Published in: Clinical Orthopaedics and Related Research® 12/2014

01-12-2014 | Clinical Research

Modern Trunnions Are More Flexible: A Mechanical Analysis of THA Taper Designs

Authors: David A. Porter, MD, Robert M. Urban, PhD, Joshua J. Jacobs, MD, Jeremy L. Gilbert, PhD, José A. Rodriguez, MD, H. John Cooper, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 12/2014

Login to get access

Abstract

Background

There is renewed concern surrounding the potential for corrosion at the modular head-neck junction to cause early failure in contemporary THAs. Although taper corrosion involves a complex interplay of many factors, a previous study suggested that a decrease in flexural rigidity of the femoral trunnion may be associated with an increased likelihood of corrosion at retrieval.

Questions/purposes

By analyzing a large revision retrieval database of femoral stems released during a span of three decades, we asked: (1) how much does flexural rigidity vary among different taper designs; (2) what is the contribution of taper geometry alone to flexural rigidity of the femoral trunnion; and (3) how have flexural rigidity and taper length changed with time in this group of revised retrievals?

Methods

A dual-center retrieval analysis of 85 modular femoral stems released between 1983 and 2012 was performed, and the flexural rigidity and length of the femoral trunnions were determined. These stems were implanted between 1991 and 2012 and retrieved at revision or removal surgery between 2004 and 2012. There were 10 different taper designs made from five different metal alloys from 16 manufacturers. Digital calipers were used to measure taper geometries by two independent observers.

Results

Median flexural rigidity was 228 N-m2; however, there was a wide range of values among the various stems spanning nearly an order of magnitude between the most flexible (80 N-m2) and most rigid (623 N-m2) trunnions, which was partly attributable to the taper geometry and to the material properties of the base alloy. There was a negative correlation between flexural rigidity and length of the trunnion and release date of the stem.

Conclusions

There is wide variability in flexural rigidity of various taper designs, with a trend toward trunnions becoming shorter and less rigid with time.

