01-12-2017 | Clinical Research

Polyethylene Damage Increases With Varus Implant Alignment in Posterior-stabilized and Constrained Condylar Knee Arthroplasty

Authors: Zhichang Li, MD, Christina I. Esposito, PhD, Chelsea N. Koch, BS, Yuo-yu Lee, MS, Douglas E. Padgett, MD, Timothy M. Wright, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 12/2017

Login to get access

Abstract

Background

Implant malalignment in primary TKA has been reported to increase stresses placed on the bearing surfaces of implant components. We used a longitudinally maintained registry coupled with an implant retrieval program to consider whether preoperative, postoperative, or prerevision malalignment was associated with increased risk of revision surgery after TKA.

Questions/purposes

(1) What is the relative polyethylene damage on medial and lateral compartments of the tibial plateaus from revised TKAs? (2) Does coronal TKA alignment affect implant performance, such that TKAs aligned in varus are predisposed to experience increased polyethylene damage? (3) Does TKA alignment differ between postoperative and prerevision radiographs, and if so, what does this difference suggest about the mechanical contact load placed on a knee with a TKA?

Methods

Between 2007 and 2012, we performed 18,065 primary TKAs at our institution. By March 2016, 178 of those TKAs (1%) were revised at our center at least 2 years after primary surgery at our institution. Eighteen of those TKAs were excluded from this analysis because the tibial insert was not explanted during revision surgery, and four more were excluded because the inserts were lost or returned to the patient before the study was initiated, leaving 156 retrieved polyethylene tibial inserts (in 153 patients) revised at greater than 2 years after the primary TKA for this retrospective study. Patients who underwent revision surgery elsewhere were not considered here, since this study depended on having retrieved components. Polyethylene damage modes of burnishing, pitting, scratching, delamination, surface deformation, abrasion, and third-body debris were subjectively graded on a scale of 0 to 3 to reflect the extent and severity of each damage mode. On preoperative, postoperative, and prerevision radiographs, overall alignment, femoral alignment, and tibial alignment in the coronal plane were measured according to the protocol recommended by the Knee Society.

Results

Knees with more overall varus alignment after TKA had increased total damage on the retrieved tibial inserts (Spearman’s rank correlation coefficients of −0.3 [95% CI, −0.4 to −0.1; p = 0.001]). We also found revised TKAs tended to drift back into greater varus before revision surgery, with a mean (SD) of 3.6° ± 4.0° valgus for postoperative alignment compared with 1.7° ± 6.4° prerevision (p = 0.04).

Conclusions

Despite surgical efforts to achieve neutral mechanical alignment, remaining varus alignment places an increased contact load on the polyethylene articular surfaces. The drift toward further varus alignment postoperatively is consistent with the knee adduction moment remaining high after surgery.

