastor L (2005) Is magnetic resonance imaging safe in cardiac pacemaker recipients? Pacing Clin Electrophysiol 28(4):274–278CrossRefPubMed
8.
go back to reference Naehle CP, Meyer C, Thomas D et al (2008) Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology 249(3):991–1001CrossRefPubMed Naehle CP, Meyer C, Thomas D et al (2008) Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology 249(3):991–1001CrossRefPubMed
10.
go back to reference Beinart R, Nazarian S (2013) Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice. Circulation 128(25):2799–2809CrossRefPubMedPubMedCentral Beinart R, Nazarian S (2013) Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice. Circulation 128(25):2799–2809CrossRefPubMedPubMedCentral
11.
go back to reference Luechinger R, Zeijlemaker VA, Pedersen EM et al (2005) In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J 26(4):376–383CrossRefPubMed Luechinger R, Zeijlemaker VA, Pedersen EM et al (2005) In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J 26(4):376–383CrossRefPubMed
12.
go back to reference Dempsey MF, Condon B, Hadley DM (2001) Investigation of the factors responsible for burns during MRI. J Magn Reson Imaging 13(4):627–631CrossRefPubMed Dempsey MF, Condon B, Hadley DM (2001) Investigation of the factors responsible for burns during MRI. J Magn Reson Imaging 13(4):627–631CrossRefPubMed
13.
go back to reference Mattei E, Triventi M, Calcagnini G et al (2008) Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations. Biomed Eng Online 7:11CrossRefPubMedPubMedCentral Mattei E, Triventi M, Calcagnini G et al (2008) Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations. Biomed Eng Online 7:11CrossRefPubMedPubMedCentral
14.
go back to reference Fontaine JM, Mohamed FB, Gottlieb C, Callans DJ, Marchlinski FE (1998) Rapid ventricular pacing in a pacemaker patient undergoing magnetic resonance imaging. Pacing Clin Electrophysiol 21:1336–1339CrossRefPubMed Fontaine JM, Mohamed FB, Gottlieb C, Callans DJ, Marchlinski FE (1998) Rapid ventricular pacing in a pacemaker patient undergoing magnetic resonance imaging. Pacing Clin Electrophysiol 21:1336–1339CrossRefPubMed
15.
go back to reference Jacob S, Panaich SS, Maheshwari R, Haddad JW, Padanilam BJ, John SK (2011) Clinical applications of magnets on cardiac rhythm management devices. Europace 13(9):1222–1230CrossRefPubMed Jacob S, Panaich SS, Maheshwari R, Haddad JW, Padanilam BJ, John SK (2011) Clinical applications of magnets on cardiac rhythm management devices. Europace 13(9):1222–1230CrossRefPubMed
16.
go back to reference Bottomley PA et al (2010) Designing passive MRI-safe implantable conducting leads with electrodes. Med Phys 37(7):3828–3843 Medtronic For Healthcare Professionals: Pacing LeadsCrossRefPubMed Bottomley PA et al (2010) Designing passive MRI-safe implantable conducting leads with electrodes. Med Phys 37(7):3828–3843 Medtronic For Healthcare Professionals: Pacing LeadsCrossRefPubMed
17.
go back to reference Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfield TM, Bluemke DA et al (2004) Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe. In vitro and invivo assessment of safety and function at 1.5T. Circulation 110:475–482CrossRefPubMedPubMedCentral Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfield TM, Bluemke DA et al (2004) Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe. In vitro and invivo assessment of safety and function at 1.5T. Circulation 110:475–482CrossRefPubMedPubMedCentral
18.
go back to reference Crossley GH, Brinker JA, Reynolds D et al (1995) Steroid elution improves the stimulation threshold in an active-fixation atrial permanent pacing lead. A randomized, controlled study. Model 4068 investigators. Circulation 92(10):2935–2939CrossRefPubMed Crossley GH, Brinker JA, Reynolds D et al (1995) Steroid elution improves the stimulation threshold in an active-fixation atrial permanent pacing lead. A randomized, controlled study. Model 4068 investigators. Circulation 92(10):2935–2939CrossRefPubMed
19.
go back to reference Schaldach M, Hubmann M, Hardt R, Weikl A (1989) Titanium nitride cardiac pacemaker electrodes. Biomed Tech (Berl) 34(7–8):185–190CrossRef Schaldach M, Hubmann M, Hardt R, Weikl A (1989) Titanium nitride cardiac pacemaker electrodes. Biomed Tech (Berl) 34(7–8):185–190CrossRef
21.
go back to reference Kanal E, Barkovich AJ, Bell C et al (2007) ACR guidance document for safe MR practices: 2007. AJR Am J Roentgenol 188:1447–1474CrossRefPubMed Kanal E, Barkovich AJ, Bell C et al (2007) ACR guidance document for safe MR practices: 2007. AJR Am J Roentgenol 188:1447–1474CrossRefPubMed
22.