Clinical Relevance

This temporal trend may partly explain why taper corrosion is being seen with increasing frequency in modern THAs.
Literature
1.
go back to reference Barrack RL, Butler RA, Laster DR, Andrews P. Stem design and dislocation after revision total hip arthroplasty: clinical results and computer modeling. J Arthroplasty. 2001;16(8 suppl 1):8–12.PubMedCrossRef Barrack RL, Butler RA, Laster DR, Andrews P. Stem design and dislocation after revision total hip arthroplasty: clinical results and computer modeling. J Arthroplasty. 2001;16(8 suppl 1):8–12.PubMedCrossRef
2.
go back to reference Branovacki G. Ortho Atlas: Hip Arthroplasty US Femoral Implants 1938–2008. Chicago, IL: Ortho Atlas Publishing, Inc; 2008. Branovacki G. Ortho Atlas: Hip Arthroplasty US Femoral Implants 1938–2008. Chicago, IL: Ortho Atlas Publishing, Inc; 2008.
3.
go back to reference Brown SA, Abera A, D’Onofrio M, Fleming C. Effects of neck extension, coverage and frequency on the fretting corrosion of modular THR bore and cone interface. In: Marlow DE, Parr JE, Mayor MB, eds. Modularity of Orthopedic Implants. ASTM Special Technical Publication 1301. West Conshohocken, PA: American Society for Testing and Materials; 1997:189–198. Brown SA, Abera A, D’Onofrio M, Fleming C. Effects of neck extension, coverage and frequency on the fretting corrosion of modular THR bore and cone interface. In: Marlow DE, Parr JE, Mayor MB, eds. Modularity of Orthopedic Implants. ASTM Special Technical Publication 1301. West Conshohocken, PA: American Society for Testing and Materials; 1997:189–198.
4.
go back to reference Brown SA, Flemming CA, Kawalec JS, Placko HE, Vassaux C, Merritt K, Payer JH, Kraay MJ. Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995;61:19–26.CrossRef Brown SA, Flemming CA, Kawalec JS, Placko HE, Vassaux C, Merritt K, Payer JH, Kraay MJ. Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995;61:19–26.CrossRef
5.
go back to reference Burroughs BR, Muratoglu OK, Bragdon CR, Wannomae KK, Christensen S, Lozynsky AJ, Harris WH. In vitro comparison of frictional torque and torsional resistance of aged conventional gamma-in-nitrogen sterilized polyethylene versus aged highly crosslinked polyethylene articulating against head sizes larger than 32 mm. Acta Orthop. 2006;77:710–718.PubMedCrossRef Burroughs BR, Muratoglu OK, Bragdon CR, Wannomae KK, Christensen S, Lozynsky AJ, Harris WH. In vitro comparison of frictional torque and torsional resistance of aged conventional gamma-in-nitrogen sterilized polyethylene versus aged highly crosslinked polyethylene articulating against head sizes larger than 32 mm. Acta Orthop. 2006;77:710–718.PubMedCrossRef
6.
go back to reference Clyburn TA. Pseudotumor in metal-on-polyethylene total hip arthroplasty: a case report. Reconstr Rev. 2013;3:18–21.CrossRef Clyburn TA. Pseudotumor in metal-on-polyethylene total hip arthroplasty: a case report. Reconstr Rev. 2013;3:18–21.CrossRef
7.
go back to reference Collier JP, Mayor MB, Jensen RE, Surprenant VA, Surprenant HP, McNamar JL, Belec L. Mechanisms of failure of modular prostheses. Clin Orthop Relat Res. 1992;285:129–139.PubMed Collier JP, Mayor MB, Jensen RE, Surprenant VA, Surprenant HP, McNamar JL, Belec L. Mechanisms of failure of modular prostheses. Clin Orthop Relat Res. 1992;285:129–139.PubMed
8.
go back to reference Collier JP, Mayor MB, Williams IR, Surprenant VA, Surprenant HP, Currier BH. The tradeoffs associated with modular hip prostheses. Clin Orthop Relat Res. 1995;311:91–101.PubMed Collier JP, Mayor MB, Williams IR, Surprenant VA, Surprenant HP, Currier BH. The tradeoffs associated with modular hip prostheses. Clin Orthop Relat Res. 1995;311:91–101.PubMed
9.
go back to reference Collier JP, Surprenant VA, Jensen RE, Mayor MB. Corrosion at the interface of cobalt-alloy heads on titanium-alloy stems. Clin Orthop Relat Res. 1991;271:305–312.PubMed Collier JP, Surprenant VA, Jensen RE, Mayor MB. Corrosion at the interface of cobalt-alloy heads on titanium-alloy stems. Clin Orthop Relat Res. 1991;271:305–312.PubMed
