Published in: Clinical Orthopaedics and Related Research® 1/2015

01-01-2015 | Symposium: 2014 Knee Society Proceedings

How Precise Is Computer-navigated Gap Assessment in TKA?

Authors: James B. Stiehl, MD, David A. Heck, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 1/2015

Login to get access

Abstract

Background

Methods to improve gap balancing in total knee arthroplasty (TKA) include the development of calibrated distractors and various devices to determine the distances of the gaps. However, few studies have validated the accuracy or precision of computer navigation to determine these measurements, especially gaps created after bone cuts have been made; doing so would be important, because optimal surgical technique relies on appropriate gap spacing.

Questions/purposes

We investigated the ability of a new image-free computer navigation surface registration protocol to measure gap distances in TKA.

Methods

Eight embalmed cadaveric specimens of the lower extremity were used. A surface registration software protocol defined the most distal and posterior surface points of the femoral condyles and the navigation system measured the distance of the most distal femoral condyle point to the surface of the tibia after tibial resection. The tibial resection was perpendicular to the mechanical axis and was cut with a 7° posterior slope. The navigation system measured gaps spaced by modular spacing blocks at 5° intervals from full extension to 120° of flexion. Repeatability assessed repeated measures by one surgeon. Reproducibility was assessed by performing the same measurements after complete reregistration of the computer protocol to the cadaver bones.

Results

The gaps measured by the computer were statistically the same as those assessed with the use of blocks with a maximum measurement error of 1 mm. Reregistration did introduce error into the measurement. The gaps changed with position of knee flexion, and there was gradual and significant stretching of the gaps with repeated measurements.

Conclusions

Preliminary testing shows that computer navigation can reproduce static measurements reliably and with equal accuracy as spacer blocks. We have not demonstrated that this could be applied in a dynamic setting.