23.
go back to reference Greenberg KL, Weinreb J, Shellock FG (2011) "MR conditional" respiratory ventilator system incident in a 3-T MRI environment. Magn Reson Imaging 29:1150–1154CrossRefPubMed Greenberg KL, Weinreb J, Shellock FG (2011) "MR conditional" respiratory ventilator system incident in a 3-T MRI environment. Magn Reson Imaging 29:1150–1154CrossRefPubMed
24.
go back to reference Nasr VG et al (2012) Performance validation of a modified magnetic resonance imaging-compatible temperature probe in children. Anesth Analg 114:1230–1234CrossRefPubMed Nasr VG et al (2012) Performance validation of a modified magnetic resonance imaging-compatible temperature probe in children. Anesth Analg 114:1230–1234CrossRefPubMed
25.
go back to reference Verma A, Ha AC, Dennie C et al (2014) Canadian Heart Rhythm Society and Canadian Association of Radiologists consensus statement on magnetic resonance imaging with cardiac implantable electronic devices. Can J Cardiol 30(10):1131–1141CrossRefPubMed Verma A, Ha AC, Dennie C et al (2014) Canadian Heart Rhythm Society and Canadian Association of Radiologists consensus statement on magnetic resonance imaging with cardiac implantable electronic devices. Can J Cardiol 30(10):1131–1141CrossRefPubMed
26.
go back to reference Jung W, Zvereva V, Hajredini B, Jäckle S (2012) Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directions. Europace 14(5):631–637CrossRefPubMed Jung W, Zvereva V, Hajredini B, Jäckle S (2012) Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directions. Europace 14(5):631–637CrossRefPubMed
27.
go back to reference Brignole M, Auricchio A, Baron-Esquivias G et al (2013) 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Europace 15:1070–1118CrossRefPubMed Brignole M, Auricchio A, Baron-Esquivias G et al (2013) 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Europace 15:1070–1118CrossRefPubMed
28.
go back to reference Russo RJ (2013) Determining the risks of clinically indicated nonthoracic magnetic resonance imaging at 1.5 T for patients with pacemakers and implantable cardioverter-defibrillators: rationale and design of the MagnaSafe Registry. Am Heart J 165:266–272CrossRefPubMed Russo RJ (2013) Determining the risks of clinically indicated nonthoracic magnetic resonance imaging at 1.5 T for patients with pacemakers and implantable cardioverter-defibrillators: rationale and design of the MagnaSafe Registry. Am Heart J 165:266–272CrossRefPubMed
29.
go back to reference Russo RJ, Costa HS, Silva PD et al (2017) Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med 376(8):755–764CrossRefPubMed Russo RJ, Costa HS, Silva PD et al (2017) Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med 376(8):755–764CrossRefPubMed
30.
go back to reference Moss AJ, Kutyifa V (2015) Safe MRI in patients with an upgraded (conditional) implantable cardioverter-defibrillator: the beneficial tip of a troublesome iceberg∗. J Am Coll Cardiol 65(24):2589–2590CrossRefPubMed Moss AJ, Kutyifa V (2015) Safe MRI in patients with an upgraded (conditional) implantable cardioverter-defibrillator: the beneficial tip of a troublesome iceberg∗. J Am Coll Cardiol 65(24):2589–2590CrossRefPubMed
31.
go back to reference Ferreira AM, Costa F, Tralhão A, Marques H, Cardim N, Adragão P (2014) MRI-conditional pacemakers: current perspectives. Med Devices (Auckl) 7:115–124CrossRef Ferreira AM, Costa F, Tralhão A, Marques H, Cardim N, Adragão P (2014) MRI-conditional pacemakers: current perspectives. Med Devices (Auckl) 7:115–124CrossRef
32.
go back to reference Gimbel JR, Bailey SM, Tchou PJ, Ruggieri PM, Wilkoff BL (2005) Strategies for the safe magnetic resonance imaging of pacemaker-dependent patients. Pacing Clin Electrophysiol 28:1041–1046CrossRefPubMed Gimbel JR, Bailey SM, Tchou PJ, Ruggieri PM, Wilkoff BL (2005) Strategies for the safe magnetic resonance imaging of pacemaker-dependent patients. Pacing Clin Electrophysiol 28:1041–1046CrossRefPubMed
35.