10.
go back to reference Collier JP, Surprenant, VA, Jensen RE, Mayor MB, Suprenant HP. Corrosion between the components of modular femoral hip prostheses. J Bone Joint Surg Br. 1992;74:511–517.PubMed Collier JP, Surprenant, VA, Jensen RE, Mayor MB, Suprenant HP. Corrosion between the components of modular femoral hip prostheses. J Bone Joint Surg Br. 1992;74:511–517.PubMed
11.
go back to reference Cook RB, Bolland BJ, Wharton JA, Tilley S, Latham JM, Wood RJ. Pseudotumour formation due to tribocorrosion at the taper interface of large diameter metal on polymer modular total hip replacements. J Arthroplasty. 2013;28:1430–1436.PubMedCrossRef Cook RB, Bolland BJ, Wharton JA, Tilley S, Latham JM, Wood RJ. Pseudotumour formation due to tribocorrosion at the taper interface of large diameter metal on polymer modular total hip replacements. J Arthroplasty. 2013;28:1430–1436.PubMedCrossRef
12.
go back to reference Cook SD, Barrack RL, Baffes GC, Clemow AJ, Serekian P, Dong N, Kester MA. Wear and corrosion of modular interfaces in total hip replacements. Clin Orthop Relat Res. 1994;298:80–88.PubMed Cook SD, Barrack RL, Baffes GC, Clemow AJ, Serekian P, Dong N, Kester MA. Wear and corrosion of modular interfaces in total hip replacements. Clin Orthop Relat Res. 1994;298:80–88.PubMed
13.
go back to reference Cook SD, Barrack R, Clemow AJ. Corrosion and wear at the modular interface of uncemented femoral stems. J Bone Joint Surg Br. 1994;76:68–72.PubMed Cook SD, Barrack R, Clemow AJ. Corrosion and wear at the modular interface of uncemented femoral stems. J Bone Joint Surg Br. 1994;76:68–72.PubMed
14.
go back to reference Cooper HJ, Della Valle CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM, Jacobs JJ. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am. 2012;94:1655–1661.PubMed Cooper HJ, Della Valle CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM, Jacobs JJ. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am. 2012;94:1655–1661.PubMed
15.
go back to reference Cooper HJ, Della Valle CJ, Jacobs JJ. Biologic implications of taper corrosion in total hip arthroplasty. Semin Arthroplasty. 2012;23:273–278.CrossRef Cooper HJ, Della Valle CJ, Jacobs JJ. Biologic implications of taper corrosion in total hip arthroplasty. Semin Arthroplasty. 2012;23:273–278.CrossRef
16.
go back to reference Dyrkacz RM, Brandt JM, Ojo OA, Turgeon TR, Wyss UP. The influence of head size on corrosion and fretting behaviour at the head-neck interface of artificial hip joints. J Arthroplasty. 2013;28:1036–1040.PubMedCrossRef Dyrkacz RM, Brandt JM, Ojo OA, Turgeon TR, Wyss UP. The influence of head size on corrosion and fretting behaviour at the head-neck interface of artificial hip joints. J Arthroplasty. 2013;28:1036–1040.PubMedCrossRef
17.
go back to reference Geetha M, Singh AK, Asokamani R, Gogia AK. Ti based biomaterials, the ultimate choice for orthopaedic implants: a review. Prog Mater Sci. 2009;54:397–425.CrossRef Geetha M, Singh AK, Asokamani R, Gogia AK. Ti based biomaterials, the ultimate choice for orthopaedic implants: a review. Prog Mater Sci. 2009;54:397–425.CrossRef
18.
go back to reference Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations: the effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993;27:1533–1544.PubMedCrossRef Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations: the effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993;27:1533–1544.PubMedCrossRef
19.
go back to reference Gilbert JL, Buckley CA, Jacobs JJ, Bertin KC, Zernich MR. Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems: a failure analysis of two implants. J Bone Joint Surg Am. 1994;76:110–115.PubMed Gilbert JL, Buckley CA, Jacobs JJ, Bertin KC, Zernich MR. Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems: a failure analysis of two implants. J Bone Joint Surg Am. 1994;76:110–115.PubMed
