Abstract

Purpose

A stylet is usually necessary when using the GlideScope® videolaryngoscope for orotracheal intubation. A special stylet, the GlideRite® Rigid Stylet (GRS), was designed for this purpose. A previous trial involving experienced operators showed that the GRS offered no performance improvement vs a standard malleable stylet (SMS). In our trial, we compared the performance of the GRS with that of the SMS in terms of time to intubation and ease of intubation when used by novice GlideScope® operators.

Methods

Sixty patients with normal-appearing airways requiring orotracheal intubation for elective surgery were randomly allocated to be intubated by novice operators with the GlideScope®, using either the GRS or the SMS. Time to intubation was assessed by a blinded observer, and the operators were blinded until just prior to tracheal intubation. Ease of intubation was assessed by a five-point ordinal scale (from 1- easy to 5 -difficult). Intubation attempts/failures, glottic grades, and usage of external laryngeal manipulation were recorded.

Results

There were no significant differences between the GRS and the SMS in terms of the median time to intubation (60 sec, interquartile range [IQR] 48-75 vs 61 sec, IQR 49-75, respectively; P = 0.94) and the ease of intubation (GRS median score: 1.5, IQR 1-2 vs SMS median score: 1, IQR 1-2; P = 0.94). There were no other significant differences between groups.

Conclusion

The GRS and the SMS have similar performance characteristics when used by novice operators for GlideScope®-assisted orotracheal intubation. (Registered at ClinicalTrials.gov: NCT00884754).
Footnotes
1
Verathon Medical Inc. GlideScope Video Intubation System - Operator and Service Manual, 2003.
 
Literature
1.
go back to reference Cooper RM, Pacey JA, Bishop MJ, McCluskey SA. Early clinical experience with a new videolaryngoscope (GlideScope) in 728 patients. Can J Anesth 2005; 52: 191-8.PubMedCrossRef Cooper RM, Pacey JA, Bishop MJ, McCluskey SA. Early clinical experience with a new videolaryngoscope (GlideScope) in 728 patients. Can J Anesth 2005; 52: 191-8.PubMedCrossRef
2.
go back to reference Sun DA, Warriner CB, Parsons DG, Klein R, Umedaly HS, Moult M. The GlideScope(R) Video Laryngoscope: randomized clinical trial in 200 patients. Br J Anaesth 2005; 94: 381-4.PubMedCrossRef Sun DA, Warriner CB, Parsons DG, Klein R, Umedaly HS, Moult M. The GlideScope(R) Video Laryngoscope: randomized clinical trial in 200 patients. Br J Anaesth 2005; 94: 381-4.PubMedCrossRef
3.
go back to reference Jones PM, Armstrong KP, Armstrong PM, et al. A comparison of GlideScope videolaryngoscopy to direct laryngoscopy for nasotracheal intubation. Anesth Analg 2008; 107: 144-8.PubMedCrossRef Jones PM, Armstrong KP, Armstrong PM, et al. A comparison of GlideScope videolaryngoscopy to direct laryngoscopy for nasotracheal intubation. Anesth Analg 2008; 107: 144-8.PubMedCrossRef
4.
go back to reference Jones PM, Turkstra TP, Armstrong KP, et al. Effect of stylet angulation and endotracheal tube camber on time to intubation with the GlideScope(R). Can J Anesth 2007; 54: 21-7.PubMedCrossRef Jones PM, Turkstra TP, Armstrong KP, et al. Effect of stylet angulation and endotracheal tube camber on time to intubation with the GlideScope(R). Can J Anesth 2007; 54: 21-7.PubMedCrossRef
6.
go back to reference Cooper RM. Complications associated with the use of the GlideScope videolaryngoscope. Can J Anesth 2007; 54: 54-7.PubMedCrossRef Cooper RM. Complications associated with the use of the GlideScope videolaryngoscope. Can J Anesth 2007; 54: 54-7.PubMedCrossRef
7.
go back to reference Malik AM, Frogel JK. Anterior tonsillar pillar perforation during GlideScope video laryngoscopy. Anesth Analg 2007; 104: 1610-1.PubMedCrossRef Malik AM, Frogel JK. Anterior tonsillar pillar perforation during GlideScope video laryngoscopy. Anesth Analg 2007; 104: 1610-1.PubMedCrossRef
8.
go back to reference van Zundert A, Maassen R, Lee R, et al. A Macintosh laryngoscope blade for videolaryngoscopy reduces stylet use in patients with normal airways. Anesth Analg 2009; 109: 825-31.PubMedCrossRef van Zundert A, Maassen R, Lee R, et al. A Macintosh laryngoscope blade for videolaryngoscopy reduces stylet use in patients with normal airways. Anesth Analg 2009; 109: 825-31.PubMedCrossRef
9.
go back to reference Bader SO, Heitz JW, Audu PB. Tracheal intubation with the GlideScope(R) videolaryngoscope, using a “J” shaped endotracheal tube. Can J Anesth 2006; 53: 634-5.PubMedCrossRef Bader SO, Heitz JW, Audu PB. Tracheal intubation with the GlideScope(R) videolaryngoscope, using a “J” shaped endotracheal tube. Can J Anesth 2006; 53: 634-5.PubMedCrossRef
10.
go back to reference Turkstra TP, Harle CC, Armstrong KP, et al. The GlideScope-specific rigid stylet and standard malleable stylet are equally effective for GlideScope use. Can J Anesth 2007; 54: 891-6.PubMedCrossRef Turkstra TP, Harle CC, Armstrong KP, et al. The GlideScope-specific rigid stylet and standard malleable stylet are equally effective for GlideScope use. Can J Anesth 2007; 54: 891-6.PubMedCrossRef
11.
go back to reference Mallampati SR, Gatt SP, Gugino LD, et al. A clinical sign to predict difficult tracheal intubation: a prospective study. Can Anaesth Soc J 1985; 32: 429-34.PubMedCrossRef Mallampati SR, Gatt SP, Gugino LD, et al. A clinical sign to predict difficult tracheal intubation: a prospective study. Can Anaesth Soc J 1985; 32: 429-34.PubMedCrossRef
12.
13.
go back to reference Lehmann EL, D’Abrera HJ. Nonparametrics: Statistical Methods Based on Ranks. San Francisco: Holden-Day; 1975 . Lehmann EL, D’Abrera HJ. Nonparametrics: Statistical Methods Based on Ranks. San Francisco: Holden-Day; 1975 .
14.
go back to reference Perneger TV. What’s wrong with Bonferroni adjustments. BMJ 1998; 316: 1236-8.PubMed Perneger TV. What’s wrong with Bonferroni adjustments. BMJ 1998; 316: 1236-8.PubMed
Metadata
Title
A randomized comparison of the GlideRite® Rigid Stylet to a malleable stylet for orotracheal intubation by novices using the GlideScope®
Authors
Philip M. Jones, MD
Francis L. C. Loh, MBBS
Hesham N. Youssef, MBBCh
Timothy P. Turkstra, MD
Publication date
01-03-2011
Publisher
Springer-Verlag
Published in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Issue 3/2011
Print ISSN: 0832-610X
Electronic ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-010-9440-z

Other articles of this Issue 3/2011

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 3/2011 Go to the issue