01-03-2015 | Symposium: Advances in UHMWPE Biomaterials

Multidirectional Wear and Impact-to-wear Tests of Phospholipid-polymer-grafted and Vitamin E-blended Crosslinked Polyethylene: A Pilot Study

Authors: Masayuki Kyomoto, PhD, Toru Moro, MD, Yoshio Takatori, MD, Sakae Tanaka, MD, Kazuhiko Ishihara, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 3/2015

Login to get access

Abstract

Background

Modifying the surface and substrate of a crosslinked polyethylene (CLPE) liner may be beneficial for high wear resistance as well as high oxidative stability and excellent mechanical properties, which would be useful in contributing to the long-term performance of orthopaedic bearings. A grafted poly(2-methacryloyloxyethyl phosphorylcholine) (PMPC) layer on a vitamin E-blended crosslinked PE (HD-CLPE[VE]) surface may provide hydrophilicity and lubricity without compromising the oxidative stability or mechanical properties.

Questions/purposes

(1) Will the modifications (PMPC grafting and vitamin E blending) affect the lubrication characteristics of the CLPE surface? (2) Will the modifications affect wear resistance? (3) Will the modifications affect fatigue resistance?

Methods

We investigated the effects of surface and substrate modifications (PMPC grafting and vitamin E blending) on the wear and fatigue fracture of thin CLPE samples. For each of the untreated and PMPC-grafted CLPE surfaces with and without vitamin E blended (four groups), wettability and lubricity surface analyses were conducted as well as multidirectional wear and impact-to-wear tests using a pin-on-disk testing machine.

Results

The water wettability and lubricity (CLPE [mean ± 95% confidence interval]: 23.2° ± 1.8°, 0.005 ± 0.001; HD-CLPE[VE]: 26.0° ± 2.3°, 0.009 ± 0.003) of the PMPC-grafted surfaces were greater (p < 0.001) than that (CLPE: 90.3° ± 1.2°, 0.067 ± 0.015; HD-CLPE[VE]: 90.8° ± 2.0°, 0.063 ± 0.008) of the untreated surface regardless of vitamin E additives. It was observed that the PMPC grafting (CLPE: 0.23 ± 0.06 mg; HD-CLPE[VE]: 0.05 ± 0.10 mg) was associated with reduced gravimetric wear (CLPE: 0.53 ± 0.08 mg, p = 0.004 HD-CLPE[VE]: 0.23 ± 0.07 mg, p = 0.038) in the multidirectional wear test. The PMPC-grafted surface characteristics did not appear to affect the impact fatigue resistance regardless of vitamin E blending.

Conclusions

PMPC grafting improved the surface hydrophilicity and lubricity, and it reduced the gravimetric wear in terms of multidirectional sliding. It did not result in differences in terms of the impact-to-unidirectional sliding regardless of vitamin E blending. Further research is needed to evaluate the wear resistance of PMPC-grafted HD-CLPE(VE) in long-term hip simulator tests under normal and severe conditions, which may offer useful clues to the possible performance of these materials in vivo.