20.
go back to reference Gilbert JL, Mali S. Medical implant corrosion: electrochemistry at metallic biomaterial surfaces. In: Eliaz N, ed. Degradation of Implant Materials. New York, NY: Springer; 2012:1–28.CrossRef Gilbert JL, Mali S. Medical implant corrosion: electrochemistry at metallic biomaterial surfaces. In: Eliaz N, ed. Degradation of Implant Materials. New York, NY: Springer; 2012:1–28.CrossRef
21.
go back to reference Gilbert JL, Mehta M, Pinder B. Fretting crevice corrosion of stainless steel stem-CoCr femoral head connections: comparisons of materials, initial moisture, and offset length. J Biomed Mater Res B Appl Biomater. 2009;88:162–173.PubMedCrossRef Gilbert JL, Mehta M, Pinder B. Fretting crevice corrosion of stainless steel stem-CoCr femoral head connections: comparisons of materials, initial moisture, and offset length. J Biomed Mater Res B Appl Biomater. 2009;88:162–173.PubMedCrossRef
22.
go back to reference Goldberg JR, Gilbert JL. In vitro corrosion testing of modular hip tapers. J Biomed Mater Res B Appl Biomater. 2003;64:78–93.PubMedCrossRef Goldberg JR, Gilbert JL. In vitro corrosion testing of modular hip tapers. J Biomed Mater Res B Appl Biomater. 2003;64:78–93.PubMedCrossRef
23.
go back to reference Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002;401:149–161.PubMedCrossRef Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002;401:149–161.PubMedCrossRef
24.
go back to reference Jacobs JJ, Skipor AK, Patterson LM, Hallab NJ, Paprosky WG, Black J, Galante JO. Metal release in patients who have had a primary total hip arthroplasty: a prospective, controlled, longitudinal study. J Bone Joint Surg Am. 1998;80:1447–1458.PubMed Jacobs JJ, Skipor AK, Patterson LM, Hallab NJ, Paprosky WG, Black J, Galante JO. Metal release in patients who have had a primary total hip arthroplasty: a prospective, controlled, longitudinal study. J Bone Joint Surg Am. 1998;80:1447–1458.PubMed
25.
go back to reference Jacobs JJ, Urban RM, Gilbert JL, Skipor AK, Black J, Jasty M, Galante JO. Local and distant products from modularity. Clin Orthop Relat Res. 1995;319:94–105.PubMed Jacobs JJ, Urban RM, Gilbert JL, Skipor AK, Black J, Jasty M, Galante JO. Local and distant products from modularity. Clin Orthop Relat Res. 1995;319:94–105.PubMed
26.
go back to reference Jauch SY, Huber G, Hoenig E, Baxmann M, Grupp TM, Morlock MM. Influence of material coupling and assembly condition on the magnitude of micromotion at the stem–neck interface of a modular hip endoprosthesis. J Biomech. 2011;44:1747–1751.PubMedCrossRef Jauch SY, Huber G, Hoenig E, Baxmann M, Grupp TM, Morlock MM. Influence of material coupling and assembly condition on the magnitude of micromotion at the stem–neck interface of a modular hip endoprosthesis. J Biomech. 2011;44:1747–1751.PubMedCrossRef
27.
go back to reference Lavernia CJ, Baerga L, Barrack RL, Tozakoglou E, Cook SD, Lata L, Rossi MD. The effects of blood and fat on Morse taper disassembly forces. Am J Orthop (Belle Mead NJ). 2009;38:187–190.PubMed Lavernia CJ, Baerga L, Barrack RL, Tozakoglou E, Cook SD, Lata L, Rossi MD. The effects of blood and fat on Morse taper disassembly forces. Am J Orthop (Belle Mead NJ). 2009;38:187–190.PubMed
28.
go back to reference Lieberman JR, Rimnac CM, Garvin KL, Klein RW, Salvati EA. An analysis of the head-neck taper interface in retrieved hip prostheses. Clin Orthop Relat Res. 1994;300:162–167.PubMed Lieberman JR, Rimnac CM, Garvin KL, Klein RW, Salvati EA. An analysis of the head-neck taper interface in retrieved hip prostheses. Clin Orthop Relat Res. 1994;300:162–167.PubMed
29.
go back to reference Lindgren JU, Brismar BH, Wikstrom AC. Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis. J Bone Joint Surg Br. 2011;93:1427–1430.PubMedCrossRef Lindgren JU, Brismar BH, Wikstrom AC. Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis. J Bone Joint Surg Br. 2011;93:1427–1430.PubMedCrossRef