Clinical Relevance

While we found a predisposition toward recurrence of the preoperative varus deformity, we did not find increased medial as opposed to lateral polyethylene damage, which may be explained by the curve-on-curve toroidal design of the articulating surfaces of the TKA implants in this study.
Literature
1.
go back to reference Altman D. Practical Statistics for Medical Research. London, UK: Chapman and Hall; 1991. Altman D. Practical Statistics for Medical Research. London, UK: Chapman and Hall; 1991.
2.
go back to reference Anderson JA, Baldini A, MacDonald JH, Tomek I, Pellicci PM, Sculco TP. Constrained condylar knee without stem extensions for difficult primary total knee arthroplasty. J Knee Surg. 2007;20:195–198.CrossRefPubMed Anderson JA, Baldini A, MacDonald JH, Tomek I, Pellicci PM, Sculco TP. Constrained condylar knee without stem extensions for difficult primary total knee arthroplasty. J Knee Surg. 2007;20:195–198.CrossRefPubMed
3.
go back to reference Andrews M, Noyes FR, Hewett TE, Andriacchi TP. Lower limb alignment and foot angle are related to stance phase knee adduction in normal subjects: a critical analysis of the reliability of gait analysis data. J Orthop Res. 1996;14:289–295.CrossRefPubMed Andrews M, Noyes FR, Hewett TE, Andriacchi TP. Lower limb alignment and foot angle are related to stance phase knee adduction in normal subjects: a critical analysis of the reliability of gait analysis data. J Orthop Res. 1996;14:289–295.CrossRefPubMed
4.
go back to reference Andriacchi TP. Biomechanics and gait analysis in total knee replacement. Orthop Rev. 1988;17:470–473.PubMed Andriacchi TP. Biomechanics and gait analysis in total knee replacement. Orthop Rev. 1988;17:470–473.PubMed
5.
go back to reference Berend ME, Ritter MA, Meding JB, Faris PM, Keating EM, Redelman R, Faris GW, Davis KE. Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res. 2004;428:26–34.CrossRef Berend ME, Ritter MA, Meding JB, Faris PM, Keating EM, Redelman R, Faris GW, Davis KE. Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res. 2004;428:26–34.CrossRef
6.
go back to reference Bonner TJ, Eardley WG, Patterson P, Gregg PJ. The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 years. J Bone Joint Surg Br. 2011;93:1217–1222.CrossRefPubMed Bonner TJ, Eardley WG, Patterson P, Gregg PJ. The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 years. J Bone Joint Surg Br. 2011;93:1217–1222.CrossRefPubMed
7.
go back to reference Centers for Disease Control and Prevention, National Center for Health Statistics. National Hospital Discharge Survey. Available at: https://www.cdc.gov/nchs/nhds/. Accessed May 31, 2017. Centers for Disease Control and Prevention, National Center for Health Statistics. National Hospital Discharge Survey. Available at: https://​www.​cdc.​gov/​nchs/​nhds/​.​ Accessed May 31, 2017.
8.
go back to reference Collier MB, Engh CA Jr, McAuley JP, Engh GA. Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am. 2007;89:1306–1314.CrossRefPubMed Collier MB, Engh CA Jr, McAuley JP, Engh GA. Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am. 2007;89:1306–1314.CrossRefPubMed
9.
go back to reference D’Lima DD, Chen PC, Colwell CW. Polyethylene contact stresses, articular congruity, and knee alignment. Clin Orthop Relat Res. 2001;392:232–238.CrossRef D’Lima DD, Chen PC, Colwell CW. Polyethylene contact stresses, articular congruity, and knee alignment. Clin Orthop Relat Res. 2001;392:232–238.CrossRef
10.
go back to reference D’Lima DD, Hermida JC, Chen PC, Colwell CW Jr. Polyethylene wear and variations in knee kinematics. Clin Orthop Relat Res. 2001;392:124–130.CrossRef D’Lima DD, Hermida JC, Chen PC, Colwell CW Jr. Polyethylene wear and variations in knee kinematics. Clin Orthop Relat Res. 2001;392:124–130.CrossRef
11.
go back to reference Donaldson WF 3rd, Sculco TP, Insall JN, Ranawat CS. Total condylar III knee prosthesis: long-term follow-up study. Clin Orthop Relat Res. 1988;226:21–28. Donaldson WF 3rd, Sculco TP, Insall JN, Ranawat CS. Total condylar III knee prosthesis: long-term follow-up study. Clin Orthop Relat Res. 1988;226:21–28.
12.
go back to reference Erhart JC, Mündermann A, Elspas B, Giori NJ, Andriacchi TP. A variable-stiffness shoe lowers the knee adduction moment in subjects with symptoms of medial compartment knee osteoarthritis. J Biomech. 2008;41:2720–2725.CrossRefPubMed Erhart JC, Mündermann A, Elspas B, Giori NJ, Andriacchi TP. A variable-stiffness shoe lowers the knee adduction moment in subjects with symptoms of medial compartment knee osteoarthritis. J Biomech. 2008;41:2720–2725.CrossRefPubMed