Clinical Relevance

Our preliminary in vitro findings suggest that some improvement in the wear performance of crosslinked polyethylene acetabular liners in total hip arthroplasty could be obtained using PMPC grafting. Further research is needed to evaluate the wear resistance of PMPC-grafted HD-CLPE(VE) in long-term hip simulator tests under normal and severe conditions, which may offer useful clues to the possible performance of these materials in vivo.
Literature
1.
go back to reference Ast MP, John TK, Labbisiere A, Robador N, Valle AG. Fractures of a single design of highly cross-linked polyethylene acetabular liners: an analysis of voluntary reports to the United States Food and Drug Administration. J Arthroplasty. 2014;29:1231–1235.PubMedCrossRef Ast MP, John TK, Labbisiere A, Robador N, Valle AG. Fractures of a single design of highly cross-linked polyethylene acetabular liners: an analysis of voluntary reports to the United States Food and Drug Administration. J Arthroplasty. 2014;29:1231–1235.PubMedCrossRef
2.
go back to reference ASTM International. ASTM F732-00. Standard test method for wear testing of polymeric materials used in total joint prostheses. West Conshohocken, PA, USA: ASTM International; 2011. ASTM International. ASTM F732-00. Standard test method for wear testing of polymeric materials used in total joint prostheses. West Conshohocken, PA, USA: ASTM International; 2011.
3.
go back to reference Baykal D, Siskey RS, Haider H, Saikko V, Ahlroos T, Kurtz SM. Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers. J Mech Behav Biomed Mater. 2014;31C:117–134.CrossRef Baykal D, Siskey RS, Haider H, Saikko V, Ahlroos T, Kurtz SM. Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers. J Mech Behav Biomed Mater. 2014;31C:117–134.CrossRef
4.
go back to reference Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:128–133.PubMedCrossRef Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:128–133.PubMedCrossRef
6.
go back to reference Chang DC, Goh JC, Teoh SH, Bose K. Cold extrusion deformation of UHMWPE in total knee replacement prostheses. Biomaterials. 1995;16:219–223.PubMedCrossRef Chang DC, Goh JC, Teoh SH, Bose K. Cold extrusion deformation of UHMWPE in total knee replacement prostheses. Biomaterials. 1995;16:219–223.PubMedCrossRef
7.
go back to reference Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am. 2007;89:2023–2029.PubMedCrossRef Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am. 2007;89:2023–2029.PubMedCrossRef
8.
go back to reference Czichos H. Introduction and background. In: Czichos H, ed. Tribology: A Systems Approach to the Science and Technology of Friction, Lubrication and Wear. New York, NY, USA: Elsevier; 1978:1–13. Czichos H. Introduction and background. In: Czichos H, ed. Tribology: A Systems Approach to the Science and Technology of Friction, Lubrication and Wear. New York, NY, USA: Elsevier; 1978:1–13.
9.
go back to reference Goldberg R, Schroeder A, Silbert G, Turjeman K, Barenholz Y, Klein J. Boundary lubricants with exceptionally low friction coefficients based on 2D close-packed phosphatidylcholine liposomes. Adv Mater. 2011;23:3517–3521.PubMedCrossRef Goldberg R, Schroeder A, Silbert G, Turjeman K, Barenholz Y, Klein J. Boundary lubricants with exceptionally low friction coefficients based on 2D close-packed phosphatidylcholine liposomes. Adv Mater. 2011;23:3517–3521.PubMedCrossRef
10.
go back to reference Harris WH. The problem is osteolysis. Clin Orthop Relat Res. 1995;311:46–53.PubMed Harris WH. The problem is osteolysis. Clin Orthop Relat Res. 1995;311:46–53.PubMed