30.
go back to reference Lucas LC, Buchanan RA, Lemons JE. Investigations on the galvanic corrosion of multialloy total hip prostheses. J Biomed Mater Res. 1981;15:731–747.PubMedCrossRef Lucas LC, Buchanan RA, Lemons JE. Investigations on the galvanic corrosion of multialloy total hip prostheses. J Biomed Mater Res. 1981;15:731–747.PubMedCrossRef
31.
go back to reference Mao X, Tay GH, Godbolt DB, Crawford RW. Pseudotumor in a well-fixed metal-on-polyethylene uncemented hip arthroplasty. J Arthroplasty. 2012;27:493.e13–17. Mao X, Tay GH, Godbolt DB, Crawford RW. Pseudotumor in a well-fixed metal-on-polyethylene uncemented hip arthroplasty. J Arthroplasty. 2012;27:493.e13–17.
32.
go back to reference Meftah M, Nicolaou N, Rodriguez JA. Metal allergy response to femoral head-neck corrosion after total hip replacement. Curr Orthop Pract. 2010;21:530–536.CrossRef Meftah M, Nicolaou N, Rodriguez JA. Metal allergy response to femoral head-neck corrosion after total hip replacement. Curr Orthop Pract. 2010;21:530–536.CrossRef
33.
go back to reference Rodriguez JA, Rathod PA. Large diameter heads: is bigger always better? J Bone Joint Surg Br. 2012;94(11 suppl A):52–54. Rodriguez JA, Rathod PA. Large diameter heads: is bigger always better? J Bone Joint Surg Br. 2012;94(11 suppl A):52–54.
34.
go back to reference Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86:420–428.PubMedCrossRef Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86:420–428.PubMedCrossRef
35.
go back to reference Svensson O, Mathiesen EB, Reinholt FP, Blomgren G. Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty: a case report. J Bone Joint Surg Am. 1988;70:1238–1242.PubMed Svensson O, Mathiesen EB, Reinholt FP, Blomgren G. Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty: a case report. J Bone Joint Surg Am. 1988;70:1238–1242.PubMed
37.
go back to reference Urban RM, Tomlinson MJ, Hall DJ, Jacobs JJ. Accumulation in liver and spleen of metal particles generated at nonbearing surfaces in hip arthroplasty. J Arthroplasty. 2004;19(8 suppl 3):94–101.PubMedCrossRef Urban RM, Tomlinson MJ, Hall DJ, Jacobs JJ. Accumulation in liver and spleen of metal particles generated at nonbearing surfaces in hip arthroplasty. J Arthroplasty. 2004;19(8 suppl 3):94–101.PubMedCrossRef
38.
go back to reference Volz RG. Orthopaedic Device Reference. Breckenridge, CO: Association of Bone and Joint Surgeons; 1998. Volz RG. Orthopaedic Device Reference. Breckenridge, CO: Association of Bone and Joint Surgeons; 1998.
39.
go back to reference Walsh AJ, Nikolaou VS, Antoniou J. Inflammatory pseudotumor complicating metal-on-highly cross-linked polyethylene total hip arthroplasty. J Arthroplasty. 2012;27:324.e5–8. Walsh AJ, Nikolaou VS, Antoniou J. Inflammatory pseudotumor complicating metal-on-highly cross-linked polyethylene total hip arthroplasty. J Arthroplasty. 2012;27:324.e5–8.
Metadata
Title
Modern Trunnions Are More Flexible: A Mechanical Analysis of THA Taper Designs
Authors
David A. Porter, MD
Robert M. Urban, PhD
Joshua J. Jacobs, MD
Jeremy L. Gilbert, PhD
José A. Rodriguez, MD
H. John Cooper, MD
Publication date
01-12-2014
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 12/2014
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-014-3965-3

Other articles of this Issue 12/2014

Clinical Orthopaedics and Related Research® 12/2014 Go to the issue

Orthopaedic Case of the Month

Thigh Pain in an 18-year-old Man

Symposium: Award Papers from Turkish Society of Orthopaedics and Traumatology 2013

Apical and Intermediate Anchors Without Fusion Improve Cobb Angle and Thoracic Kyphosis in Early-onset Scoliosis

Symposium: Award Papers from Turkish Society of Orthopaedics and Traumatology 2013

Time-dependent Changes After Latissimus Dorsi Transfer: Tenodesis or Tendon Transfer?