13.
go back to reference Fang DM, Ritter MA, Davis KE. Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty. 2009;24(6 suppl):39–43.CrossRefPubMed Fang DM, Ritter MA, Davis KE. Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty. 2009;24(6 suppl):39–43.CrossRefPubMed
14.
go back to reference Faris PM, Ritter MA, Davis KE, Priscu HM. Ten-year outcome comparison of the Anatomical Graduated Component and Vanguard total knee arthroplasty systems. J Arthroplasty. 2015;30:1733–1735.CrossRefPubMed Faris PM, Ritter MA, Davis KE, Priscu HM. Ten-year outcome comparison of the Anatomical Graduated Component and Vanguard total knee arthroplasty systems. J Arthroplasty. 2015;30:1733–1735.CrossRefPubMed
15.
go back to reference Hood RW, Wright TM, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res. 1983;17:829–842.CrossRefPubMed Hood RW, Wright TM, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res. 1983;17:829–842.CrossRefPubMed
16.
go back to reference Johnson F, Leitl S, Waugh W. The distribution of load across the knee: a comparison of static and dynamic measurements. J Bone Joint Surg Br. 1980;62:346–349.PubMed Johnson F, Leitl S, Waugh W. The distribution of load across the knee: a comparison of static and dynamic measurements. J Bone Joint Surg Br. 1980;62:346–349.PubMed
17.
go back to reference Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R. Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty. 2009;24:570–578.CrossRefPubMed Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R. Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty. 2009;24:570–578.CrossRefPubMed
18.
go back to reference Meneghini RM, Mont MA, Backstein DB, Bourne RB, Dennis DA, Scuderi GR. Development of a modern Knee Society radiographic evaluation system and methodology for total knee arthroplasty. J Arthroplasty. 2015;30:2311–2314.CrossRefPubMed Meneghini RM, Mont MA, Backstein DB, Bourne RB, Dennis DA, Scuderi GR. Development of a modern Knee Society radiographic evaluation system and methodology for total knee arthroplasty. J Arthroplasty. 2015;30:2311–2314.CrossRefPubMed
19.
go back to reference Morgan SS, Bonshahi A, Pradhan N, Gregory A, Gambhir A, Porter ML. The influence of postoperative coronal alignment on revision surgery in total knee arthroplasty. Int Orthop. 2008;32:639–642.CrossRefPubMed Morgan SS, Bonshahi A, Pradhan N, Gregory A, Gambhir A, Porter ML. The influence of postoperative coronal alignment on revision surgery in total knee arthroplasty. Int Orthop. 2008;32:639–642.CrossRefPubMed
20.
go back to reference Orishimo KF, Kremenic IJ, Deshmukh AJ, Nicholas SJ, Rodriguez JA. Does total knee arthroplasty change frontal plane knee biomechanics during gait? Clin Orthop Relat Res. 2012;470:1171–1176.CrossRefPubMed Orishimo KF, Kremenic IJ, Deshmukh AJ, Nicholas SJ, Rodriguez JA. Does total knee arthroplasty change frontal plane knee biomechanics during gait? Clin Orthop Relat Res. 2012;470:1171–1176.CrossRefPubMed
21.
go back to reference Padgett DE, Cottrell J, Kelly N, Gelber J, Farrell C, Wright TM. Retrieval analysis of nonmodular constrained tibial inserts after primary total knee replacement. Orthop Clin North Am. 2012;43:e39–e43.CrossRefPubMed Padgett DE, Cottrell J, Kelly N, Gelber J, Farrell C, Wright TM. Retrieval analysis of nonmodular constrained tibial inserts after primary total knee replacement. Orthop Clin North Am. 2012;43:e39–e43.CrossRefPubMed
22.
go back to reference Pang HN, Jamieson P, Teeter MG, McCalden RW, Naudie DD, MacDonald SJ. Retrieval analysis of posterior stabilized polyethylene tibial inserts and its clinical relevance. J Arthroplasty. 2014;29:365–368.CrossRefPubMed Pang HN, Jamieson P, Teeter MG, McCalden RW, Naudie DD, MacDonald SJ. Retrieval analysis of posterior stabilized polyethylene tibial inserts and its clinical relevance. J Arthroplasty. 2014;29:365–368.CrossRefPubMed
23.
go back to reference Parratte S, Pagnano MW, Trousdale RT, Berry DJ. Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surg Am. 2010;92:2143–2149.CrossRefPubMed Parratte S, Pagnano MW, Trousdale RT, Berry DJ. Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surg Am. 2010;92:2143–2149.CrossRefPubMed
24.