11.
go back to reference International Organization for Standardization. ISO 15989. Plastics-film and Sheeting-measurement of Water-contact Angle of Corona-treated Films. Geneva, Switzerland: ISO; 2004. International Organization for Standardization. ISO 15989. Plastics-film and Sheeting-measurement of Water-contact Angle of Corona-treated Films. Geneva, Switzerland: ISO; 2004.
12.
go back to reference Ishikawa Y, Hiratsuka K, Sasada T. Role of water in the lubrication of hydrogel. Wear. 2006;261:500–504.CrossRef Ishikawa Y, Hiratsuka K, Sasada T. Role of water in the lubrication of hydrogel. Wear. 2006;261:500–504.CrossRef
13.
go back to reference Iwasaki Y, Ishihara K. Phosphorylcholine-containing polymers for biomedical applications. Anal Bioanal Chem. 2005;381:534–546.PubMedCrossRef Iwasaki Y, Ishihara K. Phosphorylcholine-containing polymers for biomedical applications. Anal Bioanal Chem. 2005;381:534–546.PubMedCrossRef
14.
go back to reference Kirk TB, Wilson AS, Stachowiak GW. The morphology and composition of the superficial zone of mammalian articular cartilage. J Orthop Rheumatol. 1993;6:21–28. Kirk TB, Wilson AS, Stachowiak GW. The morphology and composition of the superficial zone of mammalian articular cartilage. J Orthop Rheumatol. 1993;6:21–28.
15.
go back to reference Kurtz SM, ed. UHMWPE Biomaterials Handbook. 2nd ed. New York, NY, USA: Elsevier; 2009. Kurtz SM, ed. UHMWPE Biomaterials Handbook. 2nd ed. New York, NY, USA: Elsevier; 2009.
16.
go back to reference Kyomoto M, Iwasaki Y, Moro T, Konno T, Miyaji F, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. High lubricious surface of cobalt-chromium-molybdenum alloy prepared by grafting poly(2-methacryloyloxyethyl phosphorylcholine). Biomaterials. 2007;28:3121–3130.PubMedCrossRef Kyomoto M, Iwasaki Y, Moro T, Konno T, Miyaji F, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. High lubricious surface of cobalt-chromium-molybdenum alloy prepared by grafting poly(2-methacryloyloxyethyl phosphorylcholine). Biomaterials. 2007;28:3121–3130.PubMedCrossRef
17.
go back to reference Kyomoto M, Moro T, Iwasaki Y, Miyaji F, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Superlubricious surface mimicking articular cartilage by grafting poly(2-methacryloyloxyethyl phosphorylcholine) on orthopaedic metal bearings. J Biomed Mater Res A. 2009;91:730–741.PubMedCrossRef Kyomoto M, Moro T, Iwasaki Y, Miyaji F, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Superlubricious surface mimicking articular cartilage by grafting poly(2-methacryloyloxyethyl phosphorylcholine) on orthopaedic metal bearings. J Biomed Mater Res A. 2009;91:730–741.PubMedCrossRef
18.
go back to reference Kyomoto M, Moro T, Konno T, Takadama H, Kawaguchi H, Takatori Y, Nakamura K, Yamawaki N, Ishihara K. Effects of photo-induced graft polymerization of 2-methacryloyloxyethyl phosphorylcholine on physical properties of cross-linked polyethylene in artificial hip joints. J Mater Sci Mater Med. 2007;18:1809–1815.PubMedCrossRef Kyomoto M, Moro T, Konno T, Takadama H, Kawaguchi H, Takatori Y, Nakamura K, Yamawaki N, Ishihara K. Effects of photo-induced graft polymerization of 2-methacryloyloxyethyl phosphorylcholine on physical properties of cross-linked polyethylene in artificial hip joints. J Mater Sci Mater Med. 2007;18:1809–1815.PubMedCrossRef
19.