go back to reference Savouré A, Mechulan A, Burban M, Olivier A, Lazarus A (2015) The Kora pacemaker is safe and effective for magnetic resonance imaging. Clin Med Insights Cardiol 9:85–90CrossRefPubMedPubMedCentral Savouré A, Mechulan A, Burban M, Olivier A, Lazarus A (2015) The Kora pacemaker is safe and effective for magnetic resonance imaging. Clin Med Insights Cardiol 9:85–90CrossRefPubMedPubMedCentral
36.
go back to reference Cronin EM, Wilkoff BL (2012) Magnetic resonance imaging conditional pacemakers: rationale, development and future directions. Indian Pacing Electrophysiol J 12(5):204–212CrossRefPubMedPubMedCentral Cronin EM, Wilkoff BL (2012) Magnetic resonance imaging conditional pacemakers: rationale, development and future directions. Indian Pacing Electrophysiol J 12(5):204–212CrossRefPubMedPubMedCentral
37.
go back to reference Alandete Germán SP, Isarria Vidal S, Domingo Montañana ML, Dela Vía Oraá E, Vilar Samper J (2015) Pacemakers and implantable cardioverter defibrillators, unknown to chest radiography: review, complications and systematic reading. Eur J Radiol 84(3):499–508CrossRefPubMed Alandete Germán SP, Isarria Vidal S, Domingo Montañana ML, Dela Vía Oraá E, Vilar Samper J (2015) Pacemakers and implantable cardioverter defibrillators, unknown to chest radiography: review, complications and systematic reading. Eur J Radiol 84(3):499–508CrossRefPubMed
38.
go back to reference Costelloe CM, Murphy WA, Gladish GW, Rozner MA (2012) Radiography of pacemakers and implantable cardioverter defibrillators. AJR Am J Roentgenol 199(6):1252–1258CrossRefPubMed Costelloe CM, Murphy WA, Gladish GW, Rozner MA (2012) Radiography of pacemakers and implantable cardioverter defibrillators. AJR Am J Roentgenol 199(6):1252–1258CrossRefPubMed
39.
go back to reference Aguilera AL, Volokhina YV, Fisher KL (2011) Radiography of cardiac conduction devices: a comprehensive review. Radiographics 31(6):1669–1682CrossRefPubMed Aguilera AL, Volokhina YV, Fisher KL (2011) Radiography of cardiac conduction devices: a comprehensive review. Radiographics 31(6):1669–1682CrossRefPubMed
40.
go back to reference Keller J, Neužil P, Vymazal J et al (2015) Magnetic resonance imaging in patients with a subcutaneous implantable cardioverter-defibrillator. Europace 17(5):761–766CrossRefPubMedPubMedCentral Keller J, Neužil P, Vymazal J et al (2015) Magnetic resonance imaging in patients with a subcutaneous implantable cardioverter-defibrillator. Europace 17(5):761–766CrossRefPubMedPubMedCentral
41.
go back to reference Gimbel JR, Zarghami J, Machado C, Wilkoff BL (2005) Safe scanning, but frequent artifacts mimicking bradycardia and tachycardia during magnetic resonance imaging (MRI) in patients with an implantable loop recorder (ILR). Ann Noninvasive Electrocardiol 10(4):404–408CrossRefPubMed Gimbel JR, Zarghami J, Machado C, Wilkoff BL (2005) Safe scanning, but frequent artifacts mimicking bradycardia and tachycardia during magnetic resonance imaging (MRI) in patients with an implantable loop recorder (ILR). Ann Noninvasive Electrocardiol 10(4):404–408CrossRefPubMed
43.
go back to reference Langman DA et al (2011) Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 15 tesla MRI. J Magn Reson Imaging 33:426–431CrossRefPubMed Langman DA et al (2011) Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 15 tesla MRI. J Magn Reson Imaging 33:426–431CrossRefPubMed
44.
go back to reference PG Poh, Liew JY, Poh CC. Imaging Cyborgs: Safe MRI of Patients With Cardiovascular Implantable Electronic Devices. Poster Presented at: American Roentgen Ray Society Annual Meeting 2014, Toronto. (E2621) PG Poh, Liew JY, Poh CC. Imaging Cyborgs: Safe MRI of Patients With Cardiovascular Implantable Electronic Devices. Poster Presented at: American Roentgen Ray Society Annual Meeting 2014, Toronto. (E2621)
Metadata
Title
Cardiovascular implantable electronic devices: a review of the dangers and difficulties in MR scanning and attempts to improve safety
Authors
Pei Ghim Poh
Charlene Liew
Colin Yeo
Le Roy Chong
Andrew Tan
Angeline Poh
Publication date
01-08-2017
Publisher
Springer Berlin Heidelberg
Published in
Insights into Imaging / Issue 4/2017
Electronic ISSN: 1869-4101
DOI
https://doi.org/10.1007/s13244-017-0556-3