Clinical Relevance

This computer navigation system has sufficient precision to warrant investigation in the clinical setting for measuring gaps created during the surgical procedure.
Literature
1.
go back to reference Bellemans J, D’Hooghe P, Vandenneucker H, Van Damme G, Victor J. Soft tissue balance in total knee arthroplasty: does stress relaxation occur perioperatively? Clin Orthop Relat Res. 2006;452:49–52PubMedCrossRef Bellemans J, D’Hooghe P, Vandenneucker H, Van Damme G, Victor J. Soft tissue balance in total knee arthroplasty: does stress relaxation occur perioperatively? Clin Orthop Relat Res. 2006;452:49–52PubMedCrossRef
2.
go back to reference Davis ET, Pagkalos J, Gallie PA, Macgroarty K, Waddell JP, Schemitsch EH. Defining the errors in the registration process during imageless computer navigation in total knee arthroplasty: a cadaveric study. J Arthroplasty. 2014;29:698–701PubMedCrossRef Davis ET, Pagkalos J, Gallie PA, Macgroarty K, Waddell JP, Schemitsch EH. Defining the errors in the registration process during imageless computer navigation in total knee arthroplasty: a cadaveric study. J Arthroplasty. 2014;29:698–701PubMedCrossRef
3.
go back to reference Hamelynk KJ. The rheumatoid knee. In: Hamelynk KJ, Stiehl JB, eds. LCS ® Mobile Bearing Knee Arthroplasty: A 25 Years Worldwide Experience. Heidelberg, Germany: Springer-Verlag; 2002CrossRef Hamelynk KJ. The rheumatoid knee. In: Hamelynk KJ, Stiehl JB, eds. LCS ® Mobile Bearing Knee Arthroplasty: A 25 Years Worldwide Experience. Heidelberg, Germany: Springer-Verlag; 2002CrossRef
4.
go back to reference Insall J, Scott WN, Ranawat CS. The total condylar knee prosthesis: report of 220 cases. J Bone Joint Surg Am. 1979;61:173–180PubMed Insall J, Scott WN, Ranawat CS. The total condylar knee prosthesis: report of 220 cases. J Bone Joint Surg Am. 1979;61:173–180PubMed
5.
go back to reference Matsumoto T, Kuroda R, Kubo S, Muratsu H, Muzuno K, Kurosaka M. The intra-operative joint gap in cruciate-retaining compared with posterior-stabilized total knee replacement. J Bone Joint Surg Br. 2009;91:475–480PubMedCrossRef Matsumoto T, Kuroda R, Kubo S, Muratsu H, Muzuno K, Kurosaka M. The intra-operative joint gap in cruciate-retaining compared with posterior-stabilized total knee replacement. J Bone Joint Surg Br. 2009;91:475–480PubMedCrossRef
6.
go back to reference Nowakowski AM, Majewski M, Muller-Gerbl M, Valderrabano V. Development of a force-determining tensor to measure ‘physiologic knee ligament gaps’ without bone resection using a total knee arthroplasty approach. J Orthop Sci. 2011;16:56–63PubMedCrossRef Nowakowski AM, Majewski M, Muller-Gerbl M, Valderrabano V. Development of a force-determining tensor to measure ‘physiologic knee ligament gaps’ without bone resection using a total knee arthroplasty approach. J Orthop Sci. 2011;16:56–63PubMedCrossRef
7.
go back to reference Nowakowski AM, Majewski M, Muller-Gerbl M, Valderrabano V. Measurement of knee joint gaps without bone resection: ‘physiologic’ extension and flexion gaps in total knee arthroplasty are asymmetric and unequal and anterior and posterior ligament resections produce different gap changes. J Orthop Res. 2012;30:522–527PubMedCrossRef Nowakowski AM, Majewski M, Muller-Gerbl M, Valderrabano V. Measurement of knee joint gaps without bone resection: ‘physiologic’ extension and flexion gaps in total knee arthroplasty are asymmetric and unequal and anterior and posterior ligament resections produce different gap changes. J Orthop Res. 2012;30:522–527PubMedCrossRef
8.
go back to reference Perrin N, Stindel E, Roux C. BoneMorphing versus freehand localization of anatomical landmarks: consequences for the reproducibility of implant positioning in total knee arthroplasty. Comput Aided Surg. 2005;10:301–309PubMedCrossRef Perrin N, Stindel E, Roux C. BoneMorphing versus freehand localization of anatomical landmarks: consequences for the reproducibility of implant positioning in total knee arthroplasty. Comput Aided Surg. 2005;10:301–309PubMedCrossRef
9.
go back to reference Ranawat AS, Ranawat CS, Elkus M, Rasquinha VJ, Rossi R, Babhulkar S. Total knee arthroplasty for severe valgus deformity. J Bone Joint Surg Am. 2005;87:271–284PubMed Ranawat AS, Ranawat CS, Elkus M, Rasquinha VJ, Rossi R, Babhulkar S. Total knee arthroplasty for severe valgus deformity. J Bone Joint Surg Am. 2005;87:271–284PubMed
10.
go back to reference Siston RA, Goodman SB, Delp SL, Giori NJ. Coronal plane stability before and after total knee arthroplasty. Clin Orthop Relat Res. 2007;463:43–49PubMed Siston RA, Goodman SB, Delp SL, Giori NJ. Coronal plane stability before and after total knee arthroplasty. Clin Orthop Relat Res. 2007;463:43–49PubMed
11.
go back to reference Siston RA, Goodman SB, Patel JJ, Delp SL, Giori NJ. The variability of tibial rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2006;452:65–69PubMedCrossRef Siston RA, Goodman SB, Patel JJ, Delp SL, Giori NJ. The variability of tibial rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2006;452:65–69PubMedCrossRef
12.
go back to reference Stindel E, Briard JL, Merloz P, Plaweski S, Dubrana F, Lefevre C, Troccaz J. Bone morphing: 3D morphological data for total knee arthroplasty. Comput Aided Surg. 2002;7:156–168PubMedCrossRef Stindel E, Briard JL, Merloz P, Plaweski S, Dubrana F, Lefevre C, Troccaz J. Bone morphing: 3D morphological data for total knee arthroplasty. Comput Aided Surg. 2002;7:156–168PubMedCrossRef
13.
go back to reference Tanaka K, Muratsu H, Mizuno K, Kuroda R, Yoshiya S, Kurosaka M. Soft tissue balance measurement in anterior cruciate ligament-resected knee joint: cadaveric study as a model for cruciate-retaining total knee arthroplasty. J Orthop Sci. 2007;12:149–153PubMedCrossRef Tanaka K, Muratsu H, Mizuno K, Kuroda R, Yoshiya S, Kurosaka M. Soft tissue balance measurement in anterior cruciate ligament-resected knee joint: cadaveric study as a model for cruciate-retaining total knee arthroplasty. J Orthop Sci. 2007;12:149–153PubMedCrossRef
14.
go back to reference Van Damme G, Defoort K, Ducoulombier Y, Van Glabbeek F, Bellemans J, Victor J. What should the surgeon aim for when performing computer assisted total knee arthroplasty. J Bone Joint Surg Am. 2005;87(Suppl 2):52–58PubMedCrossRef Van Damme G, Defoort K, Ducoulombier Y, Van Glabbeek F, Bellemans J, Victor J. What should the surgeon aim for when performing computer assisted total knee arthroplasty. J Bone Joint Surg Am. 2005;87(Suppl 2):52–58PubMedCrossRef
15.
go back to reference Yau WP, Leung A, Liu KG, Yan CH, Wong LL, Chiu KY. Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation assisted total knee arthroplasty. J Arthroplasty. 2007;22:1150–1161PubMedCrossRef Yau WP, Leung A, Liu KG, Yan CH, Wong LL, Chiu KY. Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation assisted total knee arthroplasty. J Arthroplasty. 2007;22:1150–1161PubMedCrossRef
Metadata
Title
How Precise Is Computer-navigated Gap Assessment in TKA?
Authors
James B. Stiehl, MD
David A. Heck, MD
Publication date
01-01-2015
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 1/2015
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-014-3785-5

Other articles of this Issue 1/2015

Clinical Orthopaedics and Related Research® 1/2015 Go to the issue