go back to reference Petersen TL, Engh GA. Radiographic assessment of knee alignment after total knee arthroplasty. J Arthroplasty. 1988;3:67–72.CrossRefPubMed Petersen TL, Engh GA. Radiographic assessment of knee alignment after total knee arthroplasty. J Arthroplasty. 1988;3:67–72.CrossRefPubMed
25.
go back to reference Prodromos CC, Andriacchi TP, Galante JO. A relationship between gait and clinical changes following high tibial osteotomy. J Bone Joint Surg Am. 1985;67:1188–1194.CrossRefPubMed Prodromos CC, Andriacchi TP, Galante JO. A relationship between gait and clinical changes following high tibial osteotomy. J Bone Joint Surg Am. 1985;67:1188–1194.CrossRefPubMed
26.
go back to reference Ritter MA. The Anatomical Graduated Component total knee replacement: a long-term evaluation with 20-year survival analysis. J Bone Joint Surg Br. 2009;91:745–749.CrossRefPubMed Ritter MA. The Anatomical Graduated Component total knee replacement: a long-term evaluation with 20-year survival analysis. J Bone Joint Surg Br. 2009;91:745–749.CrossRefPubMed
27.
go back to reference Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA. The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am. 2011;93:1588–1596.CrossRefPubMed Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA. The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am. 2011;93:1588–1596.CrossRefPubMed
28.
go back to reference Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME, Ritter MA, Nunley RM. Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplasty. 2013;28(8 suppl):116–119.CrossRefPubMed Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME, Ritter MA, Nunley RM. Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplasty. 2013;28(8 suppl):116–119.CrossRefPubMed
29.
go back to reference Sheehy L, Felson D, Zhang Y, Niu J, Lam YM, Segal N, Lynch J, Cooke TD. Does measurement of the anatomic axis consistently predict hip-knee-ankle angle (HKA) for knee alignment studies in osteoarthritis? Analysis of long limb radiographs from the multicenter osteoarthritis (MOST) study. Osteoarthritis Cartilage. 2011;19:58–64.CrossRefPubMed Sheehy L, Felson D, Zhang Y, Niu J, Lam YM, Segal N, Lynch J, Cooke TD. Does measurement of the anatomic axis consistently predict hip-knee-ankle angle (HKA) for knee alignment studies in osteoarthritis? Analysis of long limb radiographs from the multicenter osteoarthritis (MOST) study. Osteoarthritis Cartilage. 2011;19:58–64.CrossRefPubMed
30.
go back to reference Srivastava A, Lee GY, Steklov N, Colwell CW Jr, Ezzet KA, D’Lima DD. Effect of tibial component varus on wear in total knee arthroplasty. Knee. 2012;19:560–563.CrossRefPubMed Srivastava A, Lee GY, Steklov N, Colwell CW Jr, Ezzet KA, D’Lima DD. Effect of tibial component varus on wear in total knee arthroplasty. Knee. 2012;19:560–563.CrossRefPubMed
31.
go back to reference Vandekerckhove PT, Matlovich N, Teeter MG, MacDonald SJ, Howard JL, Lanting BA. The relationship between constitutional alignment and varus osteoarthritis of the knee. Knee Surg Sports Traumatol Arthrosc. 2016 Feb 1. [Epub ahead of print] Vandekerckhove PT, Matlovich N, Teeter MG, MacDonald SJ, Howard JL, Lanting BA. The relationship between constitutional alignment and varus osteoarthritis of the knee. Knee Surg Sports Traumatol Arthrosc. 2016 Feb 1. [Epub ahead of print]
32.
go back to reference Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG. Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res. 1994;299:31–43. Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG. Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res. 1994;299:31–43.
33.
go back to reference Wright TM, Bartel DL. The problem of surface damage in polyethylene total knee components. Clin Orthop Relat Res. 1986;205:67–74. Wright TM, Bartel DL. The problem of surface damage in polyethylene total knee components. Clin Orthop Relat Res. 1986;205:67–74.
Metadata
Title
Polyethylene Damage Increases With Varus Implant Alignment in Posterior-stabilized and Constrained Condylar Knee Arthroplasty
Authors
Zhichang Li, MD
Christina I. Esposito, PhD
Chelsea N. Koch, BS
Yuo-yu Lee, MS
Douglas E. Padgett, MD
Timothy M. Wright, PhD
Publication date
01-12-2017
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 12/2017
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-017-5477-4

Other articles of this Issue 12/2017

Clinical Orthopaedics and Related Research® 12/2017 Go to the issue