A comparative study of radiation doses between phantom and patients via CT angiography of the intra-/extra-cranial, pulmonary, and abdominal/pelvic arteries

Authors: M. K. A. Karim, A. Sabarudin, N. A. Muhammad, K. H. Ng

Published in: Radiological Physics and Technology | Issue 4/2019

Login to get access

Abstract

This study aimed to evaluate effective dose and size-specific dose estimate (SSDE) of computed tomography angiography (CTA) examination using an anthropomorphic phantom. We included three CTA examination protocols to evaluate the intra- and extra-cranial arteries, pulmonary artery (CTPA), and abdominal vessels. Patient SSDEs were measured retrospectively to estimate patient dose, relative to the bodyweight of the patient and volume CT dose index (CTDIvol). Our findings revealed that the highest dose was absorbed by the left lobe of the thyroid gland during intra-/extra-cranial CTA and CTPA, that is, 14.11 ± 0.24 mGy and 16.20 ± 3.95 mGy, respectively. However, the highest absorbed dose in abdominal/pelvic CTA was the gonads (8.98 ± 0.30 mGy), while other radiosensitive organs in intra- and extra-cranial CTA, CTPA, and abdominal/pelvic CTA did not demonstrate significant differences between organs/structures with p value 0.88, 0.11, and 0.54, respectively. The estimated effective dose in intra-/extra-cranial CTA was lower in patients (0.80 ± 0.60 mSv) than in the phantom (0.83 mSv), but it was the opposite for CTPA, with the effective dose being higher in patients (7.54 ± 3.09 mSv) than in the phantom (6.68 mSv). Similar to the effective dose, only CTPA SSDEs were significantly higher in men than in women (19.74 ± 4.79 mGy versus 7.9 mGy). Effective dose and SSDE are clinically relevant parameters that can help estimate a more accurate patient dose based on a patient’s size.
Literature
1.
go back to reference Jayakrishnan VK, White PM, Aitken D, Crane P, McMahon AD, Teasdale EM. Subtraction helical CT angiography of intra- and extracranial vessels: technical considerations and preliminary experience. AJNR Am J Neuroradiol. 2003;24:451–5.PubMedPubMedCentral Jayakrishnan VK, White PM, Aitken D, Crane P, McMahon AD, Teasdale EM. Subtraction helical CT angiography of intra- and extracranial vessels: technical considerations and preliminary experience. AJNR Am J Neuroradiol. 2003;24:451–5.PubMedPubMedCentral
8.
go back to reference Karim MKA, Hashim S, Sabarudin A, Bradley DA, Bahruddin NA. Evaluating organ dose and radiation risk of routine CT examinations in Johor Malaysia. Sains Malays. 2016;45:567–73. Karim MKA, Hashim S, Sabarudin A, Bradley DA, Bahruddin NA. Evaluating organ dose and radiation risk of routine CT examinations in Johor Malaysia. Sains Malays. 2016;45:567–73.
15.
go back to reference Ma C-M, Seuntjens JP. Mass-energy absorption coefficient and backscatter factor ratios for kilovoltage X-ray beams. Phys Med Biol. 1999;44:131–43.CrossRefPubMed Ma C-M, Seuntjens JP. Mass-energy absorption coefficient and backscatter factor ratios for kilovoltage X-ray beams. Phys Med Biol. 1999;44:131–43.CrossRefPubMed
16.
go back to reference ICRP. Compendium of Dose Coefficients based on ICRP publication 60. ICRP Publication 119. Ann ICRP. 2012;41(Suppl.). ICRP. Compendium of Dose Coefficients based on ICRP publication 60. ICRP Publication 119. Ann ICRP. 2012;41(Suppl.).
20.
go back to reference Commission E. European guidelines on quality criteria for computed tomography. EUR 16262 EN. 1997. Commission E. European guidelines on quality criteria for computed tomography. EUR 16262 EN. 1997.
26.
go back to reference Sabarudin A, Mustafa Z, Nassir KM, Hamid HA, Sun Z. Radiation dose reduction in thoracic and abdomen–pelvic CT using tube current modulation: a phantom study. J Appl Clin Med Phys. 2015;16:319–28.CrossRef Sabarudin A, Mustafa Z, Nassir KM, Hamid HA, Sun Z. Radiation dose reduction in thoracic and abdomen–pelvic CT using tube current modulation: a phantom study. J Appl Clin Med Phys. 2015;16:319–28.CrossRef
Metadata
Title
A comparative study of radiation doses between phantom and patients via CT angiography of the intra-/extra-cranial, pulmonary, and abdominal/pelvic arteries
Authors
M. K. A. Karim
A. Sabarudin
N. A. Muhammad
K. H. Ng
Publication date
01-12-2019
Publisher
Springer Singapore
Keyword
CT Angiography
Published in
Radiological Physics and Technology / Issue 4/2019
Print ISSN: 1865-0333
Electronic ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-019-00532-8

Other articles of this Issue 4/2019

Radiological Physics and Technology 4/2019 Go to the issue

Acknowledgments

Acknowledgment