01-12-2009 | Original Article

Reliability in Interpretation of Median Somatosensory Evoked Potentials in the Setting of Coma: Factors and Implications

Authors: Kevin Hakimi, Greg Kinney, George Kraft, Paula Micklesen, Lawrence Robinson

Published in: Neurocritical Care | Issue 3/2009

Login to get access

Abstract

Background

Median nerve short-latency somatosensory evoked potentials (SSEPs) are useful in determining prognosis for awakening after coma following hypoxic ischemic encephalopathy, but reliability of interpretation is unclear.

Objective

To measure inter- and intra-observer reliability of determining presence or absence of SSEPs in comatose patients following hypoxic ischemic encephalopathy and to identify factors that enhance reliability.

Design

Retrospective review by four readers (experience ranging from 2–27 years) of SSEP recordings in 95 comatose patients. Twenty waveforms were presented twice and five waveforms were presented pre- and post-neuromuscular junction (NMJ) blockade.

Setting

Academic teaching hospital and level 1 trauma center.

Patients

Inpatients in coma due to hypoxic ischemic encephalopathy.

Measurements and Main Results

Kappa values among the four readers, reflecting inter-observer reliability, ranged from 0.39 to 0.79 (mean 0.60, SD 0.16), a level of agreement interpreted as “moderate.” Better inter-observer reliability was seen when: cortical amplitudes were ≥0.7 μV, baseline noise was smaller, and when experience level between reviewers was most similar. In cases examined both pre- and post-NMJ blockade, average kappa values of inter-observer reliability increased significantly from 0.03 pre-NMJ blockade to 0.41 post-NMJ blockade. Intra-observer reliability was higher than that for inter-observer, with a “substantial” average kappa of 0.84 (range 0.79–0.89, SD .06).

Conclusions

Inter- and intra-observer reliability of SSEP interpretation in comatose patients varies from moderate to substantial, respectively. In order to reliably interpret the presence of small cortical responses, NMJ blockade should be used when baseline noise is excessive.
Literature
1.
go back to reference Robinson LR, Micklesen PJ, Tirschwell DL, Lew HL. Predictive value of somatosensory evoked potential (SEPs) for awakening from coma. Crit Care Med. 2003;31:960–7.CrossRefPubMed Robinson LR, Micklesen PJ, Tirschwell DL, Lew HL. Predictive value of somatosensory evoked potential (SEPs) for awakening from coma. Crit Care Med. 2003;31:960–7.CrossRefPubMed
2.
go back to reference Zandbergen EGJ, deHaan RJ, CP Stoutenbeek, Koelman JHTM, Hijdra A. Systematic review of early prediction of poor outcome in anoxic ischemic coma. Lancet. 1998;352:1808–12.CrossRefPubMed Zandbergen EGJ, deHaan RJ, CP Stoutenbeek, Koelman JHTM, Hijdra A. Systematic review of early prediction of poor outcome in anoxic ischemic coma. Lancet. 1998;352:1808–12.CrossRefPubMed
3.
go back to reference Robinson LR, Micklesen PJ. Somatosensory evoked potentials in coma prognosis. Phys Med Rehabil Clin N Am. 2004;15:43–61.PubMed Robinson LR, Micklesen PJ. Somatosensory evoked potentials in coma prognosis. Phys Med Rehabil Clin N Am. 2004;15:43–61.PubMed
4.
go back to reference Feinstein AR. Clinimetrics. London: Yale University Press; 1987. p. 186. Feinstein AR. Clinimetrics. London: Yale University Press; 1987. p. 186.
5.
go back to reference Longstreth WT Jr, Koepsell TD, van Belle G. Clinical neuroepidemiology I. Diagnosis. Arch Neurol. 1987;44:1091–9.PubMed Longstreth WT Jr, Koepsell TD, van Belle G. Clinical neuroepidemiology I. Diagnosis. Arch Neurol. 1987;44:1091–9.PubMed
6.
go back to reference Fleiss JL. The design and analysis of clinical experiment. New York: Wiley; 1986. Fleiss JL. The design and analysis of clinical experiment. New York: Wiley; 1986.
7.
go back to reference Fleiss J. Statistical methods for rates and proportions, Chapter 13. New York: Wiley; 1981. Fleiss J. Statistical methods for rates and proportions, Chapter 13. New York: Wiley; 1981.
8.
go back to reference Zandbergen EGJ, Hijdra A, deHaan RJ, et al. Inter-observer variation in the interpretation of SSEPs in anoxic-ischaemic coma. Clin Neurophysiol. 2006;117:1529–35.CrossRefPubMed Zandbergen EGJ, Hijdra A, deHaan RJ, et al. Inter-observer variation in the interpretation of SSEPs in anoxic-ischaemic coma. Clin Neurophysiol. 2006;117:1529–35.CrossRefPubMed
Metadata
Title
Reliability in Interpretation of Median Somatosensory Evoked Potentials in the Setting of Coma: Factors and Implications
Authors
Kevin Hakimi
Greg Kinney
George Kraft
Paula Micklesen
Lawrence Robinson
Publication date
01-12-2009
Publisher
Humana Press Inc
Published in
Neurocritical Care / Issue 3/2009
Print ISSN: 1541-6933
Electronic ISSN: 1556-0961
DOI
https://doi.org/10.1007/s12028-009-9251-7

Other articles of this Issue 3/2009

Neurocritical Care 3/2009 Go to the issue