go back to reference Kyomoto M, Moro T, Konno T, Takadama H, Yamawaki N, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Enhanced wear resistance of modified cross-linked polyethylene by grafting with poly(2-methacryloyloxyethyl phosphorylcholine). J Biomed Mater Res A. 2007;82:10–17.PubMedCrossRef Kyomoto M, Moro T, Konno T, Takadama H, Yamawaki N, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Enhanced wear resistance of modified cross-linked polyethylene by grafting with poly(2-methacryloyloxyethyl phosphorylcholine). J Biomed Mater Res A. 2007;82:10–17.PubMedCrossRef
20.
go back to reference Kyomoto M, Moro T, Miyaji F, Hashimoto M, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Effect of 2-methacryloyloxyethyl phosphorylcholine concentration on photo-induced graft polymerization of polyethylene in reducing the wear of orthopaedic bearing surface. J Biomed Mater Res A. 2008;86:439–447.PubMedCrossRef Kyomoto M, Moro T, Miyaji F, Hashimoto M, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Effect of 2-methacryloyloxyethyl phosphorylcholine concentration on photo-induced graft polymerization of polyethylene in reducing the wear of orthopaedic bearing surface. J Biomed Mater Res A. 2008;86:439–447.PubMedCrossRef
21.
go back to reference Kyomoto M, Moro T, Miyaji F, Hashimoto M, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Effects of mobility/immobility of surface modification by 2-methacryloyloxyethyl phosphorylcholine polymer on the durability of polyethylene for artificial joints. J Biomed Mater Res A. 2009;90:362–371.PubMedCrossRef Kyomoto M, Moro T, Miyaji F, Hashimoto M, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Effects of mobility/immobility of surface modification by 2-methacryloyloxyethyl phosphorylcholine polymer on the durability of polyethylene for artificial joints. J Biomed Mater Res A. 2009;90:362–371.PubMedCrossRef
22.
go back to reference Kyomoto M, Moro T, Miyaji F, Konno T, Hashimoto M, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Enhanced wear resistance of orthopaedic bearing due to the cross-linking of poly(MPC) graft chains induced by gamma-ray irradiation. J Biomed Mater Res B Appl Biomater. 2008;84:320–327.PubMedCrossRef Kyomoto M, Moro T, Miyaji F, Konno T, Hashimoto M, Kawaguchi H, Takatori Y, Nakamura K, Ishihara K. Enhanced wear resistance of orthopaedic bearing due to the cross-linking of poly(MPC) graft chains induced by gamma-ray irradiation. J Biomed Mater Res B Appl Biomater. 2008;84:320–327.PubMedCrossRef
23.
go back to reference Kyomoto M, Moro T, Saiga K, Hashimoto M, Ito H, Kawaguchi H, Takatori Y, Ishihara K. Biomimetic hydration lubrication with various polyelectrolyte layers on cross-linked polyethylene orthopedic bearing materials. Biomaterials. 2012;33:4451–4459.PubMedCrossRef Kyomoto M, Moro T, Saiga K, Hashimoto M, Ito H, Kawaguchi H, Takatori Y, Ishihara K. Biomimetic hydration lubrication with various polyelectrolyte layers on cross-linked polyethylene orthopedic bearing materials. Biomaterials. 2012;33:4451–4459.PubMedCrossRef
24.
go back to reference Kyomoto M, Moro T, Takatori Y, Kawaguchi H, Ishihara K. Cartilage-mimicking, high-density brush structure improves wear resistance of crosslinked polyethylene: a pilot study. Clin Orthop Relat Res. 2011;469:2327–2336.PubMedCentralPubMedCrossRef Kyomoto M, Moro T, Takatori Y, Kawaguchi H, Ishihara K. Cartilage-mimicking, high-density brush structure improves wear resistance of crosslinked polyethylene: a pilot study. Clin Orthop Relat Res. 2011;469:2327–2336.PubMedCentralPubMedCrossRef
25.
go back to reference Kyomoto M, Moro T, Takatori Y, Kawaguchi H, Nakamura K, Ishihara K. Novel self-initiated surface grafting with poly(2-methacryloyloxyethyl phosphorylcholine) on poly(ether-ether-ketone) biomaterials. Biomaterials. 2010;31:1017–1024.PubMedCrossRef Kyomoto M, Moro T, Takatori Y, Kawaguchi H, Nakamura K, Ishihara K. Novel self-initiated surface grafting with poly(2-methacryloyloxyethyl phosphorylcholine) on poly(ether-ether-ketone) biomaterials. Biomaterials. 2010;31:1017–1024.PubMedCrossRef
26.
go back to reference Kyomoto M, Moro T, Yamane S, Hashimoto M, Takatori Y, Ishihara K. Poly(ether-ether-ketone) orthopedic bearing surface modified by self-initiated surface grafting of poly(2-methacryloyloxyethyl phosphorylcholine). Biomaterials. 2013;34:7829–7839.PubMedCrossRef Kyomoto M, Moro T, Yamane S, Hashimoto M, Takatori Y, Ishihara K. Poly(ether-ether-ketone) orthopedic bearing surface modified by self-initiated surface grafting of poly(2-methacryloyloxyethyl phosphorylcholine). Biomaterials. 2013;34:7829–7839.PubMedCrossRef
27.
go back to reference Kyomoto M, Moro T, Yamane S, Watanabe K, Hashimoto M, Takatori Y, Tanaka S, Ishihara K. Poly(2-methacryloyloxyethyl phosphorylcholine) grafting and vitamin E blending for high wear resistance and oxidative stability of life-long orthopedic bearings. Biomaterials. 2014;35:6677–6686.PubMedCrossRef Kyomoto M, Moro T, Yamane S, Watanabe K, Hashimoto M, Takatori Y, Tanaka S, Ishihara K. Poly(2-methacryloyloxyethyl phosphorylcholine) grafting and vitamin E blending for high wear resistance and oxidative stability of life-long orthopedic bearings. Biomaterials. 2014;35:6677–6686.PubMedCrossRef
28.
go back to reference MacDonald D, Sakona A, Ianuzzi A, Rimnac CM, Kurtz SM. Do first-generation highly crosslinked polyethylenes oxidize in vivo? Clin Orthop Relat Res. 2011;469:2278–2285.PubMedCentralPubMedCrossRef MacDonald D, Sakona A, Ianuzzi A, Rimnac CM, Kurtz SM. Do first-generation highly crosslinked polyethylenes oxidize in vivo? Clin Orthop Relat Res. 2011;469:2278–2285.PubMedCentralPubMedCrossRef
29.
go back to reference Moro T, Takatori Y, Ishihara K, Konno T, Takigawa Y, Matsushita T, Chung UI, Nakamura K, Kawaguchi H. Surface grafting of artificial joints with a biocompatible polymer for preventing periprosthetic osteolysis. Nat Mater. 2004;3:829–837.PubMedCrossRef Moro T, Takatori Y, Ishihara K, Konno T, Takigawa Y, Matsushita T, Chung UI, Nakamura K, Kawaguchi H. Surface grafting of artificial joints with a biocompatible polymer for preventing periprosthetic osteolysis. Nat Mater. 2004;3:829–837.PubMedCrossRef
30.
go back to reference Moro T, Takatori Y, Kyomoto M, Ishihara K, Hashimoto M, Ito H, Tanaka T, Oshima H, Tanaka S, Kawaguchi H. Long-term hip simulator testing of the artificial hip joint bearing surface grafted with biocompatible phospholipid polymer. J Orthop Res. 2014;32:369–376.PubMedCrossRef Moro T, Takatori Y, Kyomoto M, Ishihara K, Hashimoto M, Ito H, Tanaka T, Oshima H, Tanaka S, Kawaguchi H. Long-term hip simulator testing of the artificial hip joint bearing surface grafted with biocompatible phospholipid polymer. J Orthop Res. 2014;32:369–376.PubMedCrossRef
31.
go back to reference Okubo Y, Hamada D, Yamamoto K, Mori SI, Ikeuchi K, Tomita N. Load-dependent frictional performance of vitamin E-blended ultrahigh molecular weight polyethylene in serum lubricant. Tribology Online. 2010;5:96–101.CrossRef Okubo Y, Hamada D, Yamamoto K, Mori SI, Ikeuchi K, Tomita N. Load-dependent frictional performance of vitamin E-blended ultrahigh molecular weight polyethylene in serum lubricant. Tribology Online. 2010;5:96–101.CrossRef
32.
go back to reference Oral E, Godleski Beckos C, Malhi AS, Muratoglu OK. The effects of high dose irradiation on the cross-linking of vitamin E-blended ultrahigh molecular weight polyethylene. Biomaterials. 2008;29:3557–3560.PubMedCentralPubMedCrossRef Oral E, Godleski Beckos C, Malhi AS, Muratoglu OK. The effects of high dose irradiation on the cross-linking of vitamin E-blended ultrahigh molecular weight polyethylene. Biomaterials. 2008;29:3557–3560.PubMedCentralPubMedCrossRef
33.
go back to reference Rodriguez JA, Rathod PA. Large diameter heads: is bigger always better? J Bone Joint Surg Br. 2012;94(Suppl A):52–54.PubMedCrossRef Rodriguez JA, Rathod PA. Large diameter heads: is bigger always better? J Bone Joint Surg Br. 2012;94(Suppl A):52–54.PubMedCrossRef
34.
go back to reference Shibata N, Tomita N. The anti-oxidative properties of alpha-tocopherol in gamma-irradiated UHMWPE with respect to fatigue and oxidation resistance. Biomaterials. 2005;26:5755–5762.PubMedCrossRef Shibata N, Tomita N. The anti-oxidative properties of alpha-tocopherol in gamma-irradiated UHMWPE with respect to fatigue and oxidation resistance. Biomaterials. 2005;26:5755–5762.PubMedCrossRef
35.
go back to reference Takatori Y, Moro T, Ishihara K, Kamogawa M, Oda H, Umeyama T, Kim YT, Ito H, Kyomoto M, Tanaka T, Kawaguchi H, Tanaka S. Clinical and radiographic outcomes of total hip replacement with poly(2-methacryloyloxyethyl phosphorylcholine)-grafted highly cross-linked polyethylene liners: three-year results of a prospective consecutive series. Mod Rheumatol. 2014:1–6 [Epub ahead of print]. Takatori Y, Moro T, Ishihara K, Kamogawa M, Oda H, Umeyama T, Kim YT, Ito H, Kyomoto M, Tanaka T, Kawaguchi H, Tanaka S. Clinical and radiographic outcomes of total hip replacement with poly(2-methacryloyloxyethyl phosphorylcholine)-grafted highly cross-linked polyethylene liners: three-year results of a prospective consecutive series. Mod Rheumatol. 2014:1–6 [Epub ahead of print].
36.
go back to reference Takatori Y, Moro T, Kamogawa M, Oda H, Morimoto S, Umeyama T, Minami M, Sugimoto H, Nakamura S, Karita T, Kim J, Koyama Y, Ito H, Kawaguchi H, Nakamura K. Poly(2-methacryloyloxyethyl phosphorylcholine)-grafted highly cross-linked polyethylene liner in primary total hip replacement: one-year results of a prospective cohort study. J Artif Organs. 2013;16:170–175.PubMedCrossRef Takatori Y, Moro T, Kamogawa M, Oda H, Morimoto S, Umeyama T, Minami M, Sugimoto H, Nakamura S, Karita T, Kim J, Koyama Y, Ito H, Kawaguchi H, Nakamura K. Poly(2-methacryloyloxyethyl phosphorylcholine)-grafted highly cross-linked polyethylene liner in primary total hip replacement: one-year results of a prospective cohort study. J Artif Organs. 2013;16:170–175.PubMedCrossRef
Metadata
Title
Multidirectional Wear and Impact-to-wear Tests of Phospholipid-polymer-grafted and Vitamin E-blended Crosslinked Polyethylene: A Pilot Study
Authors
Masayuki Kyomoto, PhD
Toru Moro, MD
Yoshio Takatori, MD
Sakae Tanaka, MD
Kazuhiko Ishihara, PhD
Publication date
01-03-2015
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 3/2015
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-014-3995-x

Other articles of this Issue 3/2015

Clinical Orthopaedics and Related Research® 3/2015 Go to the issue

Symposium: 2013 Meetings of the Musculoskeletal Tumor Society and the International Society of Limb Salvage

Frequent Complications and Severe Bone Loss Associated With the Repiphysis Expandable Distal Femoral Prosthesis

Symposium: Advances in UHMWPE Biomaterials

Editorial Comment: Advances in UHMWPE Biomaterials